Im, Lee, Min, Bae, Lee, and Sohn: Revision Total Hip Arthroplasty after Ceramic Bearing Fractures in Patients Under 60-years Old; Mid-term Results

Abstract

Purpose

Although advances in technology have reduced the risk of ceramic implant fractures in total hip arthroplasty, these injuries do occur and their treatment remains challenging. There is a lack of studies reporting on the effectiveness of ceramic components in revision hip arthroplasty after ceramic bearing fracture. The aim of this study is to evaluate clinical and radiologic outcomes of revision surgery with ceramic-on-ceramic components after ceramic bearing fractures in young (i.e., under 60 years old) and active patients.

Materials and Methods

Eight patients who, from May 2004 to November 2011, underwent ceramic-on-ceramic revision surgery following a ceramic component fracture and had more than 6 years follow up were enrolled in this study. All eight patients were male with mean ages at first and revision surgeries of 39 years (range, 31–50 years) and 43.8 years (range, 33–60 years), respectively. There were 6 and 2 cases of ceramic liner and ceramic head fractures, respectively. The average time from the first operation to revision surgery was 54.3 months (range, 9–120 months), and the average follow up period was 9.7 years (range, 6–13.3 years).

Results

At the last follow up, all patients showed improvement in Harris hip score and pain relief and there were no cases of loosening or osteolysis.

Conclusion

Revision total hip arthroplasty using ceramic-on-ceramic components after ceramic component fracture is a feasible and appropriate surgical option in young and active patients.

INTRODUCTION

Total hip arthroplasty (THA) has been successfully used for the treatment of end-stage hip disease; however, THA failures have occurred in the long term, most commonly due to aseptic loosening caused by periprosthetic osteolysis1). Since the leading cause of loosening has been identified as polyethylene wear debris of acetabular prostheses after THA using metal-on-polyethylene articulations, studies on alternative bearing surfaces are actively being pursued2).
Aseptic loosening and subsequent periprosthetic osteolysis have emerged as serious problems, particularly in young and active patients who have a longer life expectancy and use an excessive range of hip motion345). More wear-resistant ceramic-on-ceramic (CoC) articulations and alumina-on-alumina articulations with a lower coefficient of friction (0.09) compared with metal-on-polyethylene articulations (0.21) have drawn attention as attractive new alternatives for younger and more active patients based on their ability to avoid or significantly reduce problems associated with wear debris particles from polyethylene components67). Although the fracture rate of third-generation ceramic components has been reduced to 0.004%, concerns about the safety of CoC bearings remain because fractures of ceramic liners or heads have been reported after THA using CoC articulations8910).
For management of ceramic component fractures, complications (e.g., articulation wear and metallosis) have been reported when the fractured ceramic heads are replaced with metal-on-polyethylene bearings. A new stem and ceramic bearing can be implanted after removal of a well-fixed femoral stem, or a ceramic head can be inserted with a metal adapter1112). Although controversy about treatment of ceramic liner fractures remains, revision surgery using CoC bearings is recommended1112). Ceramic bearing fractures are rare and most previous studies have involved older patients11) with short-term follow-up; only a few mid- and long-term studies on treatment outcomes of revision THA following ceramic bearing fractures in young and active patients have been reported13).
The aim of this study is to evaluate the clinical and radiologic outcomes of revision THA using third-generation CoC articulation after ceramic bearing fractures in active patients under 60-years old.

