Shon, Park, R, Park, Im, and Yun: Total Hip Arthroplasty: Past, Present, and Future. What Has Been Achieved?

Abstract

In the 1960s, Sir John Charnley introduced to clinical practice his concept of low-friction total hip arthroplasty (THA). Although early designs were plagued by poor performance and even failure, there have been steady advances in implant designs, biomaterials, surgical techniques and an understanding of the biomechanical restoration of the hip; these advances have contributed to improvements in implant survival and clinical outcomes of THA in the past three decades. With improved wear resistance and mechanical reliability, a potential to last for at least 25 to 30 years are now available for THA. In this review, we focus on the evolution of THA and review current controversies and future directions of this procedure based on a single surgeon's 29-year of experience at a single institution.

INTRODUCTION

The first published reports of hip arthroplasty date as far back as the 1940s12). In the 1960s, Sir John Charnley introduced to clinical practice his concept of low-friction total hip arthroplasty (THA)34). The early results of THA with and without bone cement were disappointing56789101112) due to the poor design of implants, undersized femoral components, poor cementing technique, periprosthetic osteolysis and excessive wear of the polyethylene liner. However, there have been steady advances in THA over the last three decades (e.g., improved understanding of cementing techniques, better acetabular and femoral component design, refined implantation surgical techniques) leading to significant improvements in implant survival and clinical outcomes. THA is considered a landmark surgery in modern medical history and the clinical results over the last 30 years have proven that it is one of the most effective and successful surgical techniques for treating various pathological hip conditions13).
This review summarizes the history, current controversies and future directions of THA in order to highlight and address issues that are key to: i) improving clinical outcomes today and ii) guide future advancements and improvements on the success of this surgical technique. The information presented here is based on the experience of the author and his more than 29 years of experience performing THAs at a single institution.

