Journal List > J Korean Med Sci > v.32(10) > 1108406

Lee, Lim, and Kim: Validation of the World Health Organization Enzyme-Linked Immunosorbent Assay for the Quantitation of Immunoglobulin G Serotype-Specific Anti-Pneumococcal Antibodies in Human Serum

Abstract

The World Health Organization (WHO) enzyme-linked immunosorbent assay (ELISA) guideline is currently accepted as the gold standard for the evaluation of immunoglobulin G (IgG) antibodies specific to pneumococcal capsular polysaccharide. We conducted validation of the WHO ELISA for 7 pneumococcal serotypes (4, 6B, 9V, 14, 18C, 19F, and 23F) by evaluating its specificity, precision (reproducibility and intermediate precision), accuracy, spiking recovery test, lower limit of quantification (LLOQ), and stability at the Ewha Center for Vaccine Evaluation and Study, Seoul, Korea. We found that the specificity, reproducibility, and intermediate precision were within acceptance ranges (reproducibility, coefficient of variability [CV] ≤ 15%; intermediate precision, CV ≤ 20%) for all serotypes. Comparisons between the provisional assignments of calibration sera and the results from this laboratory showed a high correlation > 94% for all 7 serotypes, supporting the accuracy of the ELISA. The spiking recovery test also fell within an acceptable range. The quantification limit, calculated using the LLOQ, for each of the serotypes was 0.05–0.093 μg/mL. The freeze-thaw stability and the short-term temperature stability were also within an acceptable range. In conclusion, we showed good performance using the standardized WHO ELISA for the evaluation of serotype-specific anti-pneumococcal IgG antibodies; the WHO ELISA can evaluate the immune response against pneumococcal vaccines with consistency and accuracy.

Graphical Abstract

jkms-32-1581-abf001

INTRODUCTION

Streptococcus pneumoniae are encapsulated bacteria characterized by a polysaccharide (PS) capsule, which is a major virulence factor (1). S. pneumoniae is a predominant cause of many diseases, including meningitis, sepsis, pneumonia, and otitis media in young infants and the elderly (12). With the significant increase in multi-drug resistant S. pneumoniae worldwide, many efforts have been made to promote prevention of the disease and development of effective vaccines.
The 14-valent pneumococcal PS vaccine (PPSV) was licensed in 1977 (3) and thereafter, the 23-valent PPSV was developed in 1983. These vaccines induce T-cell-independent immune responses and are more effective in adults than in young children, who are the highest risk population for pneumococcal infection (4). To induce immunity in young children, the 7-valent pneumococcal conjugate vaccine (PCV7) containing serotypes 4, 6B, 9V, 14, 18C, 19F, and 23F was developed. PCV7 was shown to be highly effective for the prevention of invasive pneumococcal disease in infants and children (56).
The efficacy of a vaccine is determined by the difference in the incidence of specific diseases among vaccinated and unvaccinated subjects for the disease. However, once a vaccine is proven to be effective, efficacy trials are no longer ethical or feasible to perform. Since protection against pneumococcal infection is mediated by antibodies specific to the capsular PS serotype, an enzyme-linked immunosorbent assay (ELISA) that measures immunoglobulin G (IgG) levels is used in the evaluation of pneumococcal vaccines as a surrogate assay. The World Health Organization (WHO) has provided a laboratory guideline for the ELISA to quantify IgG antibodies specific to pneumococcal PS (PnPS) since 2000 (78). In addition, the National Institute for Biological Standards Control (NIBSC) in the UK assembled 12 calibration serum samples with assigned antibody titers to assist laboratories working on the evaluation of pneumococcal vaccines (9).
The goal of this study was to validate the WHO ELISA for the serotype-specific anti-pneumococcal IgG at the Ewha Center for Vaccine Evaluation and Study. The results were anticipated to serve as evidence of the capability and accuracy of the evaluation of pneumococcal vaccines at our center.

MATERIALS AND METHODS

Materials

Study design

This was a validation study of the WHO ELISA for the detection of the serotype-specific anti-pneumococcal IgG at the Ewha Center for Vaccine Evaluation and Study.

