Journal List > Prog Med Phys > v.28(2) > 1098588

Back, Park, and Kim: Assessment of Radiation Dose from Radioactive Wedge Filters during High-Energy X-Ray Therapy

Abstract

This paper evaluated the amount of radiation generated by wedge filters during radiation therapy using a high-energy linear accelerator, and the dose to the worker during wedge replacement. After 10-MV photon beam was irradiated with wedge filter, the wedge was removed from the linear accelerator, and the dose rate and energy spectrum were measured. The initial measurement was approximately 1 uSv/h, and the radiation level was reduced to 0.3 uSv/h after 6 min. The effective half-life derived from the dose rate measurement was approximately 3.5 min, and the influence of AI-28 was about 53%. From the energy spectrum measurements, a peak of 1,799 keV was measured for AI-28, while the peak for Co-58 was not measured in the control room. The peaks for Au-106 and Cd-105 were found only measurement was done without wedge removement from the linear accelerator. The additional doses received by the radiation worker during wedge replacement were estimated to be 0.08−0.4 mSv per year.

REFERENCES

1. NCRP Report 79. Neutron contamination from medical electron accelerator. 1984.
2. AAPM TG Report 27. Neutron measurements around high energy X-ray radiotherapy machines. 1986.
3. Chilbani O, Ma C. Photoneutron dose calculations for high energy photon beams from Siemens and Varian LINACS. Med Phys 1990–2000. 2003; 30.
4. Gudowska I, Brahme A, Andreo P, et al. Calcluation of absorbed dose and biological effectiveness from photoneutron reactions in bremsstrahlungbeam of end point 50 MeV. Phys in Med Biol. 1999; 44:2099.
5. Roig M, Panettieri V, Ginjaume M, et al. Photonuclear isotope characterization of a Siemens KDS 18 MeV LINAC head. Phys in Med Biol. 2004; 49:N242–6.
6. Borchardt IM, Patterson JR, Beddoe AH. An investigation of photonuclear reactions in Cerrobend eutectic material with an 18 MV LINAC. Phys in Med Biol. 1991; 36:649–53.
crossref
7. Petrovic N, Krestic-Vesovic J, Stojanovic D, et al. Contribution of activation products to occupational exposure following treatment using high energy photons in radiotherapy. Rad Prot Dos. 2011; 143:109–12.
8. Fischer HW, Tabot BE, Poppe B. Activation processes in a medical linear accelerator and spatial distribution of activation products. Phys Med Biol. 2006; 51:N461.
crossref
9. Fischer HW, Tabot B, Poppe B. Comparison of activation products and induced dose rates in diferent high-energy medical linear accelerators. Health Physics. 2008; 94:272–8.
10. Israngkul-Na-Ayuthaya I, Suriyapee S, Pengvanich P. Evaluation of equivalent dose from neutrons and activation products from a 15 MV x-ray LINAC. J of Rad Research. 2015; 56:919–26.

Fig. 1.
Measurement of dose rate using GM detector and activated wedge filter.
pmp-28-45f1.tif
Fig. 2.
Measurement of energy spectrum using NaI scintillator detector.
pmp-28-45f2.tif
Fig. 3.
Measured dose rate of activated wedge after 5-Gy irradiation at various field sizes.
pmp-28-45f3.tif
Fig. 4.
Measured energy spectrum from activated wedge after 5-Gy irradiation.
pmp-28-45f4.tif
Table 1.
Exposure dose to workers owing to removal of activate wedge after X-ray treatment.
Number of patients persons/day Number of wedge remove times/patient Exposure, uSv
20 1 83
  2 167
30 1 125
  2 250
40 1 167
  2 333
50 1 208
  2 417
TOOLS
Similar articles