Abstract
Purpose
We studied early and delayed contrast-enhanced CT to determine the effects of the degree of luminal enhancement on the measurement of plaque size compared to T1-weighted MRI.
Materials and Methods
T1-weighted MRI and a two-phase contrast-enhanced CT was performed in 5 New Zealand white rabbits with atherosclerosis. Early-phase images were acquired during an expected peak enhancement period of the lumen; delayed-phase images were acquired 240 sec after administration of the contrast media. Anteroposterior and lateral luminal diameters (APD, LD), luminal area (LA), total vessel area (TVA), and plaque area (PA) of the aorta were measured on MRI and CT, respectively and compared to each other.
Results
A total of 78 slices of the aorta were analyzed. PA, measured on T1-weighted MR images, was significantly greater than PA for both early-phase and delayed-phase CT (p < 0.001). The mean difference in PA was 12.13 ± 5.52 mm2 (p < 0.001), 11.51 ± 4.37 mm2 (p < 0.001), -0.62 ± 3.87 mm2 (p = 0.159) for MRI vs. early-phase CT, MRI vs. delayed-phase CT, and early-phase CT vs. delayed-phase CT, respectively.
With the emergence of multi-detector computed tomography (MDCT), contrast-enhanced coronary computed tomographic (CT) angiography has been widely used for detecting obstructive coronary artery disease (123). A unique advantage of CT compared to other noninvasive imaging modalities for the evaluation of coronary artery disease is the fact that contrast-enhanced CT shows not only calcified, but also non-calcified atherosclerotic plaques. Therefore many investigators have studied coronary atherosclerotic plaques by CT (45678). However, the detection and measurement of the size and density of plaques are influenced by the strong enhancement of the arterial lumen, which is required for detecting obstructive coronary artery disease (9).
Compared to CT, magnetic resonance imaging (MRI) is superior in terms of plaque characterization with multi-contrast imaging (101112). The measurement of plaque size by T1-weighted imagery is not influenced by luminal intensity. However, the current technology of MRI is limited in the evaluation of atherosclerotic plaques in human coronary arteries because of limited spatial and temporal resolution. Therefore, thus far, most studies on atherosclerotic plaques with MRI have used larger plaques in human carotid arteries or atherosclerotic animal models (1213141516).
A rabbit's atherosclerotic aorta is a well-established experimental model for imaging plaques representing those in human coronary arteries because the diameter of rabbit's aorta is similar to that of human coronary arteries and the method for developing an atherosclerotic rabbit model, consisting of a high cholesterol diet and balloon injury technique, has been proven as being efficient (817).
In this study, using an atherosclerotic rabbit model, we investigated the effect of the degree of luminal enhancement on the CT measurement of plaque size on early and delayed CT images compared with T1-weighted MRI.
New Zealand white rabbits (n = 5, weight 3.0 to 3.5 kg, all male) were used for the animal model of atherosclerosis. Atherosclerotic aortic lesions were induced as previously described (18). The study protocol was approved by our Animal Care and Use Committee.
All MR imaging was performed under general anesthesia of the animals by administering intramuscular tiletamine (20 mg/kg, Zoletil; Virbac, France) and xylazine (5 mg/kg, Rompun; Bayer, Korea). The rabbits were imaged in the craniocaudal direction and in the supine position with a 1.5 tesla MRI unit (Achieva, Philips Medical Systems, City, The Netherlands) using a 5-channel cardiac coil. Using a fast spin-echo sequence, sequential 2-dimensional axial T1-weighted images of the aorta measuring 3-mm thick, were obtained from the arch to the iliac bifurcation. The typical parameters included the following parameters: TR/TE= 500/12 msec, field of view = 9 cm, matrix 256 × 256, signal average = 4, and a slice spacing of 9 mm. Fat suppression was applied and the in-plane resolution was 0.35 × 0.35 mm.
