Journal List > Lab Med Online > v.10(3) > 1145155

Performance Evaluation of a Point of Care SelexOnTM B-Type Natriuretic Peptide Immunoassay

Abstract

Background

This study was conducted to evaluate the analytical performance of the SelexOnTM B-type natriuretic peptide (BNP) assay (Osang Healthcare Inc., Korea), a new rapid lateral flow immunoassay for point of care (POC) testing using whole blood.

Methods

The imprecision, linearity, and method comparison of SelexOnTM BNP assay were evaluated. Two commercial BNP assays, the ADVIA Centaur® BNP (Siemens Health Care diagnostics Inc., USA) and the Triage® BNP assays (Alere, USA), were included for method comparison using 100 whole blood samples from patients. The reference interval was verified using 120 residual samples from health examination participants.

Results

The SelexOn BNP had total CVs of 20.3%, 13.3%, and 10.3% in BNP concentrations of 89.44 pg/mL, 480.71 pg/mL, and 1,201.84 pg/mL of control materials, respectively. Linearity was observed from 56 pg/mL to 1544 pg/mL. The SelexOn BNP (y) regression equation was y=0.9706x-21.68 with Centaur BNP (x) (r=0.930) and y=0.7600x+0.0506 with Triage BNP (x) (r=0.845), respectively. The predicted mean difference (%) of the SelexOn BNP at the clinical decision levels (100 pg/mL) was up to 25% lower than the two comparative methods. The SelexOn BNP levels were below 50 pg/mL in 114 (95%) of the 120 samples.

Conclusions

The SelexOn BNP using EDTA was developed as a POC test for differential diagnosis or treatment monitoring for acute heart failure. However, clinical decision values must be improved to be compatible with other BNP methods.

초록

배경

이 연구는 전혈을 이용한 현장검사용 새로운 신속 측방유동면역분석법인 SelexOnTM B형 나트륨이뇨펩티드(B-type natriuretic peptide, BNP) 검사(Osang Healthcare, Korea)의 분석적 성능을 평가하기 위해 수행되었다.

방법

SelexOnTM BNP 검사의 비정밀도, 직선성 및 검사법 비교를 평가하였다. 방법 비교를 위해서는 두 가지 상업용 BNP 분석법인 ADVIA Centaur® BNP (Siemens Healthcare diagnostics Inc., USA)와 Triage® BNP (Alere, USA) 검사를 100명의 환자 전혈을 이용하여 실시하였다. 참고구간은 건강검진 참가자 120명의 잔여 검체로 검정하였다.

결과

SelexOn BNP는 BNP 농도가 각각 89.44 pg/mL, 480.71 pg/mL 및 1,201.84 pg/mL인 경우 총 변이계수(CV)가 20.3%, 13.3% 및 10.3%였다. 직선성은 평가된 56 pg/mL-1,544 pg/mL범위에서 관찰되었다. SelexOn BNP (y)의 회귀 방정식은 각각 Centaur BNP (x)와 y=0.9706x-21.68 (r=0.930), Triage BNP (x)와 y=0.7600x+0.0506 (r=0.845)이었다. 임상적 의사결정농도(100 pg/mL)에서 SelexOn BNP의 예측 값은 두 개의 비교 가능한 분석법보다 최대 25% 낮았다. 120 검체 중 114 (95%) 검체에서 SelexOn BNP 수치가 50 pg/mL 미만이었다.

결론

SelexOn BNP가 EDTA 전혈을 사용하여 신속하게 급성 심부전의 감별 진단과 치료 모니터링을 위한 현장검사로 사용되기 위해서는 다른 BNP 분석법과 호환가능한 임상적 의사결정농도를 갖도록 더욱 향상되어야 할 것이다.

