Journal List > Ann Clin Microbiol > v.17(2) > 1078512

Uh, Jang, Park, Kim, Seo, Yoon, Choi, Kim, and Kim: Factors Influencing the False Positive Signals of Continuous Monitoring Blood Culture System

초록

Background

The false positive signals of a continuous monitoring blood culture system (CMBCS) in-crease the reporting time and laboratory cost. This study aimed to determine the highly relevant variables that discriminate false positive signals from true positive signals in a CMBCS.

Methods

Among 184,363 blood culture sets (aerobic and anaerobic), the signal-positive samples according to a BACTEC FX system (Plus Aerobic/F, BDA; Plus Anaerobic/F, BDN) and BacT/Alert 3D system (Standard Aerobic, BSA; Standard Anaerobic, BSN) between April 2010 and November 2013 were classi-fied into two groups: false positive or true positive signals. The data of 15 parameters between the two groups were then statistically compared.

Results

Among total blood cultures, the positive rates of CMBCS signals according to BDA, BDN, BSA, and BSN were 4.9%, 2.8%, 3.8%, and 3.2%, respectively. The false positive rates of CMBCS signals according to BDA, BDN, BSA, and BSN were 0.6%, 0.1%, 0.1%, and 0.1%, respectively. The blood volume, detection time, time interval between admission and test, C-reactive protein concentration, leukocyte count, delta neutrophil index, and mean peroxidase index showed statistically significant differences between the two groups.

Conclusion

There were no variables with diagnostic sensitivity and specificity for discriminating the two groups. Therefore, analysis of bacterial growth curves produced by CMBCS is needed for early and effec-tive detection of false positive signals.

REFERENCES

1.Reimer LG., Wilson ML., Weinstein MP. Update on detection of bacteremia and fungemia. Clin Microbiol Rev. 1997. 10:444–65.
crossref
2.Petti CA., Weinstein MP., Carroll KC. Systems for Detection and Identification of Bacteria and Yeasts. Versalovic J, Carroll KC, editors. Manual of Clinical Microbiology. 10th ed.Washington, DC: ASM Press;2011. p. 15–26.
3.Kim S. Development of blood culture and quality improvement. Ann Clin Microbiol. 2013. 16:153–61.
crossref
4.Adler H., Baumlin N., Frei R. Evaluation of acridine orange staining as a replacement of subcultures for BacT/ALERT-positive, gram stain-negative blood cultures. J Clin Microbiol. 2003. 41:5238–9.
crossref
5.Shim H., Kim K., Uh Y., Seo D., Kim H., Yoon Y. The development and evaluation of blood volume measuring system for blood culture quality improvement. J Test Eval. 2012. 40:521–6.
crossref
6.Nahm CH., Choi JW., Lee J. Delta neutrophil index in automated immature granulocyte counts for assessing disease severity of patients with sepsis. Ann Clin Lab Sci. 2008. 38:241–6.
7.Jin HK., Jang JY., Uh Y., Kwon O., Yoon KJ., Kim HY, et al. The value of serum procalcitonin level for differentiation between contaminants and pathogens in bacteremia. Korean J Clin Microbiol. 2011. 14:7–12.
crossref
8.Nolte FS., Williams JM., Jerris RC., Morello JA., Leitch CD., Matushek S, et al. Multicenter clinical evaluation of a continuous monitoring blood culture system using fluorescent-sensor techn-ology (BACTEC 9240). J Clin Microbiol. 1993. 31:552–7.
crossref
9.Dreyer AW., Ismail NA., Nkosi D., Lindeque K., Matthews M., van Zyl DG, et al. Comparison of the VersaTREK blood culture system against the Bactec9240 system in patients with suspected bloodstream infections. Ann Clin Microbiol Antimicrob. 2011. 10:4.
crossref
10.Wilson ML., Weinstein MP., Reimer LG., Mirrett S., Reller LB. Controlled comparison of the BacT/Alert and BACTEC 660/730 nonradiometric blood culture systems. J Clin Microbiol. 1992. 30:323–9.
crossref
11.Rohner P., Pepey B., Auckenthaler R. Comparison of BacT/Alert with Signal blood culture system. J Clin Microbiol. 1995. 33:313–7.
crossref
12.Mirrett S., Hanson KE., Reller LB. Controlled clinical comparison of VersaTREK and BacT/ALERT blood culture systems. J Clin Microbiol. 2007. 45:299–302.
crossref
13.Levi MH., Gialanella P., Motyl MR., McKitrick JC. Rapid detection of positive blood cultures with the BACTEC NR-660 does not re-quire first-day subculturing. J Clin Microbiol. 1988. 26:2262–5.
crossref
14.Zhu LW., Yang XM., Xu XQ., Xu J., Lu HJ., Yan LX. Analysis of false-positive reaction for bacterial detection of blood samples with the automated microbial detection system BacT/ALERT 3D. Zhongguo Shi Yan Xue Ye Xue Za Zhi. 2008. 16:1207–10.

