Journal List > J Dent Rehabil Appl Sci > v.33(4) > 1159995

A comparative study on the user satisfaction between two different piezoelectric engines

Abstract

Purpose:

The aim of this study is to compare the performance of two piezoelectric engine systems by surveying satisfaction from dental clinicians.

Materials and Methods:

Two piezoelectric systems were evaluated: TRAUS XUS10 (Saeshin), PIEZOSURGERY touch (Mectron). For this study, 20 dentists responded to the 11 questionnaires in which 5 point Likert-type scale was used. The two devices were operated for 10 seconds and measured 5 times to compare the maximum noise values. In heat emission test, the handpiece was operated for 3 minutes and heat was measured at three positions each.

Results:

TRAUS XUS10 had higher satisfaction level on motor noise (P < 0.05). About function key and handpiece heat generation, PIEZOSURGERY touch showed higher satisfaction (P < 0.05) than TRAUS XUS10. The maximum noise level for each of the devices was confirmed to be 56.6 dB for the TRAUS XUS10 and 56.0 dB for PIEZOSURGERY touch. The two piezoelectric engines satisfied the safety standards with an operation temperature below 41°C after having been operated for 3 minutes.

Conclusion:

Except for the function key and handpiece heat emission, TRAUS XUS10 has comparable performance with PIEZOSURGERY touch.

초록

목적:

치과의사의 사용 만족도 비교를 통해 두 가지 피에조 엔진의 성능을 비교평가 하고자 한다.

연구 재료 및 방법:

본 연구에서는 두 가지의 피에조 엔진을 평가하였다:TRAUS XUS10 (Saeshin), PIEZOSURGERY touch (Mectron). 20명의 치과의사가 11개 평가항목에 답하였고, 평가결과는 Likert의 5점 척도를 적용하여 수치화하였다. 핸드피스 소음평가는 10초 동안 5회 측정하여 최고소음을 비교하였다. 핸드피스 발열평가는 세 부위에서 3분동안의 발열량을 측정하였다.

결과:

평가항목 중 ‘주수펌프 모터의 소음’에서 TRAUS XUS10이 PIEZOSURGERY touch보다 유의하게 만족도가 높았다. 반면, ‘기능key의 만족도’와 ‘핸드피스 발열정도’는 PIEZOSURGERY touch가 더 우수했다. 소음 최대값은 TRAUS XUS10이 56.6 dB, PIEZOSURGERY touch는 56.0 dB였다. 두 기기 모두 3분 작동 후 41°C 미만의 열을 발생시킴으로써 안전기준을 만족하였다.

결론:

TRAUS XUS10는 ‘기능 key’ 및 ‘핸드피스 발열정도’에서 약간의 개선을 한다면 그 외 대부분의 평가항목에서는 PIEZOSURGERY touch과 견줄만한 성능을 보인다.

Introduction

Piezoelectric surgical devices, which started to be developed in 1998, are composed of piezoelectric handpiece, foot pedal, main unit, irrigation tube, handpiece stand and liquid solution hanger. Piezoelectric surgical devices enable more precise surgical cutting when performing external oral surgeries than drill handpieces. Piezoelectric surgical devices are also receiving attention for the ability to conduct safer surgical procedures and superior hemostasis characteristics in surgical procedure. For safe cutting through nerves, blood vessels and soft tissue, piezoelectric surgical devices operate using micro vibration at a rate of 60 - 200 mm/sec to prevent damage. The increase of cutting accuracy leads to minimize the surgical site and shorten the period of post-operation healing time by leaving a clean cross section.1-5
The application fields of piezoelectric surgery include periodontology, endodontic surgery, implantology, ontological surgery, neurosurgery, orthopedic and hand surgery.6-8
For the sinus elevation, which is considered when RBH, residual bone height, is less than 6 mm, using a rotary instrument is apt to cause sinus membrane perforations that usually lead to post-operation complications. Risk of damaging the soft tissue and blood vessel bundles is also high. The piezosurgery engine is gaining popularity as an alternate surgery technique that reduces this risks.9 It is expected that the areas where piezoelectric devices can be utilized and applied will expand with continuous growth of the implant market.10
In this research study, the user satisfaction from using two different piezosurgery engines was quantified. By comparing the user satisfaction between the globally recognized PIEZOSURGERY Touch (Mectron, Carasco, Italy) system and TRAUS XUS10 (Saeshin, Daegu, Korea), the performance of domestic piezoelectric surgery engine, TRAUS XUS10, was indirectly evaluated.

