Journal List > Ann Lab Med > v.38(4) > 1091627

Jeon, Jung, Kim, Kwon, and Kim: Allergen Microarrays for In Vitro Diagnostics of Allergies: Comparison with ImmunoCAP and AdvanSure

Abstract

Background

In vitro detection of the allergen-specific IgE antibody (sIgE) is a useful tool for the diagnosis and treatment of allergies. Although multiple simultaneous allergen tests offer simple and low-cost screening methods, these platforms also have limitations with respect to multiplexibility and analytical performance. As an alternative assay platform, we developed and validated a microarray using allergen extracts that we termed “GOLD” chip.

Methods

Serum samples of 150 allergic rhinitis patients were used in the study, and the diagnostic performance of the microarray was compared with that of AdvanSure (LG Life Sciences, Daejun, Korea) and ImmunoCAP (Phadia, Uppsala, Sweden). Standard IgE samples were used for the quantitative measurement of sIgEs.

Results

The microarray-based assay showed excellent performance in the quantitative measurement of sIgEs, demonstrating a linear correlation within the range of sIgE concentrations tested. The limit of detection (LOD) was lower than 0.35 IU/mL, which is the current standard for the LOD cut-off. The assay also provided highly reproducible sets of data. The total agreement percentage of positive and negative calls was 92.2% compared with ImmunoCAP. Moreover, an outstanding correlation was observed between the microarray and the ImmunoCAP results, with Cohen's kappa and Pearson correlation coefficient values of 0.80 and 0.79, respectively.

Conclusions

The microarray-based in vitro diagnostic platform offers a sensitive, reproducible, and highly quantitative method to detect sIgEs. The results showed strong correlations with that of ImmunoCAP. These results suggest that the new allergen microarray can serve as a useful alternative to current screening platforms, ultimately becoming a first-line screening method.

INTRODUCTION

Allergic diseases have become a major public health concern worldwide. Allergic reactions are mediated by allergen-specific IgE antibodies (sIgEs) that trigger the release of inflammatory mediators. Therefore, the detection of sIgEs in patient sera is necessary for the diagnosis and appropriate treatment of allergies. Since it was first reported by Helmtraud Ebruster in 1959 [1], the skin prick test (SPT) has been the first-line method to detect IgE-mediated type I (immediate response) allergic reactions, and is still considered a reliable standard for the diagnosis of allergies by many clinicians, given the rapid delivery of strong evidence for sensitization at relatively low costs. Nevertheless, the SPT is a rather invasive method because each allergen has to be delivered to the patient by pricking the skin with a needle or a pin carrying a small amount of the allergen that can cause allergic symptoms, including a skin rash, swelling, nausea, and even anaphylaxis [12]. Moreover, the SPT results are not quantitative and are difficult to standardize [3]. For these reasons, the in vitro detection of sIgEs has emerged as an important tool for screening allergic diseases.
Indeed, substantial progress has been made in the development of in vitro sIgE detection assays that provide highly reproducible and sensitive methods. The currently available tests include the radioallergosorbent test (RAST), multiple allergen simultaneous test (MAST), and ImmunoCAP system (Thermo Scientific, Uppsala, Sweden). The once market-leading RAST method is no longer popular, since this is considered to be a rather inefficient and expensive method compared with others [4]. Thus, this assay has been replaced with the ImmunoCAP system as the generally accepted reference method as it provides an accurate and quality-controlled assay for the in vitro detection and measurement of sIgEs [567]. The CAP system uses a pool of a cellulose sponge matrix coated with mixture of allergen extracts or component allergens to capture sIgEs in the patient's serum, which are then detected using an enzyme-linked anti-IgE antibody via an enzymatic reaction [8]. This system is conferred with improved performance by maximizing the surface area of the solid phase to be able to bind all clinically relevant proteins of allergen extracts. As a downside, the CAP system requires an individual assay to be conducted for each allergen of interest, increasing the costs [9]. Thus, despite offering a robust assay system, ImmunoCAP is based on a singleplex assay format and is not suitable for first-line screening because of the restricted number of allergens that can be screened simultaneously. In contrast, the MAST system uses a panel of multiple allergen extracts, enabling screening of patient serum against multiple allergens simultaneously, even with a limited amount of serum in a cost-efficient manner [10]. The representative commercialized MAST platforms include AdvanSure Allergy Screen from LG Life Sciences (Daejun, Korea), Advia Centaur Immunoassay Systems from Siemens (München, Germany), and Polycheck from Biocheck GmbH (Munster, Germany). Although useful, the results obtained from MAST are often only semi-quantitative, and relatively low sensitivity and accuracy have been reported as the main drawbacks [1112].
To overcome these limitations of current screening methods, we developed an allergen microarray, termed “GOLD chip,” that enables quantitative and simultaneous detection of sIgEs in patient sera against multiple allergens. We aimed to achieve high sensitivity and specificity, comparable to those of ImmunoCAP. We conducted assay performance tests and concordance analysis in comparison with the AdvanSure Allergy Screen and ImmunoCAP.

