Journal List > Transl Clin Pharmacol > v.26(2) > 1101033

Jeong, Park, Hyun, Park, Shin, and Shin: Determination of donepezil in human plasma using ultra performance liquid chromatography-tandem mass spectrometry

Abstract

An ultra performance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS) method was developed and validated for the quantification of donepezil in human plasma. Donepezil and donepezil-D4 were extracted from human plasma by liquid-liquid extraction using a mixture of hexane and ethyl acetate (70:30 v/v). The extracted samples were analyzed using a Thermo Hypersil Gold C18 column with 5% acetic acid in 20 mM ammonium acetate buffer (pH 3.3) and 100% acetonitrile as a mobile phase with the 60:40 (v:v) isocratic method, at a flow rate of 0.3 mL/min. The injection volume was 3 µL, and the total run time was 3 min. Inter- and intra-batch accuracies ranged from 98.0% to 110.0%, and the precision was below 8%. The developed method was successfully applied to the quantification of donepezil in human plasma. The mean (standard deviation) maximum concentration and the median (range) time to maximum concentration were 8.6 (2.0) ng/mL and 2.0 h (1.0~5.0 h), respectively, in healthy Koreans after oral administration of 5 mg donepezil.

Introduction

Donepezil is a non-competitive, reversible, acetylcholinesterase (AChE) inhibitor and is used as a selective treatment for Alzheimer's disease.[1] Alzheimer's disease is a neurodegenerative disorder and causes memory dysfunction, cognitive impairment, and other symptoms. The use of AChE inhibitors or N-methyl-D-aspartate (NMDA)-receptor inhibitors helps relieve the symptoms and slows the progression of Alzheimer's disease. [2] Previously, tacrine was widely used as an AChE-based treatment; however, due to its hepatotoxicity, it has been replaced by other drugs, such as donepezil, rivastigmine, and galantamine. Among these drugs, donepezil displayed a relatively long half-life (approximately 70 to 80 h, whereas those of the other AChE inhibitors were 0.3 to 12 h) and had the appropriate potency to relieve symptoms for ataxia, cognitive impairment and other neurodegenerative symptoms caused by Alzheimer's disease. Consequently, it is widely used for vascular dementia in addition to Alzheimer's disease.[3]
Previously, quantitative analysis of donepezil was performed mainly by using high-performance liquid chromatography (HPLC) with an ultraviolet or fluorescence detector.[456] These methods have several merits. For example, samples can be easily prepared and analyzed; however, these methods also have some drawbacks. These methods require a long run time (approximately 20 min), they show low sensitivity, and they require a large injection volume (approximately 37 µL).[7] Recently, high performance or ultra performance liquid chromatographytandem mass spectrometers (HPLC-MS/MS or UPLC-MS/MS) have been used more frequently for chemical analysis, including the quantification of drugs in human plasma for various pharmacokinetic studies and for therapeutic drug monitoring.[8910] The UPLC-MS/MS system has a higher sensitivity than conventional UV or fluorescence detector methods, requiring a smaller sample volume than conventional analysis methods, and it has a shorter run time and a higher reproducibility.[6]
This study aimed to develop a donepezil analytical method with high reproducibility and sensitivity using UPLC-MS/MS and a UPLC column. Furthermore, the developed method was applied to the quantification of donepezil in human plasma.

Methods

Materials and reagents

Molecular structures of (A) donepezil and (B) donepezil-D4 (IS) are represented in Figure 1 (A and B). Donepezil hydrochloride (standard, purity 100.0%) and donepezil-D4 hydrochloride (Internal standard, purity 99.0%) were purchased from the Tokyo Chemical Industry (Tokyo, Japan) and CDN Isotopes (Quebec, Canada), respectively. Acetonitrile, methanol, ethyl acetate, and hexane were purchased from Sigma-Aldrich (St. Louis, MO, USA). All the reagents obtained were HPLC grade or extra-pure grade. Deionized water was purified using a Millipore Milli-Q system at 18.2 MΩ (MA, USA). Drug-free pooled plasma containing sodium heparin was purchased from Biochemed (VA, USA). Pooled plasma and six individual plasmas were stored at −80℃ until analysis.

