Abstract
Background/Aims
Methods
Notes
Author Contributions
Conceptualization: Martin Scherrer
Data curation: Vanessa M Eichel, MS
Formal analysis: VME, MS
Investigation: VME, MS
Methodology: VME, MS
Project administration: Samy Unser
Resources: Jonas M Jabs, SU
Supervision: Nico T Mutters, MS
Validation: NTM, MS
Visualization: VME, JMJ, SU, MS
Writing-original draft: VME
Writing-review&editing: NTM
ACKNOWLEDGMENTS
REFERENCES
Table 1.
Pre-cleaning | Immerse the distal end in cleaning solutiona) |
Suck through until no more contamination is visible in the rinsing solution, but at least 30 seconds | |
Variable time interval of storage at room temperature | |
Reprocessing | Place the test object in the basin with cleaning solutiona) |
Rinse the test object with cleaning solutiona) | |
Brush the inner lumen with a flexible disposable brush until the brush is free of impurities | |
Rinse the test object in a basin with demineralised water, rinse the inner lumen | |
Remove of residual water with medical compressed air | |
Connect the test object to the loading trolley of the EWD | |
Start the machine-based reprocessing process in the EWDb) | |
Dry the test object with medical compressed air |
a) Protocol 1: Hartmann Bodedex forte 2%; Protocol 2: Hartmann Bodedex forte 1%; Protocols 3 and 4: Dr. Weigert Neodisher Endo DIS active.
b) Protocol 1: Olympus ETD4+GA, Program 2 Standard TR, Hartmann Korsolex Endo Cleaner + Hartmann Korsolex Endo Disinfectant; Protocol 2: Olympus ETD4+GA (Program 1 Standard) + Hartmann Korsolex Endo Cleaner + Hartmann Korsolex Endo Disinfectan; Protocol 3: EWD Wassenburg Adaptascope (normal program) + Dr. Weigert Endo Clean + Dr. Weigert Endo Sept GA; Protocol 4: Olympus ETD3+PAA, Program 6 Standard, Olympus EndoDET + Olympus EndoDIS.
Table 2.
Pre-cleaning | Wipe the distal end of the endoscope with a non-sterile compress |
Immerse the distal end in cleaning solutiona) | |
Aspirate with at least 200 ml cleaning solution; Keep suction valve pressed until no more contamination is visible; but at least 30 sec | |
Flush the biopsy channel and, if necessary, the additional flushing channel with cleaning solution using a 20 ml syringe until no more visible particles are flushed out | |
Multiple actuation and subsequent removal of the air/water valveb) | |
Insert and actuate the cleaning valve several timesb) | |
Remove the distal end of the endoscope from cleaning solution | |
Separate the endoscope from the light source, suction and optic rinsing bottle | |
Variable time interval of storage at room temperature | |
Reprocessing | Connection to the leak tester and performance of the leak test |
Immerse the endoscope in the cleaning solutiona) | |
Remove the valves; Disposal of the biopsy channel valve; Manual brushing of the aspiration and air/water valve; Brush and rinse the valves in deionised water in the secondary basin; Remove water residues with medical compressed air | |
Flush all channels manually with cleaning solutiona) | |
Brush manually with a flexible disposable cleaning brush; Pull the brush several times completely through all accessible channels until the brush emerges free of visible dirt | |
Wipe the entire endoscope; Pay attention to areas that are difficult to access (operating wheels, working channel entrance, distal end with optical lens, connections) | |
Use the control wheels to detect micro lesions | |
Remove the endoscope from the cleaning solution while leaving it connected to the leak tester; Immerse in the secondary tank with demineralised water; Rinse all channels to avoid mixing of cleaning and disinfection agents | |
Blow through the channels with medical compressed air to remove any remaining deionized water | |
Release the pressure from the leak tester and disconnect the endoscope | |
Properly fit the rinsing basket and connect the channels and the leak test according to the manufacturer’s instructionsc); Close the EWD and start the program | |
Check the correct and complete procedure after the end of the reprocessing program | |
Perform a hygienic hand disinfection | |
Pull out the rinsing baskets from the EWD; Detach the rinsing adapter from the endoscope; Remove the endoscope from the rinsing basket | |
Dry the endoscope with medical compressed air |