Journal List > Lab Med Online > v.7(4) > 1057354

Yu, Lim, Kwon, Woo, and Park: Evaluation of Cobas b 101 HbA1c Analyzer Performance for Point-of-Care Testing

Abstract

Background

The use of point-of-care (POC) devices for evaluating HbA1c is increasing; accordingly, comparisons between these devices and central laboratory methods are important. In the present study, we evaluated the analytical performance of the cobas b 101 analyzer for POC HbA1c testing.

Methods

The analytical quality of the cobas b 101 system was assessed based on repeatability, within-laboratory precision, linearity, and lot-to-lot reproducibility. Two specimen types, i.e., EDTA whole blood and capillary blood, were examined using the cobas b 101 system and the Variant II Turbo instrument.

Results

The coefficient of variation for within laboratory precision was 5.22% for a normal HbA1c level and 2.56% for a higher HbA1c level. The method showed good linearity, with a coefficient of correlation of 0.990. In a comparison of two different HbA1c disk lots, a strong correlation (r=0.986) and a mean %difference of −2.9% were observed. The cobas b 101 results using EDTA whole blood were strongly correlated with the Variant II Turbo results (r=0.958), with a mean %difference of 0.8%; the cobas b 101 results using capillary blood were strongly correlated with the Variant II Turbo results, using EDTA whole blood (r=0.976), with a mean %difference of 2.0%. A comparison between HbA1c levels in EDTA whole blood and capillary blood obtained using the cobas b 101 showed a strong correlation (r=0.985) and a mean %difference of 1.3%.

Conclusions

The cobas b 101 analyzer is convenient for the measurement of HbA1c levels for diabetes management.

INTRODUCTION

The HbA1c level indicates a patient's average glucose levels for the past 3 months and facilitates the long-term management of blood glucose levels by clinicians [1]. HbA1c levels are highly standardized worldwide owing to the development of reference measurement procedures and primary reference materials. In addition, test results are standardized according to the International Federation of Clinical Chemistry (IFCC) Working Group on HbA1c Standardization and the National Glycohemoglobin Standardization Program (NGSP) HbA1c Harmonization Program [23]. Various studies have reported that immediate feedback on HbA1c levels improves glycemic control in patients with diabetes mellitus (DM) [456]. Furthermore, studies have shown that HbA1c levels and intensive care based on HbA1c monitoring are correlated with the risk of developing DM-associated complications [789].
Among laboratory examination methods, point-of-care (POC) testing is the quickest and most convenient method used by doctors to make clinical decisions [10]. Recently, various HbA1c testing devices for capillary blood or EDTA whole blood have shown positive results [1011121314]. Furthermore, several studies have established the effectiveness of HbA1c monitoring by POC testing for glycemic control and the prevention of complications in patients with DM [1516].
In this study, we evaluated the analytical performance of the cobas b 101 (Roche Diagnostics, Mannheim, Germany) analyzer for estimating HbA1c levels. The cobas b 101 analyzer was compared with the standard central laboratory method, which uses the Variant II Turbo instrument.

MATERIALS AND METHODS

1. Study design

This study was performed from July to November 2015 in the Department of Laboratory Medicine at our institute. All analyses were reviewed and approved by the Institutional Review Board (IRB) at our institute (KBSMC 2015-01-049). Patient samples were obtained from residual EDTA samples or from EDTA and capillary samples collected immediately before the test. All participants provided written informed consent prior to blood collection. EDTA whole blood specimens were taken from the antecubital vein after at least 8 hours of fasting and analyzed promptly within 1 hour. Simultaneously, capillary blood collection was performed by finger pricking to obtain 2 µL of whole blood. Participant data were subjected to a de-identification process and coded with serial numbers for the test.