Preparation of quality control samples and calibration standards

Donepezil and donepezil-D4 stock solutions (1 mg/mL) were prepared in deionized water:methanol (50:50, v/v, 50% methanol). A working solution was prepared by serial dilution using 50% methanol final concentrations of 1, 2, 5, 10, 50, 100, and 500 ng/mL, and the working solution was spiked with plasma at a ratio of 1:9. The internal standard (IS) was prepared by dilution to 10 µg/mL using 50% methanol. All the analytes and IS working solutions were stored at 4℃ until analysis.
Calibration standards and quality control (QC) samples were prepared using drug-free pooled plasma. Three different concentrations of the QC samples were prepared by adding 20 µL of working solution to 180 µL of pooled plasma (low, medium and high concentrations of 0.3, 3.5, and 35 ng/mL, respectively). Seven calibration standards were prepared by spiking the working solution into drug-free pooled plasma (final concentrations were 0.1, 0.2, 0.5, 1, 5, 10, and 50 ng/mL, respectively).

Sample preparation

The plasma preserved at −80℃ was thawed at room temperature (18~21℃) and was slightly vortexed for a short time. Donepezil extraction from plasma was conducted by liquid-liquid extraction (LLE). A total of a 200 µL sample of plasma was transferred into a polypropylene tube and was spiked with 20 µL of IS (donepezil-D4, 10 µg/mL in 50% methanol). Then, the sample was extracted for 15 min after adding hexane:ethyl acetate (70:30, v/v) and was vortexed for 3 min prior to centrifugation at 4,000 rpm for 5 min. The supernatant was transferred to a new polypropylene tube and was kept for 60 min at −80℃. Then, the organic solution layer was transferred to a new polypropylene tube. The transferred supernatant was concentrated at 45℃ for 60 min using a vacuum concentrator (Labconco Corp., MO, USA) and was re-dissolved in 150 µL of a mobile phase (5% acetic acid in 20 mM ammonium acetate:100% acetonitrile = 60:40, v/v, pH 3.3). The dissolved sample was centrifuged at 13,200 rpm for 5 min, and 3 µL of the supernatant was injected into the UPLC-MS/MS system for analysis.

Chromatographic conditions

The UPLC-ESI-MS/MS system consisted of an ACQUITY™ UPLC (Waters, MA, USA) and a triple quadrupole tandem Xevo TQ-MS mass spectrometer (Waters, MA, USA). The analyte separation was performed on a Hypersil GOLD C18 column (150 × 2.1 mm, 1.9 µm, Thermo, MA, USA) using 5% acetic acid in 20 mM ammonium acetate buffer (pH 3.3) (A) and 100% acetonitrile (B) by the isocratic method at a ratio of 60:40 (v/v). The total run time was 3 minutes. The autosampler and column oven temperature were maintained at 10℃ and 40℃, respectively. MS/MS detection was performed in the positive ion mode for donepezil and IS. Multiple reaction monitoring (MRM) mode was used for quantification, with a m/z of 380.6 → 91.1 for donepezil and a m/z of 384.2 → 245.1 for IS (donepezil-D4). All the quantified data was processed using Masslynx version 4.1 (Waters, MA, USA). Product ion spectra of (A) donepezil and (B) donepezil-D4 (IS) by collision induced dissociation of corresponding parent ions is presented in Figure 2 (A and B).

Method validation

The validation of the quantification method was performed based on the “Guideline on Bioanalytical Method Validation” published by the Korean Ministry of Food and Drug Safety (MFDS) and US Food and administration (FDA).[1112] The following items were evaluated in this study: linearity, sensitivity, accuracy, precision, selectivity, carry-over, recovery, matrix effect, dilution integrity and stability under various conditions.

Linearity

The calibration standard consisted of a blank (without analyte and IS), a zero (without analyte and with IS), and seven-point samples (with analyte and IS). Linearity was assessed at donepezil concentrations of 0.1~50 ng/mL ranges in human plasma. In addition, linear regression, with a weighting factor of 1/x2, was used to assess whether the coefficient of correlation was higher than 0.997.

Sensitivity

Sensitivity was recognized only by the signal to noise (S/N) being 10 or more on the lower limit of quantification (LLOQ), and the LLOQ was set equal to the lowest concentration of the curve (0.1 ng/mL).