Study supplies

The serum reference assay standard, 89-SF, was obtained from Dr. C. Frasch (Center for Biologics Evaluation and Research, Food and Drug Administration, Bethesda, MD, USA). The 7 serotype specific PS antigens (4, 6B, 9V, 14, 18C, 19F, and 23F) were purchased from the American Type Culture Collection (ATCC, Rockville, MD, USA). Cell wall PS (C-PS) used for the elimination of nonspecific reactivity was purchased from Statens Serum Institute, Copenhagen, Denmark. The PnPS 22F, which was selected for better absorption of the common nonspecific epitopes, was purchased from ATCC. For the test sera, 5 out of 12 pneumococcal ELISA calibration sera (748, 752, 754, 760, and 764, NIBSC, Hertfordshire, UK), which had been provisionally assigned for the detection of IgG concentrations for the 7 serotypes, were used.
The serum reference assay standard, 89-SF, was assigned as 4.1 μg/mL for PnPS 4, 16.9 μg/mL for PnPS 6B, 6.9 μg/mL for PnPS 9V, 27.8 μg/mL for PnPS 14, 4.5 μg/mL for PnPS 18C, 13.0 μg/mL for PnPS 19F, and 8.1 μg/mL for PnPS 23F (7). 89-SF for each of the serotypes was prepared at a 1:1,000 dilution with absorption solutions including only C-PS. The calibration sera 748, 752, 754, 760, and 764 were prepared at 1:250 dilutions with absorption solutions with C-PS and PS type 22F. The diluted sera were used at 2.5-fold serial dilutions up to 7 times with dilution ranges from 1/250 to 1/24,414. Each PnPS was prepared at a concentration of 2 mg/mL and was stored at −20°C. These were used in a concentration of a 5 μg/mL with phosphate-buffered saline buffer (containing both 0.02% sodium azide [NaN3] and 0.05% Tween-20).

ELISA

The standardized ELISA method for quantification of anti-PnPS antibodies, was conducted as previously reported, which can be located at http://www.vaccine.uab (89). Briefly, optimal concentrations of the 7 PnPS (1 μg/mL for 14 and 18C; 2 μg/mL for 4 and 9V; 10 μg/mL for 6B, 19F, and 23F) were coated at 37°C for 5 hours per 100 μL in 96 well medium binding polystyrene microtiter plate (Corning & Costar, Corning, NY, USA). Sera at 1:1,000 dilutions of 89-SF and 1:250 dilutions of calibration sera were pre-absorbed for 30 minutes at room temperature with each prepared solution, and were then diluted by 2.5-fold serial dilution. Coated plates with each PnPS were washed 5-times using a 30 to 60 seconds soak during the first wash with Tris-buffered solution (TBS) including 0.1% Brij™ 35 solution (Thermo Fisher Scientific, Waltham, MA, USA). Fifty microliters of all sera were added to the washed plates and incubated for 2 hours at room temperature. To each microtiter plate well, 100 μL of 1:2,000 diluted goat anti-human IgG alkaline phosphates conjugate (Southern Biotech, Birmingham, AL, USA) were added and incubated for 2 hours at room temperature. Next, 100 μL of a 1 mg/mL concentration of P-nitrophenyl phosphate in diethanolamine-MgCl2 buffer, pH 9.8, was added to all wells for 1 hour. The reaction was stopped by the addition of 50 μL of 3M sodium hydroxide (NaOH) per well. Absorbance readings were obtained with dual wavelength readings at 405 nm (experimental) and 690 nm (reference) on a microtiter plate reader.

ELISA validation test

All the tests for the validation were done in concordance with the bioanalytical method validation of Center for Drug Evaluation and Research (CDER) and the pneumococcal ELISA standardization-calibration sera of the NIBSC, UK written in April 26, 2001 (8). The overall acceptable ranges were defined as laboratory-determined antibody values of ELISA calibration sera with no more than 40% error related to the assigned values.

Specificity

Specificity was defined as the ability to selectively and precisely evaluate the anti-pneumococcal IgG antibodies regardless of the presence of other components in the testing sample. To evaluate specificity, 8 plates were analyzed; 7 plates coated with 1 of 7 PnPS (4, 6B, 9V, 14, 18C, 19F, and 23F) and 1 plate with no coated antigen. The 89-SF and calibration sera were analyzed for the 8 plates and optical density (OD) was analyzed at different sera dilutions; the results were compared between the antigen-coated plates and antigen-free plate for each PnPS.