METHODS

1. Serum samples and allergy testing

Serum samples of 150 patients with allergic rhinitis (AR) were collected at the Gil Medical Center, Gachon University, Incheon, Korea (Table 1). Each serum sample was divided into three aliquots, and each aliquot was used for AdvanSure, ImmunoCAP, and GOLD (Gil Hospital, Won Medical, LabGenomics, Dongguk University) chip-based screening. This study was approved by the Institutional Review Boards of Gachon University Gil Medical Center (IRB no. GBIRB2015-46) and Dongguk University (IRB no. DUIRB-20151127-013) (Seoul, Korea). Informed consent was obtained from each patient.

2. Allergens and serological parameters

sIgE was measured against 42 allergens using AdvanSure Allergy Screen according to the manufacturer's protocol. The results were scored in classes from 0 to 6 as follows: 0 (< 0.35 IU/mL), 1 (0.35–0.7 IU/mL), 2 (0.7–3.5 IU/mL), 3 (3.5–17.5 IU/mL), 4 (17.5–50 IU/mL), 5 (50–100 IU/mL), and 6 (> 100 IU/mL).
The concentrations of sIgEs produced in response to selected allergens (based on the AdvanSure screening results) were also measured using ImmunoCAP, according to the manufacturer's protocol. Each serum sample was screened against a maximum of six allergens. The concentrations of each sIgE along with the class (0–6, as defined above) were reported. In both cases, any result of class 2 or more was defined as positive.

3. Construction of allergen microarrays

Allergen microarrays were constructed by spotting allergens onto an H slide using OmniGrid 100 microarray (GeneMachines, CA, USA) (Fig. 1A). Lyophilized allergens were first solubilized in phosphate-buffered saline (PBS, pH 7.4; Sigma, St. Louis, MO, USA), whereas the dissolved allergens were used in their presented form. The tested allergen sources included Dermatophagoides pteronyssinus, Alternaria alternata, dog dander, oxeye daisy, ragweed, rye, and oak extracts from Greer (Lenoir, NC, USA); Dermatophagoides farina, birch-alder mix, cockroach, ash, dandelion, goldenrod, Acacia, Cladosporium herbarum, Bermuda grass, and hazelnut extracts from Allergon (Välingevägen, Sweden); cat dander and reed extracts from Squarix (Marl, Germany); pigweed and Timothy grass extracts from Bencard (München, Germany); Japanese hop extracts from Yonsei University (Seoul, Korea); and mugwort and Aspergillus fumigatus extracts from Lofarma (Milano, Italy), as well as egg white, soybean, peach, milk, crab, shrimp, Japanese cedar, sycamore mix, orchard grass, ash mix, dandelion, sallow willow, Russian thistle, Penicillium notatum, poplar mix, sweet vernal grass, pine, and mackerel. PBS and biotinylated goat anti-human IgE antibody (ε chain-specific; VectorLab, Burlingame, CA, USA) were used as negative and positive controls, respectively. Allergen solutions were mixed with the same volume of 3X saline sodium citrate buffer (Schott, Mainz, Germany). The selected allergens and controls were printed in triplicate on a glass slide (Nexterion® slide H; Schott, Mainz, Germany) coated with hydrophilic polymers presenting N-hydroxysuccinimide (NHS) ester residues (Fig. 1B). The microarrays were generated by spotting 0.2 ng of the allergens in arrays of 200-µm-diameter spots at a spot-to-spot distance of 300 µm. The size of an array was 5×3 mm. Four arrays were printed on each glass slide and were separated by a hybridization chamber (Fig. 1A, center). After spotting, the allergen chips were stored at 4℃ for at least 24 hours before use.

4. Immunoassay

The allergen microarray was exposed to 2% bovine serum albumin (BSA; BioBasic, Markham, ON, Canada) in PBST [PBS containing 0.1% Tween-20 (Amresco, Solon, OH, USA)] for 1 hour, washed with PBST, and incubated with 30 µL of human serum diluted in the same volume of PBST containing 2% BSA for 1 hour using the hybridization chamber. The slide was washed with PBST, incubated with a solution of detection antibody (biotinylated anti-human IgE, 0.5 µg/mL) in PBST containing 0.2% BSA for 1 hour, and then incubated with streptavidin-Cy3 (SA-Cy3) conjugate (1 µg/mL; Sigma-Aldrich, St. Louis, MO, USA) in PBST containing 0.2% BSA for 15 minutes in the dark. The slide was washed with PBST and distilled water sequentially, and then dried using compressed air. All steps were carried out at 25℃. The binding of allergen sIgE on spotted allergens was then detected by sequential treatment with the detection antibody and SA-Cy3 conjugate. sIgEs were visualized using a fluorescence scanner.