Accuracy and precision

For the evaluation of the accuracy and precision, five consecutive (intra-batch) analyses were performed in three different batches (inter-batch) at LLOQ and low, medium, and high concentrations of QC. An acceptance criterion of accuracy was within ±20% for LLOQ and ±15% for other concentrations. The acceptance criterion of precision also was within a coefficient of variation (CV, %) of less than 20% for LLOQ and less than 15% for other concentrations.

Selectivity

Selectivity was assessed using six individual human plasmas. We prepared four selectivity samples: (1) double blank, (2) spiked IS only, (3) spiked LLOQ concentration working solution only, and (4) spiked LLOQ concentration working solutions including IS. Then, we analyzed and compared the four selectivity samples to determine if they had interference peaks. The influence of interfering substances should be lower than 20% of the analyte peak area and lower than 5% of the IS peak area at LLOQ of the sample.

Carry-over

The carry-over effect was evaluated whether the analyte (donepezil) was higher than 20% of the LLOQ and 5% for the IS in the blank sample. The blank plasma was injected immediately after the injection of the upper limit of quantification (ULOQ) samples; this process was repeated five times.

Recovery

The recovery rate was calculated by comparing the area ratio of the spike-pre-extraction samples and the spike-post-extraction samples.

Matrix effect

Matrix effects were evaluated to assess whether endogenous substances present in the plasma interfered with analyte or IS. The matrix effect was tested with QC samples of low and high concentrations using six different individual plasmas as well as the analyte peak area and IS peak area after sample preparation.

Dilution integrity

The dilution integrity was used to evaluate the effect of sample dilution on the analysis. A sample of 100 ng/mL (2-fold of ULOQ) was diluted 1/10 and 1/50. Accuracy and precision were within 15%.

Stability

The stability test was performed under various conditions. The re-injection was compared with the results obtained by three-replicate analyses of the LoQC and HiQC samples using the samples analyzed the day before (test) as well as fresh samples (reference). Processed sample stability was determined by comparing the sample that was stored in the autosampler for 24 h. Freeze-thaw stability was assessed under three freeze-thaw cycles at −80℃. Short-term stability was measured after the samples were kept at 18~21℃ for 20 h. To assess stock stability at laboratory temperature (18~21℃) for 24 h and 4℃ for 24 days, the analyte to IS peak area ratio of the samples was compared to those of freshly prepared stock solution.
The acceptance criterion of the stability was that the % change of all the stability samples was within 15% of the nominal concentration.

Pharmacokinetic study

A clinical study was conducted in 18 healthy adult males receiving a single oral 5 mg dose of a donepezil hydrochloride tablet (Donep Tab, Daewoong Pharmaceutical, Co. Ltd.) in the Clinical Trial Center of Gachon University Gil Medical Center. The clinical study protocol was approved by the Institutional Review Board of Gachon University Gil Medical Center (IRB No: GBIRB2016-036). All the participants were informed about the purpose and content of the study and voluntarily agreed to the study. Blood samples were collected at pre-dose as well as 1, 2, 3, 3.5, 4, 4.5, 5, 5.5, 6, 8, 12, 24, 48, 72, 120, 168 h after drug administration in a tube containing sodium heparin. The collected blood was immediately centrifuged at 3,000 rpm for 10 min at 4℃ or lower, and the plasma obtained was divided into two polypropylene tubes of 0.7 mL each and stored at −80℃ until analysis.

Results

UPLC-MS/MS

The separation was performed using a Thermo Hypersil GOLD C18 column (150 × 2.1 mm, 1.9 µm) with 5% acetic acid in 20 mM ammonium acetate (a) and 100% acetonitrile (b) using the isocratic method (60:40, v/v). The results showed a good peak shape, and the analytes were well separated.
The extraction method was performed according to Philli et al's study. Donepezil was extracted using an n-hexane and ethyl acetate mixture in plasma. Liquid-liquid extraction (LLE) is a widely used technique for biological sample analysis. In several studies, a pH modifier, such as NaOH or phosphoric acid, is added for stable extraction. In this study, we attempted to extract donepezil in plasma without a pH modifier, and results showed a recovery rate of approximately 60%, which was reproducible even though it was a lower recovery than other references.