Precision

The precision was defined as the closeness of test results obtained by the values of the same analytes in a short period under the same conditions, evaluated using reproducibility and intermediate precision. For reproducibility, all 5 calibration sera were evaluated in 5 separate assays for all 7 serotypes and the coefficient of variability (CV) was assessed between the 5 assays. For intermediate precision, each serum was analyzed 7 times in a single plate for each serotype and the CV was assessed between the results of the assays.

Accuracy

Accuracy was defined as the degree of agreement of a measured quantity to its actual value. The antibody titers for each serotype were compared with provisional assignments of ELISA calibration sera. Accuracy was evaluated by the results of 24 tests analyses for 7 anti-PnPS.

Spiking recovery test

The spiking recovery test, a method to evaluate the accuracy, was analyzed by adding a known amount of analyte to the reference serum. Recovery was calculated by comparing the observed concentration as determined by ELISA and the expected concentration by summation. Acceptable ranges for recovery were defined as 100% ± 40% between results of multiple analyses at different dilution concentrations.
In this analysis, calibration serum 742 was used. The assigned values of calibration serum 742 were 6.2 μg/mL for PnPS 4, 9.9 μg/mL for PnPS 6B, 2.1 μg/mL for PnPS 9V, 7.4 μg/mL for PnPS 14 and 18C, 11.5 μg/mL for PnPS 19F, and 0.9 μg/mL for PnPS 23F. The calibration sera (serum 742) were diluted with 3 different dilutions of 1/40, 1/80, and 1/160. The diluted sera at each different dilution were mixed with a selected concentration of calibration sera, which was the lowest antibody concentration serum among 5 different calibration sera for each serotype.

Lower limit of quantification (LLOQ)

The LLOQ evaluated the accuracy and precision of the lowest concentration of serum that could be quantified. The quantitation limit for the type specific serotypes was calculated with the detection limit multiplied by 3, which was estimated in every analysis.

Stability

The stability of analytes evaluated the consistency of results in conditions likely to be encountered during sample handling and analysis. For the test, all 5 calibration sera were used.
The freeze-thaw stability of analytes was tested to evaluate the effect of the freeze thaw cycles of the analytes. The 5 calibration sera repeated the freeze thaw cycle from 1 to 4 times where sera were kept at room temperature for 30 minutes after stored in −20°C for more 12 hours.
The short-term temperature stability of analytes was tested to evaluate the effect of the kept time of analytes. The 5 calibration sera were kept at room temperature for 1, 2, and 4 hours.

Statistical analysis

OD was measured using SoftMax Pro 4.8 program of Molecular Devices Corporation (Sunnyvale, CA, USA). The OD was converted to antibody concentrations with a computer program for pneumococcal antibody ELISA that was developed by Mr. Brian Plikaytis at the Center for Disease Control and Prevention (CDC) in Atlanta, Georgia. The CV, standard deviation (SD), and mean OD for the standard of the each plate were calculated using standardized curve-fitting 4 parameter logistic equation (4PL equation). The CV, recovery, and the relativity of antibody concentrations were calculated to evaluate the validity.
The results were considered acceptable if they were within 40% error of the assigned value for the sera. For each of the validation assessments, acceptable ranges for reproducibility and intermediate precision were defined as CV ≤ 15% and CV ≤ 20%, respectively. Spiking recovery tests were analyzed by comparing observed concentrations with expected concentrations; they were deemed acceptable if test results were within 100% ± 40%. Stability (freeze thaw stability and short term temperature stability) was considered acceptable when CV ≤ 20%.

Ethics statement

This was a validation study on an analytical lab test (anti-IgG pneumococcal ELISA), using reference and calibration serum samples which are widely distributed by the FDA and NIBSC, UK. As the objective of this study was to validate results of the assay, and not directly analyze serum samples, this work did not receive an Institutional Review Board approval. Also as the serum samples were not directly obtained from human subjects in our study, there was no need for informed consent.

RESULTS

Specificity

The specificity of the pneumococcal ELISA was evaluated for 7 serotypes to PnPS antigens with optimal coating concentrations and with a blank plate with no antigen coating. The 89-SF and 5 calibration sera showed that the OD decreased with dilution. The value of the 5 calibration sera corresponded to each provisional IgG value. The non-coating wells were estimated with a similar level of blank wells regardless of the dilution fold and PnPS antigen serotype (Fig. 1).
Fig. 1
The specificity of anti-PnPS ELISA with the serum reference assay standard, 89-SF for 7 anti-PnPS types. Non-coating plate reference serum, 89-SF wells were coated with coating solution only.
ELISA = enzyme-linked immunosorbent assay, PnPS = pneumococcal polysaccharide, OD = optical density.
jkms-32-1581-g001