5. Scanning

The Cy3 fluorescence intensity was measured using a GenePix 4000B fluorescence scanner (Axon Instruments, Union City, CA, USA) at an excitation wavelength of 532 nm and an emission wavelength of 570 nm. The image files were analyzed using GenePix Pro Microarray Analysis Software (Axon Instruments). Three intensity values for each allergen were obtained from triplicate runs. For calibration, a negative control experiment was conducted to determine the background fluorescence intensity. The average fluorescence intensity was calculated and used for analysis. The fluorescence intensity corresponding to 0 and 100 IU/mL for each allergen was calculated from the least-squares fitting and used for unit conversion when needed.

6. Reproducibility test

CVs were determined by intra-assay, inter-assay, and batch-to-batch comparisons. For the determination of intra-assay precision, the mean CVs were calculated on the basis of five replicates. Inter-assay precision was determined as the mean CV obtained from three assays performed in separate chambers on the same slide. The batch-to-batch comparison was conducted using slides manufactured on different days. CV values were calculated on the basis of results of five independent assays, and each assay was carried out in triplicate. The CVs were calculated using the following equation:
CV=[SD (σ)/mean (µ)]×100

7. Statistical analysis

Correlations between the ImmunoCAP results (0–100 kU/L), AdvanSure class (0–6), and GOLD chip signal intensity (0–66000 AU) were analyzed using Pearson's correlation test. Cohen's kappa was calculated to determine the rate of agreement between assays. We categorized kappa values (κ) as almost perfect (0.8–1.0), substantial (0.6–0.8), moderate (0.4–0.6), fair (0.2–0.4), and poor (< 0.2). We then calculated the positive and total agreement percentages for each assay according to a previously reported method [13] as follows:
totalnumberofresultsnumberofdiscrepanciestotalnumberofresults×100alm-38-338-e001.jpg
Prism software (GraphPad, La Jolla, CA, USA) was used for the statistical analysis and plotting of the correlation curves.

RESULTS

1. Construction and calibration of allergen microarrays

Among various platforms for the in vitro detection of target biomolecules, microarrays provide a unique opportunity as they allow for the high-throughput screening of multiple targets simultaneously with increased sensitivity and selectivity [1415]. To construct the allergen microarray, we selected 24 allergens that are considered to be the main causal agents of AR and other allergic symptoms. The list of allergens was determined on the basis of the frequency of causative allergens from reported cases [161718], including house dust mites, various pollens, as well as pet dander (Fig. 1B). Each slide included negative (1×PBS buffer) and positive (biotinylated anti-human IgE antibody) controls. The selected allergens and controls were spotted on an activated substrate presenting NHS ester and a hydrophilic polymer (Fig. 1C). The microarray was designated “GOLD chip.”

2. Evaluation of microarray-based detection of sIgEs: qualitative detection and quantitative measurement

The assay to detect allergen-specific IgE in patient sera was performed as described elsewhere, with the necessary modification and optimization [1920212223]. In brief, the allergen microarray slide was incubated with PBST buffer containing 2% BSA before the assay to minimize the non-specific adsorption of random proteins.
Under the optimized condition, we first performed a quantitative evaluation using standard anti-D. pteronyssinus human IgE of known concentration [24]. The test samples were prepared by diluting the standard sIgE to obtain a series of eight concentrations ranging from 0.3 to 50 IU/mL. Each sample was analyzed with the microarray presenting D. pteronyssinus extract. The experiment was performed using two concentrations of detection antibody (0.5 and 1 µg/mL). The plot of fluorescence intensity versus sIgE concentration showed a positive linear correlation within the range of concentrations tested when 0.5 µg/mL of detection antibody was used (Fig. 2A). When the concentration of detection antibody increased to 1 µg/mL, a similar trend was observed; however, signal saturation occurred when the concentration of sIgE exceeded 50 IU/mL. The limit of detection (LOD) was lower than 0.3 IU/mL in both cases, which satisfies the requirement for the commercial allergy IVD assay kits.
Quantitative evaluation was then carried out using a pool of patient sera containing sIgEs to D. pteronyssinus and D. farinae. The sample was prepared by pooling serum samples from five patients who scored class 4+ in the ImmunoCAP serology test; the concentrations of sIgEs to D. pteronyssinus and D. farinae in the pooled sera were determined to be 72 IU/mL and 93 IU/mL by ImmunoCAP, respectively. The test samples were prepared in nine-fold geometric dilutions of pooled sera, starting at 1:2, and screened with the microarray presenting D. pteronyssinus and D. farinae spots. The plots of fluorescence intensity versus sIgE concentration to each allergen showed the expected positive linear correlation (See Supplemental Data Table S1, Fig. 2B, C), suggesting that the allergen microarray can provide uantitative information on the concentration of sIgEs. The result also showed that our system has an LOD<0.35 IU/mL.