INTRODUCTION

Heart failure is a major global cause of disease and death [1]. B-type natriuretic peptide (BNP) is currently regarded as a clinically 6338useful biomarker for diagnosis of heart failure and monitoring treatment [2]. BNP is a biologically active hormone composed of 32 amino acids produced in the myocardium [2]. BNP increases in response to an increased myocardial wall stress [3]. Plasma BNP levels of 100 pg/mL and 400 pg/mL are helpful for exclusion and consideration of heart failure in untreated patients with symptoms of a heart failure, respectively [1, 3].
Various commercial BNP tests are performed using automated analyzers in many clinical laboratories [2]. The point of care (POC) test continues to be needed because it can reduce the clinical potential decision time in emergency situations by providing test results within approximately one hour [1]. However, the POC test should be verified to have a good correlation and precision compared to those in the clinical laboratory assays [1].
A new SelexOnTM B-type natriuretic peptide assay (SelexOn BNP) (Osang Healthcare Inc., Anyang, Korea) was recently developed for POC testing. It consists of a single-use, disposable SelexOnTM cartridge and a portable automated analyzer system. The principle for quantifying BNP is a rapid sandwich immunoassay using gold particles conjugated with a specific antibody on a test cartridge using lateral ow with K2EDTA-anticoagulated whole blood. The camera inside the analyzer detects the signal line intensity. The integrated software converts the signal intensity to a quantitative result by comparing with an internal calibration curve (Fig. 1).

MATERIALS AND METHODS

We evaluated the analytical performance of the SelexOn BNP compared to two BNP assays based on the Clinical and Laboratory Standards Institute (CLSI) guidelines. This study was approved by the Catholic Medical Center Institutional Review Board (OC16-DISI0012).

1. Imprecision

Imprecision was evaluated according to the CLSI guideline EP15-A3 [4]. Low, medium, and high levels of the quality control materials were provided by the manufacturer. The control materials were measured in five replicates a day for five days. Repeatability and within-laboratory imprecision at each level were assessed. Within-laboratory imprecision was compared to the desirable imprecision specification (15%) recommended by Apple et al. [5]. In addition, two K2EDTA-anticoagulated whole blood samples near a clinical decision level of 100 pg/mL were measured in 20 replicates a day per sample to obtain repeatability since BNP should be measured within 4 h of collection at room temperature [5].

2. Linearity

Linearity was evaluated according to CLSI guideline EP6-A [6]. The claimed measuring interval ranged from 50 pg/mL to 2,000 pg/mL. Two patient samples with low (L) and high levels (H) were mixed as follows: 0.75*L+0.25*H, 0.5*L+0.5*H, 0.25*L+0.75*H. Each sample was measured three times using the SelexOn, and the mean values were compared to the expected values. Linear fit, second order, and third order polynomial regression analyses were performed. The regression equation was obtained (y, measured concentration; x, expected concentration) using a linear fit model.

3. Method comparison

Method comparison was performed according to CLSI guideline EP9-A3 [7]. Two BNP assays, the ADVIA Centaur® XPT BNP (Centaur BNP) (Siemens Health Care Diagnostics Inc., Tarrytown, NY, USA) and the Triage® BNP test (Triage BNP) (Alere, San Diego, CA, USA), were selected. The Centaur BNP is an automated sandwich immunoassay using EDTA plasma for central laboratory testing. The Triage BNP is a uorescence immunoassay using EDTA plasma or whole blood for POC testing. K2EDTA-anticoagulated whole blood in a plastic tube was used to perform the SelexOn BNP and the Triage BNP, and the corresponding EDTA plasma was used to perform the Centaur BNP. Regression equations between the compared assays were obtained using the Passing-Bablok fit.

4. Reference interval verification

A total of 120 EDTA whole blood residual samples obtained from individuals (aged from 22 years to 75 years; mean, 57 years), including 48 males (aged from 24 years to 75 years; mean, 60 years) and 72 females (aged from 22 years to 74 years; mean, 54 years) who participated in a health examination program without a medical history of cardiovascular disease or renal disease, were included to verify the reference interval (RI) of 100 pg/mL as a clinical decision point claimed by the manufacturer [1, 3, 8].

5. Statistics

Analyze-it version 5.20 (Analyze-it Software, Ltd., Leeds, UK) was used to analyze imprecision, linearity, and method comparison.