Materials and Methods

1. User’s satisfaction survey

After testing the two implant engines against 20 dentists at the Kyungpook National University School of Dentistry in Korea from June, 15th to July, 15th of 2015, they were given a self-administered survey with 11 categories of evaluation developed in this research study to evaluate their user satisfaction. 11 The performance of TRAUS XUS10 and PIEZOSURGERY touch was compared in this research study (Table 1). To exclude any influence caused by the brand and product name of the piezosurgery engine on user satisfaction during the test, blind tests were conducted without providing any information on the piezosurgery engine (Fig. 1).
Table 1
Product specifications of the two piezosurgery engines
 TRAUS XUS10  PIEZOSURGERY Touch
Operation frequency ofthe ultrasonic wave 27 ± 3 KH Automatic scan
From 24 KHz to 36 KHz
Output type of the ultrasonic wave 3 step system (15 - 59 VA) ENDO
PERIO
SPECIAL
CANCELLOUS
CORTICAL
IMPLANT
(values are unknown)
Output voltage of ultrasonic wave AC 220 - 240 V 100 - 240 Vac 50/60 Hz
Maximum pump capacity Max.90 ± 20%mL Adjustable on the touch screen:
ENDO/PERIO - 7 flow levels: from 0 to 6 (from 0 to approximately 75 mL/min)
SPECIAL/CANCELLOUS/CORTICAL/IMPLANT6 capacity levels: from 1 to 6(from 8 to approximately 75 mL/min)
Fig. 1
TRAUS XUS10, Saeshin (Left), PIEZOSURGERY Touch (Right)
JDRAS_33_269_fig_1.tif
11 categories of evaluation were presented in the survey. To evaluate question No.4 of the survey, cutting ability of hard tissue and adequate bone cutting properties, all clinicians were given two bone cutting tip specialized to each of the companies, and answered the question after conducting osteotomy on a pig rib specimen.12-15 The TRAUS XUS10 used Sohn’s Saw (Saeshin) and the PIEZOSURGERY touch used the OT7 (Mectron) for cutting (Fig. 2).
Fig. 2
Cutting the pig bone to evaluate the ‘cutting ability of hard tissue, adequate bone cutting properties’. (A) OT7 (PIEZOSURGERY Touch, Mectron), (B) Sohn’s Saw (TRAUS XUS10, Saeshin), (C) Pig rib sawing operation
JDRAS_33_269_fig_2.tif
By using the Likert 5-point scale to evaluate and convert each question on the scale of ‘very dissatisfied’ at 1 point, ‘dissatisfied’ at 2 points, ‘average’ at 3 points, ‘satisfied’ at 4 points, and ‘very satisfied’ at 5 points, the levels of satisfaction of the clinicians were quantified.

2. Evaluation of the noise from the handpiece

By applying the configuration for each of the devices that resulted in the most similar output values, the two devices were operated for 10 seconds without irrigation (Table 2). The level of noise was measured 5 times for each device using sound & vibration solutions TSVA Ver 3.1 (Signallink, Suwon, Korea) to compare the maximum noise values.
Table 2
Configuration values for each of the two piezosurgery engines
TRAUS XUS10 (Saeshin) PIEZOSURGERY Touch (Mectron)
Power avg. 43.5 W avg. 41 W
(36 - 51 W) (36 - 46 W)
(configuration: Boost2, P-Power3) (configuration: Cortical)
Irrigation rate 65 mL/min (3 step) 62 mL/min (6 step)