2. Precision and linearity

Repeatability and within-laboratory precision were determined using two quality control (QC) materials with different HbA1c levels, supplied by the manufacturer of the cobas HbA1c control. Both QC materials were assayed in duplicate twice daily at 9 am and 3 pm for a total of 20 days according to the 2014 Clinical and Laboratory Standards Institute (CLSI) guideline EP05-A3 [17]. The linearity of the assay was evaluated for specimens with five different HbA1c levels with duplication by mixing venous EDTA blood samples from two patients according to the CLSI guideline EP06-A [18].

3. Lot-to-lot reproducibility

The lot-to-lot reproducibility was evaluated using two different HbA1c disk lots (#434022-01 and #435021-01), and 20 residual venous EDTA blood samples treated in the same manner by the same user were evaluated for 5 days, with 4 samples evaluated in duplicate each day. The test results for 20 samples using the first lot were in the range of 5.2–8.9% HbA1c.

4. Comparative analysis

Linear regression and %differences in estimated HbA1c levels between the cobas b 101 and Variant II Turbo were determined according to the 2013 CLSI guideline EP09-A3 protocol [19]. Two types of blood specimens were obtained immediately before the tests from the 40 enrolled participants: capillary blood and EDTA whole blood. The capillary finger prick samples and EDTA whole blood samples were analyzed using the cobas b 101 POC system in the phlebotomy room and the EDTA whole blood samples were analyzed using the Variant II Turbo in the central laboratory. The HbA1c levels of samples spanned the clinically relevant range of 4.8% to 8.6%.

5. Laboratory method

HbA1c was measured at the central laboratory of our hospital using the Variant II Turbo (Bio-Rad Laboratories, Hercules, CA, USA), which is based on high-performance liquid chromatography (HPLC). The HbA1c program of the reference device was certified by the NGSP with documented traceability to the Diabetes Control and Complications Trial (DCCT) reference method. This test participated in the quality assurance survey from the College of American Pathologists (CAP), and all five of the challenges were within the acceptable range, with a mean bias of 0.06% for HbA1c levels. QC materials with low (5.10±0.2%) and high (9.80±0.2%) levels were used. The coefficient of variation was 0.96–3.03% for the low-HbA1c QC materials and 1.18–2.90% for the high-HbA1c QC materials during the study period.

6. POC analyzer methods

The cobas b 101 POC analyzer (Roche Diagnostics) is based on a photometric transmission measurement method using a latex agglutination inhibition immunoassay. The test uses 2 µL of EDTA venous whole blood or 2 µL of capillary whole blood. The POC devices were designed to operate with ready-to-use HbA1c disks and provide results in 340 seconds. This device can measure HbA1c levels in the 4–14% range. The test method was certified by the NGSP and standardized or traceable to the DCCT reference assay. HbA1c values obtained using both devices are expressed in NGSP units as % HbA1c.

7. Statistical analysis

Test imprecision was analyzed based on repeatability and within-laboratory precision. Each formula was obtained from EP05-A3 [17]. The lot-to-lot reproducibility and comparative results were evaluated based on Pearson's correlation coefficients (r). The results are displayed in scatter plots. %Differences between each POC test result and the reference method were analyzed by calculating the percentage of HbA1c reporting unit differences and a Bland-Altman plot was generated. The acceptable standard was a difference of ±6% according to the HbA1c acceptable limit established by NGSP, which was applied in the CAP GH2 survey [20]. All statistical analyses were implemented in Microsoft Excel 2010 and IBM SPSS version 18.0 (IBM, New York, NY, USA) and statistical significance was defined as P<0.05.

RESULTS

1. Precision and linearity

Imprecision was analyzed using QC materials with two levels of HbA1c for 20 days; the mean levels were 5.1% (normal) and 9.5% (pathological). The standard deviation (SD) and coefficient of variation (CV) were 0.25% and 4.83% for repeatability and 0.27% and 5.22% for within-laboratory precision tests for the normal level, and 0.21% and 2.22% for repeatability and 0.24% and 2.56% for within-laboratory precision for the pathological level, respectively (Table 1). This method showed good linearity between HbA1c levels of 4.6% and 13.3%. The estimated slope and intercept of the regression were 1.014 and 0.392, respectively, with a coefficient of correlation of 0.990.