3. Precision and repeatability of GOLD chip-based assays

As the GOLD chip assay was successfully applied for the quantitative detection of sIgEs to D. pteronyssinus and D. farinae, it was further tested for the detection of sIgEs to other allergens. The assay was conducted using patient sera with different sensitization patterns, resulting in different microarray patterns when visualized by fluorescence scanning (Fig. 2D). Before analyzing the diagnostic result, we first examined assay precision and repeatability by determining the CVs to assess intra-assay variability (0.4–3.8%), which is the variation between three spots in one array, inter-assay variability (2.4–11.4%), which reflects variation between three arrays in one slide, and batch-to-batch variability (3.6–25%), which reflects variation between three different slides (Table 2). Assay variation often results from irregularity in chip production and from the instability of assay protocols. The loss of transiently immobilized allergens during the assay, as well as the protein fouling on the microarray surface, could also be the main causes of observed variation. Nevertheless, the CV values for the microarray assays were considerably lower than those for previously reported assay platforms [2325], indicating improved precision and repeatability.

4. Comparison of the GOLD chip with ImmunoCAP and AdvanSure

We then compared the data obtained from the GOLD chip with those obtained from ImmunoCAP and AdvanSure. ImmunoCAP was selected because it is currently considered one of the best-performing IVD platforms for allergic diseases. AdvanSure was selected as the representative MAST platform because it is the most widely used MAST platform in Korea.
Serum samples from 150 AR patients were analyzed against 24 allergen extracts using the GOLD chip, ImmunoCAP, and AdvanSure systems (Fig. 3 and Table 3). First, the call agreement rate was determined for each pair of assays, i.e., GOLD vs ImmunoCAP, ImmunoCAP vs AdvanSure, and GOLD vs AdvanSure. We used class 2 as the cut-off level of positive results for both ImmunoCAP and AdvanSure, as suggested by the manufacturers, in calculating the agreement percentage for positive and negative calls. The cut-off for a positive result from the GOLD chip was determined as the fluorescence intensity corresponding to 0.7 IU/mL of IgE for each allergen extract [262728]. The overall total agreement rates were 92.2% for GOLD vs Immuno-CAP, 66.1% for ImmunoCAP vs AdvanSure, and 66.1% for GOLD vs AdvanSure (Table 3). The average kappa value was 0.80 for GOLD vs ImmunoCAP for all allergens tested, which suggests substantial correlation between the two assay platforms. In detail, 12 of the 24 allergens tested (Japanese hop, dog dander, cat dander, cockroach, ash, dandelion, oxeye, Timothy grass, acacia, reed, oak, and hazelnut) showed perfect correlation, seven allergens (D. pteronyssinus, D. farinae, Alternaria, goldenrod, rye, A. fumigatus, and Cladosporium) showed substantial correlation, four allergens (birch, mugwort, pigweed, and ragweed) showed moderate correlation, and Bermuda grass showed fair correlation. In contrast, the kappa values for ImmunoCAP vs AdvanSure and GOLD vs AdvanSure were both 0.19, indicating a weak correlation.
We then calculated Pearson's correlation coefficient (r) after taking the log of the sIgE concentrations and fluorescence intensities for each pair of assays. The calculated r values were 0.79 for GOLD vs ImmunoCAP, 0.46 for ImmunoCAP vs AdvanSure, and 0.43 for GOLD vs AdvanSure. These results suggested that there is a much stronger correlation between GOLD and ImmunoCAP than between ImmunoCAP and AdvanSure. For GOLD vs ImmunoCAP, the r values were 0.23–0.94, and 21 out of 24 screened allergens scored over 0.60. Moreover, the correlation for D. pteronyssinus and D. farinae, which are the most frequent causal agents of AR, was excellent (r >0.93).
To visually examine the correlations, we selected seven of the most frequent allergens for testing with more than 20 positive serum samples and generated scatter plots showing the concentrations of sIgEs measured by the various assay platforms (Fig. 3). The r values for ImmunoCAP vs AdvanSure ranged between −0.47 and 0.90. For all analyses, the highest agreement and correlations were found between GOLD and ImmunoCAP. These results demonstrate that the GOLD chip-based assays can not only distinguish the positive and negative serum samples with improved accuracy but also provide a precise and quantitative measurement of sIgE concentrations that correlate well with the results of ImmunoCAP.