3. Evaluation of the heat emissions of the handpiece

After fixing the handpiece to the Jig, Thermocoupling was attached to 3 separate positions: Position1 (P1): the tip attachment part, Position2 (P2): the part that comes into contact with the thug and index fingers when holding the handpiece, and Position3 (P3): the rest part when holding the handpiece. The handpiece was operated for 3 minutes and the heat generated from each of the parts was measured using Graphtec GL240 Midi Data Logger (GRAPHTEC, Yokohama, Japan).

4. Statistical analysis

A score from ‘very dissatisfied’ at 1 point, ‘dissatisfied’ at 2 points, ‘average’ at 3 points, ‘satisfied’ at 4 points, and ‘very satisfied’ at 5 points was assigned to each of the 11 categories designed for the survey in this research study. By applying a score distribution according to the Likert point scale, the average and standard deviation values were recorded (Table 3, 4). The results of executing Kolmogorov-Smirnov and Shapiro-Wilk to verify the normality of the data confirmed that both sets of data did not satisfy the conditions for normality (P < 0.05). To verify the statistical significance of the two products, Mann-Whitney U test, which is a nonparametric test, was conducted using IBM SPSS Version 20 for Windows (SPSS Inc., Chicago, USA). The results confirmed that the statistical significance was 0.05.
Table 3
Likert scale point conversion for questions on clinical test satisfaction (TRAUS XUS10)
Question n (%) Mean (SD)
Very dissatisfied  Dissatisfied   Average   Satisfied  Very Satisfied
1* 0 (0) 1 (5) 12 (60) 7 (35) 0 (0) 3.3 (0.6)
2* 0 (0) 0 (0) 15 (75) 5 (25) 0 (0) 3.3 (0.4)
3* 0 (0) 0 (0) 15 (75) 4 (20) 1 (5) 3.3 (0.6)
4* 0 (0) 1 (5) 9 (45) 9 (45) 1 (5) 3.5 (0.7)
5* 0 (0) 0 (0) 10 (50) 10 (50) 0 (0) 3.5 (0.5)
6* 0 (0) 1 (5) 6 (30) 13 (65) 0 (0) 3.6 (0.6)
7* 0 (0) 3 (15) 12 (60) 5 (25) 0 (0) 3.1 (0.6)
8* 0 (0) 2 (10) 14 (70) 4 (20) 0 (0) 3.1 (0.6)
9* 0 (0) 0 (0) 17 (85) 3 (15) 0 (0) 3.2 (0.4)
10* 1 (5) 1 (5) 15 (75) 2 (10) 1 (5) 3.1 (0.8)
11* 0 (0) 6 (30) 13 (65) 1 (5) 0 (0) 2.8 (0.6)

* 1, ease of attachment and detachment of Tip; 2, weight and balance of the ultrasonic handpiece; 3, appropriateness of the form and shape (design) of the handpiece; 4, cutting ability of hard tissue and adequate bone cutting properties; 5, wide strength band; 6, wide irrigation area; 7, noise level of irrigation pump motor; 8, satisfaction with regards to function keys; 9, level of noise; 10, amount of vibration; 11, heat emission from handpiece

Table 4
Likert scale point conversion for questions on clinical test satisfaction (PIEZOSURGERY Touch)
Question n (%) Mean (SD)
Very dissatisfied  Dissatisfied   Average   Satisfied  Very Satisfied
1* 3 (15) 6 (30) 11 (55) 3.4 (0.8)
2* 1 (5) 10 (50) 9 (45) 3.4 (0.6)
3* 2 (10) 11 (55) 7 (35) 3.3 (0.6)
4* 5 (25) 12 (60) 3 (15) 3.9 (0.6)
5* 1 (5) 2 (10) 10 (50) 5 (25) 2 (10) 3.3 (1.0)
6* 1 (5) 7 (35) 12 (60) 3.6 (0.6)
7* 1 (5) 9 (45) 7 (35) 3 (15) 2.6 (0.8)
8* 1 (5) 5 (25) 14 (70) 3.7 (0.6)
9* 6 (30) 11 (55) 3 (15) 2.9 (0.7)
10* 1 (5) 12 (60) 6 (30) 1 (5) 3.4 (0.7)
11* 1 (5) 13 (65) 5 (25) 1 (5) 3.3 (0.7)

