Journal List > Anesth Pain Med > v.14(1) > 1158425

Chung, Lee, Koo, Cho, Lee, Park, and Kim: Sustained erroneous near-infrared cerebral oxygen saturation in alert icteric patient with vanishing bile duct syndrome during and after liver transplantation - A case report -

Abstract

Monitoring cerebral oxygenation using a near infrared spectroscopy (NIRS) device is useful for estimating cerebral hypoperfusion and is available during liver transplantation (LT). However, high serum bilirubin concentration can interfere with NIRS because bilirubin absorbs near infrared light. We report a patient who underwent LT with a diagnosis of vanishing bile duct syndrome, whose regional cerebral oxygen saturation (rSO2) remained below 15% even with alert mental status and SpO2 value of 99%. The rSO2 values were almost fixed at the lowest measurable level throughout the intra- and postoperative period. We report a case of erroneously low rSO2 values during the perioperative period in a liver transplant recipient which might be attributable to skin pigmentation rather than higher serum bilirubin concentration.

In healthy people, cerebral blood flow (CBF) is maintained steadily as long as the mean arterial pressure (MAP) and partial arterial CO2 pressure (PaCO2) are within a certain range. The autoregulation of CBF is often impaired in patients with acute or chronic liver failure. In these patients, a change in MAP or PaCO2 will elicit either seriously increased or reduced CBF [1]. Factors likely to affect CBF or cerebral oxygenation include hemodynamic fluctuation, bleeding, and reperfusion syndrome during LT. However, monitoring methods for evaluating cerebral function such as cerebral perfusion pressure, jugular venous oximetry, and intracranial pressure are somewhat invasive and can lead to severe complications in patients with impaired hemostasis [2,3].
Near-infrared spectroscopy (NIRS) allows non-invasive and real time continuous monitoring of regional cerebral oxygen saturation (rSO2) [4]. Monitoring cerebral oxygenation using a NIRS device is useful for estimating cerebral hypoperfusion during LT [2,5]. Unfortunately, high serum bilirubin concentration can interfere with NIRS because the bilirubin absorbs near infrared light. Therefore readings may be incorrect in patients with high serum bilirubin. Despite this limitation, it is used clinically as it reflects important trends in most patients [2,6]. We report a patient who underwent LT with a diagnosis of vanishing bile duct syndrome whose rSO2 remained below 15% even with alert mental status. We could not observe even the trend of rSO2 values throughout the LT. This state remained even after bilirubin decreased.

CASE REPORT

A 79-year-old, 163 cm, 56 kg, and dark skin-colored male was admitted for deceased donor LT due to vanishing bile duct syndrome. His past medical history consisted of hypertension diagnosed 10 years previously. Six months earlier, he suffered from nausea and right upper quadrant abdominal pain, and was diagnosed with acute cholecystitis. At that time, his total bilirubin value was 13.28 mg/dl and it remained in the range 10–20 mg/dl in spite of several interventions including endoscopic retrograde bile drainage. His final diagnosis was vanishing bile duct syndrome and LT was recommended.
Preoperative laboratory values were as follows: total bilirubin 20.7 mg/dl, prothrombin time (international normalized ratio) 1.9, and hemoglobin 8.9 g/dl. When he entered the operating room, he was alert and was capable of ordinary conversation with the medical staff.
Anesthesia was induced with intravenous propofol 70 mg and rocuronium 50 mg. Maintenance of anesthesia was achieved with desflurane, 50% oxygen/air mixture (total flow 3 L/min), continuous remifentanil (target controlled infusion dose 0–3 ng/ml), and continuous rocuronium infusion (5 µg/kg/min). Right radial and femoral artery were cannulated. An advanced venous access catheter (AVA, Edwards Lifesciences, USA) and 7 Fr triple lumen catheter (Arrow guard blue, Arrow international, USA) were inserted into the right internal jugular vein. A pulmonary artery catheter (Swan-Ganz CCOmbo V, Edwards Lifesciences, USA) was inserted via AVA catheter for cardiac output monitoring. Cerebral oximeter (INVOS 5100c, Medtronic, USA) and BIS (Covidien, USA) were also monitored. The rSO2 values before induction of anesthesia were 15% on both sides when the SpO2 value was 99% (Fig. 1).
Fig. 1
Cerebral oximeter values before induction of anesthesia with alert mental status.
APM_14_063_fig_1.tif
Total anesthesia time was 340 minutes and all patient monitoring data was recorded in vital recorder program [7]. No remarkable events occurred. MAP was well maintained above 70 mmHg with continuous norepinephrine infusion (0.1–0.3 µg/kg/min). Hemoglobin value was above 7.8 g/dl throughout the operation. Total crystalloid 2,750 ml, colloid (albumin) 750 ml, and salvaged blood by cell saver machine (AUTOLOG®, Medtronic, USA) 89 ml were infused. No blood products were administered during the operation. The rSO2 values did not change on the left side, but a slight elevation from 15 to 30% was observed on the right side before portal vein reperfusion (Fig. 2).
Fig. 2
A changing figure of regional cerebral oxygen saturation values during operation (recorded by vital recorder program).
APM_14_063_fig_2.tif
The patient was transferred to surgical intensive care unit for postoperative recovery. The rSO2 was not monitored when the patient was in the intensive care unit. For further evaluation of the possible cause of abnormally low rSO2 values during surgery, we visited the patient three days after operation when his total bilirubin level was decreased to 9.82 mg/dl. He was alert, responded correctly to our questions about his body condition, and his vital signs were stable; blood pressure 141/63 (mean, 91) mmHg, and heart rate 53 beats/min. Hemoglobin level was 7.5 g/dl. When we monitored NIRS again, the rSO2 values were still recorded as 15% in both sides. He was discharged without any complication on postoperative day 21.