MATERIALS AND METHODS

This study involved a total of 8 patients younger than 60 years of age who: i) underwent revision surgery with CoC bearings for management of a third-generation ceramic bearing fracture from May 2004 to November 2011 and ii) had a minimum follow-up of 6 years. The patients were all male. The mean ages at the time of the initial and revision operations were 39 years (range, 31–50 years) and 43.8 years (range, 33–60 years) years old, respectively. All patients were classified as American Society of Anesthesiologists (ASA) grade 1 before revision and the average follow-up period after revision was 9.7 years (range, 6–13.3 years). The most common reason for the initial THA was avascular necrosis of the femoral head (Table 1).
Third-generation ceramic bearings were used in the first surgery for all patients. The liners used were a sandwich liner (n=4) and a full ceramic liner (n=4). The femoral head diameters were 28 mm (n=7) and 32 mm (n=1). The femoral neck lengths were short (n=4), medium (n=2), and long (n=2). There were 2 cases of ceramic head fracture and 6 cases of ceramic liner fracture. The causes for fracture of ceramic components were determined based on patient interview and medical records. Of the 2 patients with ceramic head fractures, fractures occurred after a fall (n=1) and without any trauma (n=1); a short neck was used in the initial surgery in both patients. Of the 6 patients with ceramic liner fracture, fracture appeared to be caused by trauma (n=1), improper position of the ceramic liner (n=1), and impingement associated with frequent stable squatting and sitting cross-legged positions (n=4).
Clinical evaluations involved: i) comparisons of pre- and post-operative Harris hip scores (HHS)14), ii) assessment of ambulatory ability according to the Koval classification15) preoperatively and at the last follow-up, and iii) pain assessments. Individuals were classified according to HHS as excellent (greater than 90 points), good (between 81 and 90 points), fair (between 71 and 80 points), or poor (less than 70 points). For radiologic evaluation, femoral osteolysis was assessed by dividing the proximal femur into seven Gruen zones16); progressive and continuous periprosthetic lucencies greater than 2 mm thick on postoperative follow-up radiographs were considered as indicating osteolysis. Patients were deemed to have acetabular osteolysis when continuous radiolucency lines greater than 2 mm thick were detected in DeLee and Charnley zones17). The presence of other complications was also examined.
Statistical analyses were performed using SPSS ver. 12.0 (SPSS Inc., Chicago, IL, USA). Data were analyzed using the Mann-Whitney U-test and Wilcoxon signed-rank test. Differences were considered statistically significant at P<0.05. This study was performed after gaining Institutional Review Board approval from our institution (2015-04-020).

RESULTS

A thorough synovectomy was performed at the time of revision surgery to prevent possible complications (e.g., third-body wear caused by residual ceramic particles and metallosis). Irrigation was repeated several times to remove any ceramic particles remaining in the adjacent soft tissue and bearing surfaces. One of the two patients with ceramic head fracture underwent revision using the fourth-generation ceramic bearing with metal adapter concerning plastic deformation of the Morse taper junction (Fig. 1); the other patient with head fracture underwent revision of the cup. The 6 patients with ceramic liner fractures underwent revision using the third-generation CoC bearings from the same manufacturer by replacing the original component to the femoral head with a longer neck.
Clinical evaluations revealed that the average HHS improved from 63.2 points (range, 41–82 points) to 91.8 points (range, 86–96 points) (P=0.01). Favorable HHS outcomes were achieved in all cases (excellent, 5; good, 3). Although all 8 patients complained of mild pain after revision, pain was relieved with conservative management. Progressive improvement was noted and at the final follow-up visit, no patient complained of pain that influenced their ability to perform daily activities. The waling ability of all patients was evaluated as grade 1 (independent community ambulatory) according to the Koval classification15), and there were no cases with diminished ambulatory ability at the last follow-up visit.
Radiologic assessment revealed no instances of loosening or osteolysis around the femoral or acetabular component and bony stability was achieved in all patients at the final follow up (Fig. 2). During the follow-up, one patient experienced heterotopic ossification in the absence of overt symptoms.