THE PAST

In 1990, the author began clinical practice using hybrid THA14)–a variation of Charnley's low-friction THA3) that employs cemented and cementless fixation of the femoral and acetabular components, respectively. Initially, the majority of implants were first-generation15161718) hemispherical porous acetabular cups (Harris-Galante; Zimmer, Warsaw, IN, USA) and Precoat1419) femoral stems (Harris Precoat Plus; Zimmer) (Fig. 1) with conventional non-cross-linked ultra-high molecular weight polyethylene (UHMWPE) liners2021). Implants were selected based on several factors: i) despite excellent mid-term clinical results of cemented cups, these components were associated with unacceptably high failure rates upon long-term follow-up2223), ii) the size of Precoat stems was suitable for the majority of the smaller Asian population, such as Koreans, and iii) conventional non-cross-linked UHMWPE2021) was the most widely used liner biomaterial during that time.
A review of the author's early series of 76 primary hybrid THAs with a mean follow-up of 15.5 years (range, 14-19.5 years)14), there were 23 (30.3%) acetabular reoperations (isolated acetabular liner exchange, 12 cases; cup revision, 11 cases) with an average duration to revision of 11.5 years. The survival rate of the acetabular component was 91.5% at 10 years, 72.0% at 15 years, and 45.3% at 19 years, results which are similar to other analyses of first-generation cementless cups15161718). Based on studies of cementless acetabular cups, press-fit fixation of the cup was recommended. The most common reason for acetabular reoperations was excessive polyethylene liner wear and osteolysis secondary to polyethylene particulate debris (Fig. 2), which were the most notable factors endangering the long-term survivorship of THA2124). Although the cementless hemispherical cup had good fixation ability, its ability to survive for 20 years was difficult due to wear and wear-related osteolysis (Fig. 3).
Of the 76 Precoat femoral stems16), 24 femoral stems (31.6%) experienced aseptic loosening with an average duration of eight years until need for revision surgery even with the use of the modern third-generation cementing techniques (i.e., pulsatile lavage, cement restrictor, retrograde canal filling and pressurization with a cement gun, reduced cement porosity by vacuum mixing)25). Twenty-one out of 24 cases of femoral stem aseptic loosening demonstrated poor C2 cement grade (P<0.001) (Fig. 4). The mantles of cement in the femur were graded on early postoperative radiographs, according to previously reported criteria14). Scoring is as follows: the cementing technique resulted in incomplete filling of the proximal portion of the medullary canal of the diaphysis, a so-called ‘white-out’ at the cement-bone interface (Grade A); the distribution of cement is nearly complete, but it is possible to distinguish cortical bone from cement in some areas (Grade B); extensive radiolucent line (more than 50 percent of the cement-bone interface) or voids in the cement (Grade C1); either a thin (less than one millimeter) mantle of cement at any site or a defect in the mantle of cement, with the metal in direct contact with the cortical bone (Grade C2) and any gross deficiencies in the mantle of cement, such as the presence of no cement, distal to the tip of the stem or major defects in the mantle of cement or multiple large voids (Grade D). Early failure of the cemented femoral stems in the author's early series was mainly due to an inefficient cementing technique and the use of Precoat stems1419). It was also observed that patients having a cement grade of C2 had more subsidence when compared to patients with other cement grades. Furthermore, most stem failures occurred at the cement-bone interface despite the fact that Precoat stems strengthens the stemcement interface. This was a so-called “cement-bone interface failure” because both the stem and cement became a lump and subsided in the medullary cavity of the femur. However, Precoat stems with a good cement mantle, have survived for more than 20 to 25 years, even in young patients (Fig. 5).
In 2007–2008, the author received a research grant from the Korea Food and Drug Administration (KFDA) and conducted research focused on the issues of polyethylene wear and wear-related osteolysis26272829). The author noted that the Harris-Galante cups revealed clear lysis in the hip, which was observed in anteroposterior and both oblique radiographs, at an average of 8 years after surgery; the annual wear rate of conventional non-cross-linked UHMWPE was found to be 0.175 mm/yr. The correlation between osteolysis volume in computed tomography (CT) and the area of osteolysis on a radiograph and polyethylene linear wear were also assessed by the author revealing that the sensitivity and specificity of anteroposterior radiographs for detection of osteolysis were 57.6% and 92.9%, respectively. The addition of oblique radiographs increased the sensitivity to 64.4% without changing the specificity. The sensitivity of simple radiographs increased to 92.8% for a lesion more than 1,000 mm3. In the author's cases, there was pronounced pelvic lysis of about 25% at 8 years, 50% at 10 years, and 75% at 11 to 12 years after surgery (Fig. 6). In a comparative study between radiographs and 3-dimensional CT29), plain radiographs were found to be equally useful when screening for clinically significant osteolysis. In other CT-based studies2730), the authors suggested the extent of the pelvic osteolytic volume on CT (Fig. 7) as a guide of the decision-making for treatment in patients who present with evidence of pelvic osteolysis which was noted to be accelerated in active young patients in the early stage.
In summary, the authors' past experiences using hybrid THAs have been disappointing because of poorly designed implants, undersized femoral components, poor cementing techniques, periprosthetic osteolysis, and excessive wear of the polyethylene liner.