RESULTS

1. Imprecision

The repeatabilities were 15.8%, 9.0%, and 8.1% for three mean BNP control levels of 89 pg/mL, 481 pg/mL, and 1,202 pg/mL, respectively (Table 1). The total CV was 20.3% at 89.44 pg/mL, exceeding the desirable imprecision specification (15%), 9.0% at 480.71 pg/mL and 8.1% 1201.84 pg/mL, which met the desirable imprecision specification advised by Apple et al. [5]. However, mean BNP levels in two clinical samples were 91 pg/mL and 116 pg/mL, with repeatabilities of 4.6% and 6.9%, which were lower than 15.8% of the 89 pg/mL for the control material.

2. Linearity

Linearity was observed from 56 pg/mL to 1544 pg/mL (Fig. 2). The regression equation between the measured (y) and the assigned values (x) was y=16.38 + 0.9728x based on a linear fit (R2=0.992, recovery rate: 95.8%-112.6%). The expected bias for non-linearity between the linear fit and the nonlinear (third order polynomial) fit was -20.6% (95% CI, -78.3% to 37.1%), 8.5% (0.1% to 16.8%), -1.1% (-4.6% to 2.4%), -3.9% (-6.9% to -0.8%), and 1.9% (-0.2% to 4.0%) at 56 pg/mL, 428 pg/mL, 800 pg/mL, 1,172 pg/mL, and 1,544 pg/mL, respectively.

3. Method comparison

A total of 100 patient specimens within the measuring range instructed by the SelexOn BNP manufacturer (50 pg/mL to 2,000 pg/mL) were available for method comparison (Table 2 and Fig. 3). The SelexOn BNP (y) showed correlation coefficients of 0.930 with the Centaur BNP (x) and 0.845 with the Triage BNP (x). The regression equations were y=0.9706x-21.68 with the Centaur BNP and y=0.7600x+0.0506 with the Triage BNP, respectively. The predicted mean difference (%) ((y–x)/y×100) of the SelexOn BNP (y) was -24.63% at 100 pg/mL and -8.63% at 400 pg/mL, respectively, compared to the Centaur BNP (x). The predicted mean difference (%) of the SelexOn BNP (y) was -24.0% at 100 pg/mL and -24.0% at 400 pg/mL, respectively, compared to the Triage BNP (x).

4. Verification of reference interval

Out of the 120 samples, 95.0% were less than 50 pg/mL, 96.3% were less than 75 pg/mL, and 97.5% were less than 100 pg/mL. The distribution of BNP results by age and gender have been summarized in Table 3. Of the 120 individuals across all ages, 45 males and 69 females showed BNP results below 50 pg/mL.

