Abstract
In order to observe the pattern of a flow image on multisection MR imaging technique, a flow phantom experiment was preformed using a superconducting high filed 2.0 Tesla MRI scanner.
The pattren of the first section images was homogeneous round at all flow velocities until the turbulence forming level. The patterns of the second section images,however,changed into a homogeneous round shape, a ring shape, a target shape, and a small round shape as the velocity increased. When scanned at velocities higher than the trubulence forming level, the images become distored and irregular, and eventually disappeared after the cut-off velocity.
The homogeneous round image senn at the lower velocity levels in throught to be due to the overwhelming effects of fully managetized spins influxed into the imaging section during the prior repetition time(TR). Later in the higer velocity levels the effects of the partially saturated spins and fully magnetized spins influxed during the section transit time(TR/slice number) are added, and result in ring, target, and small round patterns in the second section image.