Precision

The reproducibility measures were generally within the acceptable range with the exception of the calibration serum 764 for PnPS 9V (Table 1).
Table 1

Reproducibility precision results of 5 calibration sera for 7 anti-PnPS (acceptable range: CV ≤ 15%)

jkms-32-1581-i001
Serum PnPS
4 6B 9V 14 18C 19F 23F
748 Mean 2.19 1.15 3.98 11.54 11.30 13.81 7.92
CV (%) 3.09 3.38 2.51 5.76 1.81 4.62 2.29
752 Mean 8.29 14.15 21.71 26.72 10.13 50.84 11.6
CV (%) 3.83 3.85 2.93 3.87 4.55 5.03 4.50
754 Mean 10.23 7.40 13.72 140.41 4.54 13.04 8.18
CV (%) 1.87 7.02 3.93 1.96 3.42 6.94 3.29
760 Mean 2.01 1.34 1.09 17.04 3.08 6.63 1.68
CV (%) 1.65 6.58 5.62 6.02 8.24 5.73 3.74
764 Mean 4.27 23.80 15.23 17.86 7.49 18.41 27.15
CV (%) 3.92 5.87 17.20 5.08 2.29 2.54 2.19
PnPS = pneumococcal polysaccharide, CV = coefficient of variability.
The intermediate precision measures were within the acceptable range when compared with the provisional assigned values. When comparing CV, the measures were within the acceptable range with the exceptions of calibration serum 752 for PnPS 14, calibration serum 760 for PnPS 9V, and calibration serum 764 for PnPS 4 and PnPS 6B (Table 2).
Table 2

Intermediate precision results of 5 calibration sera for 7 anti-PnPS (acceptable range: CV ≤ 20%)

jkms-32-1581-i002
Serum PnPS
4 6B 9V 14 18C 19F 23F
748 Mean 2.57 1.00 4.09 11.95 12.12 14.11 9.17
CV (%) 6.72 9.76 4.78 13.24 6.28 1.88 13.60
752 Mean 9.42 13.30 20.43 26.87 11.14 49.17 9.28
CV (%) 10.16 12.22 13.41 28.32 9.10 9.14 10.12
754 Mean 12.14 6.15 12.47 135.56 4.78 14.54 7.84
CV (%) 8.46 11.31 12.94 15.76 7.77 11.46 13.01
760 Mean 2.23 1.37 1.03 15.06 2.93 5.82 1.59
CV (%) 18.44 10.32 22.41 5.59 2.10 13.78 12.80
764 Mean 3.79 22.64 5.07 16.71 7.56 17.18 28.02
CV (%) 23.95 4.31 20.32 7.24 10.48 18.36 10.63
PnPS = pneumococcal polysaccharide, CV = coefficient of variability.

Accuracy

Accuracy was evaluated by measuring the correlation between the provisional assigned values and the obtained results from 24 runs for 7 serotypes. There were good correlations for all serotypes including PnPS 4, 14, 18C, 19F (r = 0.99), PnPS 6B (r = 0.97), PnPS 9V (r = 0.96), and PnPS 23F (r = 0.94).

Spiking recovery test

The calibration serum 742 was used to test for 7 serotypes. The results were within the acceptable range for 5 serotypes, with the exception of PnPS 6B and 19F (Table 3).
Table 3

Spiking recovery* test of calibration serum 742 with 1/250 dilution and with the lowest antibody concentration serum in the middle of 5 sera tested in this assay for each anti-PnPS (acceptable range: 100% ± 40%)

jkms-32-1581-i003
Spike PnPS
4 6B 9V 14 18C 19F 23F
Low spike (1/40) 74.1 154.5 134.8 106.2 76.9 165.6 108.6
Medium spike (1/80) 70.5 126.9 133.4 103.5 70.9 139.8 120.3
High spike (1/160) 73.7 138.8 101.1 117.4 73.4 143.1 95.9
PnPS = pneumococcal polysaccharide.
*Recovery (%) = (observed concentration/expected concentration) × 100.

LLOQ

The LLOQ was calculated by multiplying the detection limit, measured by the ELISA, by 3. The results were 0.05 μg/mL for PnPS 4, 0.086 μg/mL for PnPS 6B, 0.085 μg/mL for PnPS 9V, 0.073 μg/mL for PnPS 14, 0.058 μg/mL for PnPS 18C, 093 μg/mL for PnPS 19F, and 0.051 μg/mL for PnPS 23F.