THE PRESENT

To address the issues raised by the author in his earlier experiences, from 2000 the author began to use a second-generation acetabular cup (Trilogy; Zimmer, USA) which had: i) an improved liner locking system when compared to the first-generation acetabular cups, ii) a polished tapered femoral stem (Polished Versys Heritage; Zimmer), and iii) a highly cross-linked polyethylene liner (Longevity; Zimmer). Additionally, the author used a proximal and distal centralizer on the stem to prevent a grade C2 cement mantle defect.
In the author's study of 95 primary hybrid THAs with an average follow-up of 10 years31), the Harris hip score (HHS) improved from preoperative 58.9 points to postoperative 91.7 points. Cementing grades for the 95 hips were Grade A (n=45; 47.3%), Grade B (n=48; 50.5%), and Grade C1 (n=2; 2.1%). There was one case of definite cup loosening (1.0%) due to periprosthetic joint infection (PJI), however, periacetabular osteolysis was not observed in these 95 cases. We extended this series to 152 hips (132 patients), with an average of 10 years follow-up (range, 6 to 16 years) in 2016. The clinical results were similar to those in the previous study31) and there were no reoperations. None of the hips revealed definite lysis on simple radiographs (anteroposterior and both oblique) of the pelvis and femur in this series. The excellent results obtained in this study (Fig. 8) were attributed to the use of a second-generation acetabular cup with highly cross-linked polyethylene liner, a polished surface stem with an improved cementing technique, and an appropriately sized femoral stem. When measured using the Polyware software (Draftware Inc., North Webster, IN, USA), the annual wear rate for highly cross-linked polyethylene liner and conventional non-cross-linked polyethylene liner was found to be 0.048 mm/yr and 0.080 mm/yr, respectively. These results were similar to other studies3233), revealing that the annular linear wear for highly cross-linked UHMWP was about one-third or less than half when compared to that of the conventional polyethylene liners.
Due to the high incidence of impingement after THA34) (Fig. 9), the author started using larger femoral heads (36 mm) to prevent impingement between the rim of the acetabular cup and the neck of the femoral stem. In 2001, ceramic-on-metal (COM) bearing surfaces were first introduced by John Fisher35). The advantages of COM are the reduced possibility of liner fracture and the ability to use a femoral head of 36 mm or more. In a comparative in vitro and in vivo study36), the performance of COM bearings was found to be superior to metal-on-metal or ceramic-on-ceramic (COC) couples. The author began to use COM (36 mm Biolox delta ceramic head and Ultamer metal liner) (Fig. 10) in 2009. In the author's analysis of 162 primary COM THAs with a mean of 4 years follow-up (range, 2–7 years), there were no cases of adverse local37) or systematic38) reactions, infections, or loosening. Average HHS was 94.8 points, and more importantly, 70%, 80%, and 86% of the patients were able to naturally sit cross-legged, kneel and squat. In the author's clinical study39), the results of 140 COM THAs were compared with 94 non-COM THAs and it was noted that the COM group treated with 36 mm femoral heads had a higher HHS than the non-COM group treated with 28 mm femoral heads (P=0.013). Furthermore, the total hip joint range of motion (ROM) was significantly greater in the COM group compared with the non-COM group (P<0.001). The author also compared the metal ion levels in patients who could sit cross-legged, kneel, and squat with those who could not do the same in the COM group. It was observed that the serum cobalt ion levels in the patients who could squat were significantly higher compared with those in the patients who could not squat (P=0.033). Similarly, the serum cobalt and chromium ion levels of the patients who could kneel were significantly higher compared with those patients who could not kneel (P=0.049 and 0.031). Although the serum metal ion concentration was well below the concentration that could have a detrimental effect on the human body, significantly higher levels of cobalt and chromium concentrations were detected in the COM THA group compared with those of the non-COM THA group. Patients with a COM THA were capable of a greater ROM to suite the Asian lifestyle (i.e., the ability to sit cross-legged, kneel, and squat) were at risk of metal ion-related problems even with a 36 mm femoral head. In summary, there was no cases of metal-ion related problems observed in primary COM THA with 36 mm femoral heads. However, the author experienced a case of pseudotumor formation with acetabular cup aseptic loosening after revision COM THA40). Patients who underwent 36-mm COM THAs also experienced favorable clinical outcomes, with an improved hip function at short-term and midterm follow-ups but higher circulating Co and Cr ion levels compared with patients who underwent non-COM THAs. The serum metal ion levels and metal ion-related complications require additional investigation with a longer follow-up period, to better elucidate the influence of an Asian lifestyle on metal ion-related complications, such as muscle necrosis, osteolysis, pseudotumor, and hypersensitivity reaction.
Between 2002 and 2017, and before the author retired, data from a total of 1,557 primary THAs were available from Korea University Guro Hospital in South Korea. There were three cases (0.19%) of femoral stem revision, 30 cases (1.9%) of PJI treated with aggressive early debridement (n= 26; 1.7%), or revision surgery (n=4; 0.26%). Additionally, four cases (0.26%) of dislocation were noted along with two cases (0.13%) of periprosthetic fracture. Of the 1,557 cases in 16 years, only 41 cases (2.6%) underwent reoperations or revisions for any cause. The author's results are similar to other established studies using national joint registries4142), revealing that the incidence of 15-year survival rate in THAs' in the 1990s was about 75–80%. However, when assessed in 2017, the survival rate of THA after 1998 was greater than 90%, hence it is evident that over time, loosening and revisions related to osteolysis continue to increase in the group where conventional non-cross-linked UHMWPE liners were used, whereas it has stayed at 1% even after 15 years in the group where highly cross-linked UHMWPE liners was used. Although PJI, dislocation, and periprosthetic fracture are common reasons for reoperation after THA, advancements in surgical techniques, tribology of the implants and improvement in the fixation techniques have greatly improved the long-term survival of the implants at present. The promising results of successful implant survival and improvement in a patient-reported outcome further suggests that THA is a feasible option, even in young patients with severe symptoms at the author's institute.