DISCUSSION

We developed and tested an allergen microarray-based assay platform termed “GOLD chip” for application in in vitro detection of allergic diseases. This platform has several advantages over other platforms. First, the microarray format allows for parallel and simultaneous assays of multiple allergen extracts using a small amount of sample volume (<30 µL), with improved accuracy. Second, the use of a planar substrate allows for the easy removal of the remaining reagents, facilitating automation and shortening the turn-around time. Third, the high-density microarrays allow for high-throughput assays in various biological and medical screening purposes. For example, the microarray format has been utilized for component resolved diagnostics using over 100 component allergens in ImmunoCAP ISAC®. The ImmunoCAP ISAC® offers an improved diagnostics platform by enabling component resolved diagnostics with evident advantages for polysensitized patients with sIgE for pan-allergens and/or cross-reactive allergens. Despite this advantage, the Immuno-CAP ISAC® cannot fully substitute for the MAST platforms due to the limited sensitivity and absence of some important allergenic components. Thus, there is a need for first-line screening platforms that enable screening against multiple allergen sources prior to the identification of causative component allergens [32].
As a substrate for the microarray, we exploited a NHS-functionalized glass coated with hydrophilic polymers. The NHS ester was used to form a covalent bond with the amine functional group on the protein surface, ensuring the irreversible and stable immobilization of allergen probes. The hydrophilic polymer coating was needed to avoid non-specific adsorption of random proteins and to minimize protein fouling, lowering background signals. This platform has advantages over conventional platforms that use non-specific adsorption of allergens on threads or nitrocellulose membranes. First, the loss of immobilized proteins during the course of assays is minimal, thereby increasing the assay sensitivity. Second, the platform is compatible with various detection methods, including optical and electrical monitoring. We were able to adopt fluorescence-based detection for the assay, while other MAST platforms, such as AdvanSure and Advia, utilize chemiluminescence-based technology. Fluorescence detection is considered more sensitive and quantitative within the linear response range [33].
To gauge the performance of the GOLD chip, we performed assays using standard IgE and patient sera, and compared the results with those of ImmunoCAP. Various assay parameters of the GOLD chip showed improved sensitivity and reliability. The determined LOD was below 0.35 IU/mL, which is generally accepted for commercial assay platforms, and the measured CV values were considerably lower than those reported previously [2325]. The GOLD chip-based assays provided quantitative results over a linear concentration range for sIgEs from 0.35 to 100 IU/mL. The target specificity of the GOLD chip-based assays was confirmed by calculating the agreement percentage and Cohen's kappa analysis, showing excellent agreement with ImmunoCAP. Pearson's correlation analysis also indicated that the concordance between ImmunoCAP and GOLD chip was better than that between ImmunoCAP and AdvanSure. These results strongly suggest that the GOLD chip provides a sensitive, reproducible, quantitative, and multiplexed assay platform that can replace current platforms. We also believe the reliability of our assay could be further improved by increasing the number of negative samples, given that in the present analysis the small number of negative samples was outweighed by the large number of positive samples. The GOLD chip can also be used for pre-screening in personalized allergy immunotherapy and detection of antigen-induced human IgG for improving the diagnostics of various immune diseases [343536].

Acknowledgment

This work was supported by the Gachon University Gil Medical Center (grant number: FRD2014-07) and by SMBA (grant number: C0505957).

Notes

Authors' Disclosures of Potential Conflicts of Interest: No potential conflict of interest relevant to this article was reported.

