Journal List > Prog Med Phys > v.26(1) > 1098497

Kang, Lee, and Lee: Calculation of Dose Distribution for SBRT Patient Using Geant4 Simulation Code

Abstract

The Monte Carlo based dose calculation program for stereotactic body radiotherapy was developed in this study. The Geant4 toolkit widely used in the radiotherapy was used for this study. The photon energy spectrum of the medical linac studied in the previous research was applied for the patient dose calculations. The geometry of the radiation fields defined by multi-leaf collimators were taken into account in the PrimaryGeneratorAction class of the Geant4 code. The total of 8 fields were demonstrated in the patient dose calculations, where rotation matrix as a function of gantry angle was used for the determination of the source positions. The DicomHandler class converted the binary file format of the DICOM data containing the matrix number, pixel size, endian type, HU number, bit size, padding value and high bits order to the ASCII file format. The patient phantom was constructed using the converted ASCII file. The EGSnrc code was used to compare the calculation efficiency of the material data.

References

1. Leksell L. Stereotactic radiosurgery. J Neuosurg Psychiatr. 46:797–803. 1983.
crossref
2. Kang JK, Lee DJ. Development of Monte Carlo Code for the Dose Calculation of the Stereotactic Radiosurgery. Progress in Medical Physics. 23(4):303–308. 2012.
3. Rogers DWO. BEAM: A Monte Carlo code to simulate radiotherapy treatment units. Med Phys. 22(5):503. 1995.
crossref
4. Fippel M, Laub W, Huber B, Nüsslin F. Experimental investigation of a fast Monte Carlo photon beam dose calculation algorithm. Phys Med Biol. 44(12):3039–3054. 1999.
crossref
5. Fippel M, Kawrakow I, Friedrich K. Electron beam dose calculations with the VMC algorithm and the verification data of the NCI working group. Phys Med Biol. 42(3):501–520. 1997.
crossref
6. Fippel M, Haryanto F, Dohm O, Nüsslin F, Kriesen S. A virtual photon energy fluence model for Monte Carlo dose calculation. Med Phys. 30(3):301. 2003.
crossref
7. Fippel M. Efficient particle transport simulation through beam modulating devices for Monte Carlo treatment planning. Med Phys. 31(5):1235. 2004.
crossref
8. Kawrakow I, Fippel M. Investigation of variance reduction techniques for Monte Carlo photon dose calculation using XVMC. Phys Med Biol. 45(8):2163–2183. 2000.
crossref
9. Schneider W, Bortfeld T, Schlegel W. Correlation between CT numbers and tissue parameters needed for Monte Carlo simulations of clinical dose distributions. Phys Med Biol. 45(2):459–478. 2000.
crossref
10. Kim HD, Kim BY, Kim EC, Yun SM, Kang JK, Kim SK. Comparison of dose distributions for Hounsfield number conversion methods in GEANT4. J Korean Phys Soc. 64(12):1912–1918. 2014.
crossref
11. ICRU Report 46. Photon, Electron, Proton and Neutron Interaction Data for Body Tissues, Bethesda, MD (. 1992.

Fig. 1.
The isodose distributions for the center slice of the CT data. (a) dose distributions from iPLan, (b) dose distributions in this study.
pmp-26-36f1.tif
Fig. 2.
Dose distributions of the lateral slices, (a) dose distributions from iPLan, (b) dose distributions in this study.
pmp-26-36f2.tif
Table 1.
Compositions by fractional weight of the materials used in the BEAMnrc default method and their mass densit ranges.
# Material ρ (gcm−3) range Fraction by weight
H C N O Na Mg P S Cl Ar K Ca Zn
0 Air 0.044     75.5 23.2           1.3      
1 Lung 0.302 10.3 10.5 3.1 74.9 0.2   0.2 0.3 0.3   0.2    
2 Soft 1.101 10.1 11.1 2.6 76.2                  
3 Bone 2.088 4.72 14.43 4.2 44.61   0.22 10.5 0.32       20.99 0.01
Table 2.
The treatment planning parameters for a patient in the iPlan system.
Object name Object type Dose/fraction No. of fraction Total dose Normalization mode
Tumor L4 PTV 10.0 Gy 4 80.0%=40.00 Gy 93.5% covers 99.5% of
          target volume
Table 3.
The field parameters for a SBRT patient.
Name Depth equiv. (mm) Abs. dose (Gy) Rel. dose (%) MU
IMRT Beam1 46.9 7.18 14.4 4,288.0
IMRT Beam2 119.8 5.07 10.1 4,076.0
IMRT Beam3 63.0 8.26 16.5 3,776.0
IMRT Beam4 52.4 8.16 16.3 3,848.0
IMRT Beam5 85.7 7.63 15.3 3,524.0
IMRT Beam6 111.2 5.88 11.8 3,620.0
IMRT Beam7 48.7 3.94 7.9 3,736.0
IMRT Beam8 77.4 6.74 13.5 3,992.0
TOOLS