The CT was performed with a 64-slice multi-detector CT scanner (LightSpeed VCT XT, GE Healthcare). A 21-gauge intravenous access was placed in the ear vein. CT and MRI was performed within 48 hours of each other.
We used a 1:2 dilution of the contrast agent (320mg iodine/mL) with saline, yielding a concentration of 100-110 mg iodine per milliliter. With this concentration, the aortic luminal attenuation was similar to that of routine human coronary artery CT angiography (250 to 350 HU).
General anesthesia using intramuscular tiletamine (20 mg/kg, Zoletil; Virbac, France) and xylazine (5 mg/kg, Rompun; Bayer, Korea) was performed for imaging. A CT scan was performed in the craniocaudal direction from the aortic arch to iliac bifurcation in the supine position.
The scan start for early-phase of the study was determined using a real-time bolus-tracking technique. The region of interest (ROI) was drawn in the descending thoracic aorta to monitor a threshold of 100 HU above baseline attenuation. The time delay to start a scan was 4 sec after a threshold trigger of 100 HU was attained. A total of 15 mL of the diluted contrast was administered at a rate of 0.5 mL/sec using an automated infusion pump (Medrad Envision CT, Medrad Inc, Indianola, U.S.A.). Delayed-phase scanning began 240 sec after the administration of contrast media. The 240-sec time point for the delayed phase was selected with the preliminary scans of three rabbits, which was visually optimal in distinguishing the lumen and plaques.
The imaging parameters were as follows: axial mode, collimation 0.625 mm × 64, 120 kV, 180 mA, and a rotation time 0.5s. The total acquisition time ranged from 14 to 17 sec. Axial images were reconstructed with a field of view of 160 × 160 mm, a 512 × 512 matrix, and a slice thickness of 3 mm with a soft convolution kernel compared with a 3-mm thick MRI.
The MR and CT images were transferred to a dedicated workstation (Aquarius, Terrarecon Inc, San Maeto, U.S.A.) for analysis. CT images were reformatted with a 3-mm thickness in order to match the 3-mm thick MR images. We measured anterior-posterior diameter (APD), lateral diameter (LD), luminal area (LA), and total vessel area (TVA). APD was defined as anterior-posterior diameter between the inner margins of the vessel wall parallel to the anterior-posterior axis of vertebrae. LD was defined as the orthogonal diameter of APD between the inner margins of the vessel wall. Measurements for APD, LD, LA, and TVA were manually traced (Fig. 1). Plaque area (PA) was calculated by subtracting LA from TVA. An experienced radiologist performed all the measurements. Measurement on CT was performed with a window setting of 450 and a level setting of 70 for the standardized comparison. Finally the measurements on MR, early-phase CT, and delayed-phase CT images were compared.
The agreement between measurements for APD, LD, LA, and TVA by CT and MRI was evaluated with the Bland-Altman test. The statistical significance of the differences of each measurement of plaque according to different imaging methods was assessed using Student's t-test for paired samples. P-values less than 0.05 were considered to be statistically significant. MedCalc software (Version 8.2.0.1, Mariakerke, Belgium) was used for the statistical evaluation.
The number of analyzed slices from the five rabbit aortas was 78, while the mean value of the luminal density was 291.8 ± 10.6 HU in early phase CT and 145.4 ± 9.2 HU in delayed phase CT.
Table 1 summarized each measurement from different imaging methods. Significant agreement between MRI and CT measurements was noted for the vessel diameters and areas (Figs. 2, 3). The APD, LD, and LA measured on T1-weighted MR images were significantly smaller than those on early-phase CT and delayed-phase CT (p < 0.001). TVA and PA measured on T1-weighted MR images were significantly greater than those on early-phase CT and delayed-phase CT (p < 0.001). Compared to both early-phase and delayed-phase CT, MRI tended to provide smaller values for APD, LD, LA and larger values for TVA, especially when the dimension was larger.