Table 1.
Number (%) of signal positive blood cultures by continuous monitoring blood culture systems
Blood culture results No. (%) of blood culture bottle
BACTEC FX system BacT/Alert 3D system
  Aerobic Anaerobic Aerobic Anaerobic
Total number 63,344 (100) 63,344 (100) 121,519 (100) 121,519 (100)
Signal negative 60,266 (95.1) 61,584 (97.2) 116,928 (96.2) 117,665 (96.8)
Signal positive 3,078 (4.9) 1,760 (2.8) 4,591 (3.8) 3,854 (3.2)
 True positive 2,739 (4.3) 1,675 (2.7) 4,437 (3.7) 3,697 (3.1)
 False positive 339 (0.6) 85 (0.1) 154 (0.1) 157 (0.1)
Table 2.
Data comparison of false positive and true positive signals obtained from BACTEC FX system and BacT/Alert 3D system by according to parameters
Parameters BACTEC FX system BacT/Alert 3D system
Aerobic Anaerobic Aerobic Anaerobic
False positive True positive P value AUC False positive True positive P value AUC False positive True positive P value AUC False positive True positive P value AUC
Gender (F/M) 145/194 1,094/1,645 0.316 47/38 688/987 0.010 NT 57/97 1,931/2,506 0.109 64/93 1,703/1,994 0.192
Age (yr) 60.5 61.2 0.761 56.5 58.9 0.487 52.7 60.2 0.001 0.615 53.1 60.1 0.003 0.551
Blood volume (mL) 3.5 3.3 0.018 0.531 2.7 2.8 0.664 3.0 2.1 0.000 0.584 2.9 2.2 0.001 0.545
Detection time (min) 2,448 1,553 0.000 0.660 2,241 1,773 0.072 1,830 1,559 0.108 1,961 1,470 0.002 0.541
Interval time (days) 12.6 17.3 0.004 0.530 15.8 12.2 0.238 13.2 11.2 0.432 5.4 9.3 0.000 0.513
Entry§ (D/N) 173/166 1,384/1,355 0.863 35/50 868/807 0.059 72/82 2,170/2,267 0.623 61/96 1,852/1,845 0.007 NT
CRP (mg/dL) 6.8 9.0 0.003 0.583 9.7 8.8 0.566 6.4 9.2 0.002 0.596 7.6 9.6 0.078
Leukocyte count (106/L) 10,181 11,490 0.019 0.564 10,054 1,160 0.176 12,150 11,479 0.341 14,969 11,562 0.027 0.553
Neutrophil count (106/L) 7,734 9,440 0.000 0.579 7,591 8,863 0.096 8,327 9,213 0.127 9,593 9,156 0.446
LUC (%) 2.3 2.0 0.218 2.2 2.0 0.589 2.8 1.8 0.091 4.0 1.9 0.049 0.577
DNI 1.7 4.6 0.000 0.644 2.8 4.8 0.002 0.533 3.0 4.6 0.001 0.548 4.2 5.5 0.118
MPXI 2.4 1.7 0.037 0.576 2.9 1.7 0.079 1.2 1.4 0.621 2.9 1.7 0.019 0.545
pH 7.4 7.4 0.832 7.4 7.4 0.680 7.4 7.4 0.326 7.4 7.4 0.643
PaO2 (mmHg) 96.8 99.2 0.455 110.8 98.1 0.039 0.630 103.4 97.7 0.170 100.0 96.5 0.365
PaCO2 (mmHg) 33.3 33.8 0.596 36.1 32.4 0.019 0.649 33.5 33.5 0.983 32.6 32.5 0.903
Abbreviations: AUC, area under the relative operating characteristic curve; F, female; M, male; NT, not tested; D, day work time (working hours; 08:00-17:00); N, night work time (working hours; 17;00-08:00); CRP, C-reactive protein; LUC, large unstained cell; DNI, delta neutrophil index; MPXI, mean peroxidase index.

