Cho, Jung, and Kang: A Clinical Study on Intracompartmental Pressure of Leg Using Slit Catheter

Abstract

Intracompartmental pressure of leg was measured by the slit catheter technique in one hundred fifty compartments of one hundred cases; fifty were from healthy young male adults, the others were from tibial fractured patients during three years from March 1983 to Fobruary 1986. The results were as follows; 1. Slit-catheter technique was found as a accurate method for continuous pressure monitoring. 2. The range of normal pressure was from zero to six millimeters of mercury(1.45mmHg±0.85). 3. Tissue pressure was 0–15mmHg(4.03mmHg±1.83) in unaffected side of fractured patients, 4–35mmHg(14.65mmHg±2.05) in affected side of fractured patients.