Highly significant agreement was achieved between early-phase and delayed-phase CT measurements of vessel diameters and areas (Fig. 4). APD and LA measured on early-phase CT was significantly greater than those on delayed phase CT (p < 0.001 and p = 0.021, respectively). However, no significant difference in the measurement of LD, TVA, and PA between early-phase CT and delayed-phase CT (p=0.905, 0.641, 0.159, respectively).
The mean difference of PA was 12.13 ± 5.52 mm2(p < 0.001), 11.51 ± 4.37 mm2 (p < 0.001), -0.62 ± 3.87 mm2 (p = 0.159) for MRI vs. early-phase CT, MRI vs. delayed-phase CT, and early-phase CT vs. delayed-phase CT, respectively.
With introduction of 64 multi-detector computed tomography, contrast-enhanced coronary CT angiography enables the detection of obstructive coronary artery disease but also noncalcified plaque (12345678). However, the accuracy of coronary CT to detect non-calcified coronary plaque has not been well known.
Many investigators have studied plaque detection and size measurement of coronary arteries on coronary CT angiography compared with intravascular ultrasonography (IVUS) as the standard reference. The sensitivities and specificities of CT for the detection of coronary artery segments with plaque have been reported as quite high between 80 and 90% and there was close correlation between CT and IVUS for a plaque cross-sectional area, volume of single plaques, and plaque volume per coronary segment according to the literature (671920). In the previous study at our institute, while the correlation coefficient for measurements of non-calcified plaques between CT and IVUS were relatively high (r = 0.708), CT significantly underestimated the plaque area compared with IVUS (6). The limited spatial resolution of CT accounts for the difficulty in accurately quantifying plaque area by CT. CT may not measure plaques accurately, especially when their thickness is relatively low. This means that early plaque with diffuse thickening of the coronary arterial wall could not be detected by CT.
Furthermore, the detection and measurement of the size of plaques can be influenced by strong enhancement of the arterial lumen, which is required for detecting obstructive coronary artery disease. Cademartiri et al. (9) reported that the intracoronary attenuation significantly modified the attenuation of plaques assessed by CT. However, there has been no report describing that intracoronary attenuation affects the measurement of the plaque area as a result of a CT. All the previous studies regarding the measurement of plaque area with CT adopted the coronary artery angiography protocol, which targeted peak enhancement of the arterial lumen to maximize contrast between the lumen and plaque or surrounding tissue, in order to easily detect obstructive coronary lesions. In the present study, we first selected a 240-sec delayed phase that showed visually optimal arterial luminal enhancement that distinguished the lumen and plaques. Then, we compared early phase and delayed phase to identify the influence of luminal density on the measurement of plaque areas.
The anteroposterior luminal diameter and luminal area measured on early-phase CT were greater than those on delayed-phase CT. Consequently, the plaque areas measured on early phase were smaller than those measured on the delayed phase. However, the difference was not statistically significant. Compared with MRI, both phases of CT underestimated the plaque area. Therefore, both the early-phase CT using the usual coronary CT angiography protocol and the delayed-phase CT, significantly underestimates plaque area compared with MRI.
MDCT is now widely used for the detection of coronary artery disease including obstructive lesions and non-obstructive non-calcified lesions in clinical practice. The low incidence of non-calcified plaques in asymptomatic patients with a high risk of coronary artery disease may be explained by the underestimation of non-calcified plaque with CT (67). In particular, the potential utility of CT to monitor plaque change by statin therapy may be limited by the low sensitivity of CT for detecting non-calcified plaque when it is small.
Limitations of this study were as follows: MRI might underestimate the luminal area because of a blood flow artifact at the peripheral lumen near to the vessel wall, which might be a factor in overestimating the plaque area by MRI. We did not use a pathologic specimen as a standard reference because shrinkage of the vessel wall during formalin fixation after sacrificing the animals was substantial as compared with in vivo images. We did not consider the possibility and influence of plaque enhancement in the delayed phase CT, in which the influence seemed to be minimal in terms of measuring the plaque area. The CT protocol in this study was not the same as the coronary CT angiography in clinical practice. Interobserver and intraobserver agreement were not obtained which was known to be the cause of ~30% variability for plaque volume in other studies (672122). Finally, we did not use control animals to compare the measurement of the vessel wall under normal and atherosclerotic conditions.
