Vadiati Saberi, Khosravifard, and Nourzadeh: Effect of slice inclination and object position within the field of view on the measurement accuracy of potential implant sites on cone-beam computed tomography

Abstract

Purpose

The purpose of this study was to evaluate the accuracy of linear measurements in the horizontal and vertical dimensions based on object position and slice inclination in cone-beam computed tomography (CBCT) images.

Materials and Methods

Ten dry sheep hemi-mandibles, each with 4 sites (incisor, canine, premolar, and molar), were evaluated when either centrally or peripherally positioned within the field of view (FOV) with the image slices subjected to either oblique or orthogonal inclinations. Four types of images were created of each region: central/cross-sectional, central/coronal, peripheral/cross-sectional, and peripheral/coronal. The horizontal and vertical dimensions were measured for each region of each image type. Direct measurements of each region were obtained using a digital caliper in both horizontal and vertical dimensions. CBCT and direct measurements were compared using the Bland-Altman plot method. P values <0.05 were considered to indicate statistical significance.

Results

The buccolingual dimension of the incisor and premolar areas and the height of the incisor, canine, and molar areas showed statistically significant differences on the peripheral/coronal images compared to the direct measurements (P<0.05). Molar area height in the central/coronal slices also differed significantly from the direct measurements (P<0.05). Cross-sectional images of either the central or peripheral position had no marked difference from the gold-standard values, indicating sufficient accuracy.

Conclusion

Peripheral object positioning within the FOV in combination with applying an orthogonal inclination to the slices resulted in significant inaccuracies in the horizontal and vertical measurements. The most undesirable effect was observed in the molar area and the vertical dimension.

Introduction

Cone-beam computed tomography (CBCT) has gained versatile applications due to its sub-millimeter spatial resolution, relatively low cost, and low radiation dose administered compared to conventional computed tomography. Linear CBCT measurements are mainly used for pre-surgical implant planning to determine the available bone dimensions and distance from the vital anatomical structures.12 Orthodontic analyses and determination of lesion dimensions in the jaws also require linear measurements.345
Studies of the accuracy of CBCT-based measurements have produced conflicting results. While most studies have indicated the presence of measurement errors of less than 1 mm, some have reported errors as high as 6 mm. A dimensional inaccuracy of one-tenth of a millimeter has been reported to be sufficient to jeopardize the outcomes of surgical procedures such as implant insertion.6789 Injury to the neurovascular bundle and resultant neuroma, paresthesia, anesthesia, and hemorrhage are the most serious consequences of errors originating from inaccurate measurements.10
Several factors, including slice thickness, voxel size, and head orientation, have been proposed as contributors to the amount of distortion in linear CBCT measurements.1112 Contemporary devices provide various choices for the field-of-view (FOV) size as well as the position of the region of interest (ROI) within the FOV. It is assumed that central or peripheral positioning of the ROI in the FOV influences the magnitude of inaccuracy in the final image.1314 Moreover, most CBCT software allows slices to be obtained in any orthogonal and non-orthogonal direction, which is likely to provide varying measurement results.10 The purpose of the present study was to determine the concomitant influence of object positioning within the FOV and slice inclination on the accuracy of linear measurements obtained from CBCT images of potential implant sites.

Materials and Methods

This in vitro experimental study received approval from the research ethics committee of Guilan University of Medical Sciences (Approval ID: IR.GUMS. REC.1397.076). Dry sheep hemi-mandibles of similar dimensions were used to simulate the alveolar bone.13 All hemi-mandibles in this study contained 4 distinct areas: incisor, canine, premolar, and molar. Furthermore, the buccal and lingual cortical plates of all samples were intact. Hemi-mandibles with disrupted cortices, unusual bony undercuts, or evidence of previous bony lesions were excluded from the investigation.

