This article has been
cited by other articles in ScienceCentral.
Dear Editor,
Severe muscle injuries can lead to the release of myoglobin into the bloodstream, subsequently filtering into the urine via the kidneys [
1,
2]. Elevated myoglobin levels can be detrimental to the kidneys, causing renal damage or failure. Therefore, urinary myoglobin measurement is crucial; however, existing commercially available kits are designed for serum or plasma [
3,
4]. High myoglobin concentrations in serum or plasma can induce a high-dose hook effect [
5,
6]. Using a serum or plasma myoglobin assay for urine specimens poses a risk of a hook effect owing to high myoglobin concentrations that may be present in urine from patients with severe muscle injuries [
1,
7]. Therefore, the predilution of urine specimens would address high concentrations and potential hook effect but may impact accuracy as diluted specimens might approach the assay sensitivity [
4].
Various urinary myoglobin dilutions were evaluated using the Elecsys Myoglobin assay (Cat#. 12178214160; Roche Diagnostics; Indianapolis, IN, USA) on a cobas e602 analyzer, which has an analytical measuring interval (AMI) of 1.18–168.54 nmol/L (21–3,000 ng/mL). Results were reported in ng/mL and converted to the International System of Units based on the molecular weight of myoglobin (17.8 kD) [
8]. Residual, de-identified, frozen (−20°C) urine specimens were used (IRB protocol #00007275; informed consent waived). Specimens were pH-adjusted to 8–9 at the time of collection using Na
2CO
3. Specimens were diluted onboard with Elecsys Diluent Universal (Roche Diagnostics), and dilutions of 1:400 were previously investigated in accordance with CLSI guidelines (CLSI EP34) [
9].
As a preliminary assessment, urine specimens (N=7; 56.19–1,516.85 nmol/L) were retested without dilution (neat) and at a 1:50 dilution. Of these, one produced a suspected hook effect, where the neat result was 100.00 nmol/L but was 1,496.46 nmol/L at a 1:50 dilution. This suggests that a hook effect in urine may occur at lower analyte concentrations than specified for the serum matrix by the manufacturer, as a hook effect does not occur for serum myoglobin concentrations up to 1,685.39 nmol/L [
4]. For all remaining specimens (N=6; 23.76–68.03 nmol/L), the results were elevated by approximately 2.5-fold when using a 1:50 dilution, suggesting that the specimens had concentrations low enough to impede a dilution of that magnitude, potentially affecting measurement accuracy.
To investigate whether a lower dilution factor would allow the detection of a hook effect while providing more accurate results, additional urine specimens (N=35; 56.19–33,539.33 nmol/L) were reanalyzed neat, 1:10, and 1:50 dilutions. Specimens below the AMI (<168.54 nmol/L) were categorized as not exhibiting a hook effect. Specimens that produced results above the AMI after a 1:50 dilution (>8,426.97 nmol/L) were further diluted at 1:400. A hook effect was detected in 30 of the 35 specimens at both the 1:10 and 1:50 dilutions, with several specimens requiring a 1:400 dilution to obtain a result (
Table 1).
Table 1
Detection of a high-dose hook effect using 1:10, 1:50, and 1:400 dilutions
Sample number |
Neat (nmol/L) |
1:10 dilution (nmol/L) |
Hook effect detected (1:10; >168.54 nmol/L) |
1:50 dilution (nmol/L) |
Hook effect detected (1:50; >168.54 nmol/L) |
1:400 dilution (nmol/L) |
1 |
48.89 |
125.79 |
No hook effect |
154.83 |
No hook effect |
|
2 |
41.63 |
1,049.21 |
Detected |
2,412.81 |
Detected |
|
3 |
10.15 |
274.04 |
Detected |
2,901.46 |
Detected |
|
4 |
85.34 |
74.78 |
No hook effect |
85.45 |
No hook effect |
|
5 |
55.85 |
197.98 |
Detected |
247.36 |
Detected |
|
6 |
64.27 |
>1,685.40 |
Detected |
2,429.38 |
Detected |
|
7 |
42.74 |
>1,685.40 |
Detected |
>8,426.97 |
Detected |
21,166.57 |
8 |
25.99 |
80.62 |
No hook effect |
115.79 |
No hook effect |
|
9 |
110.28 |
1,140.84 |
Detected |
1,316.69 |
Detected |
|
10 |
43.51 |
>1,685.40 |
Detected |
6,093.54 |
Detected |
|
11 |
20.08 |
732.08 |
Detected |
5,770.34 |
Detected |
|
12 |
89.61 |
111.97 |
No hook effect |
110.79 |
No hook effect |
|
13 |
20.31 |
282.58 |
Detected |
530.56 |
Detected |
|
14 |
30.52 |
>1,685.40 |
Detected |
>8,426.97 |
Detected |
31,265.56 |
15 |
34.17 |
459.78 |
Detected |
824.10 |
Detected |
|
16 |
28.51 |
>1,685.40 |
Detected |
>8,426.97 |
Detected |
12,656.46 |
17 |
34.17 |
>1,685.40 |
Detected |
>8,426.97 |
Detected |
15,379.49 |
18 |
9.56 |
901.91 |
Detected |
>8,426.97 |
Detected |
21,466.80 |
19 |
28.60 |
1,326.57 |
Detected |
6,596.07 |
Detected |
|
20 |
>168.54 |
>1,685.40 |
Detected |
3,221.52 |
Detected |
|
21 |
31.82 |
1,133.48 |
Detected |
5,771.01 |
Detected |
|
22 |
48.18 |
952.13 |
Detected |
2,553.48 |
Detected |
|
23 |
16.34 |
384.89 |
Detected |
3,048.03 |
Detected |
|
24 |
61.57 |
1,084.78 |
Detected |
1,499.72 |
Detected |
|
25 |
8.31 |
302.87 |
Detected |
2,155.17 |
Detected |
|
26 |
28.19 |
413.99 |
Detected |
613.09 |
Detected |
|
27 |
>168.54 |
>1,685.40 |
Detected |
2,070.11 |
Detected |
|
28 |
7.66 |
265.34 |
Detected |
390.22 |
Detected |
|
29 |
10.99 |
75.96 |
No hook effect |
152.87 |
No hook effect |
|
30 |
63.43 |
>1,685.40 |
Detected |
6,177.75 |
Detected |
|
31 |
96.24 |
395.45 |
Detected |
431.01 |
Detected |
|
32 |
31.87 |
552.81 |
Detected |
1,724.72 |
Detected |
|
33 |
77.53 |
1,501.52 |
Detected |
2,090.51 |
Detected |
|
34 |
108.71 |
224.44 |
Detected |
259.21 |
Detected |
|
35 |
>168.54 |
1,316.69 |
Detected |
1,617.02 |
Detected |
|

