Park, Park, Kim, Han, Han, Cho, and Jang: The New Sysmex XN-2000 Automated Blood Cell Analyzer More Accurately Measures the Absolute Number and the Proportion of Hematopoietic Stem and Progenitor Cells Than XE-2100 When Compared to Flow Cytometric Enumeration of CD34+ Cells
The gold standard to count circulating hematopoietic stem and progenitor cells (HPCs) is to enumerate CD34+ cells by using flow cytometry [1, 2]. XE-2100 (Sysmex, Kobe, Japan), a currently used hematology analyzer, can be used to quantify HPCs, generating counts that reportedly correlate with those obtained using flow cytometric analysis of CD34+ cells; however, XE-2100 HPC counts were 1.86-fold higher than those calculated by flow cytometry [3]. The recently launched XN-2000 (Sysmex) can also be used to quantify HPCs, but its performance on HPC counts has not been evaluated. We assessed the HPC counting performance of XE-2100 and XN-2000 and compared them to CD34+ cell counts by flow cytometry.
A total of 120 peripheral blood samples collected from 31 patients who had undergone stem cell transplantation (SCT) at Asan Medical Center, Seoul, Korea between June 2013 and October 2013 were randomly selected. The CD34+ cell proportion (%) was measured in each sample by using flow cytometric enumeration of CD34+ HPCs [1], performed using FACScanto II (Becton-Dickinson, Sunnyvale, CA, USA) and phycoerythrin (PE)-conjugated CD34 and fluorescein isothiocyanate (FITC)-conjugated CD45 (Becton-Dickinson). CD34+ cell numbers obtained using XE-2100 and XN-2000 analyzers were calculated by multiplying their respective white blood cell (WBC) counts with their CD34+ cell proportion (%). As a reference for the HPC counts, the CD34+ cell counts, defined as the means of XE-2100 and XN-2000 CD34+ cell numbers, were generated. HPC numbers and proportion (%) were measured by both XE-2100 (immature myeloid information [IMI] channel, XE-2100 HPC numbers, and proportion [%]) and XN-2000 (white precursor cell [WPC] channel, XN-2000 HPC numbers and proportion [%]). Linear regression and Bland-Altman analyses were applied to compare XE-2100 and XN-2000 HPC proportion (%) and the CD34+ cell proportion obtained using flow cytometry. An identical analysis was performed comparing XE-2100 and XN-2000 HPC numbers and CD34+ cell counts.
HPC count-related parameters for 120 samples obtained using the two automated blood cell analyzers were compared. XN-2000 HPC numbers were significantly lower than those obtained using XE-2100 (0.005×109/L vs. 0.007×109/L, P<0.001). The regression analysis of XE-2100 HPC and CD34+ cell proportion (%) showed poor correlation (γ=0.187). In contrast, the proportion (%) of XN-2000 HPCs yielded correlations that were significantly improved (γ=0.590). XE-2100 HPC and CD34+ cell proportion (%) showed a mean bias of 0.2%, suggesting that HPC proportion (%) measured using XE-2100 is overestimated by a mean of 0.2% when compared to those measured using flow cytometry (Fig. 1A). In contrast, HPC proportion (%) measured using XN-2000 produced an improved mean bias of 0.0% (Fig. 1B).
Subsequent analysis of HPC numbers confirmed that XN-2000 showed improved performance over XE-2100. The regression analysis between HPC numbers measured using the two automated blood cell analyzers and CD34+ cell numbers showed that XN-2000 had improved correlation (γ=0.652) compared to XE-2100 (γ=0.548). The analysis also showed that bias was reduced when XN-2000 was used instead of XE-2100 (0.00×109/L vs. 0.04×109/L) (Fig. 1C and D). The cutoff point in the CD34+ cell counts for deciding stem cell collection time in SCT is reported to be 0.01×109/L-0.02×109/L [4, 5, 6]. Thus, identical stem cell collection cutoff counts in HPC numbers for XN-2000 can be calculated as 0.017×109/L-0.021×109/L by using the regression equation determined in our study (XN-2000 HPC count=0.4006×CD34+ cell count+0.0133, γ=0.652).
In our study, correlations (γ=0.548) between XE-2100 HPC numbers and those obtained using flow cytometry were weaker than those obtained in a previous study (γ=0.7482) [3]. This discrepancy might be attributed to the small sample size in our study (120 vs. 236), or higher proportion of samples with low or high WBC counts (<5.0×109/L or >40.0×109/L) in our study (40.8%) than in the previous study (34.3%), since the discrepancy between XE-2100 HPC numbers and those obtained using flow cytometry were more frequently reported in samples with low or high WBC counts than in normal samples [3]. Because our study population included patients with hematologic malignancy, i.e., unhealthy donors for allogeneic SCT, selection bias may also contribute to this discrepancy. In this study, we demonstrated that HPC counts are overestimated when XE-2100 was used when compared to flow cytometry, and the degree of overestimation is significantly reduced by using XN-2000 instead. Therefore, these results suggest that XN-2000 generates more accurate HPC counts than XE-2100, when compared with flow cytometric enumeration of CD34+ cells.
In conclusion, XN-2000 can provide more accurate HPC-related data than XE-2100, providing more accurate information to clinicians in terms of the timing of stem cell collection for SCT.

