Journal List > Lab Med Online > v.14(2) > 1516086820

Performance Evaluation of a Modified Version of the Norudia Glycated Albumin Assay that Reduces Negative Interference

Abstract

Background

Glycated albumin (GA) is a biomarker of short-term glycemic status. Several cases of extremely low GA levels measured by Norudia GA assay kits (Sekisui Medical Co., Ltd., Tokyo, Japan) were inconsistent with clinical status. This study evaluated the analytical performance of the modified version of the Norudia GA that claims to have resolved the false negative issue.

Methods

Precision and linearity were evaluated following the guidance of the Clinical & Laboratory Standards Institute (CLSI) EP05-A3 and EP06-A, respectively. A comparison study was performed against the Lucica GA-L (Asahi Kasei Pharma Corporation, Tokyo, Japan) based on CLSI EP09-A3. The temperature stability of GA was also assessed. The reference interval was verified following CLSI EP28-A3C.

Results

Coefficients of variation in the precision analysis were all acceptable. Linearity assessment demonstrated a coefficient of determination (R2) of 0.998. The comparison study showed a high correlation coefficient (r) of 0.973 relative to the Lucica GA-L. Stability analysis revealed GA tended to increase with storage duration. The transference of the reference interval was verified. Negative interference was reduced in the modified version of the Norudia GA.

Conclusions

The modified version of the Norudia GA assay showed comparable performance in measuring GA as well as reduced interference in samples that had shown false negative results with the original version of the Norudia GA. The modified assay may avoid negative interference putatively caused by anti-oxidative agents in clinical laboratories using the original version of the Norudia GA assay.

INTRODUCTION

Glycated albumin (glycoalbumin, GA) is a marker of glycemic status of the previous two to four weeks, whereas HbA1c reflects that of the previous one to two months [1, 2]. Compared to HbA1c, GA reflects rapid changes in glycemic control and is not affected by hematologic diseases [2]. While GA has been widely adopted in Asian countries, its clinical significance in Caucasians was only recently validated [3]. Our institute has been measuring GA with the Norudia GA assay (Sekisui Medical Co., Ltd., Tokyo, Japan) since 2018. When using the original version of the Norudia GA, the researchers noted certain cases with extremely low GA results inconsistent with the glycemic status of the corresponding patients. This phenomenon led to the development of a modified version of the Norudia GA assay. This study evaluated the analytical performance of the modified version of the Norudia GA assay (Sekisui Medical Co., Ltd.) that aimed to resolve false negative results.