Data was expressed as mean value.

Levene test showed equal variance (P>0.05).

Time interval between admission day and request day of blood culture.

§ Entry time of blood culture bottles to continuous monitoring blood culture systems.

Table 3.
Number (%) of false positive and true positive signals obtained from BACTEC FX system and BacT/Alert 3D system according to blood volume and detection time
Parameters BACTEC FX system BacT/Alert 3D system
Aerobic Anaerobic Aerobic Anaerobic
False positive True positive False positive True positive False positive True positive False positive True positive
Blood volume (mL)
 ≤0.5 126 (7.7) 382 (14.1) 13 (15.3) 342 (20.5) 38 (24.7) 1,440 (32.6) 48 (30.6) 1,295 (35.0)
 >0.5-10 310 (91.7) 2,295 (84.4) 71 (83.5) 1,293 (77.6) 111 (72.1) 2,947 (66.7) 106 (67.5) 2,370 (64.2)
 >10 2 (0.6) 41 (1.5) 1 (1.2) 32 (1.9) 5 (3.2) 30 (0.7) 3 (1.9) 29 (0.8)
 Total 338 (100) 2,718 (100) 85 (100) 1,667 (100) 154 (100) 4,417 (100) 157 (100) 3,694 (100)
Detection time
 Day (s) 1 115 (33.9) 1,706 (62.3) 46 (54.1) 985 (58.8) 93 (60.4) 2,750 (62.0) 83 (52.9) 2,404 (65.0)
  2 120 (35.4) 732 (26.7) 14 (16.5) 395 (23.6) 26 (16.9) 1,223 (27.6) 30 (19.1) 947 (25.6)
  3 52 (15.3) 198 (7.2) 7 (8.2) 113 (6.7) 13 (8.4) 308 (6.9) 23 (14.6) 185 (5.0)
  4 31 (9.2) 61 (2.2) 9 (10.6) 104 (6.2) 8 (5.2) 109 (2.4) 6 (3.8) 90 (2.5)
  5 21 (6.2) 42 (1.6) 9 (10.6) 78 (4.7) 14 (9.1) 47 (1.1) 15 (9.6) 71 (1.9)
  Total 339 (100) 2,739 (100) 85 (100) 1,675 (100) 154 (100) 4,437 (100) 157 (100) 3,697 (100)
 Hour (s) ≤1 0 (0) 2 (0.1) 10 (11.8) 1 (0.1) 1 (0.6) 0 (0) 0 (0) 0 (0)
  >1-2 0 (0) 14 (0.5) 2 (2.4) 26 (1.6) 1 (0.6) 1 (0) 4 (2.5) 0 (0)
  >2-4 5 (1.5) 58 (2.1) 4 (4.7) 41 (2.4) 43 (28.0) 37 (0.8) 32 (20.4) 25 (0.7)
  >4-8 1 (0.3) 24 (0.9) 8 (9.4) 17 (1.0) 1 (0.6) 30 (0.7) 0 (0) 34 (0.9)
  >8-16 46 (13.6) 910 (33.2) 9 (10.6) 598 (35.7) 32 (20.8) 1,283 (28.9) 35 (22.3) 1,455 (39.4)
  >16-32 108 (31.8) 1,072 (39.1) 16 (18.8) 509 (30.4) 22 (14.3) 2,035 (45.9) 24 (15.3) 1,473 (39.8)
  >32-64 116 (34.2) 506 (18.5) 16 (18.8) 272 (16.2) 28 (18.2) 826 (18.6) 37 (23.6) 497 (13.4)
