Journal List > J Korean Soc Echocardiogr > v.10(1) > 1075062

J Korean Soc Echocardiogr. 2002 Jun;10(1):18-23. Korean.
Published online June 30, 2002.  https://doi.org/10.4250/jkse.2002.10.1.18
Copyright © 2002 Korean Society of Echocardiography
Intraoperative assessment of Native Coronary Artery and Bypass Graft Flow Using a 15 MHz Linear Array Transducer
Eun Ju Cho, Ho Joong Youn, Hae Ok Chung, Chul Soo Park, Hui Kyung Jeon, Wook Sung Chung, Chong Jin Kim, Jae Hyung Kim, Kyu Bo Choi, Soon Jo Hong, Sung Bo Sim, Sun Hee Lee and Moon Sub Kwack
Department of Internal Medicine, College of Medicine,The Catholic University of Korea, Seoul, Korea.
Department of Thoracic and Cardiovascular Surgery, College of Medicine,The Catholic University of Korea, Seoul, Korea.
Abstract

BACKGROUND

The internal mammary artery graft (IMAG) showed a gradual transition in the phasic flow pattern from predominant systolic velocity proximally at the origin from the subclavian artery to the predominant diastolic velocity distal to the anastomosis with coronary artery. IMAG without significant stenosis showed a typical pulsed Doppler flow pattern similar to that of coronary artery characterized by a predominant diastolic component. On the other hand, patients with occluded IMAGs or IMAGs with severe stenosis, low velocity profiles were recorded during diastole with an increase in the systolic component.

OBJECTIVES

Identifying abnormal graft flow intraoperatively could allow for immediate graft revision. The aim of this study was to test the feasibility of using a new ultrasound 15 MHz linear array transducer.

METHODS

In six patients (M:F=4:2, mean age=69+/-11 yrs), a 15 MHz linear array probe with a sterile cover was placed directly on native coronary arteries and grafts after anatomosis. 2-dimensional image, color and pulsed Doppler signals of native coronary arteries, IMA, saphenous vein and penetrating intramyocardial coronary arteries (PICA) were observed.

RESULTS

1) 2-dimensional image and color flow signals of all grafts and native coronary arteries were visualized. 2) The flow pattern of pulsed Doppler signals of all native coronary arteries including PICA were diastolic dominant. 3) The diastolic dominant pattern typical of a patent grafts on pulsed Doppler were obtained at anastomosis site of graft vessels. 4) The intimal thickening of right coronary artery was also visualized on 2-dimensional images.

CONCLUSION

A new echo Doppler probe can be useful for intraoperative assessment of graft flow during bypass surgery.

Keywords: Monitoring, intraoperative; Coronary artery bypass; Blood flow velocity; Echocardiography

TOOLS
Similar articles

Native Coronary Artery and Grafted Artery Spasm Just after Coronary Artery Bypass Grafting: A Case Report

A Study on Graft Angiography and Patency after Coronary Artery Bypass Graft

Totally Occluded Grafted Right Internal Mammary Artery to Anomalously Originated Right Coronary Artery

Hemodynamic effects of the geometric dimensions of graft vessels in coronary artery bypass graft models

The Angiographic Evaluation of Patency after Coronary Artery Bypass Graft