THE FUTURE

Several areas are expected to play larger roles in the future of THAs. Current THAs with alternative bearings (highly cross-linked polyethylene or ceramics) have provided excellent mid-term and long-term results. Hence, we can now focus on functional outcome improvements following THAs, especially in the Asian population–a group whose lifestyle varies greatly compared with the western population. Therefore, hip surgeons must be aware of advances in surgical techniques and the evidence behind implant selection as well as the continual improvement in the functional outcomes of this procedure to maximize longevity of the prosthesis. In Asia, these surgeries must be conducted with lifestyles of the population in mind (e.g., sitting on the floor, sitting cross-legged, kneeling) as these behaviors directly relate to a patient quality of life, thus serving as an important parameter in the clinical outcomes after THA. Hence, findings such as improvement in ROM resulting from the use of 36 mm heads compared with 28 mm heads (Table 1) and a significant difference in the clinical results between the groups for sitting cross-legged, squatting, and kneeling (Table 2) must be duly noted.
However, if patients are performing activities that increase ROM after THAs, impingement may occur between the cup liner and femoral stem, in current THA designs (Fig. 11). Impingement in THAs was observed in 27–84% of THA patients in retrieval studies (Fig. 12), which was well within the range observed in the author's retrieval study34). Even though the position of the cup and stem had been well placed within the safe zone, rim damage due to impingement was observed in over 50% of cases The high incidence of impingement after THA implies that the functional outcomes of current THA designs lacks the ability to perform all daily physical activities which are needed to satisfy the demands of an Asian population. Changes in the implant design (e.g., avoiding skirted heads, maximizing the head-to-neck ratio, using a chamfered acetabular liner and a trapezoidal neck cross-section) may help reduce the prevalence of impingement34).
With regard to bearing surfaces from a global perspective, the recent trends and epidemiology of bearing surface usage in primary THAs in South Korea vary compared those in other countries434445), including the use of a COC bearing surface–the most prevalent articulation regardless of patient's age and hospital amounting to up to 80% of total bearing surface cases46). This is in contrast to the trend in the USA4344), where the COC bearing is only used in 4% of cases, whereas the polyethylene liner is used most often (92% of cases). This tendency is similar to what has been reported in Europe45). The choice of bearing surface may be affected by many factors, which includes the nation's medical delivery system, payment type, disease pattern and age distribution of patients that undergo THA. In the future, the results of a large-scale multi-national study on bearing surfaces must be reported. A more thorough understanding of the kinematics of the hip joint after THA is required since we do not completely understand this parameter in dynamic post-THA situation. Ongoing insights in hip kinematics4748) after THA may improve the prosthetic function and patient satisfaction as well as simplify bearing surface decisions while increasing component lifespan. Recent studies on the spinopelvic motion49505152) based on hip-spine relations53) also suggest that our conventional knowledge of risk factors for dislocation may need to be revisited. Acetabular cup implantation ideally must be adapted depending on spinopelvic interactions, thereby lowering the rate of impingement and subsequent dislocation.
The current THA design model is tailored to the lifestyle of patients in Western countries, and most research on health-related quality of life of THA patients is from Western countries. Future research should be focused on assessing the living conditions of THA patients and their lifestyles to improve health-related quality of life in Asian countries545556). Furthermore, a new scoring system, which can assess the clinical, functional and quality of life after THA should be addressed.

