Choi, Lee, Oh, and Kong: Navigation-Assisted Balloon Eustachian Tuboplasty for Eustachian Tube Dilatory Dysfunction

Abstract

Objectives.

Balloon Eustachian tuboplasty (BET) is a novel treatment method for Eustachian tube dilatory dysfunction (ETD). However, surgeons cannot identify the insertion depth of the catheter during BET, resulting in potential risks such as internal carotid artery (ICA) injury. Therefore, we developed an image-guided navigation balloon catheter to identify the insertion depth of the catheter and to establish awareness of the proximity of the ICA. This study aimed to evaluate the technical feasibility of this image-guided navigation balloon catheter system in patients with ETD.

Methods.

Twenty-nine patients (38 ears; nine bilateral; 21 right ears, and 17 left ears) diagnosed with ETD were assessed. All patients who showed no improvement despite medical therapy with topical steroids, anti-reflux medication, and the Valsalva maneuver for a minimum of 6 weeks received image-guided navigation-assisted BET. The 7-item Eustachian Tube Dysfunction Questionnaire (ETDQ-7) score and Valsalva maneuver were used to evaluate patients’ symptoms preoperatively and at the postoperative follow-up.

Results.

Image-guided navigation-assisted BET was safely performed in all patients. The mean total ETDQ-7 score was 25.4±7.1 preoperatively, 17.5±6.2 at 1 month, and 15.2±7.0 at 6 months (P<0.001). In total, a Valsalva maneuver was possible for 28 of 38 ears (73.7%) at the time of the patient’s final visit at 6 months post-procedure.

Conclusion.

Image-guided navigation balloon catheters are a potentially valuable tool in patients with ETD. Their use is also technically feasible and safe when performing BET to treat ETD.

INTRODUCTION

Balloon dilatation of the Eustachian tube (balloon Eustachian tuboplasty [BET]) is a novel treatment method for Eustachian tube dilatory dysfunction (ETD) [1-3]. Since Ockermann et al. [4] introduced endoscopic BET in 2010, evidence of its shortterm and long-term benefits has accumulated [4-7]. However, the potential risks of BET include development of a circumferential stricture, fracture of the medial cartilaginous lamina, hemorrhage into the pterygoid space, and excessive widening causing a patulous Eustachian tube (ET) [8]. To date, it has not been possible for surgeons to identify the insertion depth of the catheter during BET. Excessively deep insertion of the guide wire could cause the wire to pass into the middle ear, injuring the ossicles or tympanic membrane [8]. The catheter may also become contorted in the ET during insertion and fail to dilate to the correct position. In addition, there is a risk that the catheter could injure the internal carotid artery (ICA) because of the close anatomical relationship between the bony ET and the ICA [9]. Carotid injury has been reported during BET procedures in a cadaveric study [10]. Additionally, previous studies have anecdotally reported catastrophic injuries and deaths during patulous ET injection procedures [11]. To avoid causing a devastating injury, it is clinically important to recognize the proximity of the ICA. Thus, we developed an image-guided navigation balloon catheter that could be used during BET for ETD to identify the insertion depth of the catheter and to establish awareness of the proximity of the ICA. In this study, we evaluated the technical feasibility of this image-guided navigation balloon catheter system in patients with ETD.

MATERIALS AND METHODS

This study was approved by the Institutional Review Board of the Pusan National University Hospital (IRB No. H-1810-008-071). All patients provided written informed consent before the study.

Patients

This study was performed at Pusan National University Hospital between January 2017 and August 2018. The inclusion criteria for BET were patients age ≥18 years who could not do a Valsalva maneuver for more than 6 months and who suffered from persistent symptoms of: (1) unilateral or bilateral otitis media with effusion, (2) tympanic membrane retraction with type B or C tympanogram, or (3) were diagnosed with ETD. ETD was diagnosed when patients presented with 3 or more ET dysfunction symptoms (ear pain, ear pressure, tinnitus, cracking or popping in ears, muffled hearing, feeling that ears are clogged). All patients failed to improve despite medical therapy with topical steroids, anti-reflux medication and auto-inflation devices as minimum of 6 weeks. The exclusion criteria were: (1) history of prior ET surgery, (2) ear tubes in place or an unhealed perforation, (3) follow-up duration less than 6 months. Clinical examinations were performed 1 week, 1 month, 3 months, and 6 months after the procedure. To describe symptoms, we used the 7-item Eustachian Tube Dysfunction Questionnaire (ETDQ-7) score [12] in Table 1 and subjective reporting of the ability to perform the Valsalva maneuver (yes or no).

