Lee, Jung, and Lee: Research of Efficiency for Gas Electron Multiplier Detector to Monitor Low Energy Gamma-Ray and Beta-Ray

Abstract

Radiation monitoring is one of the most important process in all places where radioactive material is used including hospital. In this preliminary study, we made GAS electron multiplier (GEM) detector and acquired relative efficiencies in order to see if GEM detector can be useful in radiation monitoring system. The relative efficiency was acquired by using the ratio of GEM detector efficiency to CdTe detector efficiency. The relative efficiency of 72% and 4% was acquired for beta-ray and gamma-ray respectively.

References

1. 박인규: 방사선 검출기와 국제연구.물리학과 첨단 기술. 25–30. 2011.
2. Simon Frank: Commissioning of the G EM Detectors in the COMPAS Exp erime nt. Dip loma The sis. Physik Depa rtment, Technische Unive rsitat Munc hen (. 2011.
3. Buzulutskov A. Radiation dete ctors ba sed on gas e lec tron multipliers (R evi ew s). Instr um Exp Tech. 50(3):287–310. 2007.
4. 곽정원, 신정욱, 김진성 등. 전립선암에 대한 양성자치료에서 금마커에 의한 방사선 선량분포의 영향.의학물리. 21(3):291–297. 2010.
5. 전경수, 오영기, 백종근 등. 이온전리함 및 TLD법을 이용한 multi-detector comp uted tomogra phy의 흡수선량 및 체표면 선 량 평가.의학물리. 24(1):35–40. 2013.
6. Tsyganov Edward, Antich Peter, Parkey Robert, et al. Gas electron multiplying detectors for medical applications. Nucl Instru and Me th. 597(2):257–265. 2008.
crossref
7. Fraga F, Margato L, Fetal S, et al. Luminescence a nd imaging with gas e lec tron multiplie rs. Nuc l Instru and Me th. 513(1):379–387. 2003.
8. Fabio Sauli: Development and applications of the gas electron multiplier. N uc l Instru and Me th. 513(1):115–119. 2001.
9. Peskov V, Fonte R, Danielsson M, et al. The study and optimization of new micropattern gaseous detectors for high rate applications. IEEE Tra ns Nuc l Sci 48 (4) (. 2001.
10. Yong HS, IG Kim, Hie HC, et al. Energy resolution of a single G EM dete ctor. Journal of the Kore an Physica l S ocie ty. 55(6):2366–2371. 2009.
11. Lepy Marie-Christine: Detection e fficie ncy, IAEA -ALMERA Technical Visit. (. 2010.

Fig. 1.
(a) GEM detector, (b) GEM detector system.
pmp-25-95f1.tif
Fig. 2.
GEM 검출기 내부 구조.
pmp-25-95f2.tif
Table 1.
Sr-90의 상대적 검출 효율.
CdTe GEM
Count Background Net count Count Background Net count 상대적 검출효율
10,927±309 115±25 10,812±304 129,048±2,088 90±11 128,958±2,091 71.6±2.65

Sr-90의 평균적인 상대적 검출효율 71.6±2.65 (mean±s.d.)

CdTe의 540 KeV영역의 검출효율 6%

Table 2.
Cs-137의 상대적 검출 효율.
CdTe GEM
Count Background Net count Count Background Net count 상대적 검출효율
82,974±3,292 110±7 82,864±3,292 52,401±1,461 91±14 52,310±1,467 3.79±0.18

Cs-137의 평균적인 상대적 검출효율 3.79±0.18 (mean±s.d.)

CdTe의 660 KeV영역의 검출효율 6%