Acknowledgments

The Sysmex Company (Kobe, Japan) provided the Sysmex XN-2000 automated blood cell analyzer used in this study.

Notes

No potential conflicts of interest relevant to this article were reported.

References

1. Gratama JW, Orfao A, Barnett D, Brando B, Huber A, Janossy G, et al. Flow cytometric enumeration of CD34+ hematopoietic stem and progenitor cells. European Working Group on Clinical Cell Analysis. Cytometry. 1998; 34:128–142. PMID: 9696157.
2. Elliott C, Samson DM, Armitage S, Lyttelton MP, McGuigan D, Hargreaves R, et al. When to harvest peripheral-blood stem cells after mobilization therapy: prediction of CD34-positive cell yield by preceding day CD34-positive concentration in peripheral blood. J Clin Oncol. 1996; 14:970–973. PMID: 8622047.
crossref
3. Letestu R, Marzac C, Audat F, Belhocine R, Tondeur S, Baccini V, et al. Use of hematopoietic progenitor cell count on the Sysmex XE-2100 for peripheral blood stem cell harvest monitoring. Leuk Lymphoma. 2007; 48:89–96. PMID: 17327950.
crossref
4. Vaughn ML, Waller EK. Monitoring blood for CD34+ cells to determine timing of hematopoietic progenitor cells apheresis. In : Kolonin MG, Simmons PJ, editors. Stem cell mobilization: methods and protocols, methods in molecular biology, vol. 904. New York: Springer;2012. p. 79–83.
5. Chapple P, Prince HM, Quinn M, Bertoncello I, Juneja S, Wolf M, et al. Peripheral blood CD34+ cell count reliably predicts autograft yield. Bone Marrow Transplant. 1998; 22:125–130. PMID: 9707018.
crossref
6. Linn YC, Heng KK, Rohimah S, Goh YT. Peripheral blood progenitor cell mobilization in three groups of subjects: a comparison of leukapheresis yield and timing. J Clin Apher. 2000; 15:217–223. PMID: 11124688.
crossref
Fig. 1
Bland-Altman analysis results for HPC proportion (%) measured using XE-2100 (A), XN-2000 (B) automated blood cell analyzers, and the CD34+ cell proportion (%) measured using flow cytometry are shown. Bland-Altman analysis results comparing absolute HPC numbers measured using XE-2100 (C) and XN-2000 (D) analyzers and CD34+ cell counts are also shown.
Abbreviations: CD, cluster of differentiation; HPC, hematopoietic stem and progenitor cells.
alm-35-146-g001