CONCLUSION

Although the early results of THAs have been disappointing, there have been steady advances in the last three decades. Currently, THA is a landmark surgery in modern medical history and clinical results from the last 30 years have proven THA to be one of the most successful surgical techniques for treating degenerative conditions of the hip. Our results support the use of current THA as a feasible option, even in young patients with severe symptoms.
The future of THA must not only focus on improvement in the physical function, but also the daily quality of life, including that of the Asian population. Dynamic kinematics after THA should be better understood, including its effect on each bearing surface. A new scoring system is needed, which can assess the clinical, functional and quality of life to improve outcomes after THA in the Asian population.

Figures and Tables

Fig. 1

Various cemented femoral stems. The Precoat Plus stem is located in the most left-hand corner of the image.

hp-31-179-g001
Fig. 2

(AD) Radiograph and computed tomography images 12-year post hybrid total hip arthroplasty in a 47-year-old male patient revealing excessive wear and pelvic lysis.

hp-31-179-g002
Fig. 3

Computed tomography image and anteroposterior hip radiography 24-years post hybrid total hip arthroplasty in a 55-year-old male patient revealing excessive polyethylene liner wear and periacetabular lysis in the left hip joint.

hp-31-179-g003
Fig. 4

(AE) Postoperative radiographic images. Arrows show distal cement mantle defects with osteolysis.

hp-31-179-g004
Fig. 5

Anteroposterior hip radiography 27-year post hybrid total hip arthroplasty in a 48-year-old female patient revealing excellent radiological results.

hp-31-179-g005
Fig. 6

Lysis free rate with time.

hp-31-179-g006
Fig. 7

A screenshot revealing osteolysis volume measurement on computed tomography using Rapidia 2.8 software.

hp-31-179-g007
Fig. 8

Anteroposterior hip radiography 17-year post bilateral hybrid total hip arthroplasty in a 50-year-old male patient revealing excellent radiological results.

hp-31-179-g008
Fig. 9

(AF) Retrieved polyethylene liners revealing rim damage extending across the entire rim.

hp-31-179-g009
Fig. 10

Anteroposterior hip radiography 4-year post ceramic-on-metal bearing total hip arthroplasty in a 51-year old male patient revealing excellent radiological results.

hp-31-179-g010
Fig. 11

(A, B) Retrieved acetabular and femoral components from a 43-year-old male patient revealing polyethylene liner damage by impingement between the rim of polyethylene liner and neck of the femoral stem.

hp-31-179-g011
Fig. 12

Hip anteroposterior radiographs (A, B) and a clinical photo (C) of a 32-year-old male patient revealing polyethylene liner wear and damages by impingement.

hp-31-179-g012
Table 1

The Range of Motion and Postoperative HHS between 36 mm COM Group and 28 mm Non-COM Group

hp-31-179-i001
Variable 36 mm COM group 28 mm non-COM group P-value
Flexion (°) 123.1±16.8 116.0±16.2 0.002
Abduction (°) 43.8±10.6 39.9±9.6 0.018
Adduction (°) 33.5±8.6 29.7±9.5 0.006
External rotation (°) 39.9±8.9 35.6±8.7 0.009
Internal rotation (°) 32.4±10.0 27.2±9.8 0.001
Total range of motion (°) 272.7±30.7 248.5±32.3 <0.001
Postoperative HHS (points) 95.4±5.4 93.2±6.4 0.013

Values are presented as mean±standard deviation.

HHS: Harris hip score, COM: ceramic-on-metal.

Table 2

The Clinical Outcomes between the Groups with Sit Cross-legged, Squat, and Kneel and Those without within the Same 36 mm Head

hp-31-179-i002
Variable Number Postoperative HHS (points) P-value
Sit cross-legged 0.012
 Negative 57 91.6
 Positive 129 96.0
Squat 0.001
 Negative 40 91.8
 Positive 149 95.6
Kneel 0.002
 Negative 25 90.1
 Positive 166 95.3

HHS: Harris hip score.

Notes

CONFLICT OF INTEREST The authors declare that there is no potential conflict of interest relevant to this article.