DISCUSSION

We evaluated the analytical performance of the SelexOn BNP for imprecision, linearity, method comparison, and verification of reference interval based on CLSI guidelines.
The total imprecision in the control material ranged from 10.3% to 20.3%, and the CV (20.3%) at 89 pg/mL exceeded 15% of a desirable specification for imprecision [5, 9]. Although repeatability was only evaluated in consideration of BNP stability in whole blood [5], similar low concentration clinical samples had lower repeatability (4.6%-6.9%) than control (15.8%). This may be because the substrate of the manufacturer-provided control material was not the same as that of the clinical sample. On the basis of the intra-individual biological variation (CVI), the goal for imprecision should be less than or equal to half of the CVI [5, 10]. BNP has a very wide intra-individual biological variation (CVI, 30%-50%) like other hormones, and the total imprecision of the SelexOn BNP is considered less than the minimum requirement of analytical imprecision goal (25%) [9-12, 14].
BNP usually increases in heart failure, and patients with a higher BNP value (≥100 pg/mL) in the acute setting should be evaluated via further work-up to rule out heart failure [1, 3, 13]. If the BNP level rises above 35 pg/mL in patients with a suspected heart failure in the non-acute setting, additional echocardiography should be recommended for heart failure [13]. BNP can also increase in various cardiac and non-cardiac conditions like acute coronary syndrome, acute myocardial infarction, pulmonary embolism, hyperthyroidism, renal failure, or septic shock [11, 13]. Therefore, BNP is reportedly a useful biomarker for ruling out rather than diagnosing heart failure because BNP at clinical decision levels has high negative predictive values in both the acute and the non-acute settings [13]. In addition, High BNP during heart failure follow-up and monitoring is also reportedly associated with adverse outcomes [13].
Here, the correlation between the SelexOn BNP with the other two BNP assays was not high (r=0.930 for the Centaur BNP; r= 0.845 for the Triage BNP), and the predicted mean difference (%) of the SelexOn BNP at clinical decision levels the 100 pg/mL and 400 pg/mL was up to 25% lower than that in the two comparable assays. The substantial differences between methods agree with the previous BNP immunoassay comparison studies [1, 2, 10, 14, 15]. Similar results have been reported in previous studies comparing the Triage and the Centaur BNPs [10, 14]. The main putative cause of the non-harmonized BNP assays is that there are no primary reference materials for BNP measurements; many manufacturers use different antibodies and calibrators for BNP immunoassays [1, 2, 16]. Therefore, it is important to identify and harmonize the systematic differences in BNP assays, and more appropriate clinical decision thresholds may be necessary for the individual BNP assay [2, 14]. If the BNP assay has a negative systematic bias, it may be interpreted as a false negative in disease differential diagnosis at low concentrations near 100 pg/mL. Our results suggest that calibration adjustments may be necessary to improve the negative systematic bias of the SelexOn BNP. Variations between lots can have a significant impact on the test results for cartridge-type POT tests, but the lot-to-lot variability was not evaluated in the present study.
In the reference interval verification, most (95% of the 120 individuals) with the SelexOn BNP exhibited values below 50 pg/mL. There was no difference observed between gender and age in the ratio outside the reference interval for the clinical decision point, but further study will be needed for stratified reference intervals. The BNP reference intervals for the healthy elderly are higher than the clinical decision level used for the heart failure guidelines [13, 17, 18]. Therefore, for the elderly, data should be interpreted in relation with the clinical symptoms [17, 18].
In summary, the SelexOn BNP using EDTA whole blood was developed as a POC test for differential diagnosis or treatment monitoring for acute heart failure. However, harmonization should be further strengthened to ensure that clinical decision values are compatible with other BNP methods.

Acknowledgements

This research was supported by a grant of the Korea Health Technology R&D Project through the Korea Health Industry Development Institute (KHIDI), funded by the Ministry of Health & Welfare, Republic of Korea (grant number: HI16C0821).
This research was supported by a Grant of Translational R&D Project through the Institute for Bio-Medical Convergence, Incheon St. Mary’s Hospital, The Catholic University of Korea.
We are grateful to Sang Gil Bae, Hyeji You, and Seon-Hwa Kim for excellent technical assistance.

Notes

Conflicts of Interest

None declared.