2. Lot-to-lot reproducibility

The lot-to-lot reproducibility for cobas b 101 HbA1c test results using two different lots is shown in Fig. 1A (r=0.986, P<0.001). The mean %difference was −2.9% (range −8.7% to 0.0%) (Fig. 1B). For the first lot, there was a strong correlation between the results obtained using the cobas b 101 and Variant II Turbo; the estimated coefficient of correlation and %difference were 0.983 (P<0.001) and −1.5% (range −8.6% to 2.6%), respectively. For the second lot, there was also a strong correlation between the results obtained using the cobas b 101 and the Variant II Turbo, with a correlation coefficient of 0.985 (P<0.001) and a %difference of 1.7% (range −5.0% to 5.8%). The slope and the intercept of the regression for the two lots using the central laboratory method were 0.905 and 0.525 for the first lot and 0.834 and 0.539 for the second lot.

3. Comparative analysis

The regression of the estimates obtained using the cobas b 101 HbA1c and Variant II Turbo using EDTA whole blood showed a slope of 0.946 and an intercept of 0.374. On the scatter plot, the graph showed a strong correlation, with a correlation coefficient of 0.958 (P<0.001; Fig. 2A). The mean %difference was 0.8% (range −5.4% to 5.9%) and tended to be more highly dispersed at the higher HbA1c level (Fig. 2B). No specimen exceeded a 6% %difference.
When comparing the cobas b 101 analyzer results using capillary blood and the Variant II Turbo results using EDTA whole blood, a regression line with a slope of 0.991 and intercept of 0.179 was obtained. On the scatter plot, the graph showed a strong correlation, with a correlation coefficient of 0.976 (P<0.001; Fig. 2C). The mean %difference was 2.0% (range −4.8% to 6.8%) (Fig. 2D). Moreover, a strong correlation was observed between EDTA whole blood and capillary HbA1c levels, with a correlation coefficient of 0.985, using the cobas b 101 system (P<0.001; Fig. 3A). The mean %difference was 1.3% (range −5.5% to 7.7%) (Fig. 3B). Table 2 contains all results for the reproducibility and comparative analyses, indicating the expected device or lot value (Y) when the other device or lot (X) had a 6.5% HbA1c level.