DISCUSSION

The leading cause of aseptic loosening–a major complication of hip arthtoplasty–has been identified as polyethylene wear debris of acetabular prostheses after THA using metal-on-polyethylene articulations. Because of this, studies on alternative bearing surfaces (e.g., metal-on-metal, CoC, ceramic-on-metal) are being actively pursued18). Of these bearing types, a CoC bearing appears to be an attractive alternative for younger and more active patients based on its improved wear resistance and lower coefficient of friction67). However, concerns about the safety of CoC bearings remain because ceramic liner or ceramic head fractures have been reported after CoC THA8910). Excluding trauma, the major causes of alumina breakage are recognized to be neck-liner impingement and edge loading; these causes may be due to a reduction in head-neck ratio resulting from the use of 28 mm alumina head and circular-shaped neck, and the use of thinner, polyethylene-reinforced alumina liners. The use of a short-neck ceramic head is a known risk factor for ceramic head fracture19). Furthermore, a high rate of alumina liner breakage has been reported in patients with larger anteversion angles of the acetabular cup by Ha et al.20) and in male patients with a heavier weight by Poggie et al21). In the present study, the subjects were all male and underwent surgeries with short-neck heads. Additional risk factors were the use of sandwich-type liners (n=4) and lifestyle habits that may be contributing to impingement (n=4).
Standard clinical guidelines for the treatment of third-generation ceramic fractures have not yet been established. In cases of ceramic head fractures, a newly replaced ceramic head can be fractured by stress concentrated on the head even with only mild erosive changes of the taper. Two revision THA treatment approaches have been suggested; the first is to insert a new stem and ceramic head after removing a well-fixed femoral stem and the alternative is to exchange the original components with cobalt-chromium femoral heads on polyethylene-bearing surfaces without stem revision22).
In revision surgeries, the insertion of a new ceramic head onto a damaged Morse taper may result in weakened tensile strength within the ceramic components by concentrating areas of stress, and thus may initiate fractures in the newly implanted ceramic head. Although this potential challenge can be avoided by stem revision, the removal of a well-fixed femoral stem remains controversial and raises technical challenges (e.g., massive transfusion, prolonged operation time, and risk of complications such as damage to the remaining bone stock). Although rare, serious complications (e.g., metallosis) have been reported when using a metal head during revision THA due to ceramic particles that remain because of incomplete synovectomy112324). In a multicenter survivorship study, Allain et al.11) have reported that third-body wear can be accelerated by ceramic particles that remain in the adjacent soft tissue and bearing surfaces after revisions of ceramic head fracture. Matziolis et al.24) have recommended that if the stem is not removed, CoC bearings with a metal adapter would be used because complete removal of ceramic debris during revision is impossible in most cases and residual ceramic particles in the metal-on-polyethene bearings can lead to the wear of metal bearings. In this study, taper damage was confirmed in 1 case with ceramic head fracture and a new ceramic head with a metal adapter was used.
The treatment of ceramic liner fractures remains controversial and has been rarely described42526). Liner fractures typically have larger fragments than fractured head fragments, so the removal of the fractured fragments is easier; unless the head fracture is associated, revision with new CoC bearings can be performed because the Morse taper is not damaged. Park et al.27) have suggested that, since ceramic heads are more durable than cobalt-chrome heads, more wear-resistant CoC bearings are favorable for use in revision surgeries of liner fractures with no gross damage to the Morse taper. Moreover, Traina et al.26) and Hannouche28) have performed revision surgeries using ceramic heads in patients with ceramic liner fractures. In our study involving young (i.e., less than 60 years old) and active patients, HHS improved, and no complications (e.g., metallosis, osteolysis and prosthetic loosening) occurred at the last follow-up in any revisions using CoC bearing surfaces after CoC bearing fracture.
This study was limited by the relatively small sample size. Additionally, no revisions using fourth-generation ceramic bearings–which are more durable and have improved wear resistance compared to third generation ceramic bearings were included in this study. Also, since most patients received their first surgery in other hospitals, the authors were unable to accurately assess the position of the acetabular and femoral components that could affect ceramic bearing surface fractures. However, the relatively small sample size rather attempts to demonstrate excellent hardness, wear resistance and biocompatibility of CoC bearing surfaces. Studies regarding fractures of fourth-generation ceramic heads have been rarely reported.
This study obtained satisfactory clinical and radiologic outcomes by performing revision THA using CoC components after ceramic bearing fracture with an average follow-up of 9.7 years in active patients under 60 years old. Further studies with larger sample sizes and longer follow-ups are necessary.

CONCLUSION

Considering wear in ceramic bearing surfaces, osteolysis and biocompatibility of ceramic wear particles, revision THA using CoC components after ceramic bearing fracture resulted in satisfactory clinical and radiologic results in young and active patients during an average follow-up of 9.7 years in our study. When performing revision for ceramic bearing fracture, a thorough synovectomy is recommended to prevent third-body wear caused by residual ceramic particles and the use of CoC bearing surface is considered to be a better option than metal-on-polyethylene bearing surfaces. If removal of the well-fixed femoral stem is unnecessary during revision surgery, the use of a ceramic head with a metal adapter might be a good option to avoid the risk of damaged Morse taper.

