Shin and Lee: A New Era of Intraoperative Neuromonitoring: Beyond the Electromyography Endotracheal Tube During Thyroid Surgery
It is important to preserve the recurrent laryngeal nerve (RLN) during thyroidectomy. Although the RLN can be preserved by visual identification, intraoperative neuromonitoring (IONM) systems using an electromyography (EMG) endotracheal tube with a surface electrode have been developed to identify the functional integrity of the RLN during surgery. These systems help to check the status of the RLN by measuring the EMG signal of the vocalis muscle after weak electrical stimulation is applied to the RLN via a nerve stimulator. Recent studies have revealed that IONM reduces permanent and temporary RLN injuries during thyroidectomy [1,2].
An IONM system consists of a recording side and a stimulating side. The recording side detects muscle movements using EMG. The stimulating side provides weak electrical stimulation to the nerve via a nerve stimulator. Even if IONM is used during thyroidectomy, stimulating the nerve is necessary to confirm whether a nerve injury has occurred. However, repeated stimulation of the nerve during surgery is cumbersome and inefficient. Therefore, methods using a stimulating dissecting instrument or an attachable nerve stimulator capable of stimulating the RLN simultaneously with surgical dissection during open and remote-access thyroidectomy are being developed [3-5].
The most important problem with IONM using an EMG endotracheal tube as the recording side is that its positive predictive value is relatively low, ranging widely from about 10% to 90%, compared to its high negative predictive value, which ranges from 92% to 100% [6]. This is because the contact between the electrodes of the EMG endotracheal tube and the vocal folds is affected by the insertion direction or displacement of the EMG endotracheal tube and the patient’s postural changes during surgery. Another problem is that IONM cannot be used during thyroid surgery unless anesthesia is induced by intubating the patient with an EMG endotracheal tube before surgery. In addition, IONM is difficult to use in patients with a tracheostomy tube inserted or in patients with a narrow trachea because of a huge goiter or tumor invasion. Furthermore, EMG endotracheal tubes are relatively expensive; hence, there may be limitations in their coverage by medical insurance.
New types of EMG endotracheal tubes with surface electrodes have been developed to improve contact between the vocal folds and surface electrodes. Further, an alternative IONM system is being developed using adhesive skin electrodes and transcutaneous or transcartilage needle electrodes, which do not require an EMG endotracheal tube [7-10]. Adhesive skin electrodes have the disadvantage of a low amplitude, but they can stably measure the EMG signal of the laryngeal muscle. Transcutaneous or transcartilage needle electrodes can acquire a high amplitude with low false-positive results, but they may be exposed to the surgical field and interfere with the surgical procedure around the upper pole of the thyroid gland. They also have the disadvantage of being a relatively invasive method that can cause bleeding or hematoma in the vocal fold because of needle insertion. However, an IONM system using adhesive skin electrodes and transcutaneous or transcartilage needle electrodes has the advantage of inexpensively measuring the EMG signal of the laryngeal muscle without the EMG endotracheal tube. Furthermore, this technique is advantageous for intraoperative applications, even in patients who have not been administered anesthesia and intubated using the EMG endotracheal tube.
To confirm the state of the RLN by evaluating the movement of the laryngeal muscles, two pairs of electrode must contact the vocal folds. If either electrode has poor contact, accurate EMG measurements cannot be made. Therefore, instead of evaluating muscle movements using EMG, novel IONM systems that measure muscle movements with a new single sensor, such as a surface pressure sensor or an accelerometer based on a piezo-electric sensor, have been introduced [11,12]. In the future, these sensors can be further developed into skin adhesive sensors or endotracheal tube-attached sensors.
It is important to preserve the RLN during thyroid surgery. Several methods and technologies are being developed to overcome the disadvantages of the current IONM system using the EMG endotracheal tube. In the future, we believe that these developments will help surgeons preserve the RLN during thyroid surgery.

Notes

No potential conflict of interest relevant to this article was reported.

AUTHOR CONTRIBUTIONS

Conceptualization: BJL. Data curation: SCS. Formal analysis, Methodology, & Project administration: BJL. Visualization: SCS. Writing–original draft: all authors. Writing–review & editing: BJL.

REFERENCES

1. Bai B, Chen W. Protective Effects of intraoperative nerve monitoring (IONM) for recurrent laryngeal nerve injury in thyroidectomy: meta-analysis. Sci Rep. 2018; May. 8(1):7761.
crossref
2. Shin SC, Lee BJ. Utility of intraoperative neuromonitoring in thyroid surgery. J Clin Otolaryngol. 2018; 29(2):157–61.
crossref
3. Chiang FY, Lu IC, Chang PY, Sun H, Wang P, Lu XB, et al. Stimulating dissecting instruments during neuromonitoring of RLN in thyroid surgery. Laryngoscope. 2015; Dec. 125(12):2832–7.
crossref
4. Sung ES, Lee JC, Shin SC, Choi SW, Jung DW, Lee BJ. Development of a novel detachable magnetic nerve stimulator for intraoperative neuromonitoring. World J Surg. 2018; Jan. 42(1):137–42.
crossref
5. Sung ES, Lee JC, Kim SH, Shin SC, Jung DW, Lee BJ. Development of an attachable endoscopic nerve stimulator for intraoperative neuromonitoring during endoscopic or robotic thyroidectomy. Otolaryngol Head Neck Surg. 2018; Mar. 158(3):465–8.
crossref
6. Dralle H, Sekulla C, Lorenz K, Brauckhoff M, Machens A; German IONM Study Group. Intraoperative monitoring of the recurrent laryngeal nerve in thyroid surgery. World J Surg. 2008; Jul. 32(7):1358–66.
crossref
7. Lee HS, Oh J, Kim SW, Jeong YW, Wu CW, Chiang FY, et al. Intraoperative neuromonitoring of recurrent laryngeal nerve during thyroidectomy with adhesive skin electrodes. World J Surg. 2020; Jan. 44(1):148–54.
crossref
8. Li P, Liang QZ, Wang DL, Han B, Yi X, Wei W, et al. The transcutaneous electromyography recording method for intraoperative neuromonitoring of recurrent laryngeal nerve during minimally invasive parathyroidectomy. Sci Rep. 2020; May. 10(1):7609.
crossref
9. Wu CW, Chiang FY, Randolph GW, Dionigi G, Kim HY, Lin YC, et al. Feasibility of intraoperative neuromonitoring during thyroid surgery using transcartilage surface recording electrodes. Thyroid. 2018; Nov. 28(11):1508–16.
crossref
10. Jung SM, Tae K, Song CM, Lee SH, Jeong JH, Ji YB. Efficacy of transcartilaginous electrodes for intraoperative neural monitoring during thyroid surgery. Clin Exp Otorhinolaryngol. 2020; Nov. 13(4):422–8.
crossref
11. Sung ES, Lee JC, Shin SC, Kwon HK, Na HS, Park DH, et al. Development of a novel intraoperative neuromonitoring system using an accelerometer sensor in thyroid surgery: a porcine model study. Clin Exp Otorhinolaryngol. 2019; Nov. 12(4):420–6.
crossref
12. Sung ES, Lee JC, Shin SC, Kwon HG, Kim MS, Kim DJ, et al. Development of a novel intraoperative neuromonitoring system using a surface pressure sensor to detect muscle movement: a rabbit model study. Clinical Exp Otorhinolaryngol. 2019; May. 12(2):217–23.
crossref
TOOLS
Similar articles