REFERENCES

1. Fu S, Ping P, Zhu Q, Ye P, Luo L. 2018; Brain natriuretic peptide and its biochemical, analytical, and clinical issues in heart failure: A narrative review. Front Physiol. 9:692. DOI: 10.3389/fphys.2018.00692. PMID: 29922182. PMCID: PMC5996066.
crossref
2. Semenov AG, Tamm NN, Apple FS, Schulz KM, Love SA, Ler R, et al. 2017; Searching for a BNP standard: Glycosylated proBNP as a common calibrator enables improved comparability of commercial BNP immunoassays. Clin Biochem. 50:181–5. DOI: 10.1016/j.clinbiochem.2016.11.003. PMID: 27823960.
crossref
3. Dickstein K, Cohen-Solal A, Filippatos G, McMurray JJ, Ponikowski P, Poole-Wilson PA, et al. 2008; ESC guidelines for the diagnosis and treatment of acute and chronic heart failure 2008: The task force for the diagnosis and treatment of acute and chronic heart failure 2008 of the European Society of Cardiology. Developed in collaboration with the Heart Failure Association of the ESC (HFA) and endorsed by the European Society of Intensive Care Medicine (ESICM). Eur J Heart Fail. 10:933–89. DOI: 10.1016/j.ejheart.2008.08.005. PMID: 18826876.
4. Clinical and Laboratory Standards Institute. 2014; User veri cation of precision and estimation of bias; Approved guideline-Third edition. CLSI document EP15-A3. Clinical and Laboratory Standards Institute;Wayne, PA:
5. Apple FS, Panteghini M, Ravkilde J, Mair J, Wu AH, Tate J, et al. 2005; Quality speci cations for B-type natriuretic peptide assays. Clin Chem. 51:486–93. DOI: 10.1373/clinchem.2004.044594. PMID: 15738513.
6. Clinical and Laboratory Standards Institute. 2003; Evaluation of the linearity of quantitative measurement procedures: A statistical approach; Approved guideline. CLSI document EP06-A. Clinical and Laboratory Standards Institute;Wayne, PA:
7. Clinical and Laboratory Standards Institute. 2013; Measurement procedure comparison and bias estimation using patient samples; Approved guideline-Third edition. CLSI document EP09-A3. Clinical and Laboratory Standards Institute;Wayne, PA:
8. Clinical and Laboratory Standards Institute. 2010; Defining, establishing, and verifying reference intervals in the clinical laboratory; Approved guideline-Third edition. CLSI document EP28-A3c. Clinical and Laboratory Standards Institute;Wayne, PA:
9. Plebani M, Siest G. 2006; Natriuretic peptides and evidence-based quality specifications. Clin Chem Lab Med. 44:355–7. DOI: 10.1515/CCLM.2006.089. PMID: 16599825.
crossref
10. Wu AH, Packer M, Smith A, Bijou R, Fink D, Mair J, et al. 2004; Analytical and clinical evaluation of the Bayer ADVIA Centaur automated B-type natriuretic peptide assay in patients with heart failure: a multisite study. Clin Chem. 50:867–73. DOI: 10.1373/clinchem.2003.026138. PMID: 15010423.
crossref
11. Clerico A, Carlo Zucchelli G, Pilo A, Passino C, Emdin M. 2006; Clinical relevance of biological variation: the lesson of brain natriuretic peptide (BNP) and NT-proBNP assay. Clin Chem Lab Med. 44:366–78. DOI: 10.1515/CCLM.2006.063. PMID: 16599827.
crossref
12. Bruins S, Fokkema MR, Römer JW, Dejongste MJ, van der Dijs FP, van den Ouweland JM, et al. 2004; High intraindividual variation of B-type natriuretic peptide (BNP) and amino-terminal proBNP in patients with stable chronic heart failure. Clin Chem. 50:2052–8. DOI: 10.1373/clinchem.2004.038752. PMID: 15345664.
crossref
13. Ponikowski P, Voors AA, Anker SD, Bueno H, Cleland JGF, Coats AJS, et al. 2016; 2016 ESC Guidelines for the diagnosis and treatment of acute and chronic heart failure: The task force for the diagnosis and treatment of acute and chronic heart failure of the European Society of Cardiology (ESC). Developed with the special contribution of the Heart Failure Association (HFA) of the ESC. Eur Heart J. 37:2129–200. DOI: 10.1093/eurheartj/ehw128. PMID: 27206819.
14. Prontera C, Zaninotto M, Giovannini S, Zucchelli GC, Pilo A, Sciacovelli L, et al. 2009; Proficiency testing project for brain natriuretic peptide (BNP) and the N-terminal part of the propeptide of BNP (NT-proBNP) immunoassays: The CardioOrmocheck study. Clin Chem Lab Med. 47:762–8. DOI: 10.1515/CCLM.2009.153. PMID: 19402791.
crossref
15. Kang SY, Suh JT, Kim MH, Lee WI, Lee HJ. 2005; Evaluation of BNP assays with Access 2 and AxSYM BNP in comparison with Point-of-Care Triage BNP assay. Korean J Lab Med. 25:300–5.
16. Clerico A, Zaninotto M, Passino C, Plebani M. 2017; New issues on measurement of B-type natriuretic peptides. Clin Chem Lab Med. 56:32–9. DOI: 10.1515/cclm-2017-0433. PMID: 28809748.
crossref
17. Alehagen U, Goetze JP, Dahlström U. 2007; Reference intervals and decision limits for B-type natriuretic peptide (BNP) and its precursor (Nt-proBNP) in the elderly. Clin Chim Acta. 382:8–14. DOI: 10.1016/j.cca.2007.03.005. PMID: 17433809.
crossref
18. Keyzer JM, Hoffmann JJ, Ringoir L, Nabbe KC, Widdershoven JW, Pop VJ. 2014; Age- and gender-specific brain natriuretic peptide (BNP) reference ranges in primary care. Clin Chem Lab Med. 52:1341–6. DOI: 10.1515/cclm-2013-0791. PMID: 24781675.
crossref