  >64 63 (18.6) 153 (5.6) 20 (23.5) 211 (12.6) 26 (16.9) 225 (5.1) 25 (15.9) 213 (5.8)
  Total 339 (100) 2,739 (100) 85 (100) 1,675 (100) 154 (100) 4,437 (100) 157 (100) 3,697 (100)
Table 4.
Number (%) of false positive and true positive signals obtained from BACTEC FX system and BacT/Alert 3D system according to white blood cell count
WBC count (106/L) BACTEC FX system BacT/Alert 3D system
Aerobic Anaerobic Aerobic Anaerobic
False positive True positive False positive True positive False positive True positive False positive True positive
TLC ≤1,000 15 (4.5) 92 (3.5) 2 (2.4) 58 (3.6) 0 (0) 112 (2.6) 4 (2.6) 106 (2.9)
 >1,000-4,000 23 (7.0) 196 (7.4) 7 (8.4) 107 (6.6) 11 (7.2) 316 (7.2) 11 (7.1) 246 (6.7)
 >4,000-10,000 163 (49.2) 1,059 (39.8) 38 (45.8) 665 (40.7) 65 (43.0) 1,674 (38.4) 51 (32.9) 1,410 (38.7)
 >10,000-30,000 125 (37.8) 1,238 (46.5) 35 (42.2) 764 (46.8) 72 (47.7) 2,170 (49.8) 81 (52.2) 1,802 (49.4)
 >30,000-50,000 3 (0.9) 64 (2.4) 1 (1.2) 35 (2.1) 1 (0.7) 79 (1.8) 4 (2.6) 79 (2.2)
 >50,000-100,000 0 (0) 10 (0.4) 0 (0) 2 (0.1) 1 (0.7) 8 (0.2) 2 (1.3) 1 (0.0)
 >100,000 2 (0.6) 1 (0.0) 0 (0) 1 (0.1) 1 (0.7) 2 (0.0) 2 (1.3) 4 (0.1)
 Total 331 (100) 2,660 (100) 83 (100) 1,632 (100) 151 (100) 4,361 (100) 155 (100) 3,648 (100)
TNC ≤1,000 22 (7.3) 133 (5.4) 3 (4.1) 72 (4.7) 2 (1.5) 176 (4.3) 7 (5.0) 159 (4.6)
 >1,000-4,000 55 (18.4) 392 (15.8) 14 (18.9) 259 (17.0) 29 (21.3) 624 (15.3) 27 (19.3) 558 (16.3)
 >4,000-10,000 141 (47.2) 1,098 (44.4) 35 (47.3) 693 (45.6) 52 (38.2) 1,823 (44.8) 51 (36.4) 1,510 (44.0)
 >10,000-30,000 78 (26.1) 804 (32.5) 22 (29.7) 479 (31.5) 53 (39.0) 1,393 (34.2) 53 (37.9) 1,164 (33.9)
 >30,000-50,000 2 (0.7) 43 (1.7) 0 (0) 18 (1.2) 0 (0) 52 (1.3) 2 (1.4) 37 (1.1)
 >50,000-100,000 0 (0) 4 (0.2) 0 (0) 0 (0) 0 (0) 6 (0.1) 0 (0) 3 (0.1)
 >100,000 1 (0.3) 0 (0) 0 (0) 0 (0) 0 (0) 0 (0) 0 (0) 0 (0)
 Total 299 (100) 2,474 (100) 74 (100) 1,521 (100) 136 (100) 4,074 (100) 140 (100) 3,431 (100)

Abbreviations: WBC, white blood cell; TLC, total leukocyte count; TNC, total neutrophil count.

TOOLS