DISCUSSION

Venous blood sampling is currently the gold standard for the assessment of blood glucose levels. However, this sampling method cannot be applied to home-based glucose self-monitoring. Capillary blood glucose testing using portable POC devices has been hailed as an alternative method to venous blood sampling owing to its better compliance, rapid reporting of test results, low cost, and potential for self-monitoring [21]. Nonetheless, there are some doubts regarding its diagnostic accuracy compared to that of venous blood sampling, and a recent study has shown that many current POC HbA1c devices do not meet the analytical requirements stipulated by the NGSP [222].
Current HbA1c POC devices have shown good performance in some recent studies designed to compare the validity of capillary blood glucose and venous blood glucose testing [1011121314]. In our study, we not only evaluated POC device performance for EDTA whole blood glucose, but also compared HbA1c levels using EDTA whole blood and capillary blood as samples determined by the cobas b 101 assay to evaluate its clinical efficiency.
The NGSP has tightened the criteria several times for the certification of manufacturer methods, with the goal of improving the quality of HbA1c testing. The performance threshold for manufacturers is ±6% with respect to the relative bias of the Secondary Reference Laboratory measurements [23]. Similarly, the CAP replaced peer-group grading of the HbA1c level for the GH-2 HbA1c survey with accuracy-based grading, and has since tightened the acceptable performance limits from ±15% to ±7% in 2011–2012 and ±6% in 2013–2016 [2425]. The performance of POC tests for the determination of HbA1c levels is generally assessed based on precision; it is recommended that an imprecision of less than 3% is a desirable analytical goal for laboratory HbA1c methods based on clinical requirements, and an optimal imprecision of 2% NGSP units is now recommended by leading professional groups [226]. In this study, the cobas b 101 analyzer showed good precision at the pathological level according to the desired analytical goal, but did not meet the optimal imprecision goal. At the normal level, the within-laboratory precision did not meet the goal. Therefore, this test was considered acceptable for follow-up monitoring of HbA1c at the 7.0% HbA1c treatment goal recommended by the American Diabetes Association Standards of Medical Care in Diabetes 2017 [27]. Overall comparative analyses showed a strong correlation and all of the mean %differences were within 6%.
In a previous study, the cobas b 101 showed an acceptable imprecision result in an evaluation of seven HbA1c POC devices using patient venous blood samples [14]. Two different reagent lot numbers for the cobas b 101 met the NGSP criteria, with an intra-laboratory precision of 2.4% and 1.2% (CVs) and −0.05%–0.23% bias with three certified secondary reference measurement procedures, indicating good performance compared to that of the other six POC instruments.
Although the cobas b 101 showed promising results, this study had a few limitations. We could not evaluate a broad range of HbA1c levels owing to a lack of patient samples with extremely low or high levels. In the repeatability and within-laboratory precision studies for the cobas b 101 POC device, the 5.1% level for QC materials did not fully represent the diagnostic performance for a diabetes cutoff HbA1c level of 6.5%. We did not exclude interference from hemoglobin variants because we could not obtain fresh specimens with hemoglobinopathies, which have the potential to impact the HbA1c results.
In conclusion, comparative analyses with the reference method using the Variant II Turbo and the POC test using the cobas b 101 showed strong correlations using EDTA whole blood samples in our study. Moreover, our findings demonstrated a strong correlation between HbA1c levels obtained using EDTA whole blood and capillary samples in the cobas b 101 assay. However, exhaustive precision analyses are necessary before clinical use. Therefore, the cobas b 101 analyzer is a convenient assay for HbA1c levels and may be useful for diabetes management.

Figures and Tables

Fig. 1

Lot-to-lot reproducibility of the cobas b 101 HbA1c test. (A) Regression analysis of the cobas b 101 using EDTA venous blood from two different lots. The linear curve and 95% confidence interval [CI] are represented as red and blue lines, respectively. (B) Bland-Altman difference plot summarizing lot-to-lot comparisons.

lmo-7-182-g001
Fig. 2

Comparison of the cobas b 101 with the Variant II Turbo. (A) Regression of the cobas b 101 results using EDTA blood against the Variant II Turbo results using EDTA blood. The linear curve and 95% confidence interval [CI] are represented as red and blue lines, respectively. (B) Bland-Altman difference plot summarizing the comparison between the cobas b 101 results using EDTA blood and Variant II Turbo results using EDTA blood. (C) Regression of the cobas b 101 results using capillary blood against the Variant II Turbo results using EDTA blood. The linear curve and 95% confidence interval [CI] are represented as red and blue lines, respectively. (D) Bland-Altman difference plot summarizing the comparison between the cobas b 101 results using capillary blood and Variant II Turbo results using EDTA blood.

lmo-7-182-g002
Fig. 3

Regression analysis of the cobas b 101 using two types of samples, EDTA venous whole blood and capillary blood. (A) Regression analysis of the cobas b 101 using two types of samples, EDTA venous blood and capillary blood. The linear curve and 95% confidence interval [CI] are represented as red and blue lines, respectively. (B) Bland-Altman difference plot summarizing the comparison between the venous whole blood and capillary blood samples tested using the cobas b 101.