Method validation

Linearity

Using a seven-point calibration standard, the linearity of the method was evaluated. The method showed adequate linearity; all of the correlation coefficients (r) of the calibration standard were higher than 0.995, and the determination coefficient (r2) was also higher than 0.980 in three independent batches (Table 1).

Sensitivity

The chromatograms for the double blank sample (not contain any analytes) and for the LLOQ sample are found in Figure 3A and Figure 3B, respectively. The signal-to-noise ratio in the LLOQ sample was 154.3 (Fig. 4A), which satisfied the sensitivity criterion suggested in the bioanalytical method validation guidelines.[1112] The representative chromatogram of the sample that was obtained after oral administration of donepezil is presented in Figure 4B.

Accuracy and precision

The range of inter and intra accuracy was 98.0%~110.0% and 100.7%~104.7%, respectively. The range of inter and intra precision was 0.98%~6.43% and 2.74%~7.97%, respectively. The results of the four quality control samples showed that the LLOQ was within ±20% for all the batches and was within ±15% at all the other QC concentrations (Table 2).

Selectivity

To confirm the interference by endogenous substances, the LLOQ samples (IS spiked) and the blank plasma samples were subjected to the same preparation, and the resulting interference peaks were analyzed on the chromatograms. No interference peak was found in the blank plasma after ULOQ sample injection (data not shown).

Recovery

To assess the recovery, three replicates at low, medium, and high concentrations of QC for donepezil were prepared. The mean recovery was 58.9%, 60.5%, and 60.4%, respectively, and the CV was 1.51% among the three QC samples (Table 3).

Matrix effect

The matrix effect was evaluated using low and high QC samples prepared using six individual plasmas. The CV (%) was 0% for the low QC and was 1.02% for the high QC sample (data not shown).

Dilution integrity

A 1/10 diluted sampled represented the accuracy and was 110.0%~113.9%, and the precision was 1.58%. A 1/50 diluted sample represented the accuracy and was 108.30%~115.8%, with a precision of 2.80%. The 1/10 dilution results satisfied this criteria, but the 1/50 dilution results did not.

Stability

Re-injection, processed sample stability, freeze-thaw stability, short-term stability test and stock stability were evaluated. Three replicated LoQC and HiQC samples were used in the determination of stability. Re-injection, processed sample stability, freeze-thaw stability and short-term stability were compared, and a nominal concentration % change within 15% was considered acceptable. A stock short-term and long-term stability was calculated by comparing the peak area ratio of the test sample and the reference sample. All the stability test results are presented in Table 4.

Pharmacokinetic study

The method showed sufficient sensitivity, linearity, and suitability in the donepezil concentration ranges of 0.1~50 ng/mL for all batches. Moreover, the proper reproducibility and stability were obtained using human plasma. The mean concentration-time profile of donepezil is presented in Figure 5. The time to maximum concentration (Tmax) was observed at approximately 7 h.