References

1. Law WA. Post-operative study of vitallium mould arthroplasty of the hip joint. J Bone Joint Surg Br. 1948; 30B:76–83.
crossref
2. Smith-Petersen MN. Evolution of mould arthroplasty of the hip joint. J Bone Joint Surg Br. 1948; 30B:59–75.
crossref
3. Charnley J. Anchorage of the femoral head prosthesis to the shaft of the femur. J Bone Joint Surg Br. 1960; 42:28–30.
crossref
4. Charnley J. Arthroplasty of the hip. A new operation. Lancet. 1961; 1:1129–1132.
crossref
5. Amstutz HC, Campbell P, Kossovsky N, Clarke IC. Mechanism and clinical significance of wear debris-induced osteolysis. Clin Orthop Relat Res. 1992; (276):7–18.
crossref
6. Chandler HP, Reineck FT, Wixson RL, McCarthy JC. Total hip replacement in patients younger than thirty years old. A five-year follow-up study. J Bone Joint Surg Am. 1981; 63:1426–1434.
crossref
7. Collis DK. Cemented total hip replacement in patients who are less than fifty years old. J Bone Joint Surg Am. 1984; 66:353–359.
crossref
8. Cooper RA, McAllister CM, Borden LS, Bauer TW. Polyethylene debris-induced osteolysis and loosening in uncemented total hip arthroplasty. A cause of late failure. J Arthroplasty. 1992; 7:285–290.
crossref
9. Goetz DD, Smith EJ, Harris WH. The prevalence of femoral osteolysis associated with components inserted with or without cement in total hip replacements. A retrospective matched-pair series. J Bone Joint Surg Am. 1994; 76:1121–1129.
crossref
10. Gruen TA, McNeice GM, Amstutz HC. “Modes of failure” of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res. 1979; (141):17–27.
11. Phillips FM, Pottenger LA, Finn HA, Vandermolen J. Cementless total hip arthroplasty in patients with steroid-induced avascular necrosis of the hip A 62-month follow-up study. Clin Orthop Relat Res. 1994; (303):147–154.
12. Salvati EA, Cornell CN. Long-term follow-up of total hip replacement in patients with avascular necrosis. Instr Course Lect. 1988; 37:67–73.
13. Reese A, Macaulay W. Hybrid total hip arthroplasty: stateof-the-art in the new millennium? J South Orthop Assoc. 2003; 12:75–78.
14. Suh DH, Yun HH, Chun SK, Shon WY. Fifteen-year results of precoated femoral stem in primary hybrid total hip arthroplasty. Clin Orthop Surg. 2013; 5:110–117.
crossref
15. Smith SE, Estok DM 2nd, Harris WH. Average 12-year outcome of a chrome-cobalt, beaded, bony ingrowth acetabular component. J Arthroplasty. 1998; 13:50–60.
crossref
16. Kim YH, Kim JS, Cho SH. Primary total hip arthroplasty with a cementless porous-coated anatomic total hip prosthesis: 10- to 12-year results of prospective and consecutive series. J Arthroplasty. 1999; 14:538–548.
crossref
17. Clohisy JC, Harris WH. The Harris-Galante porous-coated acetabular component with screw fixation. An average tenyear follow-up study. J Bone Joint Surg Am. 1999; 81:66–73.
crossref
18. Maloney WJ, Galante JO, Anderson M, et al. Fixation, polyethylene wear, and pelvic osteolysis in primary total hip replacement. Clin Orthop Relat Res. 1999; (369):157–164.
crossref
19. Ong A, Wong KL, Lai M, Garino JP, Steinberg ME. Early failure of precoated femoral components in primary total hip arthroplasty. J Bone Joint Surg Am. 2002; 84:786–792.
crossref
20. Charnley J. Total hip replacement by low-friction arthroplasty. Clin Orthop Relat Res. 1970; 72:7–21.
crossref
21. Sochart DH, Porter ML. The long-term results of Charnley low-friction arthroplasty in young patients who have congenital dislocation, degenerative osteoarthrosis, or rheumatoid arthritis. J Bone Joint Surg Am. 1997; 79:1599–1617.