MATERIALS AND METHODS

GA was measured with the modified version of the Norudia GA assay using Cobas c702 (Roche Diagnostics International, Rotkreuz, Switzerland). The Norudia GA assay is capable of measuring GA levels from both serum and plasma, requiring approximately 10 minutes per test. Precision was evaluated following CLSI EP05-A3 [4]. Accordingly, two Norudia GA control materials (Lot 903RCT, Sekisui Medical Co., Ltd.) with estimated GA levels of 13.2% and 31.0% and one pooled serum sample with an estimated level of 20.0% were used. The samples were measured in duplicate, twice daily, for 30 days to assess repeatability and within-laboratory precision. The goal for imprecision was 2.6% or less, which is the allowed imprecision in the Westgard database [5]. Precision was considered acceptable if the coefficient of variation (CV) was less than the total allowable error (7.2%) listed in the Westgard database [5]. Linearity was evaluated following CLSI EP06-A [6]. To produce a high-concentration GA solution, pooled normal serum with a glucose level of 5 g/dL was created by adding D-(+)-glucose powder (Sigma-Aldrich, Saint Louis, MO, USA). The solution was incubated for three days at 37°C and then diluted to five concentration levels admixed with 0.85% saline. For method comparison, 539 serum samples, of which 39 were false negative samples, collected in a serum separating tube (Vacutainer SST II Tube 8.5 mL, #368972; Becton Dickinson, Sunnyvale, CA, USA) were used. The false negative samples were initially identified by their significant deviation from the corresponding patient’s previous GA result. To confirm false negativity, these samples were retested using the Lucica GA-L assay (Asahi Kasei Pharma Corporation, Tokyo, Japan). Negative interference was defined as a % difference between the GA results of the Norudia GA and Lucica GA-L assays of <14.4%Norudia GALucica GA-LLucica GA-L<0.144. The % differences were calculated relative to the results of the Lucica GA-L assay, rather than the mean values. Considering the total allowable error for GA (7.2%) suggested by Ricos et al. [7], the cutoff for negative interference was set as twice the total allowable error (14.4%), assuming that the error in the compared assays in the opposite direction from the ground truth would be the maximum allowable error. During the comparison study, three assays were utilized: 1) Lucica GA-L, 2) Original Norudia GA, and 3) Modified Norudia GA. The Lucica GA-L assay was used as the reference in the comparison study of both the original and modified versions of the Norudia GA assay since the Lucica GA-L assay showed results consistent with the glycemic status of each patient. Furthermore, the Diabetes Mellitus Indices Committee of the Japanese Society of Clinical Chemistry reported that the Lucica GA-L assay correlates well with the isotope dilution–tandem mass spectrometry reference method in both the reference material (JCCRM611) and patient sera [8]. The Passing-Bablok regression and Bland-Altman plots were used to compare the methods according to CLSI EP09-A3 [9]. To avoid a biased positive correlation, the % differences in the Bland-Altman plots were plotted against the results of the Lucica GA-L assay, not the mean values [10]. Sample stability at 4°C and -70°C was evaluated using seven specimens, including one false negative specimen. The initial values and those after 7, 14, and 28 days of storage were evaluated. To verify the current GA reference interval of our institute, samples from 20 healthy controls were used following CLSI EP28-A3C [11]. The healthy control group was selected from individuals who visited our institute for a routine medical check-up and had normal fasting glucose levels. All statistical analyses were conducted using Statistics 1.9.0, Polynomials 3.2.2, and GLM 1.8.3 on Julia 1.9. All plots were created using Gadfly 1.3.4 on Julia 1.9. The Institutional Review Board of Samsung Medical Center, Seoul, Korea, approved the study (IRB No. 2020-08-081-006) and waived the need for informed consent.

RESULTS

Precision analysis showed acceptable results. The ranges of repeatability and within-laboratory CV were 1.5–2.2% and 3.1–5.3%, respectively (Table 1). The repeatability CVs were within the desirable imprecision (2.6%), and the within-laboratory CVs were less than the total allowable error (7.2%). Linearity analysis demonstrated the best fit with first-order polynomial regression with a slope of 1.040 (95% CI, 1.004–1.075) and a coefficient of determination (R2) of 0.998 (Fig. 1). Method comparison of the modified version of the Norudia GA assay (Fig. 2A and 2B) and the original version of the Norudia GA assay (Fig. 2C and 2D) exhibited a correlation coefficient (r) of 0.973 and 0.892, respectively, with reference to the Lucica GA-L assay. The Bland-Altman plots revealed that the samples with falsely low results obtained using the original Norudia GA produced a higher GA result when measured with the modified version (Fig. 2B and 2D). Nonetheless, certain samples had falsely depressed results in the modified version of the Norudia GA assay compared with the results from the Lucica GA-L assay (Fig. 2B). According to the current definition of negative interference, 14 samples demonstrated negative interference in the modified version of the Norudia GA assay. Table 2 illustrates the false negative samples and their GA results obtained with all three assays. Sample stability revealed GA levels tended to increase with storage duration at both 4°C and -70°C (Fig. 3). The 20 healthy control samples showed a GA range of 12.5–15.7%, all within the reference interval (11.2–17.5%) [12], verifying transference of this reference interval (Fig. 4). No outliers existed among the 20 healthy control samples as determined by the Tukey method.