Preoperative computed tomography scan protocol

Preoperative computed tomography (CT) was used for calibration and visualization of the cartilaginous portion of ET [13]. High-resolution paranasal CT scans were conducted using a GMS Revolution CT scanner (GE Healthcare, Milwaukee, WI, USA) with the patients in supine position, with and without Valsalva maneuver. The authors explained the Valsalva maneuver and procedure to all patients prior to their CT scan. The technician was instructed to ask the patients to exhale forcefully against a closed nose immediately before scanning. The axial plane imaging acquisition was performed at 1-mm thickness, 215 mAs, and 120 kVp. The CT scan was performed only once each with and without Valsalva maneuver due to radiation exposure. The mean total amount of radiation exposure (dose-length product [DLP] and the volume CT dose index [CTDIVOL]) were 583.6±15.2 mGy·cm and 42.30±8.37 mGy.

Balloon dilatation using the image-guided navigation balloon catheter

Each patient underwent BET by using both the Naviloon (balloon and guide catheter) and the NET-Navigation (image guided navigation System; Medicore, Seongnam, Korea). The balloon catheter is designed to be inserted and inflated in the cartilaginous portion of the ET for treatment of ETD. The balloon catheter consists of a flexible proximal shaft, an actuator for advancing and retracting the balloon catheter, a rigid shaft that passes through the guide catheter, a balloon (length 16 mm, diameter 6 mm at 12 atm) at the distal end of the shaft for ET dilation, and a distal atraumatic tip that contains an electromagnetic navigation sensor (Fig. 1). The NET-Navigation target localization error for surface registration is less than 2 mm. Although the balloon catheter was equipped with a navigation system, the catheter also has a black proximal endoscopic marker to aid in positioning under direct endoscopic visualization (Fig. 2). The guide catheter is designed to provide a means to access the ET. Unlike other guide catheters, this device consists of a flexible shaft with a central lumen instead of a rigid shaft (Fig. 1). If anatomical variations such as deviated nasal septum, inferior turbinate hypertrophy, or narrow nasal cavity were present in the patient, surgeons can easily bend the flexible shaft according to the anatomical variation during BET. The catheter tip was used as a registration probe and contacted the face during the skin tracing process. All procedures were performed with endoscopic assistance (30°-view angle). A balloon was slowly passed through the ET lumen until it reached the isthmus with the cartilaginous portion of the ET, which was confirmed simultaneously by the navigation system. The balloon was inflated once with sterile water to a target pressure of up to 12 atm for 2 minutes, then deflated for 1 minute, and then reinflated for 1 more minute [14]. The details of a representative case are highlighted in Supplementary Video 1.

RESULTS

Patients’ demographic characteristics are presented in Table 2. A retrospective review found that image-guided navigation-assisted BET had been performed in 38 ears (29 patients, nine bilateral, 21 right ears and 17 left ears). Twenty of the patients were men and 9 were women, with an age range of 18 to 62 years (mean, 38.6±13.1 years). Sixteen ears had otitis media with effusion with type B tympanograms, and 22 ears had retracted tympanic membranes with type B or C tympanograms. To evaluate the effectiveness of the surgical procedure, we analyzed patients’ clinical records from the final visit at 6 months post-procedure. The mean total ETDQ-7 score was 25.4±7.1 preoperatively, 17.5±6.2 at 1 month, and 15.2±7.0 at 6 months. Statistically significant differences were found between the preoperative ETDQ-7 score and both the 1-month and 6-month follow-up scores (P<0.001, paired t-test) (Fig. 3). It was possible to perform a successful Valsalva maneuver in 28 of the 38 ears (73.7%) at the final visit. In a comparison of preoperative and postoperative CT scans taken during the Valsalva maneuver, ET patency was visually confirmed at 6 months after the operation (Fig. 4). Otomicroscopy demonstrated that 10 of the 16 ears (62.5%) with type B tympanograms appeared normal, with type A tympanograms at the time of the patients’ final visit. Additionally, 15 of the 22 (68.2%) ears with retracted tympanic membrane appeared normal; although the remaining 7 (31.2%) of these cases still had retraction, they showed a significant improvement of the total ETDQ-7 score compared with the preoperative score. There were no major complications related to balloon dilation. Two patients (two ears) had blood-tinged nasal discharge and two ears (one patient) had otalgia, which disappeared spontaneously within a day. BET was safely performed in all patients and was evaluated as technically successful. In all patients, we were able to confirm the insertion depth of the catheter and to establish awareness of the proximity of the ICA during the BET procedure (Fig. 2).