Further, we performed a recovery experiment by serially diluting a high-concentration urine specimen (48,744.33 nmol/L) with a low-concentration urine specimen (6.18 nmol/L). Expected concentrations were calculated based on the high- and low-concentration specimens used. Each serially diluted specimen was tested neat and at 1:10, 1:50, and 1:400 dilutions, and the results were compared to the expected concentrations. However, the dilutions were only performed when the anticipated concentration was within the AMI of the assay (1.18–168.54 nmol/L). The acceptable limit of recovery (80.4%–119.6%) was based on a previously reported total allowable error for myoglobin in serum [
10]. For each of the 1:10, 1:50, and 1:400 dilutions, recoveries exceeded 119.6% when the diluted specimen result (prior to correcting for the dilution factor) was 4.06–37.29 nmol/L (
Table 2). The manufacturer recommends that the serum concentration after dilution should exceed 2.81 nmol/L; nevertheless, we observed that a higher concentration may be necessary for diluted urine specimens. Diluted results were within acceptable limits of recovery when the diluted specimen produced results between 35.86–132.19 nmol/L (
Table 2). These findings demonstrate that accuracy reduces with decreasing myoglobin concentrations owing to the dilution process. Therefore, a smaller predilution factor (1:10) is recommended with additional dilutions (1:50 and 1:400) as necessary to detect a hook effect while avoiding inaccuracies for low-concentration specimens.
Table 2
Evaluation of dilutions using the Roche Myoglobin assay on the cobas e602 analyzer
Expected concentration (nmol/L) |
Result of diluted specimen* (nmol/L) |
Dilution |
Final reported result* (nmol/L) |
% Recovery |
48,744.33 |
121.86 |
1:400 |
48,744.33 |
100.0 |
24,375.28 |
68.00 |
1:400 |
27,201.07 |
111.6 |
12,190.73 |
35.86 |
1:400 |
14,342.58 |
117.7 |
6,098.48 |
132.19 |
1:50 |
6,609.66 |
108.4 |
3,052.36 |
67.31 |
1:50 |
3,365.67 |
110.3 |
1,529.27 |
37.29 |
1:50 |
1,864.61 |
121.9 |
767.75 |
84.94 |
1:10 |
849.44 |
110.6 |
386.97 |
44.03 |
1:10 |
440.34 |
113.8 |
196.57 |
23.51 |
1:10 |
235.11 |
119.6 |
101.40 |
12.90 |
1:10 |
129.04 |
127.3 |
53.82 |
6.75 |
1:10 |
67.53 |
125.5 |
30.00 |
4.06 |
1:10 |
40.56 |
135.2 |
18.09 |
Not applicable |
No dilution |
18.93 |
104.6 |
12.13 |
Not applicable |
No dilution |
12.87 |
106.1 |
9.16 |
Not applicable |
No dilution |
9.61 |
104.9 |
7.70 |
Not applicable |
No dilution |
7.87 |
102.2 |
6.97 |
Not applicable |
No dilution |
7.08 |
101.6 |