lmo-7-182-g003
Table 1

Imprecision with 95% confidence intervals (95% CI) for the cobas b 101 HbA1c analyzer based on EP05

lmo-7-182-i001
Quality control materials Pathological level (Mean 9.5% HbA1c) Pathological level (Mean 9.5% HbA1c)
SD (95% CI) %CV (95% CI) SD (95% CI) %CV (95% CI)
Repeatability 0.25 (0.20–0.31) 4.83 (3.97–6.18) 0.21 (0.17–0.27) 2.22 (1.83–2.85)
Within-Laboratory Precision 0.27 (0.23–0.32) 5.22 (4.48–6.25) 0.24 (0.21–0.29) 2.56 (2.19–3.09)
Table 2

Summary of the lot-to-lot reproducibility and comparative analyses

lmo-7-182-i002
Device or lot (sample type) Test samples (n) Equation Expected y at x=6.5% of HbA1c [95% CI] Mean % difference [95% CI]
Y X
Cobas b 101 Lot #2* (EDTA W/B) Cobas b 101 Lot #1* (EDTA W/B) 20 y = 0.948x + 0.163 (r = 0.986, P < 0.001) 6.3% [6.2–6.4] -2.9% [-4.0 to -1.8]
Cobas b 101 (EDTA W/B) Variant II Turbo (EDTA W/B) 40 y = 0.946x + 0.374 (r = 0.958, P < 0.001) 6.5% [6.5–6.6] 0.8% [-0.2–1.8]
Cobas b 101 (Capillary blood) Variant II Turbo (EDTA W/B) 40 y = 0.991x + 0.179 (r = 0.976, P < 0.001) 6.6% [6.6–6.7] 2.0% [0.9–3.2]
Cobas b 101 (Capillary blood) Cobas b 101 (EDTA W/B) 40 y = 1.035x - 0.134 (r = 0.985, P < 0.001) 6.6% [6.5–6.6] 1.3% [0.4–2.2]

*Residual EDTA whole blood (W/B) sample; Immediately drawn EDTA W/B or capillary blood sample.

Notes

This article is available from http://www.labmedonline.org

AUTHORS' DISCLOSURES OF POTENTIAL CONFLICTS OF INTEREST No potential confiicts of interest relevant to this article were reported.