Preparation of the samples

Ten dry sheep hemi-mandibles were selected for the study. The areas to be assessed in each hemi-mandible were determined according to the criteria described by Neves et al.10 which defined the incisor region to be 1 cm distal from the mid-sagittal plane, the canine region 1 cm distal from the former, and the premolar and molar regions at the mental foramen and 1-2 cm distal to the mental foramen, respectively. A total of 40 areas were available for measurement. Radiopaque gutta-percha markers (size 30; Meta Biomed, Cheongju, Korea) were fixed parallel to each other on the buccal and lingual sides of the 4 areas of each hemi-mandible. Care was taken to position the buccal and lingual markers exactly facing each other by using threads to ensure standardized regions for obtaining the measurements. Each hemi-mandible was coated with 1.5 cm of wax to simulate soft tissue density.15

Radiographic examinations

A rectangular piece of plastic foam was prepared with a 2-mm-deep longitudinal central slit for insertion of the hemi-mandibles (Fig. 1). Use of the plastic foam ensured the fixed position of each hemi-mandible during the radiographic examination.
The CBCT examinations were performed with a Pax-i 3D unit (Vatech, Hwaseong, Korea). The exposure protocol was as follows: 95 kV; 5.2 mA; FOV, 150 mm×150 mm; and voxel size, 0.2 mm. Each hemi-mandible underwent 2 CBCT exposures while centrally and peripherally positioned. For the centrally positioned exposures, the plastic foam containing the hemi-mandible was placed on a straight line passing through the center of the FOV anteroposteriorly. The horizontal and vertical positions of exposures were kept on ‘center’ and ‘mandible’ settings, respectively. Peripherally positioned exposures were acquired by placing the hemi-mandibles on an oblique line that anteriorly formed an angle of 20° with the anterior end of the central line and posteriorly was 5 cm distanced from the posterior end of the central line of the FOV (Fig. 2).

Radiographic assessments

In each of the central and peripheral exposures, image slices were created with 2 inclinations: cross-sectional (non-orthogonal) and coronal (orthogonal). Cross-sectional slices were created in the “section” part of the software (Ez3D-i, Vatech) by manually drawing a curved line through the center of the bone in the axial plane, resulting in the automatic creation of cross-sectional images in the buccolingual direction that were roughly perpendicular to the curved drawn line and therefore were oriented obliquely in relation to the buccolingual aspect of the bone. The thickness and interval of the slices were set at 1 mm.
Coronal slices were acquired from the orthogonal slices that were provided automatically by the “MPR” section of the software. These slices were of the same thickness and interval as the cross-sectional slices.
For each region of the hemi-mandibles, the coronal slice corresponding to the full length of the buccal and lingual radiopaque markers was selected and stored as a saved project. A cross-sectional slice of each region that revealed the maximum extent (not necessarily the full length) of the markers was also selected and saved as a project. Displaying the full length of the buccolingual markers in a single cross-sectional slice was not possible, since in contrast to the coronal slices, cross-sectional slices do not pass perpendicularly through the center of the bone in the buccolingual direction.
In total, 4 types of image slices were saved for each region: central/cross-sectional, central/coronal, peripheral/cross-sectional, and peripheral/coronal. The saved image slices of each region were randomly presented to an experienced maxillofacial radiologist with no knowledge of the real measurement values. The radiologist was asked to measure the buccolingual and vertical dimension (height) of the bone for each image type. The buccolingual dimension was defined as the horizontal distance between the buccolingual radiopaque markers on their crestal portion. Bone height was defined as the vertical distance from the superior crest to the inferior cortex of the bone (Fig. 3). All measurements were performed twice, with measurements separated by a 2-week interval. Finally, for each region, the mean values of the first and second observations were calculated and recorded as the buccolingual dimension and height for that specific area on the associated image type. All images were displayed on a medical LCD monitor (RadiForce MX241W; EIZO Corporation, Hakusan, Japan). As the images were stored as saved projects, neither the slices nor their inclinations could be altered by the observer. This ensured that the observer only assessed the determined areas.
Upon completion of the CBCT measurements, the buccolingual dimension and height of each region were directly measured using the gold-standard technique involving a digital caliper (SC-6; Mitutoyo Corporation, Kawasaki, Japan). The buccolingual dimension was measured as the horizontal distance between the crestal portion of the buccal and lingual markers of each region. Height was measured as the vertical distance between the superior crestal and inferior cortical portions of each area.