* 1, ease of attachment and detachment of Tip; 2, weight and balance of the ultrasonic handpiece; 3, appropriateness of the form and shape (design) of the handpiece; 4, cutting ability of hard tissue and adequate bone cutting properties; 5, wide strength band; 6, wide irrigation area; 7, noise level of irrigation pump motor; 8, satisfaction with regards to function keys; 9, level of noise; 10, amount of vibration; 11, heat emission from handpiece

Results

1. User’s satisfaction survey

For Q7 which surveyed the noise from the irrigation pump motor, TRAUS XUS10 device was confirmed to have a statistically higher level of satisfaction when compared to PIEZOSURGERY touch. On the other hand, for Q8 which surveyed the function keys and satisfaction and for Q11 which surveyed the heat emission levels of the handpiece, the results confirmed that PIEZOSURGERY touch was superior (Table 3, 4).16 There was no statistically significant difference confirmed between the remaining category questions (Fig. 3).
Fig. 3
Comparison for each survey question. * Statistically significant with P < 0.05. Q1: ease of attachment and detachment of Tip, Q2: weight and balance of the ultrasonic handpiece, Q3: appropriateness of the form and shape (design) of the handpiece, Q4: cutting ability of hard tissue and adequate bone cutting properties, Q5: wide strength band, Q6: wide irrigation area, Q7: noise level of irrigation pump motor, Q8: satisfaction with regards to function keys, Q9: level of noise, Q10: amount of vibration, Q11: heat emission from handpiece.
JDRAS_33_269_fig_3.tif

2. Noise level evaluation for two handpieces

The maximum noise level for each of the devices was confirmed to be 56.6 dB for the TRAUS XUS10 and 56.0 dB for PIEZOSURGERY touch.

3. Heat emission level evaluation for two handpieces

For the three measured positions, the heat emission level of the TRAUS XUS10 showed a higher level of heat during operation when compared to that of PIEZOSURGERY touch device (Fig. 4).
Fig. 4
The Changes in temperature according to the operation duration of the handpieces. (A) Position1: the tip attachment part, (B) Position2: the part that comes into contact with the thug and index fingers when holding the handpiece, (C) Position3: the rest part when holding the handpiece
JDRAS_33_269_fig_4.tif
According to international standards from IEC (International Electrotechnical Commission),17 the maximum permitted temperature when a device is operation is defined as 41°C. When conducting the test for just 3 minutes according to the companies recommend with a rest of 10 minutes for the TRAUS XUS10, both TRAUS XUS10 and PIEZOSURGERY touch satisfied the safety standards with an operation temperature below 41°C after having been operated for 3 minutes.