Stability

Freeze-thaw stability

The 5 reference sera were subjected to 4 freeze-thaw cycles for 7 serotypes. The results ranged from 86.61% to 115.59%, within the acceptable range when compared to result from the first cycle (Table 4).
Table 4

FT stability of 5 calibration sera for 7 anti-PnPS (acceptable range: within ± 20%)

jkms-32-1581-i004
PnPS FT cycle*
Serum 748 Serum 752
1 2 3 4 1 2 3 4
4 100.00 94.67 106.67 107.42 100.00 91.52 108.30 105.79
6B 100.00 97.88 100.31 105.56 100.00 98.19 100.29 104.48
9V 100.00 102.72 99.56 104.10 100.00 92.01 103.56 107.22
14 100.00 93.94 96.08 105.19 100.00 92.39 102.40 104.52
18C 100.00 94.86 100.90 98.58 100.00 89.58 103.13 106.08
19F 100.00 99.82 101.15 104.60 100.00 91.05 105.21 108.03
23F 100.00 97.33 99.50 106.22 100.00 93.79 106.82 100.72
FT = freeze thaw, PnPS = pneumococcal polysaccharide.
*Freeze thaw cycle: sera were subjected to 4 freeze thaw cycles. Relative to FT 1 (%) = (based upon mean value) FT 2/FT 1 × 100.

Short-term temperature stability

The 5 sera were placed at room temperature for 1, 2, and 4 hours and were then assessed using ELISA. The results ranged from 81.3% to 120.0%, within the acceptable range when compared to result of the 5 sera placed at room temperature for 1 hour (Fig. 2).
Fig. 2
Short-term temperature stability of 5 calibration sera for 7 PnPS (acceptable range: within 100% ± 20%). For analysis of short term temperature, sera were left at room temperature for 1, 2, and 4 hours prior to analysis.
PnPS = pneumococcal polysaccharide.
jkms-32-1581-g002

Percent error relative to the assigned values

When analyzing all the results from each of the validation analyses, the results were within 40% error for the majority of the antigens, with the exception of PnPS 6B and 23F for 2 and 1 calibration sera, respectively (Table 5).
Table 5

Performance of Ewha Center for Vaccine Evaluation and Study for the analysis of 5 calibration sera-shown is the number of sera whose assayed value has a percent error ≤ 40%

jkms-32-1581-i005
Value PnPS
4 6B 9V 14 18C 19F 23F All types
No. of serum samples (%) 5/5 (100) 3/5 (60) 5/5 (100) 5/5 (100) 5/5 (100) 5/5 (100) 4/5 (80) 32/35 (91)
PnPS = pneumococcal polysaccharide.