Discussion

An analytical method of donepezil detection in plasma using the UPLC-ESI-MS/MS system was developed. The method was successfully applied to extract and quantify donepezil in human plasma. The method was validated according to the standard operating procedures based on the guidelines for bioanalytical method validation published by the Korean MFDS and the US FDA.[1112] All the evaluated items met the criteria of linearity, accuracy, sensitivity, matrix effect, carry-over, stability, and reproducibility under specified conditions. Using the UPLC and a UPLC column, only a small amount of injection volume (3 µL) was needed for quantification, and the run time for analysis was relatively short (approximately 3 minutes). The intra- and inter-batch accuracies ranged from 98.0% to 110.0% and from 100.7% to 104.7%, respectively. The intra- and inter-batch precisions ranged from 0.98% to 5.59% and from 2.74% to 7.97%, respectively. The current UPLC-MS/MS method seemed to be used for accurate quantification, even at relatively low concentrations of donepezil, with sensitivity and reproducibility.
The levels of donepezil in human plasma obtained from healthy subjects after oral dosing with 5 mg of donepezil hydrochloride (Donep®) was successfully determined using the developed and validated method. The mean ± standard deviation (SD) of the maximum concentration (Cmax) was 8.6 ± 2.0 ng/mL, and the median (range) Tmax was 2.0 (1.0~5.0 h), in eighteen healthy Korean subjects after 5 mg donepezil administration (Fig. 5).
The current method showed reliable results with a liquidliquid extraction (LLE) preparation method and a C18 column with a relatively short run time of 3 min as well as a small injection volume of 3 µL compared to previous studies (Table 5).[13141516] First, as a comparison of the sample preparation method, Shah, H. J. et al. used solid-phase extraction (SPE) for sample preparation. Although SPE in known to be highly selective compared to liquid-liquid extraction (LLE) or protein precipitation (PP), it is considered to be difficult for handling, and it has a higher associated cost.[1718] Second, for the analysis column, the method developed by Park, E. J. et al. used LLE and a hydrophilic interaction chromatography (HILIC) to separate donepezil in plasma. HILIC was considered a powerful method to analyze hydrophilic or polar compounds in reverse phase.[1319] The shorter run time of 2 min was observed in the method with a HILIC column than that of our study results. However, 3 min of run time was considered relatively comparable to 1.8~2 min of run time; moreover, the smallest injection volume was needed among the compared previous methods.
In conclusion, a sensitive and simple analysis method for donepezil in human plasma was developed with the UPLC-MS/MS. Using the developed method, plasma donepezil concentrations were appropriately determined in healthy Korean subjects. The suggested method could be applied to various donepezil pharmacokinetic studies.

Acknowledgments

This research was supported by the Bio & Medical Technology Development Program of the National Research Foundation (NRF) and was funded by the Korean government (MSIP&MOHW) (No. NRF-2015M3A9E1028327).

Notes

Reviewer: This article reviewed by peer experts who are not TCP editors.

Conflict of interest: - Authors: All the authors declare no competing financial interests.