DISCUSSION

BET has emerged as a new minimally invasive intervention used for the treatment of ETD [15,16]. In this study, we developed an image-guided navigation balloon catheter to be used during BET to treat ETD. We hypothesized that image-guided navigation balloon catheter insertion may be a complementary method to the standard technique, because the balloon catheter tip can be identified during the procedure. It is critical to confirm the position of the catheter tip during BET, as a potential risk of the BET procedure is that a false passage may be created by the balloon catheter, especially when introducing the catheter tip [8,17-19]. Although cadaveric studies have documented intentional dilation of the bone on postprocedural CT scans, this does not rule out the possibility of a micro-fracture along a bony canal [4,8]. To avoid catastrophic injury, it is clinically important to recognize the proximity of the ICA.
Previous researchers have reported a variety of clinical success rates for BET. According to the recent study of Poe et al. [20], the Valsalva maneuver showed improvement at all follow-up times, but there was a decline from 100% positive Valsalva maneuvers after balloon dilation to 63% after a period of 7 to 14 months. Schroder et al. [21] reported a 1-year improvement rate of 73% in 1,076 ETs in 622 patients; 82% of the patients with initial improvement maintained their improvement for 5 years. In a study of 42 ETs, Silvola et al. [22] reported an 80% improvement rate that remained stable for 2 years. This study showed similar results, with an improvement rate of 73.7%. Moreover, all patients had improved clinical symptoms (ETDQ-7 scores) at the 6-month follow-up.
In this study, there were no procedure-related complications. BET was safely performed in all patients and considered technically successful. During all of the BET procedures, we were able to confirm the insertion depth of the catheter and establish awareness of the proximity of the ICA. There is no risk of injury during insertion posed by the surrounding surgical structures because the catheter has a distal atraumatic tip that contains an electromagnetic navigation sensor. This device was deemed technically feasible. The use of an image-guided navigation balloon catheter during BET is simple and convenient, and no additional training was required for otology surgeons who had experience with conventional BET. Additionally, except for a calibration time of roughly 10 minutes, no additional operation time is required.
Previous studies have reported that continuous visualization of the balloon catheter may be possible using a dual endoscopic transtympanic approach [13,23,24]. However, this technique has not been widely adopted because of its complexity, and it also does not provide full visualization of the balloon catheter and ET. Our results suggest that continuous visualization of the balloon catheter using the image-guided navigation balloon catheter decreases the risk of contortion, allows the surgeon to confirm the correct position of the catheter tip, and reduces the chance of insertion into a false passage, such as the carotid canal.
Compared to conventional BET without navigation, radiation exposure may be a concern because of the requirement for a preoperative CT scan; however, as measurements of the mean total amount of radiation exposure, the DLP and the CTDIVOL were 583.6±15.2 mGy·cm and 42.30±8.37 mGy, respectively, which convert to an effective dose of less than 2 mSv [25,26]. The effective dose from the CT scan is broadly comparable to that of natural background radiation (cosmic, terrestrial, and internal), which accounts for 1 mSv per year, and the average radon exposure in the United States, which is 2 mSv per year [25].
Although our preliminary study demonstrated several advantages of using image-guided navigation during BET, it has some limitations. First, the number of cases identified was small, since ETD is an uncommon disease. Second, for similar reasons, this study had no external control group and was a single-center study. Third, there is still no consensus on post-procedure evaluation methods. However, the ETDQ-7 assesses symptoms of ET dysfunction and is the only patient-reported outcomes tool to have undergone an initial validation study [27]. For these reasons, further multi-center studies with larger sample sizes are required to validate our results.
Our preliminary findings indicate that image-guided navigation balloon catheter insertion allowed us to clearly monitor the location of the catheter tip during BET. This provides the surgeon with real-time feedback on the position of the catheter tip during BET and prevents the creation of a false passage using the balloon catheter. It is a potentially valuable tool in patients with ETD, and it is technically feasible. The application of an image-guided navigation balloon catheter during BET is simple, convenient, and effective. In conclusion, although further studies are needed, the results of image-guided navigation balloon catheter insertion were encouraging.