Figures and Tables

Fig. 1

Metal adaptor was used because of taper damage.

hp-30-156-g001
Fig. 2

(A) Anteroposterior radiograph of pelvis (patient 1) shows fractured ceramic particles (arrow). (B) Intraoperative photograph shows fractured ceramic liner. (C) Radiograph taken 9 years after revision surgery shows no osteolysis or implant loosening.

hp-30-156-g002
Table 1

Basic Characteristics and Results of the Patients

hp-30-156-i001
Patient Age at revision (yr) Initial diagnosis Stem of primary THA Cup of primary THA Between prior operation and revision (mo) Reason for 2nd operation Complication after revision Follow-up after last revision (mo) Pre-HHS score Last-HHS score
1 33 ONFH Bicontact SD Stem Aesculap, PLASMA Cup 21 Liner breakage None 153 59 92
2 47 ONFH Smith-Nephew, SL-Plus Stem Smith-Nephew, EP-FIT PLUS Cup 50 Liner breakage None 114 82 96
3 46 ONFH Smith-Nephew, SL-Plus Stem Smith-Nephew, EP-FIT PLUS Cup 84 Liner breakage None 84 72 93
4 46 ONFH Smith-Nephew, SL-Plus Stem Smith-Nephew, EP-FIT PLUS Cup 9 Liner breakage None 159 41 92
5 60 ONFH Secu-Fit Stryker, Trident Cup 120 Liner breakage None 72 65 89
6 38 ONFH Bicontact SD Stem Aesculap, plasma Cup 45 Ceramic head breakage None 132 53 86
7 39 ONFH Smith-Nephew, SL-Plus Stem Smith-Nephew, EP-FIT Plus Cup 30 Liner breakage None 136 60 89
8 38 DDH sequelae DePuy, AML Depuy, Duraloc Option Cup 75 Ceramic head breakage None 78 51 93

All patients were male.

THA: total hip arthroplasty, HHS: Harris hip score, ONFH: osteonecrosis femoral head, DDH: developmental dysplasia of hip.

ACKNOWLEDGEMENTS

This research was supported by the BISA Research Grant of Keimyung University in 2017.

Notes

CONFLICT OF INTEREST The authors declare that there is no potential conflict of interest relevant to this article.