Statistical analysis

Data were imported to MedCalc statistical software version 18.9.1 (MedCalc Inc., Ostend, Belgium). Comparison of the measurements obtained from the different image slices with the gold-standard values was performed using the Bland-Altman method. P<0.05 was considered to indicate statistical significance.

Results

Measurements obtained from the cross-sectional slices in both the central and peripheral positions did not show significant discrepancies from the directly measured values in the horizontal and vertical dimensions in any of the 4 assessed regions (P>0.05).
The coronal slice measurements in the central position were also consistent with the direct values, except for the molar area height, which was overestimated by a mean of 0.34 mm (P=0.01).
The coronal slices in the peripheral position showed the greatest amount of measurement discrepancy, which manifested as overestimation of the buccolingual dimension of the incisor and premolar areas by 0.56 mm and 0.61 mm, respectively (P<0.05). Furthermore, the incisor, canine, and molar area heights were overestimated by an average of 0.34 mm, 0.67 mm, and 0.67 mm, respectively (P<0.05).
Table 1 shows the mean differences in the upper and lower limits of the buccolingual dimension and height values obtained with each type of image slice compared to the gold-standard measurements with the Bland-Altman method. In this method, upper limits were designated as cases in which the test method measurements were greater than the real measurements, while cases in which the real measurements were greater were used to designate lower limits. P<0.05 was considered to indicate statistical significance.

Discussion

In the present study, different images were created of each region by placing the hemi-mandibles in the center or periphery of the FOV and applying non-orthogonal (oblique) and orthogonal (coronal) inclinations to the slices. Measurement accuracy was subsequently assessed in the horizontal and vertical dimensions. We intended to determine the influence of simultaneously altering the object position and slice inclination on measurement accuracy, since to our knowledge, the concomitant effect of manipulating these 2 variables has not been previously investigated.
Although CBCT manufacturers recommend optimal patient positioning during image acquisition, patients with skeletal malformations and malocclusions are likely to be positioned improperly within the device.16 This prevents some portions of the jaw from being ideally positioned within the center of the FOV. However, contemporary CBCT software tools have the capability of creating image slices in any desired direction in accordance with the concept that CBCT benefits from isotropic voxel resolution.10 Given the potential of these variations to yield different results, their impact on the accuracy of the measurements must be fully investigated. Considering that CBCT images are routinely used for the pre-surgical planning of potential implant sites, measurement errors could result in faulty decisions regarding implant size and type, as well as the need for augmentation procedures.
Studies of the reliability of CBCT volumetric measurements in orthodontic patients have failed to show significant discrepancies;1718 however, investigations of linear measurements of potential implant sites have yielded conflicting results under certain circumstances.19 This may be attributed to the fact that implant site measurements are much more sensitive than orthodontic volumetric evaluations. A discrepancy of one-tenth of a millimeter is sufficient to threaten the integrity of regional cortical boundaries and vital anatomical structures such as the inferior alveolar nerve.9
Moshfeghi et al.20 found no significant difference in the linear measurements obtained from the axial and coronal slices and with different voxel sizes. This finding could be attributed to the similar nature of these 2 slice types, both of which are orthogonally oriented. However, Neves et al.10 evaluated vertical measurements in the oblique (cross-sectional) and orthoradial (coronal) slices of different jaw regions and concluded that vertical measurements of the coronal slices in the molar region lack sufficient accuracy compared to cross-sectional slices. Similarly, in the present study, a significant difference was encountered between direct measurements and measurements obtained from the coronal slices of the molar area in the central position (P<0.05). These differences were even more pronounced in the peripheral position, as coronal slices yielded inaccurate measurements in the horizontal dimension of the incisor and premolar areas, as well as the vertical dimension of the incisor, canine, and molar regions (P<0.05).
Sabban et al.9 and Visconti et al.11 assessed the accuracy of measurements made with different head orientations. They applied tilt, flexion, and extension movements with a deviation of 20°. Measurement errors ranging between - 2 mm and 3 mm were observed in different jaw regions. In the present study, peripheral positioning was obtained by placing the hemi-mandibles 20° laterally from the center of the FOV to simulate abnormal conditions such as jaw asymmetry or head rotation during patient positioning. Peripheral/coronal slices showed the greatest amount of inaccuracy, which was up to 0.63 mm for the horizontal measurements and 0.67 mm for the vertical measurements. The molar region was the area most strongly affected by measurement discrepancies, as coronal slices of this region showed significant inaccuracy even in the central position.
With regard to dimension, vertical measurements generally tended to be more severely affected than horizontal measurements. Similarly, Sabban et al.9 suggested that vertical measurements are relatively susceptible to inaccuracy since the bone volume displaced eccentrically in the image layer is greater in the vertical dimension. This is concerning, since vertical errors usually result in more undesirable outcomes in the anatomical relationship of the implants with their surrounding structures.
Within the limitations of this in vitro study, it can be concluded that simultaneous peripheral positioning of the ROI in the FOV and application of orthogonal (coronal) slices result in the greatest amount of measurement inaccuracy. The molar area and vertical dimension are at particularly high risk for overestimation by CBCT measurements, especially when not positioned ideally within the FOV.