Fig. 1
The SelexonTM B-type natriuretic peptide (BNP) assay system. (A) The disposable cartridge uses a lateral flow immunoassay to detect BNP. Control and test lines are visible on the nitro-cellulose membrane. The control line indicates that the sample has migrated across the membrane as intended. (B) The camera inside the analyzer detects the signal line intensity. The integrated software converts the signal intensity to a quantitative result and shows it on the display.
LMO-10-207-f1.tif
Fig. 2
Linearity of the SelexOnTM B-type natriuretic peptide assay. The red and green regression lines represent the linear and polynomial fits, respectively. The regression equation between the measured (y) and the assigned values (x) was y=16.38+0.9728x based on a linear fit (R2=0.992).
LMO-10-207-f2.tif
Fig. 3
Scatter plot of method comparison for the SelexOn BNP, the Centaur BNP, and the Triage BNP assays. The red line represents the linear regression equation by the Passing-Bablok fit (left) and difference plots (right). The SelexOn BNP (y) regression equation was (A) y=0.9706x-21.68 with Centaur BNP (x) (r=0.930), and (B) y=0.7600x+0.0506 with Triage BNP (x) (r=0.845).
LMO-10-207-f3.tif
Table 1
Precision of the SelexOnTM B-type natriuretic peptide (SelexOn BNP) assay
Material Mean BNP (pg/mL) Repeatability Within-laboratory


SD CV SD CV
Control 1* 89 14 15.8% 18 20.3%
Control 2 481 43 9.0% 64 13.3%
Control 3 1,202 98 8.1% 124 10.3%
Sample 1 91 4 4.6% - -
Sample 2 116 8 6.9% - -

*CLSI guideline EP15-A3 (five repeats a day for five days) for control materials.

20 repeats a day for the K2EDTA-whole blood samples.

Table 2
Method comparison of the SelexOnTM BNP with the Centaur® BNP and the Triage® BNP assays (N=100)
Methods compared (y vs. x) Intervals measured (y, pg/mL) r Passing-Bablok fit equation y-intercept (95% CI) Slope (95% CI)
SelexOn vs. Centaur 52-1995 0.930 y=0.9706x-21.68 -49.27 to -5.654 0.8969 to 1.108
SelexOn vs. Triage 52-1995 0.845 y=0.7600x+0.0506 -22.60 to 30.62 0.6217 to 0.8576
Table 3
Distribution of BNP results from non-heart failure samples participating in the health examination program (N=120)
Age group All (%) All ages (N=120) <45 years (N=22) 45-54 years (N=22) 55-64 years (N=37) 65-75 years (N=39)






BNP (pg/mL) Male Female Male Female Male Female Male Female Male Female
<50 114 (95) 45 69 3 18 9 10 14 22 19 19
50-74 1 (0.8) 1 0 - - - - - 1
75-100 2 (1.7) 1 1 1 - - 1 -
101-160 3 (2.5) 1 2 - - 1 1 - 1
TOOLS
Similar articles