crossref
22. Jones LC, Hungerford DS. Cement disease. Clin Orthop Relat Res. 1987; (225):192–206.
crossref
23. Chen FS, Di Cesare PE, Kale AA, et al. Results of cemented metal-backed acetabular components: a 10-year-average follow-up study. J Arthroplasty. 1998; 13:867–873.
24. Clarke IC, Good V, Anissian L, Gustafson A. Charnley wear model for validation of hip simulators--ball diameter versus polytetrafluoroethylene and polyethylene wear. Proc Inst Mech Eng H. 1997; 211:25–36.
crossref
25. Parks ML, Walsh HA, Salvati EA, Li S. Effect of increasing temperature on the properties of four bone cements. Clin Orthop Relat Res. 1998; (355):238–248.
crossref
26. Yun HH, Shon WY, Yoon JR, Yang JH, Lim DS. Reliability of a PowerPoint method for wear measurement after total hip arthroplasty: a retrieval study using 3-dimensional laser scanning. J Arthroplasty. 2012; 27:1530–1537.
crossref
27. Yun HH, Shon WY, Hong SJ, Yoon JR, Yang JH. Relationship between the pelvic osteolytic volume on computed tomography and clinical outcome in patients with cementless acetabular components. Int Orthop. 2011; 35:1453–1459.
crossref
28. Yun HH, Jajodia NK, Myung JS, Oh JK, Park SW, Shon WY. Use of slide presentation software as a tool to measure hip arthroplasty wear. J Arthroplasty. 2009; 24:1210–1215.
crossref
29. Shon WY, Gupta S, Biswal S, Han SH, Hong SJ, Moon JG. Pelvic osteolysis relationship to radiographs and polyethylene wear. J Arthroplasty. 2009; 24:743–750.
crossref
30. Suh DH, Han SB, Yun HH, Chun SK, Shon WY. Characterization of progression of pelvic osteolysis after cementless total hip arthroplasty: computed tomographic study. J Arthroplasty. 2013; 28:1851–1855.
crossref
31. Yi JW, Shon WY, Huh CY, Yun HH, Huh YJ. Follow-up study of the cemented polished femoral stem for more than five years. J Korean Hip Soc. 2011; 23:25–31.
crossref
32. McKellop H, Shen FW, Lu B, Campbell P, Salovey R. Development of an extremely wear-resistant ultra high molecular weight polyethylene for total hip replacements. J Orthop Res. 1999; 17:157–167.
crossref
33. Wroblewski BM, Siney PD, Fleming PA. Low-friction arthroplasty of the hip using alumina ceramic and cross-linked polyethylene. A 17-year follow-up report. J Bone Joint Surg Br. 2005; 87:1220–1221.
34. Shon WY, Baldini T, Peterson MG, Wright TM, Salvati EA. Impingement in total hip arthroplasty a study of retrieved acetabular components. J Arthroplasty. 2005; 20:427–435.
35. Firkins PJ, Tipper JL, Ingham E, Stone MH, Farrar R, Fisher J. A novel low wearing differential hardness, ceramic-on-metal hip joint prosthesis. J Biomech. 2001; 34:1291–1298.
crossref
36. Williams S, Schepers A, Isaac G, et al. The 2007 Otto Aufranc Award. Ceramic-on-metal hip arthroplasties: a comparative in vitro and in vivo study. Clin Orthop Relat Res. 2007; 465:23–32.
37. Bisschop R, Boomsma MF, Van Raay JJ, Tiebosch AT, Maas M, Gerritsma CL. High prevalence of pseudotumors in patients with a Birmingham Hip Resurfacing prosthesis: a prospective cohort study of one hundred and twenty-nine patients. J Bone Joint Surg Am. 2013; 95:1554–1560.
crossref
38. Bozic KJ, Browne J, Dangles CJ, et al. Modern metal-on-metal hip implants. J Am Acad Orthop Surg. 2012; 20:402–406.
crossref
39. Han SB, Oh JK, Jang WY, Choudhary N, Kim HG, Shon WY. Increased serum ion levels after ceramic-on-metal bearing total hip arthroplasty: influence of an Asian lifestyle. J Arthroplasty. 2018; 33:887–892.
crossref