DISCUSSION

GA has been widely adopted in clinical practice recently to estimate short-term glycemic control. Previous publications evaluating the original version of the Norudia GA assay demonstrated that the assay is comparable with the Lucica GA-L assay in measuring GA [12, 13]. While our institute had a total of 29,813 GA orders in 2022, only 39 samples obtained from 34 different patients were false negatives as per our definition of negative interference from August 2021 to February 2023. This phenomenon is estimated to occur in approximately 0.08% of the GA orders, thus capturing this during performance evaluation is practically infeasible given the scarcity of this phenomenon.
During an investigation in 2019 when the researchers first encountered an extremely low GA result, monoclonal paraprotein precipitation was hypothesized to interfere with light absorbance. Sparse literature suggests paraprotein precipitation as the cause of interference resulting in negative results of various chemistry assays [14-17]. Lithium-heparin was the proposed cause of the precipitation [15, 17]. In the present study, two patients with monoclonal paraproteinemia showed prominent turbidity in samples stored in a lithium-heparin tube. Furthermore, M-protein and lithium- heparin-induced negative interference occurred in the original version of the Norudia GA assay. However, most cases with falsely low GA results did not have monoclonal paraproteinemia, which implies that not all interferences could be explained by this phenomenon.
Another hypothesis was that anti-oxidative agents such as ascorbic acid could interfere with the Trinder reaction [18, 19]. The Norudia GA assay measures GA based on the absorbance of purple- red pigment produced through oxidization [20], and anti-oxidative agents could interfere with the reaction. Among the thirtyfour patients who provided negative interference samples, seven were taking thioctic acid, which shows anti-oxidative activity. Notably, the number of patients taking anti-oxidative agents could be underestimated since the intake of over-the-counter medications such as vitamin supplements cannot be determined with prescription records.
The manufacturer implemented a new formula for their Norudia GA assay to overcome negative interference. The revised formula increased the amount of substance that eliminates the reduction agent, decreasing negative interference. While details regarding the formula have not been disclosed by the manufacturer, the improvement in results supports the hypothesis that anti-oxidative agents interfere with the reaction. For all negatively affected samples, the modified version of the Norudia GA assay showed results higher than the original version of the Norudia GA assay. However, some samples still demonstrated negative interference in the modified version of the Norudia GA assay, albeit to a lesser degree than the original version. Given that negative interference remains despite the revised formula and that not all patients prescribed with anti-oxidative agents displayed falsely low GA levels, the exact mechanism of the negative interference is yet to be elucidated. Further research should investigate the underlying mechanism causing falsely low results and minimize the remaining negative interference.
In summary, the modified version of the Norudia GA assay demonstrated fair analytical performance. Given that it reduces the negative interference, putatively caused by anti-oxidative agents, observed in the original version of the Norudia GA assay, the modified assay would be beneficial in clinical laboratories experiencing negative interference using the original version of the Norudia GA assay.

Acknowledgments

This study was funded by Sekisui Medical Co., Ltd. (Tokyo, Japan) based on the Consignment Agreement.

Notes

Conflicts of Interest

None declared.