Overall, these findings support a predilution step when analyzing urine specimens using the Roche Myoglobin assay. A hook effect was observed in urine at lower concentrations than the serum hook effect threshold reported in the assay package insert. We conclude that to effectively use a serum myoglobin assay to measure myoglobin in urine, an appropriate predilution factor should be carefully validated to enable the detection of a high-dose hook effect while producing results with good accuracy and clinical acceptability.
ACKNOWLEDGEMENTS
The authors gratefully acknowledge Taylor M. Snow for specimen acquisition.
Go to :

Notes
Go to :

References
2. Chen-Levy Z, Wener MH, Toivola B, Daum P, Reyes M, Fine JS. 2005; Factors affecting urinary myoglobin stability
in vitro. Am J Clin Pathol. 123:432–8. DOI:
10.1309/9AQ62FR265ER3E2W. PMID:
15716240.

3. de Waard H, van 't Sant P. 2009; Use of serum myoglobin assays for urine myoglobin measurements can cause false-negative results. Clin Chem. 55:1240–1. DOI:
10.1373/clinchem.2008.118968. PMID:
19342504.

5. Kurt-Mangold M, Drees D, Krasowski MD. 2012; Extremely high myoglobin plasma concentrations producing hook effect in a critically ill patient. Clin Chim Acta. 414:179–81. DOI:
10.1016/j.cca.2012.08.024. PMID:
22960204.

6. Kikuchi H, Ohta A, Takahashi Y, Shibata A, Ohtake T, Watanabe K. 2003; High-dose hook effect in an immunochromatography-optical quantitative reader method for myoglobin. Clin Chem. 49:1709–10. DOI:
10.1373/49.10.1709. PMID:
14500606.

8. Lamb EJ, Jones GRD. Rifai N, Chiu RWK, editors. 2022. Kidney function tests. Tietz textbook of laboratory medicine. 7th ed. Elsevier;Saint-Louis, MO: p. 1531–91. DOI:
10.1515/9783111433998-058.
9. CLSI. 2018. Validating an extended measuring interval. Establishing and verifying an extended measuring interval through specimen dilution and spiking. 1st ed. CLSI EP34. Clinical and Laboratory Standards Institute;Wayne, PA: p. 46–55.
10. Ricós C, Alvarez V, Cava F, García-Lario JV, Hernández A, Jiménez CV, et al. 1999; Current databases on biological variation: pros, cons and progress. Scand J Clin Lab Invest. 59:491–500. DOI:
10.1080/00365519950185229. PMID:
10667686.

Go to :