HIGHLIGHTS

▪ An image-guided navigation balloon catheter provides surgeons with real-time feedback on the position of the catheter tip during balloon Eustachian tuboplasty.
▪ Clinical improvement—defined as the ability to perform a successful Valsalva maneuver—was found in 73.7% of ears at 6 months post-procedure.
▪ All patients had improved clinical symptoms (7-item Eustachian Tube Dysfunction Questionnaire) within 6 months, and showed no complications.
▪ Image-guided navigation balloon catheters are a potentially valuable tool, and are technically feasible in patients with Eustachian tube dilatory dysfunction.

Notes

No potential conflict of interest relevant to this article was reported.

AUTHOR CONTRIBUTIONS

Conceptualization: SKK. Data curation: SWC. Formal analysis: SHL, Methodology: SJO: Visualization: SWC. Writing–original draft: SWC. Writing–review & editing: SKK.

SUPPLEMENTARY MATERIALS

Supplementary materials can be available at https://doi.org/10.21053/ceo.2019.01305.
Supplementary Video 1.
Surgical video of navigation-assisted balloon Eustachian tuboplasty.

REFERENCES

1. Martino E, Di Thaden R, Krombach GA, Westhofen M. Function tests for the Eustachian tube: current knowledge. HNO. 2004; Nov. 52(11):1029–39.
2. Browning GG, Gatehouse S. The prevalence of middle ear disease in the adult British population. Clin Otolaryngol Allied Sci. 1992; Aug. 17(4):317–21.
crossref
3. Luukkainen V, Kivekas I, Hammaren-Malmi S, Rautiainen M, Poyhonen L, Aarnisalo AA, et al. Balloon Eustachian tuboplasty under local anesthesia: is it feasible? Laryngoscope. 2017; May. 127(5):1021–5.
crossref
4. Ockermann T, Reineke U, Upile T, Ebmeyer J, Sudhoff HH. Balloon dilation eustachian tuboplasty: a feasibility study. Otol Neurotol. 2010; Sep. 31(7):1100–3.
5. Catalano PJ, Jonnalagadda S, Yu VM. Balloon catheter dilatation of Eustachian tube: a preliminary study. Otol Neurotol. 2012; Dec. 33(9):1549–52.
6. Miller BJ, Elhassan HA. Balloon dilatation of the Eustachian tube: an evidence-based review of case series for those considering its use. Clin Otolaryngol. 2013; Dec. 38(6):525–32.
crossref
7. Randrup TS, Ovesen T. Balloon eustachian tuboplasty: a systematic review. Otolaryngol Head Neck Surg. 2015; Mar. 152(3):383–92.
8. Poe DS, Hanna BM. Balloon dilation of the cartilaginous portion of the eustachian tube: initial safety and feasibility analysis in a cadaver model. Am J Otolaryngol. 2011; Mar-Apr. 32(2):115–23.
crossref
9. Ashry Y, Kawai K, Poe D. Utility of adjunctive procedures with balloon dilation of the Eustachian tube. Laryngoscope Investig Otolaryngol. 2017; Nov. 2(6):337–43.
crossref
10. Kepchar J, Acevedo J, Schroeder J, Littlefield P. Transtympanic balloon dilatation of eustachian tube: a human cadaver pilot study. J Laryngol Otol. 2012; Nov. 126(11):1102–7.
crossref
11. Busby DR, Slemmons DH, Miller TF Jr. Fatal epistaxis via carotid aneurysm and eustachian tube. Arch Otolaryngol. 1968; Mar. 87(3):295–8.