References

1. Heinzerling L, Mari A, Bergmann KC, Bresciani M, Burbach G, Darsow U, et al. The skin prick test - European standards. Clin Transl Allergy. 2013; 3:3. PMID: 23369181.
2. Liccardi G, D'Amato G, Canonica GW, Salzillo A, Piccolo A, Passalacqua G. Systemic reactions from skin testing: literature review. J Investig Allergol Clin Immunol. 2006; 16:75–78.
3. Ownby DR. Allergy testing: in vivo versus in vitro. Pediatr Clin North Am. 1988; 35:995–1009. PMID: 3050840.
4. Berg TLO, Johansson SGO. Allergy diagnosis with the radioallergosorbent test. J Allergy Clin Immunol. 1974; 54:209–221. PMID: 4411916.
5. Gleeson M, Cripps AW, Hensley MJ, Wlodarczyk JH, Henry RL, Clancy RL. A clinical evaluation in children of the Pharmacia ImmunoCAP system for inhalant allergens. Clin Exp Allergy. 1996; 26:697–702. PMID: 8809427.
6. Kelso JM, Sodhi N, Gosselin VA, Yunginger JW. Diagnostic performance characteristics of the standard Phadebas RAST, modified RAST, and Pharmacia CAP system versus skin testing. Ann Allergy. 1991; 67:511–514. PMID: 1958005.
7. Park KH, Lee J, Sim DW, Lee SC. Comparison of singleplex specific IgE detection immunoassays: ImmunoCAP Phadia 250 and Immulite 2000 3gAllergy. Ann Lab Med. 2018; 38:23–31. PMID: 29071815.
8. Banker DD, Daftary VG, Daftary GV, Bhandari NM, Dayanand SR. Allergy diagnosis by Pharmacia Cap System. Indian J Med Sci. 1999; 53:387–389. PMID: 10710831.
9. Han MK, Seo MH, Lee D, Kim SH, Park HS, Kim HS. Optimization of critical factors affecting the performance of an allergen chip for the analysis of an allergen-specific human IgE in serum. Anal Sci. 2007; 23:545–549. PMID: 17495399.
10. Melioli G, Marcomini L, Agazzi A, Bazurro G, Tosca M, Rossi GA, et al. The IgE repertoire in children and adolescents resolved at component level: a cross-sectional study. Pediatr Allergy Immunol. 2012; 23:433–440. PMID: 22103266.
11. Rim JH, Park BG, Kim JH, Kim HS. Comparison and clinical utility evaluation of four multiple allergen simultaneous tests including two newly introduced fully automated analyzers. Pract Lab Med. 2016; 4:50–61. PMID: 28856193.
12. Park KH, Lee J, Lee SC, Son YW, Sim DW, Lee JH, et al. Comparison of the ImmunoCAP Assay and AdvanSure™ AlloScreen Advanced Multiplex Specific IgE Detection Assay. Yonsei Med J. 2017; 58:786–792. PMID: 28540992.
13. Cho JH, Suh JD, Kim JK, Hong SC, Park IH, Lee HM. Correlation between skin-prick testing, individual specific IgE tests, and a multiallergen IgE assay for allergy detection in patients with chronic rhinitis. Am J Rhinol Allergy. 2014; 28:388–391. PMID: 25198024.
14. Templin MF, Stoll D, Schrenk M, Traub PC, Vohringer CF, Joos TO. Protein microarray technology. Trends Biotechnol. 2002; 20:160–166. PMID: 11906748.
15. Jeon HJ, Lee MH, Jang WH, Kwon YK. Intein-mediated protein engineering for biosensor fabrication. Biochip J. 2016; 10:277–287.
16. Park HJ, Lee JH, Park KH, Ann HW, Jin MN, Choi SY, et al. A nationwide survey of inhalant allergens sensitization and levels of indoor major allergens in Korea. Allergy Asthma Immunol Res. 2014; 6:222–227. PMID: 24843797.
17. Choi JS, Roh JY, Lee JR. Clinical availability of component-resolved diagnosis using microarray technology in atopic dermatitis. Ann Dermatol. 2014; 26:437–446. PMID: 25143671.
18. Kim J, Hahm MI, Lee SY, Kim WK, Chae Y, Park YM, et al. Sensitization to aeroallergens in Korean children: a population-based study in 2010. J Korean Med Sci. 2011; 26:1165–1172. PMID: 21935271.
19. Lebrun SJ, Petchpud WN, Hui A, McLaughlin CS. Development of a sensitive, colorometric microarray assay for allergen-responsive human IgE. J Immunol Methods. 2005; 300:24–31. PMID: 15893759.
20. Fall BI, Eberlein-Konig B, Behrendt H, Niessner R, Ring J, Weller MG. Microarrays for the screening of allergen-specific IgE in human serum. Anal Chem. 2003; 75:556–562. PMID: 12585484.
21. Kim TE, Park SW, Cho NY, Choi SY, Yong TS, Nahm BH, et al. Quantitative measurement of serum allergen-specific IgE on protein chip. Exp Mol Med. 2002; 34:152–158. PMID: 12085991.
22. Wiltshire S, O'Malley S, Lambert J, Kukanskis K, Edgar D, Kingsmore SF, et al. Detection of multiple allergen-specific IgEs on microarrays by immunoassay with rolling circle amplification. Clin Chem. 2000; 46:1990–1993. PMID: 11106333.
23. Jahn-Schmid B, Harwanegg C, Hiller R, Bohle B, Ebner C, Scheiner O, et al. Allergen microarray: comparison of microarray using recombinant allergens with conventional diagnostic methods to detect allergen-specific serum immunoglobulin E. Clin Exp Allergy. 2003; 33:1443–1449. PMID: 14519153.
24. Ornberg RL, Harper TF, Liu H. Western blot analysis with quantum dot fluorescence technology: a sensitive and quantitative method for multiplexed proteomics. Nat Methods. 2005; 2:79–81.
25. Deinhofer K, Sevcik H, Balic N, Harwanegg C, Hiller R, Rumpold H. Microarrayed allergens for IgE profiling. Methods. 2004; 32:249–254. PMID: 14962759.
26. Kimiko T, Kunie K, Hitoshi T, Hiroaki M, Satoshi N, Eishin M. Usability of Fag e 2 ImmunoCAP in the diagnosis of buckwheat allergy. Arch Dermatol Res. 2011; 303:635–642. PMID: 21461893.
27. Lucassen R, Schulte-Pelkum J, Csuvarszki C, Kleine-Tebbe J, Fooke M, Mahler M. Evaluation of a novel rapid test system for the detection of allergic sensitization to Timothy grass pollen against established laboratory methods and skin prick test. J Allergy (Cairo). 2010; 2010:pii: 524084. DOI: 10.1155/2010/524084.
28. Saarinen KM, Suomalainen H, Savilahti E. Diagnostic value of skin-prick and patch tests and serum eosinophil cationic protein and cow's milkspecific IgE in infants with cow's milk allergy. Clin Exp Allergy. 2001; 31:423–429. PMID: 11260154.
29. Grier TJ. Laboratory methods for allergen extract analysis and quality control. Clin Rev Allergy Immunol. 2001; 21:111–140. PMID: 11725601.
30. Andri L, Senna G, Betteli C, Givanni S, Andri G, Dimitri G, et al. Local nasal immunotherapy with extract in powder form is effective and safe in grass pollen rhinitis: a double-blind study. J Allergy Clin Immunol. 1996; 97:34–41. PMID: 8568135.
31. Wagenlehner FM, Bschleipfer T, Pilatz A, Weidner W. Pollen extract for chronic prostatitis-chronic pelvic pain syndrome. Urol Clin North Am. 2011; 38:285–292. PMID: 21798390.
32. Jung JH, Kang IG, Kim ST. Comparison of component-resolved diagnosis by using allergen microarray with the conventional tests in allergic rhinitis patients: the first using in Korea. Clin Exp Otorhinolaryngol. 2015; 8:385–389. PMID: 26622959.
33. Schuurman J, Perdok GJ, Lourens TE, Parren PW, Chapman MD, Aalberse RC. Production of a mouse/human chimeric IgE monoclonal antibody to the house dust mite allergen Der p 2 and its use for the absolute quantification of allergen-specific IgE. J Allergy Clin Immunol. 1997; 99:545–550. PMID: 9111501.
34. Lupinek C, Wollmann E, Baar A, Banerjee S, Breiteneder H, Broecker BM. Advances in allergen-microarray technology for diagnosis and monitoring of allergy: the MeDALL allergen-chip. Methods. 2014; 66:106–119. PMID: 24161540.
35. Curin M, Swoboda I, Wollmann E, Lupinek C, Spitzauer S, van Hage M, et al. Microarrayed dog, cat, and horse allergens show weak correlation between allergen-specific IgE and IgG responses. J Allergy Clin Immunol. 2014; 133:918–921. PMID: 24406070.
36. Pomponi D, Di Zenzo G, Zennaro D, Calabresi V, Eming R, Zuzzi S, et al. Detection of IgG and IgE reactivity to BP180 using the ISAC® microarray system. Br J Dermatol. 2013; 168:1205–1214. PMID: 23252883.