- Reviewers: Nothing to declare

- Editors: Nothing to declare

References

1. Colović MB, Krstić DZ, Lazarević-Pašti TD, Bondžić AM, Vasić VM. Acetylcholinesterase inhibitors: pharmacology and toxicology. Curr Neuropharmacol. 2013; 11:315–335. DOI: 10.2174/1570159X11311030006. PMID: 24179466.
2. Mangialasche F, Solomon A, Winblad B, Mecocci P, Kivipelto M. Alzheimer's disease: clinical trials and drug development. Lancet Neurol. 2010; 9:702–716. DOI: 10.1016/S1474-4422(10)70119-8. PMID: 20610346.
crossref
3. Shigeta M, Homma A. Donepezil for Alzheimer's disease: pharmacodynamic, pharmacokinetic, and clinical profiles. CNS Drug Rev. 2001; 7:353–368. PMID: 11830754.
crossref
4. Yasui-Furukori N, Furuya R, Takahata T, Tateishi T. Determination of donepezil, an acetylcholinesterase inhibitor, in human plasma by high-performance liquid chromatography with ultraviolet absorbance detection. J Chromatogr B Analyt Technol Biomed Life Sci. 2002; 768:261–265. DOI: 10.1016/S1570-0232(01)00592-X.
crossref
5. Rogers SL, Friedhoff LT. Pharmacokinetic and pharmacodynamic profile of donepezil HCl following single oral doses. Br J Clin Pharmacol. 1998; 46(Suppl 1):1–6. DOI: 10.1046/j.1365-2125.1998.0460s1001.x.
crossref
6. Nakashima K, Itoh K, Kono M, Nakashima MN, Wada M. Determination of donepezil hydrochloride in human and rat plasma, blood and brain microdialysates by HPLC with a short C30 column. J Pharm Biomed Anal. 2006; 41:201–206. DOI: 10.1016/j.jpba.2005.10.017. PMID: 16321494.
crossref
7. Ponnayyan Sulochana S, Sharma K, Mullangi R, Sukumaran SK. Review of the validated HPLC and LC-MS/MS methods for determination of drugs used in clinical practice for Alzheimer's disease Bioanalytical methods for quantitation of drugs to treat Alzheimer's. Biomed Chromatogr. 2014; 28:1431–1490. DOI: 10.1002/bmc.3116. PMID: 24515838.
8. Korfmacher WA. Foundation review: Principles and applications of LC-MS in new drug discovery. Drug Discov Today. 2005; 10:1357–1367. DOI: 10.1016/S1359-6446(05)03620-2. PMID: 16253874.
crossref
9. Nováková L, Matysová L, Solich P. Advantages of application of UPLC in pharmaceutical analysis. Talanta. 2006; 68:908–918. DOI: 10.1016/j.talanta.2005.06.035. PMID: 18970409.
crossref
10. Vogeser M, Seger C. A decade of HPLC-MS/MS in the routine clinical laboratory — Goals for further developments. Clin Biochem. 2008; 41:649–662. DOI: 10.1016/j.clinbiochem.2008.02.017. PMID: 18374660.
crossref
11. FDA. Bioanalytical Method Validation, Guidance for Industry. Accessed 6 September 2017. https://www.fda.gov/downloads/drugs/guidances/ucm368107.pdf/.
12. MFDS. Guideline on Bioanalytical Method Validation. Accessed 6 September 2017. http://nifds.go.kr/_custom/nifds/_common/board/download.jsp?attach_no=18470/.
13. Park EJ, Lee HW, Ji HY, Kim HY, Lee MH, Park ES, et al. Hydrophilic interaction chromatography-tandem mass spectrometry of donepezil in human plasma: Application to a pharmacokinetic study of donepezil in volunteers. Arch Pharm Res. 2008; 31:1205–1211. DOI: 10.1007/s12272-001-1290-6. PMID: 18806965.
crossref
14. Shah HJ, Kundlik ML, Pandya A, Prajapati S, Subbaiah G, Patel CN, et al. A rapid and specific approach for direct measurement of donepezil concentration in human plasma by LC-MS/MS employing solid-phase extraction. Biomed Chromatogr. 2009; 23:141–151. DOI: 10.1002/bmc.1095. PMID: 18823072.
crossref
15. Bhateria M, Ramakrishna R, Pakala DB, Bhatta RS. Development of an LC–MS/MS method for simultaneous determination of memantine and donepezil in rat plasma and its application to pharmacokinetic study. J Chromatogr B Analyt Technol Biomed Life Sci. 2015; 1001:131–139. DOI: 10.1016/j.jchromb.2015.07.042.
crossref
16. Pilli NR, Inamadugu JK, Kondreddy N, Karra VK, Damaramadugu R, Rao JV. A rapid and sensitive LC-MS/MS method for quantification of donepezil and its active metabolite, 6-o-desmethyl donepezil in human plasma and its pharmacokinetic application. Biomed Chromatogr. 2011; 25:943–951. DOI: 10.1002/bmc.1552. PMID: 21154884.
crossref
17. Nováková L, Vlčková H. A review of current trends and advances in modern bio-analytical methods: chromatography and sample preparation. Anal Chim Acta. 2009; 656:8–35. DOI: 10.1016/j.aca.2009.10.004. PMID: 19932811.
crossref
18. Pavlovič DM, Babič S, Horvat AJM, Kaštelan-Macan M. Sample preparation in analysis of pharmaceuticals. Trac-Trends Anal Chem. 2007; 26:1062–1075. DOI: 10.1016/j.trac.2007.09.010.
19. Dejaegher B, Mangelings D, Vander Heyden Y. Method development for HILIC assays. J Sep Sci. 2008; 31:1438–1448. DOI: 10.1002/jssc.200700680. PMID: 18461571.
crossref
20. Katakam P, Kalakuntla RR, Adiki SK, Chandu BR. Development and validation of a liquid chromatography mass spectrometry method for the determination of donepezil in human plasma. J Pharm Res. 2013; 7:720–726. DOI: 10.1016/j.jopr.2013.08.021.
crossref
Figure 1

Molecular structures of (A) donepezil and (B) donepezil-D4 (IS).

tcp-26-64-g001
Figure 2

Product ion spectra of (A) donepezil and (B) donepezil-D4 (IS) by collision-induced dissociation of corresponding parent ions.

tcp-26-64-g002
Figure 3

Representative chromatograms of (A) a blank plasma, (B) human plasma spiked with 0.1 ng/mL of donepezil and internal standard (upper: donepezil, lower: IS in each chromatogram).