crossref
12. McCoul ED, Anand VK, Christos PJ. Validating the clinical assessment of eustachian tube dysfunction: The Eustachian Tube Dysfunction Questionnaire (ETDQ-7). Laryngoscope. 2012; May. 122(5):1137–41.
crossref
13. Tarabichi M, Najmi M. Visualization of the eustachian tube lumen with Valsalva computed tomography. Laryngoscope. 2015; Mar. 125(3):724–9.
crossref
14. Poe DS, Silvola J, Pyykko I. Balloon dilation of the cartilaginous eustachian tube. Otolaryngol Head Neck Surg. 2011; Apr. 144(4):563–9.
crossref
15. Hwang SY, Kok S, Walton J. Balloon dilation for eustachian tube dysfunction: systematic review. J Laryngol Otol. 2016; Jul. 130 Suppl 4:S2–6.
crossref
16. Williams B, Taylor BA, Clifton N, Bance M. Balloon dilation of the Eustachian tube: a tympanometric outcomes analysis. J Otolaryngol Head Neck Surg. 2016; Feb. 45:13.
crossref
17. Ockermann T, Reineke U, Upile T, Ebmeyer J, Sudhoff HH. Balloon dilatation eustachian tuboplasty: a clinical study. Laryngoscope. 2010; Jul. 120(7):1411–6.
crossref
18. Bowles PF, Agrawal S, Salam MA. Balloon tuboplasty in patients with Eustachian tube dysfunction: a prospective study in 39 patients (55 ears). Clin Otolaryngol. 2017; Oct. 42(5):1057–60.
crossref
19. Jufas N, Treble A, Newey A, Patel N. Endoscopically guided transtympanic balloon catheter dilatation of the eustachian tube: a cadaveric pilot study. Otol Neurotol. 2016; Apr. 37(4):350–5.
20. Poe D, Anand V, Dean M, Roberts WH, Stolovitzky JP, Hoffmann K, et al. Balloon dilation of the eustachian tube for dilatory dysfunction: a randomized controlled trial. Laryngoscope. 2018; May. 128(5):1200–6.
crossref
21. Schroder S, Lehmann M, Ebmeyer J, Upile T, Sudhoff H. Balloon Eustachian tuboplasty: a retrospective cohort study. Clin Otolaryngol. 2015; Dec. 40(6):629–38.
crossref
22. Silvola J, Kivekas I, Poe DS. Balloon dilation of the cartilaginous portion of the Eustachian tube. Otolaryngol Head Neck Surg. 2014; Jul. 151(1):125–30.
crossref
23. Jufas N, Patel N. Transtympanic balloon dilatation of the eustachian tube: systematic review. J Laryngol Otol. 2016; May. 130(5):425–30.
crossref
24. McCoul ED, Anand VK. Eustachian tube balloon dilation surgery. Int Forum Allergy Rhinol. 2012; May-Jun. 2(3):191–8.
crossref
25. Huda W, Ogden KM, Khorasani MR. Converting dose-length product to effective dose at CT. Radiology. 2008; Sep. 248(3):995–1003.
crossref
26. Christner JA, Kofler JM, McCollough CH. Estimating effective dose for CT using dose-length product compared with using organ doses: consequences of adopting International Commission on Radiological Protection publication 103 or dual-energy scanning. AJR Am J Roentgenol. 2010; Apr. 194(4):881–9.
crossref
27. Schilder AG, Bhutta MF, Butler CC, Holy C, Levine LH, Kvaerner KJ, et al. Eustachian tube dysfunction: consensus statement on definition, types, clinical presentation and diagnosis. Clin Otolaryngol. 2015; Oct. 40(5):407–11.
crossref