SUPPLEMENTARY MATERIAL

Supplemental Data Table S1

Assay sensitivity of detection D. pteronyssinus- and D. farinae-specific IgE in the serum pool using the GOLD chip
alm-38-338-s001.pdf
Fig. 1

Schematic representations of the GOLD chip and the detection of allergen-specific IgEs. (A) Allergens are spotted onto hydrophilic polymer-coated glass slides in triplicate for immobilization. The assay is performed in a silicon chamber and the results are analyzed by fluorescence scanning. (B) Schematic representation of the array. The circles represent the spotting positions of the allergen extracts (in triplicate). (C) The allergens are immobilized via covalent bond formation between the NHS ester on the activated slide and an amine on the protein surface.

alm-38-338-g001
Fig. 2

Analysis of allergen microarray-based assays. (A) The concentration of allergen-specific IgE (sIgE) was measured quantitatively using standard human IgEs. (B) Quantitative evaluation of varying amounts of sIgE for D. pteronyssinus in human sera. (C) Quantitative evaluation of varying amounts of sIgE for D. farinae in human sera. (D) Sera from four patients displaying different sIgE profiles.

alm-38-338-g002
Fig. 3

Correlation analysis of three in vitro diagnostic platforms: GOLD chip, ImmunoCAP, and AdvanSure. The sera of 150 patients with allergy were screened for seven of the most common allergens, and the amount of allergen-specific IgE (sIgE) measured by each platform was plotted for Pearson's correlation analysis.

alm-38-338-g003
Table 1

Characteristics of study participants

alm-38-338-i001
Total number of serum samples 150
Demographic characteristics
 Male/female ratio 99/51
 Age, median (yr) 31.5
 Age, range (yr) 3–73
Clinical symptoms and signs
 Allergic rhinitis (%) 100.0
 Nasal obstruction (%) 70.0
 Asthma (%) 0.7
 Atopic dermatitis (%) 6.0
 Urticaria (%) 1.3
Table 2

Reproducibility of microarray testing of allergen-specific IgE

alm-38-338-i002
CV (%) D. pteronyssinus D. farinae Mugwort Alternaria Japanese hop Birch Cat Dog Cockroach
Intra-assay 3.0 3.8 4.1 1.1 0.4 2.8 2.0 1.6 1.3
Inter-assay 2.4 7.1 11.4 3.2 5.8 4.3 5.7 3.7 3.0
Batch-to-batch 3.6 13.8 25.0 13.4 7.6 5.4 5.6 9.8 4.7

Abbreviation: CV, coefficient of variation.