References

1. Gonen B, Rubenstein A, Rochman H, Tanega SP, Horwitz DL. Haemoglobin A1: An indicator of the metabolic control of diabetic patients. Lancet. 1977; 2:734–737.
2. Weykamp C. HbA1c: a review of analytical and clinical aspects. Ann Lab Med. 2013; 33:393–400.
crossref
3. Weykamp C, John WG, Mosca A. A review of the challenge in measuring hemoglobin A1c. J Diabetes Sci Technol. 2009; 3:439–445.
crossref
4. Cagliero E, Levina EV, Nathan DM. Immediate feedback of HbA1c levels improves glycemic control in type 1 and insulin-treated type 2 diabetic patients. Diabetes Care. 1999; 22:1785–1789.
crossref
5. Ferenczi A, Reddy K, Lorber DL. Effect of immediate hemoglobin A1c results on treatment decisions in office practice. Endocr Pract. 2001; 7:85–88.
crossref
6. Miller CD, Barnes CS, Phillips LS, Ziemer DC, Gallina DL, Cook CB, et al. Rapid A1c availability improves clinical decision-making in an urban primary care clinic. Diabetes Care. 2003; 26:1158–1163.
crossref
7. International Expert Committee. International Expert Committee report on the role of the A1C assay in the diagnosis of diabetes. Diabetes Care. 2009; 32:1327–1334.
8. Diabetes Control and Complications Trial Research Group. The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus. N Engl J Med. 1993; 329:977–986.
9. UK Prospective Diabetes Study (UKPDS) Group. Intensive blood-glucose control with sulphonylureas or insulin compared with conventional treatment and risk of complications in patients with type 2 diabetes (UKPDS 33). Lancet. 1998; 352:837–853.
10. Petersen JR, Omoruyi FO, Mohammad AA, Shea TJ, Okorodudu AO, Ju H. Hemoglobin A1c: assessment of three POC analyzers relative to a central laboratory method. Clin Chim Acta. 2010; 411:2062–2066.
crossref
11. Lee JE. Alternative biomarkers for assessing glycemic control in diabetes: fructosamine, glycated albumin, and 1,5-anhydroglucitol. Ann Pediatr Endocrinol Metab. 2015; 20:74–78.
crossref
12. Sanchez-Mora C, S Rodriquez-Oliva M, Fernandez-Riejos P, Mateo J, Polo-Padillo J, Goberna R, et al. Evaluation of two HbA1c point-of-care analyzers. Clin Chem Lab Med. 2011; 49:653–657.
crossref
13. Wan Mohd Zin RM, Ahmad Kamil ZI, Tuan Soh TR, Embong M, Wan Mohamud WN. Haemoglobin A1c: comparing performance of two point of care devices with laboratory analyser. BMC Res Notes. 2013; 6:540.
crossref
14. Lenters-Westra E, Slingerland RJ. Three of 7 hemoglobin A1c point-of-care instruments do not meet generally accepted analytical performance criteria. Clin Chem. 2014; 60:1062–1072.
crossref
15. Mayega RW, Guwatudde D, Makumbi FE, Nakwagala FN, Peterson S, Tomson G, et al. Comparison of fasting plasma glucose and haemoglobin A1c point-of-care tests in screening for diabetes and abnormal glucose regulation in a rural low income setting. Diabetes Res Clin Pract. 2014; 104:112–120.
crossref
16. Brown M, Kuhlman D, Larson L, Sloan K, Ablah E, Konda K, et al. Does availability of expanded point-of-care services improve outcomes for rural diabetic patients? Prim Care Diabetes. 2013; 7:129–134.
crossref
17. Clinical and Laboratory Standards Institute (CLSI). Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline-Third Edition. CLSI document EP05-A3. Wayne, PA: Clinical and Laboratory Standards Institute;2014.
18. Clinical and Laboratory Standards Institute. Evaluation of Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline. CLSI guideline EP06-A. Wayne, PA: Clinical and Laboratory Standards Institute;2003.
19. Clinical and Laboratory Standards Institute (CLSI). Measurement Procedure Comparison and Bias Estimation Using Patient Samples; Approved Guideline-Third Edition. CLSI document EP09-A3. Wayne, PA: Clinical and Laboratory Standards Institute;2013.
20. World Health Organization. Global report on diabetes. Geneva: World Health Organization;2016.
21. International Organization for Standardization. ISO 15197: Determination of performance criteria for in vitro blood glucose monitoring systems for management of human diabetes mellitus. Geneva: International Organization for Standardization;2002.
22. Lenters-Westra E, Slingerland RJ. Six of eight hemoglobin A1c point-of-care instruments do not meet the general accepted analytical performance criteria. Clin Chem. 2010; 56:44–52.
crossref
23. National Glycohemoglobin Standardization Program (NGSP). Harmonizing Hemoglobin A1c Testing. Updated on 29 Sep 2016. http://www.ngsp.org/docs/Protocol.pdf.
24. Little RR, Rohlfing CL, Sacks DB. Status of hemoglobin A1c measurement and goals for improvement: from chaos to order for improving diabetes care. Clin Chem. 2011; 57:205–214.
crossref
25. National Glycohemoglobin Standardization Program (NGSP). College of American Pathologists (CAP) GH5 Survey Data. Updated on Sep 29 2016. http://www.ngsp.org/CAP/CAP16b.pdf.
26. Shephard MD. Analytical goals for point-of-care testing used for diabetes management in Australian health care settings outside the laboratory. Point Care. 2006; 5:177–185.
27. American Diabetes Association. 6. Glycemic Targets. Diabetes Care. 2017; 40:S48–S56.
TOOLS
Similar articles