tcp-26-64-g003
Figure 4

Representative chromatograms of (A) signal-to-noise level in the LLOQ sample and (B) human plasma after oral administration of donepezil 5 mg (upper: donepezil, lower: IS in (B)).

tcp-26-64-g004
Figure 5

The mean plasma concentration time profiles of donepezil in healthy Korean subjects after single administered donepezil 5 mg. Bars represent standard deviations. The inset graph describes the time-concentration profiles from time 0 to 26 hours.

tcp-26-64-g005
Table 1

The parameters of calibration standard in three batches

tcp-26-64-i001
Batch name Slope Intercept Correlation coefficient (r) Determination coefficient (r2)
Set 1 0.0427436 0.00258514 0.999195 0.998392
Set 2 0.0420818 0.00214402 0.997584 0.995173
Set 3 0.0389284 0.00207884 0.999781 0.999562

*Linear regression was optimized with a weighting of 1/x2

Table 2

Accuracy and precision of the validation quality control (QC) samples

tcp-26-64-i002
Concentration (ng/mL) 0.1 0.3 3.5 35
Accuracy (%) Batch 1 98.00 100.7 103.4 104.2
Batch 2 110.0 99.33 98.00 98.77
Batch 3 106.0 102.0 103.3 105.0
Inter-batch 104.7 100.7 101.6 102.7
Precision (CV, %) Batch 1 5.59 4.91 0.98 1.83
Batch 2 6.43 2.81 1.15 2.23
Batch 3 5.17 4.38 0.99 1.32
Inter-batch 7.97 2.74 3.25 3.25
Table 3

Recovery results of the quality control (QC) samples

tcp-26-64-i003
N = 3 Recovery (%) Mean Recovery (%)
LoQC (0.3 ng/mL) 62.5 58.9
56.0
58.3
MeQC (3.5 ng/mL) 60.8 60.5
58.9
62.0
HiQC (35 ng/mL) 58.3 60.4
62.2
60.8

LoQC, low concentration of quality control (QC) sample; MeQC: medium concentration of QC sample; HiQC: high concentration of QC sample.

Table 4

Summary of the stability results

tcp-26-64-i004
Donepezil concentration (ng/mL)
LoQC (0.3 ng/mL) HiQC (35 ng/mL)
Re-injecta 4.55 −0.25
Processed sampleb 2.25 1.01
Freeze-thawc 1.10 −1.20
Short-termd 0 −0.89
Stock short-terme −4.26 −1.14
Stock long-termf 0 0.43

aafter three re-injections, bafter 24 h in the autosampler at 10℃, cafter three freeze-thaw cycles, dafter 20 h at room temperature, eafter 24 h at room temperature, fafter 24 days at 4℃; LoQC, low concentration of the quality control (QC) sample; HiQC: high concentration of the QC sample; Data are represented as the mean % change.

Table 5

Comparison of the analytical method among four references and this study

tcp-26-64-i005
Studies Column Plasma volume Internal standard Sample preparation method Run time Injection volume
Current Study Results Thermo Hypersil GOLD 200 µL Donepezil-D4 LLE (using n-hexane and ethyl acetate) 3 minutes 3 µL
Park, E. J.[13] Water atlantis HILIC Silica column 200 µL Cisapride LLE (using 0.5 M NaOH and methyl tert-butyl ether) 2 minutes 7 µL
Shah, H. J. et al.[14] Thermo betabasic-C8 500 µL Escitalopram oxalate SPE 1.8 minutes 20 µL
Bhateria, M., et al.[15] Thermo syncronis-C18 column 50 µL Amantadine hydrochloride LLE (using methanol) 3 minutes 10 µL
Pilli, N.R.[16] Chromolith® speedROD (RP C18 column) 300 µL Dipyridamole LLE (using n-hexane and ethyl acetate) 2.7 minutes 20 µL
P Katakam.[20] C18 column 300 µL Donepezil-D7 LLE (using dichloromethane and hexane) 4 minutes Not provided

LLE, liquid-liquid extraction; SPE, solid-phase extraction.

TOOLS
Similar articles