DISCUSSION

In this study, we evaluated the validity of ELISA for quantitation of serotype specific anti-pneumococcal IgG antibodies at the Ewha Center for Vaccine Evaluation and Study. The specificity, precision, accuracy, spiking recovery test, LLOQ, and stability were analyzed for 7 serotypes (4, 6B, 9V, 14, 18C, 19F, 23F) included in the PCV7. The serum standard, 89-SF, and 5 calibration sera showed specificity in the detection of the IgG for the 7 PnPS antigens. The specificity is an important measure reflecting the ability to inhibit non-specific reactions and allowing for the analysis of the presence of precise anti-pneumococcal antibodies in a patient's serum (8910).
The reproducibility and intermediate precision analyses were generally within acceptable ranges (reproducibility, CV ≤ 15%; intermediate precision, CV ≤ 20%) for all 7 serotypes. The accuracy was proven by showing a high correlation when compared with the provisional antibody values of the calibration sera. The results of the spiking recovery test were also acceptable with the exception of PnPS 6B and 19F. However, previous studies have shown that there is variation within antibody values for the serotypes PnPS 6B, 9V, 19F, and 23F (11). According to the reference report of the pneumococcal ELISA standardization-calibration sera, 5 laboratories showed different results for each serotype and serum; disparity in PnPS 6B values between laboratories were speculated in that study (11). Our findings were consistent with results in the previous study. The LLOQ ranged from 0.05 to 0.093 μg/mL for the specific PS antigens, which are much lower than the surrogate for protection against pneumococcal diseases (0.2–0.35 μg/mL). The stability results were within the acceptable ranges, proving that measures were very stable during serum handling and storage, the freeze and thaw cycles, and the analytical process. Consequently, the validity for the evaluation of 7 pneumococcal specific anti-PnPS IgG using ELISA, conducted at the Ewha Center for Vaccine Evaluation and Study, was confirmed.
The ELISA is a sensitive and the plausible method in the biological field. Although the WHO ELISA protocol for anti-pneumococcal IgG antibody analysis is well described, each lab that performs the assay should be well validated. As a result, 12 research centers have suggested the need for guidelines detailing analysis methods for the antibody concentration and pneumococcal ELISA study using quality control sera (12). Provisional assignments of 12 calibration sera were established for each specific PnPS antigen. The Center for Biologics Evaluation and Research and CDC initially suggested that at least 85% of values for each serotype should have no greater than 40% error relative to the provisional assigned value (13). However, it was later agreed that when evaluating anti-pneumococcal IgG antibodies using ELISA, it is acceptable if 9 of the 12 sera are no greater than the 40% error criteria. In this study, although we used only 5 calibration sera, the results were within the acceptable criteria for most antigens with the exception of PnPS 6B and 23F. PnPS 6B relatively differs from other serotypes and the results in this study showed correspondence to results ascertained from other laboratories (111213).
After the validation assay was established at the Ewha Center for Vaccine Evaluation and Study, the IgG serotype-specific anti-pneumococcal antibody ELISA has been widely applied on various studies including infants (14) and also elderly subjects (15). This utility of ELISA for the quantitation of IgG serotype-specific anti-pneumococcal antibodies in human serum is not only limited to evaluating immunogenicity to vaccines, but also has been used to evaluate patients suspected with immune abnormalities (1617). Recently, it was used in the diagnosis of CD21 deficiency, which is a type of primary immunodeficiency; it showed impaired antibody responses to polysaccharides but intact antibody responses to protein antigens (18). Thus, there is a need for the anti-IgG pneumococcal ELISA in various settings, all of which need accurate, reproducible, and well-validated analysis methods.
This study was conducted for 7 pneumococcal capsular serotypes. Further studies are needed for more serotypes using the new reference standard serum, 007sp, which has replaced 89-SF (1920).
In conclusion, the Ewha Center for Vaccine Evaluation and Study shows good performance in the use of the standardized WHO ELISA for serotype specific anti-pneumococcal IgG antibodies. The ELISA was effective in evaluating the immune response against pneumococcal vaccines with great consistency and accuracy.

ACKNOWLEDGMENT

We thank Dr. Moon H. Nahm at the University of Alabama at Birmingham, Birmingham AL, USA for the assistance with the establishment of the ELISA and Dr. Carl Frasch at the Center for Biologics Evaluation and Research, Food and Drug Administration, Bethesda, MD. USA, for providing us with the serum reference assay standard 89-SF.

Notes

Funding This work was funded by the Ministry of Food and Drug Safety (11172MFDS360) to KHK.

DISCLOSURE The authors have no potential conflicts of interest to disclose.

Author Contributions

  • Conceptualization: Kim KH.

  • Investigation: Lim SY.

  • Supervision: Kim KH.

  • Writing - original draft: Lee H, Lim SY.