REFERENCES

1. Kisugi R, Kouzuma T, Yamamoto T, Akizuki S, Miyamoto H, Someya Y, et al. 2007; Structural and glycation site changes of albumin in diabetic patient with very high glycated albumin. Clin Chim Acta. 382:59–64. DOI: 10.1016/j.cca.2007.04.001. PMID: 17482589.
crossref
2. Koga M. 2014; Glycated albumin; clinical usefulness. Clin Chim Acta. 433:96–104. DOI: 10.1016/j.cca.2014.03.001. PMID: 24631132.
crossref
3. Bellia C, Zaninotto M, Cosma C, Agnello L, Bivona G, Marinova M, et al. 2018; Clinical usefulness of glycated albumin in the diagnosis of diabetes: results from an Italian study. Clin Biochem. 54:68–72. DOI: 10.1016/j.clinbiochem.2018.02.017. PMID: 29486186.
crossref
4. Clinical, Laboratory Standards Institute. 2014. Evaluation of precision of quantitative measurement procedures; Approved guideline-Third edition. CLSI document EP05-A3. Clinical and Laboratory Standards Institute;Wayne, PA:
5. Westgard QC. Desirable biological variation database specifications. https://www.westgard.com/biodatabase1.htm. Accessed on Jan 2023.
6. Clinical, laboratory standards institute. 2003. Evaluation of the linearity of quantitative measurement procedures: A statistical approach; Approved guideline. CLSI document EP06-A. Clinical and Laboratory Standards Institute;Wayne, PA:
7. Ricós C, Alvarez V, Cava F, García-Lario JV, Hernández A, Jiménez CV, et al. 1999; Current databases on biological variation: pros, cons and progress. Scand J Clin Lab Invest. 59:491–500. DOI: 10.1080/00365519950185229. PMID: 10667686.
crossref
8. Takei I, Hoshino T, Tominaga M, Ishibashi M, Kuwa K, Umemoto M, et al. 2016; Committee on diabetes mellitus indices of the Japan society of clinical chemistry-recommended reference measurement procedure and reference materials for glycated albumin determination. Ann Clin Biochem. 53:124–32. DOI: 10.1177/0004563215599178. PMID: 26187553.
crossref
9. Clinical, Laboratory Standards Institute. 2013. Measurement procedure comparison and bias estimation using patient samples; Approved guideline-Third edition. CLSI document EP09-A3. Clinical and Laboratory Standards Institute;Wayne, PA:
10. Krouwer JS. 2008; Why Bland-Altman plots should use X, not (Y+X)/2 when X is a reference method. Stat Med. 27:778–80. DOI: 10.1002/sim.3086. PMID: 17907247.
11. Clinical, Laboratory Standard Institute. 2010. Defining, establishing, and verifying reference intervals in the clinical laboratory; Approved guideline-Third edition. CLSI document EP28-A3c . Clinical and Laboratory Standards Institute;Wayne, PA:
12. Ha C, Oh J, Park HD. 2021; Evaluation of the analytical performance of the Norudia GA Glycoalbumin Test. Lab Med Online. 11:55–9. DOI: 10.47429/lmo.2021.11.1.55.
crossref
13. Choe W, Kim S, Chang J, Park H, Kim HN, Yoo SJ. 2020; Performance of Norudia glycated albumin assay on multiple analytical platforms and comparison to Lucica assay. Clin Lab. 66:2383–7. DOI: 10.7754/Clin.Lab.2020.200107. PMID: 33180436.
crossref
14. Berth M, Delanghe J. 2004; Protein precipitation as a possible important pitfall in the clinical chemistry analysis of blood samples containing monoclonal immunoglobulins: 2 case reports and a review of the literature. Acta Clin Belg. 59:263–73. DOI: 10.1179/acb.2004.039. PMID: 15641396.
crossref
15. Dimeski G, Carter A. 2005; Rare IgM interference with Roche/Hitachi Modular glucose and gamma-glutamyltransferase methods in heparin samples. Clin Chem. 51:2202–4. DOI: 10.1373/clinchem.2005.053561. PMID: 16179422.
16. Alberti MO, Drake TA, Song L. 2015; The pH of chemistry assays plays an important role in monoclonal immunoglobulin interferences. Pract Lab Med. 3:8–16. DOI: 10.1016/j.plabm.2015.09.001. PMID: 28932804. PMCID: PMC5597720.
crossref
17. Yukimasa N, Oboshi W, Hayashi K, Kuribayashi H, Uzawa R, Fukuchi K, et al. 2016; Interference with creatinine assay by IgM-λ monoclonal protein in lithium heparin blood collection tube from a malignant lymphoma patient. Int J Med Pharm Case Reports. 8:1–7. DOI: 10.9734/IJMPCR/2016/29258.
crossref
18. Trinder P. 1969; Determination of blood glucose using an oxidase-peroxidase system with a non-carcinogenic chromogen. J Clin Pathol. 22:158–61. DOI: 10.1136/jcp.22.2.158. PMID: 5776547. PMCID: PMC474026.
crossref
19. Ko YK, Lee HS, Lee K, Song J, Park HD. 2023; The negative role of Vitamin C in total cholesterol and triglycerides tests. Lab Med Online. 13:13–21. DOI: 10.47429/lmo.2023.13.1.13.
crossref
20. Sekisui Medical Co., Ltd. Norudia GA package insert sheet. https://www.sekisuimedical.jp/english/business/diagnostics/insert/pdf/NORUDIA_GA.pdf. Accessed on Dec 2022.