Acknowledgements

The authors would like to thank Dr. Mohammad Ebrahim Ghaffari (Department of Biostatistics, Guilan University of Medical Sciences) for the statistical consults.

Notes

Conflicts of Interest: None

References

1. Lascala CA, Panella J, Marques MM. Analysis of the accuracy of linear measurements obtained by cone beam computed tomography (CBCT-NewTom). Dentomaxillofac Radiol. 2004; 33:291–294. PMID: 15585804.
crossref
2. Kosalagood P, Silkosessak OC, Pittayapat P, Pisarnturakit P, Pauwels R, Jacobs R. Linear measurement accuracy of eight cone beam computed tomography scanners. Clin Implant Dent Relat Res. 2015; 17:1217–1227. PMID: 24673845.
crossref
3. Hassan B, van der Stelt P, Sanderink G. Accuracy of three-dimensional measurements obtained from cone beam computed tomography surface-rendered images for cephalometric analysis: influence of patient scanning position. Eur J Orthod. 2009; 31:129–134. PMID: 19106265.
crossref
4. Timock AM, Cook V, McDonald T, Leo MC, Crowe J, Benninger BL, et al. Accuracy and reliability of buccal bone height and thickness measurements from cone-beam computed tomography imaging. Am J Orthod Dentofacial Orthop. 2011; 140:734–744. PMID: 22051495.
crossref
5. Mah J, Hatcher D. Three-dimensional craniofacial imaging. Am J Orthod Dentofacial Orthop. 2004; 126:308–309. PMID: 15356493.
crossref
6. Kobayashi K, Shimoda S, Nakagawa Y, Yamamoto A. Accuracy in measurement of distance using limited cone-beam computerized tomography. Int J Oral Maxillofac Implants. 2004; 19:228–231. PMID: 15101594.
7. Frei C, Buser D, Dula K. Study on the necessity for cross-section imaging of the posterior mandible for treatment planning of standard cases in implant dentistry. Clin Oral Implants Res. 2004; 15:490–497. PMID: 15248885.
crossref
8. Ludlow JB, Laster WS, See M, Bailey LJ, Hershey HG. Accuracy of measurements of mandibular anatomy in cone beam computed tomography images. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2007; 103:534–542. PMID: 17395068.
crossref
9. Sabban H, Mahdian M, Dhingra A, Lurie AG, Tadinada A. Evaluation of linear measurements of implant sites based on head orientation during acquisition: an ex vivo study using cone-beam computed tomography. Imaging Sci Dent. 2015; 45:73–80. PMID: 26125001.
crossref
10. Neves FS, Vasconcelos TV, Oenning AC, de Azevedo-Vaz SL, de Almeida SM, Freitas DQ. Oblique or orthoradial CBCT slices for preoperative implant planning: which one is more accurate? Braz J Oral Sci. 2014; 13:104–108.
crossref
11. Visconti MA, Verner FS, Assis NM, Devito KL. Influence of maxillomandibular positioning in cone beam computed tomography for implant planning. Int J Oral Maxillofac Surg. 2013; 42:880–886. PMID: 23566433.
crossref