Discussion

The objective of this research study was to measure the user satisfaction between two different piezoelectric engines and conduct a comparative evaluation of performance between the product developed in Korea and a piezoelectric engine that has been certified overseas. The clinicians evaluated to the piezoelectric engines that had been setup in the same room. For Q4, which asked about the cutting ability of hard tissue and adequate bone cutting properties, while PIEZOSURGERY touch showed a high level of user satisfaction, the results also were confirmed that there was no statistical significance. When conducting cutting operations in this research, each company’s specialized tip for each engine was used. It has been known that the overall resonant frequencies change depending on the shape and size of the surgical tip, and the center frequency changes depending on the differences in the pressure applied during a surgical procedure. Therefore, it is not possible to totally eliminate the influences from the differences in the tip designs of the two piezoelectric engines and the pressure applied when operating the two piezoelectric engines, which is a factor dependent upon the clinician. On that point of view, we used the company’s recommended tip to minimize the predictable errors. For Q7, which asked about the noise level of irrigation pump motor, TRAUS XUS10 was confirmed to have a statistically significantly high level of user satisfaction. In general ultrasonic bone procedure, headpieces used in dental practices must operate at low frequency ranges in order to perform at a high output level. However, once the frequency drops as low as 20 kHz, which is in the audible frequency range and this causes difficulties in using the device due to noise. That is why these devices use 30 - 40 kHz frequency range.18 The frequency range used for the ultrasonic function of TRAUS XUS10 handpiece was 27 ± 3 kHz, which falls in an appropriate frequency range to be used in clinical studies. On the other hand, Piezosurgery has 24 kHz - 36 kHz frequency range. Kramer et al.19 claim it shows a similar level of sound pressure level (SPL), compared to when conducting conventional osteotomies. Even though TRAUS XUS10 showed a comparatively higher handpiece noise level, when operating this device together with irrigation pumps, TRAUS XUS10 gained more satisfaction from the users. That is considered to be because the irrigation pump noise of the PIEZOSURGERY touch was higher than that of the TRAUS XUS10. If mechanical technique of the each piezoelectric engine is informed, further research on the cause and result of the data will be explained better, however, it is not allowed now. Letho etc.20 reported that people who are relatively susceptible to hearing damage have the risk to incur gradual hearing damage even from dental drills. The sound radiated during dental procedures has high level of noise above 70 dB (A) and high frequency of 4 kHz. That is why there is a clinical demand for technology to reduce this risk of hearing damage. Technology that can reduce noise will be recognized as a one of the important factors on the level of user satisfaction when using a device.21,22
The range of the level of sound generated by dental devices when treating patients such as scaling, tooth preparation etc. and doing non-medical examination like idling is between 67.7 - 78.3 dB (A). For patients who show a shocked or uncomfortable response when exposed to sharp noise during treatment, the influence of the noise radiated from the devices cannot be ignored in terms of patient cooperation. The ISO, International Organization for Standardization, regulates noise level standards for a room and suggests working place to keep noise level at NRN (Noise Reduction Number) 60 - 70 dB (A). Dental devices have a necessity to be developed to meet the requirement.21
For Q8, function keys and satisfaction, and for Q11, heat emission levels of the handpiece, PIEZOSURGERY touch had higher level of user satisfaction. On the basis of the fact that both products provide several function keys that enable various output values and irrigation rate ranges to be configured, there could be another factor to make the survey result divergent. One possible factor is device design. PIEZOSURGERY touch is equipped with an LED touch screen unlike the button style of the TRAUS XUS10. Considering a design only, the two devices show a significant difference. These elements are expected to effect on the satisfaction level of the devices that cater to the market trends where consumers prefer touch screens in various devices. Add to this, PIEZOSURGERY touch has more subdivided function keys which allow the users to setup the mode they need more sophisticated. While TRAUS XUS10 has just several keys with roughly categorized, it does not mean that it is not acceptable for clinical usage. Well-subdivided keys allow more customized set-up, though, excessive division can result in time consuming for changing set-up during operation. Roughly dived function keys have strong point on that point of view; however, they have a risk to give inadequate operation condition as well. Considering these weak and strong points, the users favor will also affect on the satisfaction finally. For Q11, the level of satisfaction for TRAUS XUS10 was low and this can be caused by the differences in the structure of the handpiece that cannot insulate heat as well as the handpiece of PIEZOSURGERY touch. The two devices were confirmed to have an especially higher increase in temperature near the position P2 where piezoelectric elements are placed. TRAUS XUS10 showed a temperature increase in the positions P1 and P3, which can be assumed to be caused by the design of the handpiece of TRAUS XUS10, which is not as good at insulating heat as much as the handpiece of PIEZOSURGERY touch does. Because the heat generated by the piezoelectric elements’ vibration is transferred to the whole body of the handpiece, the technological know-how to insulate this heat can cause the mentioned perceived differences in heat emission. In this study, the measurement of emitted heat was limited to the heat transferred to the clinician. If future research can include examination on the increase in temperature of the tissue that contacts with the end of the surgical tip during surgery, it will be possible to investigate the overall level of user satisfaction for the devices based on rapid post-surgery recovery and a decrease in medical complication from the surgical procedure.23
Even though TRAUS XUS10 needs to be improved slightly in terms of function key and the amount of heat emitted through the handpiece, it appears that this device has a comparable level of satisfaction in most of the other evaluation categories when compared to PIEZOSURGERY touch. As the first device in Korea that combines an implant motor and an ultrasonic device, TRAUS XUS10 is expected to be competitive with PIEZOSURGERY touch with accompanied by improvement on performance and sensitive reaction to users’ opinions. In addition, when an implant surgery requires both an ultrasonic treatment device and implant drill device, all in one system of TRAUS XUS10 decreases the movement range of the operator by installing only one device. Than can decrease clinician’s fatigue and enhances the efficiency of surgery.