DISCUSSION

Vanishing bile duct syndrome is a group of disorders which result in progressive destruction and ‘vanishing’ of intrahepatic bile ducts (ductopenia). Multiple etiologies including autoimmune, toxins, genetic abnormalities, infections, and neoplasms can lead to this disorder. Ductopenia is a pathologic term which refers to a loss of intrahepatic bile ducts, and eventually leads to cholestasis [8]. The specific cause of the disease in this current patient was not clear. His total bilirubin level was persistently high (10–20 ml/dl) about 6 months before LT. This long lasting hyperbilirubinemia seems to have caused significant skin pigmentation that led to the measurement error in NIRS and did not immediately disappear even after bilirubin concentration decreased. Under these circumstances, NIRS provides unreliable values in patients with vanishing bile duct syndrome during LT and this problem persists in spite of lowered bilirubin concentration after surgery. It indicates a critical limitation of NIRS.
According to the manufacturer’s manual, the measurable range of rSO2 reading was 15–95% (updated every 5–6 seconds) in the INVOS 5100c. Therefore, if rSO2 values were < 15%, the machine displayed a result of 15%. The rSO2 values of our patient were 15% on both sides before anesthesia induction. It was hard to believe that these rSO2 values reflected the real cerebral oxygen status, because he was alert and was capable of conducting ordinary conversation at that time.
The INVOS system uses dual wavelengths of near infrared light that are absorbed by oxy/deoxy hemoglobin (730 and 810 nm). It is reported that the absorptivity of near infrared light by oxy/deoxy hemoglobin is well separated at 730 nm wavelength, but is similar at 810 nm. Unfortunately, conjugated bilirubin has a peak absorptivity at 730 nm. Bilirubin can be deposited in all tissue layers which can affect the capacity for assessing cerebral oxygenation by NIRS in patients with icterus [6,9].
Song et al. [2] reported that high total bilirubin and low hemoglobin are independent associated factors for low rSO2 values in patients with liver disease. They suggested that low rSO2 values in patients with icterus may be an artifact rather than true ischemia [2]. Madsen et al. [6] investigated 48 patients who underwent LT and reported that rSO2 was lowered but the internal jugular O2 saturation was not affected. They concluded that despite underestimating rSO2, the directional changes in rSO2 seemed to be observable even in patients with high bilirubin levels [6].
In the current case, we assume that rSO2 values were lower than the device was capable of measuring as described in the manufacturer’s manual. Therefore, we could not detect even the directional changes in rSO2. His rSO2 remained unmeasurable until 3 days after the operation with alert mental status and lowered bilirubin level. One study revealed that changes in serum bilirubin concentration do not influence the rSO2 values during LT [10]. It seems likely that tissue pigmentation deposits may not disappear immediately in some patients [6]. Bilirubin can competitively absorb near infrared light and therefore play a role as confounder for rSO2 measuring. Bilirubin itself cannot affect either cerebral blood flow or cerebral oxygen uptake [6,9]. Hemoglobin levels were maintained above 8 g/dl, with MAP above 70 mmHg. Clinical causes other than an underestimation of rSO2 related to high bilirubinemia are hard to imagine.
The relationship between high serum bilirubin and falsely lower rSO2 has been reported. However, a case in which rSO2 was below the lower limit throughout LT and was maintained even after bilirubin diminished seem to be rare.
The limitation of this case report is that we did not use other methods for estimation of CBF such as transcranial Doppler sonography. We took it for granted the low rSO2 was an artifact because the patient was alert. If we had used the transcranial Doppler, we could have definitively ruled out the possibility of cerebral hypoperfusion. Moreover, it is unclear how long it might take to normalize the rSO2 after bilirubin decreased or what level of bilirubin is adequate for reliable rSO2 in icterus patients. There is a need for further research.
In conclusion, severe hyperbilirubinemia may make it impossible to estimate cerebral oxygenation using NIRS devices and furthermore, this limitation would persist even after the bilirubin decreased. Careful attention should be paid to the interpretation of NIRS values in patients with severe hyperbilirubinemia during LT.