12. Shokri A, Khajeh S. In vitro comparison of the effect of different slice thicknesses on the accuracy of linear measurements on cone beam computed tomography images in implant sites. J Craniofac Surg. 2015; 26:157–160. PMID: 25569395.
crossref
13. Nikbin A, Dalili Kajan Z, Taramsari M, Khosravifard N. Effect of object position in the field of view and application of a metal artifact reduction algorithm on the detection of vertical root fractures on cone-beam computed tomography scans: an in vitro study. Imaging Sci Dent. 2018; 48:245–254. PMID: 30607348.
crossref
14. Alkhader M, Hudieb M, Jarab F, Shaweesh A. The visibility of mandibular canal on orthoradial and oblique CBCT slices at molar implant sites. Biotechnol Biotechnol Equip. 2016; 30:770–776.
crossref
15. Kamburoğlu K, Murat S, Kılıç C, Yüksel S, Avsever H, Farman A, et al. Accuracy of CBCT images in the assessment of buccal marginal alveolar peri-implant defects: effect of field of view. Dentomaxillofac Radiol. 2014; 43:20130332. PMID: 24645965.
crossref
16. Dawood A, Brown J, Sauret-Jackson V, Purkayastha S. Optimization of cone beam CT exposure for pre-surgical evaluation of the implant site. Dentomaxillofac Radiol. 2012; 41:70–74. PMID: 22184628.
crossref
17. Moreira CR, Sales MA, Lopes PM, Cavalcanti MG. Assessment of linear and angular measurements on three-dimensional cone-beam computed tomographic images. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2009; 108:430–436. PMID: 19386521.
crossref
18. Frongia G, Piancino MG, Bracco P. Cone-beam computed tomography: accuracy of three-dimensional cephalometry analysis and influence of patient scanning position. J Craniofac Surg. 2012; 23:1038–1043. PMID: 22777473.
19. Rokn AR, Hashemi K, Akbari S, Kharazifard MJ, Barikani H, Panjnoosh M. Accuracy of linear measurements using cone beam computed tomography in comparison with clinical measurements. J Dent (Tehran). 2016; 13:333–339. PMID: 28127327.
20. Moshfeghi M, Tavakoli MA, Hosseini ET, Hosseini AT, Hosseini IT. Analysis of linear measurement accuracy obtained by cone beam computed tomography (CBCT- NewTom VG). Dent Res J (Isfahan). 2012; 9(Suppl 1):S57–S62. PMID: 23814563.
Fig. 1

A hemi-mandible fixed in the central slit of a plastic foam block. The 4 regions are defined by gutta-percha markers on the buccal and lingual sides.

isd-50-37-g001
Fig. 2

Axial images of centrally (A) and peripherally positioned (B) exposures of a hemi-mandible.

isd-50-37-g002
Fig. 3

Various image types of the same region with buccolingual and height measurements. A. Central/coronal. B. Central/cross-sectional. C. Peripheral/coronal. D. Peripheral/cross-sectional.

isd-50-37-g003
Table 1

Mean differences (MDs) of the buccolingual dimension (BL) and height of the different image types compared to direct measurements by region (unit: mm)

isd-50-37-i001

MD: calculated from the upper and lower limits obtained using the Bland-Altman method. *: P<0.05