Conclusions

TRAUS XUS10 showed higher user satisfaction on the noise level of irrigation pump motor than PIEZOSURGERY touch. While, PIEZOSURGERY touch had higher satisfaction with regards to function key and heat emission from handpiece compared to TRAUS XUS10. TRAUS XUS10 has comparable satisfaction with that of PIEZOSURGERY touch in other survey lists. When further improvement based on the clinicians’ evaluation and its lower market price meets together, TRAUS XUS10 is considered to have enough potential competitiveness in peizosurgery market.

Acknowledgements

This work was supported by the Ministry of Trade, Industry and Energy of the Republic of Korea and Daegu Institute for Regional Program Evaluation and Daegu Technopark (Regional Key Industries Promotion Project).

References

1. Stübinger S, Kuttenberger J, Filippi A, Sader R, Zeilhofer HF. Intraoral piezosurgery: preliminary results of a new technique. J Oral Maxillofac Surg. 2005; 63:1283–7. DOI: 10.1016/j.joms.2005.05.304. PMID: 16122591.
2. Blus C, Szmukler-Moncler S, Salama M, Salama H, Garber D. Sinus bone grafting procedures using ultrasonic bone surgery:5-year experience. Int J Periodontics Restorative Dent. 2008; 28:221–9. PMID: 18605597.
3. Robiony M, Polini F, Costa F, Vercellotti T, Politi M. Piezoelectric bone cutting in multipiece maxillary osteotomies. J Oral Maxillofac Surg. 2004; 62:75961. DOI: 10.1016/j.joms.2004.01.010.
4. Ma L, Stübinger S, Liu XL, Schneider UA, Lang NP. Healing of osteotomy sites applying either piezosurgery or two conventional saw blades: a pilot study in rabbits. Int Orthop. 2013; 37:1597–603. DOI: 10.1007/s00264-013-1908-3. PMID: 23793460. PMCID: PMC3728383.
5. Eggers G, Klein J, Blank J, Hassfeld S. Piezosurgery: an ultrasound device for cutting bone and its use and limitations in maxillofacial surgery. Br J Oral Maxillofac Surg. 2004; 42:451–3. DOI: 10.1016/j.bjoms.2004.04.006. PMID: 15336773.
6. Labanca M, Azzola F, Vinci R, Rodella LF. Piezoelectric surgery: twenty years of use. Br J Oral Maxillofac Surg. 2008; 46:265–9. DOI: 10.1016/j.bjoms.2007.12.007. PMID: 18342999.
7. Seshan H, Konuganti K, Zope S. Piezosurgery in periodontology and oral implantology. J Indian Soc Periodontol. 2009; 13:155–6. DOI: 10.4103/0972-124X.60229. PMID: 20379414. PMCID: PMC2848787.
8. Vercellotti T, Dellepiane M, Mora R, Salami A. Piezoelectric bone surgery in otosclerosis. Acta Otolaryngol. 2007; 127:932–7. DOI: 10.1080/00016480601110154. PMID: 17712671.
9. Jensen OT, Shulman LB, Block MS, Iacono VJ. Report of the Sinus Consensus Conference of 1996. Int J Oral Maxillofac Implants. 1998; 13(Suppl1):1–45. PMID: 9715571.