Fig. 1.
(A) Photographs of the balloon catheter (arrow), guide catheter (arrowhead), and inflation device (asterisk). (B) Distal end of the balloon catheter, comprising an electromagnetic navigation sensor (arrow), a balloon (16 mm in length and 6 mm in diameter at 12 atm; arrowhead), and a black proximal balloon marker (asterisk). (C) Cross-sectional view of the balloon catheter consisting of an electromagnetic sensor part (0.9 mm in diameter; asterisk), inner tube (1.2 mm in diameter; arrow) and outer tube (1.6 mm in diameter; arrowhead). (D) Flexibility of the guide catheter. Neutral position (asterisks), right- and left-sided bending (arrows), and upward and downward bending (arrowheads).
ceo-2019-01305f1.tif
Fig. 2.
(A) Intraoperative view showing the insertion depth of the catheter tip as confirmed during a balloon Eustachian tuboplasty procedure. The catheter tip was advanced until its tip reached the isthmus. The on-screen crosshair provided real-time localization in three dimensions. Intraoperative view showing the internal carotid artery (asterisks). (B) Endoscopic photograph of catheter tip insertion over a black proximal endoscopic marker (denoting canalization of the cartilaginous segment of the Eustachian tube).
ceo-2019-01305f2.tif
Fig. 3.
Total 7-item Eustachian Tube Dysfunction Questionnaire (ETDQ-7) scores preoperatively, and at 1 month and 6 months post-procedure.
ceo-2019-01305f3.tif
Fig. 4.
(A) Preoperative computed tomography (CT) scans during the Valsalva maneuver, showing the nasopharyngeal end of the Eustachian tube (arrowheads). (B) Six-month postoperative CT scans during the Valsalva maneuver, showing patency of the cartilaginous Eustachian tube (arrows). (C) Six-month postoperative CT scans without the Valsalva maneuver. No micro-fracture was found along the bony canal of the internal carotid artery (asterisks).
ceo-2019-01305f4.tif
Table 1.
The 7-item Eustachian Tube Dysfunction Questionnaire [12]
Over the past 1 month, how much has each of the following been a problem for you? No problem Moderate problem Severe problem
1. Pressure in the ears? 1 2 3 4 5 6 7
2. Pain in the ears? 1 2 3 4 5 6 7
3. A feeling that your ears are clogged or “under water”? 1 2 3 4 5 6 7
4. Ear symptoms when you have a cold or sinusitis? 1 2 3 4 5 6 7
5. Crackling or popping sounds in the ears? 1 2 3 4 5 6 7
6. Ringing in the ears? 1 2 3 4 5 6 7
7. A feeling that your hearing is muffled? 1 2 3 4 5 6 7
Table 2.
Demographic and clinical characteristics of the treated patients
Patient Sex/age (yr) Side Symptom duration (yr) Previous tympanostomy tube history Post-procedural complication Total ETDQ-7 score
Preoperative Post-procedure
1 mo 6 mo
1 M/36 Right 12 Yes No 30 15 18
2 M/39 Right 14 Yes No 32 31 15
3 M/32 Left 9 Yes No 27 22 26
4 M/34 Right 7 Yes No 38 26 24
Left 7 Yes No 42 26 23
5 M/38 Right 8 Yes No 19 14 9
Left 8 Yes No 35 23 10
6 F/23 Right 1 No No 17 17 10
Left 1 No No 17 16 14
7 M/24 Right 2 Yes No 18 15 10
Left 2 Yes No 23 19 10
8 M/18 Left 3 Yes No 36 15 32
9 M/23 Left 1 No No 20 22 10
10 M/19 Right 2 No Pain (otalgia) 28 19 10
Left 2 No Pain (otalgia) 28 9 10
11 M/49 Right 10 Yes No 32 9 7
Left 10 Yes No 20 12 7
12 F/43 Right 20 Yes No 17 9 7
Left 20 Yes No 30 21 19
13 F/31 Right 16 Yes Blood-tinged nasal discharge 26 17 11
14 F/55 Right 17 Yes No 18 16 7
15 F/53 Right 12 Yes No 29 15 17
16 M/45 Left 14 Yes No 26 22 14
17 F/53 Right 5 Yes No 17 14 18
18 M/62 Left 30 Yes No 38 31 29
19 M/57 Right 18 Yes No 21 10 8
Left 18 Yes No 28 10 11
20 M/42 Right 3 No No 17 13 13
Left 3 No No 15 12 18
21 F/21 Left 7 Yes Blood-tinged nasal discharge 17 19 7
22 F/58 Right 10 Yes No 19 9 7
23 M/47 Right 10 Yes No 27 19 17
24 F/29 Right 5 No No 24 23 26
25 M/39 Right 10 Yes No 30 21 17
26 M/54 Right 15 Yes No 26 15 15
27 M/44 Right 7 No No 28 26 22
28 M/45 Left 13 Yes No 19 9 10
29 M/54 Left 4 No No 34 26 13

ETDQ-7, 7-item Eustachian Tube Dysfunction Questionnaire.