Luminal density by contrast enhancement may affect the measurement of the luminal area and plaque area but the difference is not significant. Both early-phase CT and delayed-phase CT significantly underestimate plaque areas compared to MRI. Therefore, visualization and measurement of non-calcified plaques of coronary arteries with CT in clinical practice may not accurately represent significant atherosclerotic changes in the coronary arterial wall.
Figures and Tables
Table 1
Note.─ All numbers represent the mean ± standard deviation.
The unit for numbers is mm2.
T1W-MRI = T1-weighted magnetic resonance image
eCT = early-phase CT
dCT = delayed-phase CT
APD = anterior-posterior diameter
LD = lateral diameter
LA = luminal area
TVA = total vessel area
PA - plaque area, which is calculated by subtracting luminal area from total vessel area
References
1. Mowatt G, Cook JA, Hillis GS, Walker S, Fraser C, Jia X, et al. 64-slice computed tomography angiography in the diagnosis and assessment of coronary artery disease: systematic review and meta-analysis. Heart. 2008; 94:1386–1393.
2. Budoff MJ, Dowe D, Jollis JG, Gitter M, Sutherland J, Halamert E, et al. Diagnostic performance of 64-multidetector row coronary computed tomographic angiography for evaluation of coronary artery stenosis in individuals without known coronary artery disease: results from the prospective multicenter ACCURACY (assessment by coronary computed tomographic angiography of individuals undergoing invasive coronary angiography) trial. J Am Coll Cardiol. 2008; 52:1724–1732.
3. Miller JM, Rochitte CE, Dewey M, Arbab-Zadeh A, Niinuma H, Gottlieb I, et al. Diagnostic performance of coronary angiography by 64-row CT. N Engl J Med. 2008; 359:2324–2336.
4. Schepis T, Marwan M, Pflederer T, Seltmann M, Ropers D, Daniel WG, et al. Quantification of non-calcified coronary atherosclerotic plaques with dual-source computed tomography: comparison with intravascular ultrasound. Heart. 2010; 96:610–615.
5. Yang WI, Hur J, Ko YG, Choi BW, Kim JS, Choi D, et al. Assessment of tissue characteristics of noncalcified coronary plaques by 64-slice computed tomography in comparison with integrated backscatter intravascular ultrasound. Coron Artery Dis. 2010; 21:168–174.
6. Hur J, Kim YJ, Lee HJ, Nam JE, Choe KO, Seo JS, et al. Quantification and characterization of obstructive coronary plaques using 64-slice computed tomography: a comparison with intravascular ultrasound. J Comput Assist Tomogr. 2009; 33:186–192.
7. Leber AW, Becker A, Knez A, von Ziegler F, Sirol M, Nikolaou K, et al. Accuracy of 64-slice computed tomography to classify and quantify plaque volumes in the proximal coronary system: a comparative study using intravascular ultrasound. J Am Coll Cardiol. 2006; 47:672–677.
8. Viles-Gonzalez JF, Poon M, Sanz J, Rius T, Nikolaou K, Fayad ZA, et al. In vivo 16-slice, multidetector-row computed tomography for the assessment of experimental atherosclerosis: comparison with magnetic resonance imaging and histopathology. Circulation. 2004; 110:1467–1472.
9. Cademartiri F, Mollet NR, Runza G, Bruining N, Hamers R, Somers P, et al. Influence of intracoronary attenuation on coronary plaque measurements using multislice computed tomography: observations in an ex vivo model of coronary computed tomography angiography. Eur Radiol. 2005; 15:1426–1431.
