Abstract
Purpose
To evaluate the physiologic background of aneurysms poorly visualized during 3D-TOF MRA, contrast-enhanced MRA (CEMRA) and DSA due to hemodynamic isolation.
Materials and Methods
Using handmade elastic silicon phantoms to represent terminal basilar tip aneurysms, 3D-TOF MRA, CEMRA and DSA were used to determine blood turnover. Aneurysmal neck size was 2 mm and 10 mm, and the use of a pulsatile pump also helped recreate human physiologic parameters. We compared the results with those of computational fluid dynamics.
Results
DSA images of the narrow-necked aneurysm showed that a small volume of contrast medium washed into it during the systolic phase. As the width of its neck increased, the turnover volume of fragments of contrast bolus also increased. At CEMRA, the broad-necked aneurysm was visualized as the main bolus of Gd-DT PA passed through it, and at delayed CEMRA, the narrow-necked aneurysm was visualized faintly after the passage of bolus Gd-DT PA. The results correlated closely with those of 3DTOF MRA and computational fluid dynamics.