40. Naik LG, Shon WY, Clarke IC, Moon JG, Mukund P, Kim SM. Pseudotumor and subsequent implant loosening as a complication of revision total hip arthroplasty with ceramic-on-metal bearing: a case report. Hip Pelvis. 2018; 30:276–281.
crossref
41. Pearson MJ, Grover LM, Lord JM, Jones SW, Davis ET. Bearings in hip arthroplasty: joint registries vs precision medicine: review article. HSS J. 2017; 13:20–27.
42. Kasina P, Wall A, Lapidus LJ, et al. Postoperative thromboprophylaxis with new oral anticoagulants is superior to LMWH in hip arthroplasty surgery: findings from the Swedish registry. Clin Orthop Relat Res. 2019; 477:1335–1343.
crossref
43. Heckmann ND, Sivasundaram L, Stefl MD, Kang HP, Basler ET, Lieberman JR. Total hip arthroplasty bearing surface trends in the United States from 2007 to 2014: the rise of ceramic on polyethylene. J Arthroplasty. 2018; 33:1757–1763.e1.
crossref
44. Rajaee SS, Theriault RV, Pevear ME, Smith EL. National trends in primary total hip arthroplasty in extremely young patients: a focus on bearing surface usage from 2009 to 2012. J Arthroplasty. 2016; 31:9 Suppl. 63–68.
crossref
45. Fawsitt CG, Thom HHZ, Hunt LP, et al. Choice of prosthetic implant combinations in total hip replacement: cost-effectiveness analysis using UK and Swedish hip joint registries data. Value Health. 2019; 22:303–312.
crossref
46. Yoon PW, Yoo JJ, Kim Y, Yoo S, Lee S, Kim HJ. The epidemiology and national trends of bearing surface usage in primary total hip arthroplasty in Korea. Clin Orthop Surg. 2016; 8:29–37.
crossref
47. Rivière C, Harman C, Parsons T, Villet L, Cobb J, Maillot C. Kinematic alignment versus conventional techniques for total hip arthroplasty: a retrospective case control study. Orthop Traumatol Surg Res. 2019; 105:895–905.
crossref
48. Komiyama K, Hamai S, Hara D, et al. Dynamic hip kinematics during squatting before and after total hip arthroplasty. J Orthop Surg Res. 2018; 13:162.
crossref
49. Lum ZC, Coury JG, Cohen JL, Dorr LD. The current knowledge on spinopelvic mobility. J Arthroplasty. 2018; 33:291–296.
crossref
50. Stefl M, Lundergan W, Heckmann N, et al. Spinopelvic mobility and acetabular component position for total hip arthroplasty. Bone Joint J. 2017; 99:1 Supple A. 37–45.
crossref
51. Shon WY, Sharma V, Keon OJ, Moon JG, Suh DH. Can pelvic tilting be ignored in total hip arthroplasty? Int J Surg Case Rep. 2014; 5:633–636.
crossref
52. Shon WY, Gupta S, Biswal S, et al. Validation of a simple radiographic method to determine variations in pelvic and acetabular cup sagittal plane alignment after total hip arthroplasty. Skeletal Radiol. 2008; 37:1119–1127.
crossref
53. Lazennec JY, Charlot N, Gorin M, et al. Hip-spine relationship: a radio-anatomical study for optimization in acetabular cup positioning. Surg Radiol Anat. 2004; 26:136–144.
crossref
54. Tang H, Du H, Tang Q, Yang D, Shao H, Zhou Y. Chinese patients’ satisfaction with total hip arthroplasty: what is important and dissatisfactory? J Arthroplasty. 2014; 29:2245–2250.
crossref
55. Fujita K, Kang HS, Mawatari M, Makimoto K, Lee M, Hwang J. Quality of life, effects on Asian lifestyle, and perceived satisfaction after total hip arthroplasty in Japan and Korea. Int J Orthop Trauma Nurs. 2018; 31:20–25.
crossref
56. Fujita K, Xia Z, Liu X, Mawatari M, Makimoto K. Lifestyle and health-related quality of life in Asian patients with total hip arthroplasties. Nurs Health Sci. 2014; 16:365–372.
crossref
TOOLS
ORCID iDs

Ho Hyun Yun
https://orcid.org/0000-0003-3626-9047

Similar articles