References

1. Karam JA, Tokarski AT, Ciccotti M, Austin MS, Deirmengian GK. Revision total hip arthroplasty in younger patients: indications, reasons for failure and survivorship. Phys Sportsmed. 2012; 40:96–101.
crossref
2. Willert HG, Semlitsch M. Reactions of the articular capsule to wear products of artificial joint prostheses. J Biomed Mater Res. 1977; 11:157–164.
crossref
3. Crowther JD, Lachiewicz PF. Survival and polyethylene wear of porous-coated acetabular components in patients less than fifty years old: results at nine to fourteen years. J Bone Joint Surg Am. 2002; 84-A:729–735.
4. D'Antonio J, Capello W, Manley M, Bierbaum B. New experience with alumina-on-alumina ceramic bearings for total hip arthroplasty. J Arthroplasty. 2002; 17:390–397.
5. McAuley JP, Szuszczewicz ES, Young A, Engh CA Sr. Total hip arthroplasty in patients 50 years and younger. Clin Orthop Relat Res. 2004; (418):119–125.
crossref
6. Skinner HB. Ceramic bearing surfaces. Clin Orthop Relat Res. 1999; (369):83–91.
crossref
7. Yoo JJ, Kim YM, Yoon KS, Koo KH, Song WS, Kim HJ. Alumina-on-alumina total hip arthroplasty. A five-year minimum follow-up study. J Bone Joint Surg Am. 2005; 87:530–535.
8. Han CD, Choi CJ, Yang IH, Lee DH. Total hip arthroplasty with ceramic-on-ceramic articulation: minimum 5-year follow-up results. J Korean Orthop Assoc. 2006; 41:421–427.
crossref
9. Yoo JJ, Kim YM, Yoon KS, et al. Alumina-on-alumina THA in active patients younger than 40 years old: a 5-year minimum follow-up study. J Korean Orthop Assoc. 2006; 41:404–412.
crossref
10. Traina F, De Fine M, Bordini B, Toni A. Risk factors for ceramic liner fracture after total hip arthroplasty. HIP Int. 2012; 22:607–614.
crossref
11. Allain J, Roudot-Thoraval F, Delecrin J, Anract P, Migaud H, Goutallier D. Revision total hip arthroplasty performed after fracture of a ceramic femoral head. A multicenter survivorship study. J Bone Joint Surg Am. 2003; 85-A:825–830.
12. Rambani R, Kepecs DM, Mäkinen TJ, Safir OA, Gross AE, Kuzyk PR. Revision total hip arthroplasty for fractured ceramic bearings: a review of best practices for revision cases. J Arthroplasty. 2017; 32:1959–1964.
crossref
13. Sharma V, Ranawat AS, Rasquinha VJ, Weiskopf J, Howard H, Ranawat CS. Revision total hip arthroplasty for ceramic head fracture: a long-term follow-up. J Arthroplasty. 2010; 25:342–347.
14. Mahomed NN, Arndt DC, McGrory BJ, Harris WH. The Harris hip score: comparison of patient self-report with surgeon assessment. J Arthroplasty. 2001; 16:575–580.
15. Koval KJ, Skovron ML, Aharonoff GB, Meadows SE, Zuckerman JD. Ambulatory ability after hip fracture. A prospective study in geriatric patients. Clin Orthop Relat Res. 1995; (310):150–159.
16. Bands R, Pelker RR, Shine J, Bradburn H, Margolis R, Leach J. The noncemented porous-coated hip prosthesis. A three-year clinical follow-up study and roentgenographic analysis. Clin Orthop Relat Res. 1991; (269):209–219.
17. DeLee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res. 1976; (121):20–32.
crossref
18. Archibeck MJ, Jacobs JJ, Black J. Alternate bearing surfaces in total joint arthroplasty: biologic considerations. Clin Orthop Relat Res. 2000; (379):12–21.
19. Koo KH, Ha YC, Jung WH, Kim SR, Yoo JJ, Kim HJ. Isolated fracture of the ceramic head after third-generation alumina-on-alumina total hip arthroplasty. J Bone Joint Surg Am. 2008; 90:329–336.
crossref
20. Ha YC, Kim SY, Kim HJ, Yoo JJ, Koo KH. Ceramic liner fracture after cementless alumina-on-alumina total hip arthroplasty. Clin Orthop Relat Res. 2007; 458:106–110.
crossref
21. Poggie RA, Turgeon TR, Coutts RD. Failure analysis of a ceramic bearing acetabular component. J Bone Joint Surg Am. 2007; 89:367–375.
crossref
22. Pulliam IT, Trousdale RT. Fracture of a ceramic femoral head after a revision operation. A case report. J Bone Joint Surg Am. 1997; 79:118–121.
crossref
23. Kempf I, Semlitsch M. Massive wear of a steel ball head by ceramic fragments in the polyethylene acetabular cup after revision of a total hip prosthesis with fractured ceramic ball. Arch Orthop Trauma Surg. 1990; 109:284–287.
crossref
24. Matziolis G, Perka C, Disch A. Massive metallosis after revision of a fractured ceramic head onto a metal head. Arch Orthop Trauma Surg. 2003; 123:48–50.
crossref
25. Garino JP. Modern ceramic-on-ceramic total hip systems in the United States: early results. Clin Orthop Relat Res. 2000; (379):41–47.
26. Traina F, Tassinari E, De Fine M, Bordini B, Toni A. Revision of ceramic hip replacements for fracture of a ceramic component: AAOS exhibit selection. J Bone Joint Surg Am. 2011; 93:e147.
27. Park YS, Hwang SK, Choy WS, Kim YS, Moon YW, Lim SJ. Ceramic failure after total hip arthroplasty with an alumina-on-alumina bearing. J Bone Joint Surg Am. 2006; 88:780–787.
crossref
28. Hannouche D, Nich C, Bizot P, Meunier A, Nizard R, Sedel L. Fractures of ceramic bearings: history and present status. Clin Orthop Relat Res. 2003; (417):19–26.
TOOLS
ORCID iDs

Kyung-Jae Lee
https://orcid.org/0000-0003-4811-574X

Similar articles