References

1. Brooks GF, Carroll KC, Butel JS, Morse SA. The streptococci. In : Brooks GF, Butel JS, Morse SA, editors. Jawetz, Melnick, & Adelberg's Medical Microbiology. 23rd ed. New York, NY: McGraw-Hill;2004. p. 231–247.
2. Zenni MK, Cheatham SH, Thompson JM, Reed GW, Batson AB, Palmer PS, Holland KL, Edwards KM. Streptococcus pneumoniae colonization in the young child: association with otitis media and resistance to penicillin. J Pediatr. 1995; 127:533–537.
3. Leinonen M, Säkkinen A, Kalliokoski R, Luotonen J, Timonen M, Mäkelä PH. Antibody response to 14-valent pneumococcal capsular polysaccharide vaccine in pre-school age children. Pediatr Infect Dis. 1986; 5:39–44.
4. Mufson MA, Hughey D, Lydick E. Type-specific antibody responses of volunteers immunized with 23-valent pneumococcal polysaccharide vaccine. J Infect Dis. 1985; 151:749–750.
5. Shinefield HR, Black S. Efficacy of pneumococcal conjugate vaccines in large scale field trials. Pediatr Infect Dis J. 2000; 19:394–397.
6. Oosterhuis-Kafeja F, Beutels P, Van Damme P. Immunogenicity, efficacy, safety and effectiveness of pneumococcal conjugate vaccines (1998–2006). Vaccine. 2007; 25:2194–2212.
7. Quataert SA, Kirch CS, Wiedl LJ, Phipps DC, Strohmeyer S, Cimino CO, Skuse J, Madore DV. Assignment of weight-based antibody units to a human antipneumococcal standard reference serum, lot 89-S. Clin Diagn Lab Immunol. 1995; 2:590–597.
8. Wernette CM, Frasch CE, Madore D, Carlone G, Goldblatt D, Plikaytis B, Benjamin W, Quataert SA, Hildreth S, Sikkema DJ, et al. Enzyme-linked immunosorbent assay for quantitation of human antibodies to pneumococcal polysaccharides. Clin Diagn Lab Immunol. 2003; 10:514–519.
9. Concepcion NF, Frasch CE. Pneumococcal type 22F polysaccharide absorption improves the specificity of a pneumococcal-polysaccharide enzyme-linked immunosorbent assay. Clin Diagn Lab Immunol. 2001; 8:266–272.
10. Coughlin RT, White AC, Anderson CA, Carlone GM, Klein DL, Treanor J. Characterization of pneumococcal specific antibodies in healthy unvaccinated adults. Vaccine. 1998; 16:1761–1767.
11. Center for Drug Evaluation and Research (US). Center for Veterinary Medicine (US). Guidance for Industry: Bioanalytical Method Validation. Silver Spring, MD: Food and Drug Administration;2001.
12. Frasch CE, Plikaytis B, Carlone G, Goldblatt D. Pneumococcal ELISA standardization: Calibration sera [Internet]. accessed on 24 May 2017. Available at https://www.vaccine.uab.edu/.
13. Plikaytis BD, Goldblatt D, Frasch CE, Blondeau C, Bybel MJ, Giebink GS, Jonsdottir I, Käyhty H, Konradsen HB, Madore DV, et al. An analytical model applied to a multicenter pneumococcal enzyme-linked immunosorbent assay study. J Clin Microbiol. 2000; 38:2043–2050.
14. Lee HJ, Park SE, Kim KH. Immune response to 19A serotype after immunization of 19F containing pneumococcal conjugate vaccine in Korean children aged 12–23 months. Korean J Pediatr. 2011; 54:163–168.
15. Lee H, Nahm MH, Kim KH. The effect of age on the response to the pneumococcal polysaccharide vaccine. BMC Infect Dis. 2010; 10:60.
16. Ambrosino DM. Impaired polysaccharide responses in immunodeficient patients: relevance to bone marrow transplant patients. Bone Marrow Transplant. 1991; 7:Suppl 3. 48–51.
17. Paris K, Sorensen RU. Assessment and clinical interpretation of polysaccharide antibody responses. Ann Allergy Asthma Immunol. 2007; 99:462–464.
18. van Montfrans JM, Hoepelman AI, Otto S, van Gijn M, van de Corput L, de Weger RA, Monaco-Shawver L, Banerjee PP, Sanders EA, Jol-van der Zijde CM, et al. CD27 deficiency is associated with combined immunodeficiency and persistent symptomatic EBV viremia. J Allergy Clin Immunol. 2012; 129:787–793.e6.
19. Goldblatt D, Plikaytis BD, Akkoyunlu M, Antonello J, Ashton L, Blake M, Burton R, Care R, Durant N, Feavers I, et al. Establishment of a new human pneumococcal standard reference serum, 007sp. Clin Vaccine Immunol. 2011; 18:1728–1736.
20. Belmonti S, Lombardi F, Morandi M, Fabbiani M, Tordini G, Cauda R, De Luca A, Di Giambenedetto S, Montagnani F. Evaluation and optimization of an ELISA procedure to quantify antibodies against pneumococcal polysaccharides included in the 13-valent conjugate vaccine. J Immunoassay Immunochem. 2016; 37:189–200.
TOOLS
ORCID iDs

Hyunju Lee
https://orcid.org/0000-0003-0107-0724

Soo Young Lim
https://orcid.org/0000-0003-1826-2474

Kyung-Hyo Kim
https://orcid.org/0000-0002-0333-6808

Similar articles