Table 3

Agreement and correlation analysis between GOLD chip, ImmunoCAP, and AdvanSure

alm-38-338-i003
Allergen Sera (N) GOLD vs ImmunoCAP ImmunoCAP vs AdvanSure GOLD vs AdvanSure
Total %* Positive % κ (95% CI) r§ Total %* Positive % κ (95% CI) r§ Total %* Positive % κ (95% CI) r§ Reference
Total 92.2 94.8 0.80 0.79 66.1 93.5 0.19 0.46 68.0 66.1 0.19 0.43
D. pteronyssinus 46 95.7 95.3 0.73 (0.37 to 1.00) 0.93 91.3 93.3 −0.03 (−0.09 to 0.02) 0.77 91.3 93.3 −0.03 (−0.09 to 0.02) 0.73 29
D. farinae 48 95.8 95.5 0.78 (0.48 to 1.00) 0.94 95.8 100 0.65 (0.20 to 1.00) 0.70 93.8 93.5 0.54 (0.10 to 0.99) 0.72
Alternaria 13 92.3 100 0.76 (0.31 to 1.00) 0.91 76.9 90.0 0.26 (−0.35 to 0.88) 0.90 84.6 90.9 0.41 (−0.27 to 1.00) 0.86 29
Mugwort 20 85.0 100 0.50 (0.05 to 0.95) 0.88 77.3 93.3 0.40 (−0.04 to 0.83) 0.52 80.0 93.8 0.23 (−0.29 to 0.75) 0.47 30
Japanese hop 20 100 100 1.00 0.81 95.0 95.0 0.00 0.58 95.0 95.0 0.54
Dog 20 95.0 93.8 0.86 (0.59 to 1.00) 0.89 85.0 94.1 0.32 (−0.27 to 0.90) 0.87 83.3 88.2 −0.08 (−0.19 to 0.03) 0.76 29
Cat 26 100 100 1.00 0.78 84.6 91.7 −0.08 (−0.17 to 0.00) 0.77 84.6 91.7 −0.08 (−0.17 to 0.00) 0.47
Birch 8 87.5 85.7 0.60 (−0.07 to 1.00) 0.88 75.0 85.7 −0.14 (−0.34 to 0.05) 0.21 62.5 71.4 −0.20 (−0.49 to 0.09) 0.34 31
Cockroach 14 92.9 90.9 0.81 (0.46 to 1.00) 0.80 85.7 91.7 0.42 (−0.25 to 1.00) 0.69 85.7 83.3 0.59 (0.11 to 1.00) 0.71
Ash 7 100 100 1.00 0.92 85.7 100 0.59 (−0.09 to 1.00) 0.56 85.7 100 0.59 (−0.09 to 1.00) 0.30
Dandelion 11 90.9 100 0.81 (0.47 to 1.00) 0.89 63.6 83.3 0.24 (−0.30 to 0.78) 0.71 81.8 75.0 0.62 (0.18 to 1.00) 0.74
Oxeye 10 100 100 1.00 0.88 60.0 75.0 0.23 (−0.32 to 0.78) 0.49 60.0 50.0 0.23 (−0.32 to 0.78) 0.45 29
Pigweed 9 77.8 100 0.53 (0.02 to 1.00) 0.83 55.6 100 0.25 (−0.11 to 0.61) −0.39 55.6 33.3 0.25 (−0.11 to 0.61) −0.52
Goldenrod 10 90.0 80.0 0.80 (0.44 to 1.00) 0.77 70.0 100 0.40 (−0.05 to 0.85) 0.57 50.0 42.9 0.07 (−0.43 to 0.58) 0.58
Timothy 9 100 100 1.00 0.74 11.1 100 0.00 0.07 33.3 14.3 0.07 (−0.09 to 0.23) 0.46
Acacia 6 100 100 1.00 0.72 66.7 75.0 0.25 (−0.56 to 1.00) 0.13 83.3 80.0 0.57 (−0.12 to 1.00) 0.69
Ragweed 8 71.4 100 0.46 (−0.07 to 0.99) 0.66 57.1 100 0.13 (−0.15 to 0.40 ) 0.76 57.1 60.0 0.28 (−0.15 to 0.70) 0.10 29
Rye 12 91.7 100 0.75 (0.30 to 1.00) 0.60 25.0 100 0.04 (−0.05 to 0.12) 0.09 33.3 27.3 0.06 (−0.07 to 0.19) 0.29 29
A. fumigatus 7 85.7 100 0.70 (0.17 to 1.00) 0.55 42.9 100 0.13 (−0.15 to 0.40) 0.60 57.1 50.0 0.22 (−0.20 to 0.64) 0.08
Cladosporium 13 84.6 75.0 0.64 (0.18 to 1.00) 0.24 30.8 75.0 −0.09 (−0.42 to 0.24) 0.49 23.1 20.0 −0.28 (−0.72 to 0.17) −0.06
Bermuda 14 71.4 60.0 0.38 (−0.13 to 0.88) 0.23 50.0 100 0.17 (−0.07 to 0.41) 0.08 50.0 50.0 0.00 (−0.37 to 0.37) 0.44
Reed 3 100 100 1.00 1.0 66.7 100 0.40 (−0.37 to 1.00) 0.91 66.7 50.0 0.40 (−0.37 to 1.00) 0.94
Oak 3 100 100 1.00 1.0 66.7 100 0.00 0.85 66.7 66.7 0.00 0.83 29
Hazel 3 100 100 1.00 1.0 66.7 100 0.00 −0.47 66.7 66.7 0.00 −0.54

*Total agreement percentage; Positive agreement percentage; Cohen's kappa coefficient: perfect, 0.8–1.0; substantial, 0.6–0.8; moderate, 0.4–0.6; fair, 0.2–0.4; poor, <0.2; §Pearson's correlation coefficient.

TOOLS
Similar articles