CONFLICTS OF INTEREST

No potential conflict of interest relevant to this article was reported.

REFERENCES

1. Pere P, Höckerstedt K, Isoniemi H, Lindgren L. Cerebral blood flow and oxygenation in liver transplantation for acute or chronic hepatic disease without venovenous bypass. Liver Transpl. 2000; 6:471–9. DOI: 10.1053/jlts.2000.8186. PMID: 10915171.
2. Song JG, Jeong SM, Shin WJ, Jun IG, Shin K, Huh IY, et al. Laboratory variables associated with low near-infrared cerebral oxygen saturation in icteric patients before liver transplantation surgery. Anesth Analg. 2011; 112:1347–52. DOI: 10.1213/ANE.0b013e318214b2b0. PMID: 21543778.
3. Blei AT, Olafsson S, Webster S, Levy R. Complications of intracranial pressure monitoring in fulminant hepatic failure. Lancet. 1993; 341:157–8. DOI: 10.1016/0140-6736(93)90016-A.
4. Casati A, Spreafico E, Putzu M, Fanelli G. [New technology for noninvasive brain monitoring: continuous cerebral oximetry]. Minerva Anestesiol. 2006; 72:605–25. Italian. PMID: 16865080.
5. Plachky J, Hofer S, Volkmann M, Martin E, Bardenheuer HJ, Weigand MA. Regional cerebral oxygen saturation is a sensitive marker of cerebral hypoperfusion during orthotopic liver transplantation. Anesth Analg. 2004; 99:344–9. DOI: 10.1213/01.ANE.0000124032.31843.61. PMID: 15271702.
6. Madsen PL, Skak C, Rasmussen A, Secher NH. Interference of cerebral near-infrared oximetry in patients with icterus. Anesth Analg. 2000; 90:489–93. DOI: 10.1213/00000539-200002000-00046. PMID: 10648345.
7. Lee HC, Jung CW. Vital Recorder-a free research tool for automatic recording of high-resolution time-synchronised physiological data from multiple anaesthesia devices. Sci Rep. 2018; 8:1527. DOI: 10.1038/s41598-018-20062-4. PMID: 29367620. PMCID: PMC5784161.
8. Reau NS, Jensen DM. Vanishing bile duct syndrome. Clin Liver Dis. 2008; 12:203–17. DOI: 10.1016/j.cld.2007.11.007. PMID: 18242505.
9. Murkin JM, Arango M. Near-infrared spectroscopy as an index of brain and tissue oxygenation. Br J Anaesth. 2009; 103(Suppl 1):i3–13. DOI: 10.1093/bja/aep299. PMID: 20007987.
10. Jun IG, Shin WJ, Park YS, Song JG, Kim YK, Hwang GS. Factors affecting intraoperative changes in regional cerebral oxygen saturation in patients undergoing liver transplantation. Transplant Proc. 2013; 45:245–50. DOI: 10.1016/j.transproceed.2012.10.025. PMID: 23375309.
TOOLS
Similar articles