10. Yuan C, Kerwin WS, Ferguson MS, Polissar N, Zhang S, Cai J, et al. Contrast-enhanced high resolution MRI for atherosclerotic carotid artery tissue characterization. J Magn Reson Imaging. 2002; 15:62–67.
11. Zaho XQ, Phan BA, Chu B, Bray F, Moore AB, Polissar NL, et al. Testing the hypothesis of atherosclerotic plaque lipid depletion during lipid therapy by magnetic resonance imaging: study design of carotid plaque composition study. Am Heart J. 2007; 154:239–246.
12. Cai J, Hatsukami TS, Ferguson MS, Kerwin WS, Saam T, Chu B, et al. In vivo quantitative measurement of intact fibrous cap and lipid-rich necrotic core size in atherosclerotic carotid plaque: comparison of high-resolution, contrast-enhanced magnetic resonance imaging and histology. Circulation. 2005; 112:3437–3444.
13. Briely-Saebo KC, Mulder WJ, Mani V, Hyafil F, Amirbekian V, Aguinaldo JG, et al. Magnetic resonance imaging of vulnerable atherosclerotic plaques: current imaging strategies and molecular imaging probes. J Magn Reson Imaging. 2007; 26:460–479.
14. Kerwin WS, Zhao X, Yuan C, Hatsukami TS, Maravilla KR, Underhill HR, et al. Contrast-enhanced MRI of carotid atherosclerosis: dependence on contrast agent. J Magn Reson Imaging. 2009; 30:35–40.
15. Dong L, Wang J, Yarnykh VL, Underhill HR, Neradilek MB, Polissar N, et al. Efficient flow suppressed MRI improves interscan reproducibility of carotid atherosclerosis plaque burden measurements. J Magn Reson Imaging. 2010; 32:452–458.
16. Helft G, Worthley SG, Fuster V, Zaman AG, Schechter C, Osende JI, et al. Atherosclerotic aortic component quantification by noninvasive magnetic resonance imaging: an in vivo study in rabbits. J Am Coll Cardiol. 2001; 37:1149–1154.
17. Courtman DW, Schwartz SM, Hart CE. Sequential injury of the rabbit abdominal aorta induces intramural coagulation and luminal narrowing independent of intimal mass: extrinsic pathway inhibition eliminates luminal narrowing. Circ Res. 1998; 82:996–1006.
18. Choi BW, Hur J, Lee HJ, Kim YJ, Kim TH, Choe KO. Gadolinium-enhanced magnetic resonance imaging of atherosclerotic plaques in comparison with histopathology: an in vivo study in aorta of rabbits. J Korean Soc Magn Reson Med. 2009; 13:81–87.
19. Petranovic M, Soni A, Bezzera H, Loureiro R, Sarwar A, Raffel C, et al. Assessment of nonstenotic coronary lesions by 64-slice multidetector computed tomography in comparison to intravascular ultrasound: evaluation of nonculprit coronary lesions. J Cardiovasc Comput Tomogr. 2009; 3:24–31.
20. Sun J, Zhang Z, Lu B, Yu W, Yang Y, Zhou Y, et al. Identification and quantification of coronary atherosclerotic plaques: a comparison of 64-MDCT and intravascular ultrasound. AJR Am J Roentgenol. 2008; 190:748–754.
21. Hoffmann H, Frieler K, Hamm B, Dewey M. Intra- and interobserver variability in detection and assessment of calcified and noncalcified coronary artery plaques using 64-slice computed tomography: variability in coronary artery plaque measurement using MSCT. Int J Cardiovasc Imaging. 2008; 24:735–742.
22. Pflederer T, Schmid M, Ropers D, Ropers U, Komatsu S, Daniel WG, et al. Interobserver variability of 64-slice computed tomography for the quantification of non-calcified coronary atherosclerotic plaque. Rofo. 2007; 179:953–957.