10. Seshan H, Konuganti K, Zope S. Piezosurgery in periodontology and oral implantology. J Indian Soc Periodontol. 2009; 13:155–6. DOI: 10.4103/0972-124X.60229. PMID: 20379414. PMCID: PMC2848787.
11. Lee DH, Lee KB. The comparative study of user satisfaction on various implant engine system. J Dent Rehabil Appl Sci. 2014; 30:9–15. DOI: 10.14368/jdras.2014.30.1.9.
12. Pearce AI, Richards RG, Milz S, Schneider E, Pearce SG. Animal models for implant biomaterial research in bone: a review. Eur Cell Mater. 2007; 13:1–10. DOI: 10.22203/eCM.v013a01. PMID: 17334975.
13. An YH, Freidman RJ. Animal models in orthopaedic research. Boca Raton;CRC Press;. 1998; 3957.
14. Buser D, Schenk RK, Steinemann S, Fiorellini JP, Fox CH, Stich H. Influence of surface characteristics on bone integration of titanium implants. A histomorphometric study in miniature pigs. J Biomed Mater Res. 1991; 25:889–902. DOI: 10.1002/jbm.820250708. PMID: 1918105.
15. Sun C, Huang G, Christensen FB, Dalstra M, Overgaard S, Bünger C. Mechanical and histological analysis of bone-pedicle screw interface in vivo: titanium versus stainless steel. Chin Med J (Engl). 1999; 112:456–60.
16. Davies AR, Ware JE Jr. Measuring patient satisfaction with dental care. Soc Sci Med A. 1981; 15:75160. DOI: 10.1016/0271-7123(81)90019-5.
17. IECEE. International Electrotechnical Commission. Available from: http://www.iec.ch/search/?q=Applied%20parts%20not%20intended%20to%20supply%20heat%20to%20a%20patient'. (updated 2017 Aug 14).
18. Lee JH, Oh JM, Hong Y, Kim S, Paik J, Lee YJ, Lee JB, Lee SD. Design and evaluation of ultrasonic bone surgical instruments for dental application. J Korean Inst Electr Electron Mater Eng. 2012; 25:990–5. DOI: 10.4313/JKEM.2012.25.12.990.
19. Kramer FJ, Bornitz M, Zahnert T, Schliephake H. Can piezoelectric ultrasound osteotomies result in serious noise trauma? Int J Oral Maxillofac Surg. 2015; 44:1355–61. DOI: 10.1016/j.ijom.2015.03.004. PMID: 26227860.
20. Lehto TU, Laurikainen ET, Aitasalo KJ, Pietilä TJ, Helenius HY, Johansson R. Hearing of dentists in the long run: a 15-year follow-up study. Community Dent Oral Epidemiol. 1989; 17:207–11. DOI: 10.1111/j.1600-0528.1989.tb00613.x. PMID: 2758794.
21. Ji DH, Choi MS. Characteristics of noise radiated at dental clinic. J Korean Soc Environ Eng. 2009; 31:1123–8.
22. Lee JS, Han YS, Cho YS. The noise level assessment of dental equipment. J Dent Hyg Sci. 2015; 15:603–11. DOI: 10.17135/jdhs.2015.15.5.603.
23. Walmsley AD. Potential hazards of the dental ultrasonic descaler. Ultrasound Med Biol. 1988; 14:15–20. DOI: 10.1016/0301-5629(88)90159-7.
TOOLS
Similar articles