Fig. 1
Linear and polynomial regressions of GA measured with the modified version of the Norudia GA assay.
lmo-14-2-100-f1.tif
Fig. 2
(A) Scattergram with Passing-Bablok regression and (B) Bland-Altman plot showing the results of comparison between Norudia GA (modified) and Lucica GA-L, and (C) Scattergram with Passing-Bablok regression and (D) Bland-Altman plot showing the results of comparison between Norudia GA (original) and Lucica GA-L. The solid navy blue line indicates the fitted Passing-Bablok regression, and the blue-colored area represents the 95% CI. In the Bland-Altman plots, the solid black line indicates the mean % difference, and the dashed lines represent the 95% CI. The red dots indicate the samples with negative interference demonstrating a % difference of <-14.4% between the original version of the Norudia GA and Lucica GA-L assays.
lmo-14-2-100-f2.tif
Fig. 3
Temperature stability of glycated albumin (GA) measured with the modified version of the Norudia GA assay after storage at either 4°C or -70°C for 7, 14, and 28 days. The long-dashed grey lines indicate the stability limit (±4.4%), whereas the short-dashed grey lines indicate the total allowable error (±7.2%) as designated by Westgard.
lmo-14-2-100-f3.tif
Fig. 4
Histogram of GA results from 20 healthy control samples used for reference interval transference. The values overlaid on the histogram represent the GA results.
lmo-14-2-100-f4.tif
Table 1
Precision of the modified version of the Norudia GA assay
Material Mean SD Repeatability Within-laboratory CV
QC 1 13.88 0.151 1.8% 5.3%
QC 2 31.81 0.199 2.2% 3.7%
Pooled serum 19.79 0.116 1.5% 3.1%

Abbreviations: SD, standard deviation; CV, coefficient of variation; QC, quality control.

Table 2
Results of specimens with negative interference in the original version of the Norudia GA assay
GA results (%) % Difference from Lucica GA-L
Lucica GA-L Norudia GA (original) Norudia GA (modified) Norudia GA (original) Norudia GA (modified)
8.5 -0.6 3.0 -107.1 -64.7
11.5 -0.7 6.5 -106.1 -43.5
9.4 -0.5 3.9 -105.3 -58.5
7.9 0.2 3.3 -97.5 -58.2
14.2 0.5 17.2 -96.5 21.1
11.8 0.5 11.0 -95.8 -6.8
25.6 2.0 11.1 -92.2 -56.6
13.2 1.1 10.4 -91.7 -21.2
14.5 3.9 12.3 -73.1 -15.2
16.0 4.5 13.7 -71.9 -14.4
16.2 5.2 16.5 -67.9 1.9
16.3 5.4 18.1 -66.9 11.0
20.1 6.9 23.4 -65.7 16.4
16.3 5.8 14.1 -64.4 -13.5
19.1 6.8 16.5 -64.4 -13.6
19.1 7.1 16.9 -62.8 -11.5
23.0 8.6 17.5 -62.6 -23.9
27.1 10.7 20.3 -60.5 -25.1
19.1 8.2 17.1 -57.1 -10.5
20.6 9.5 17.4 -53.9 -15.5
15.7 7.5 15.4 -52.2 -1.9
18.3 9.0 17.3 -50.8 -5.5
20.4 10.6 19.2 -48.0 -5.9
16.1 9.2 17.5 -42.9 8.7
17.9 10.4 17.2 -41.9 -3.9
17.3 10.3 17.8 -40.5 2.9
19.7 12.2 15.7 -38.1 -20.3
18.6 11.6 17.0 -37.6 -8.6
20.5 13.4 20.4 -34.6 -0.5
17.7 11.6 13.7 -34.5 -22.6
21.5 14.2 16.2 -34.0 -24.7
18.3 12.3 17.2 -32.8 -6.0
17.2 11.7 16.1 -32.0 -6.4
16.2 11.9 17.6 -26.5 8.6
47.7 36.8 40.5 -22.9 -15.1
15.8 12.5 16.3 -20.9 3.2
21.0 16.7 22.2 -20.5 5.7
21.7 17.8 20.8 -18.0 -4.1
17.1 14.5 17.0 -15.2 -0.6

The % differences with a negative interference according to our definition (<-14.4%) were indicated in bold.

Abbreviations: GA, glycated albumin.

TOOLS
Similar articles