Kim, Ji, Moon, Hur, and Yun: Analytical Evaluation of the DiaSys Albumin in Urine/CSF FS Kit for Urine Albumin Measurement Using a JEOL BioMajesty JCA-BM6010/C Analyzer

Abstract

Background

High albuminuria is defined as albumin excretion of >30 mg/24 hr or an albumin-to-creatinine ratio of 30 mg/g in a random urine sample. We assessed the analytical performance of the Albumin in Urine/CSF FS kit (DiaSys Inc., UK) using a BioMajesty JCA-6010/C analyzer (JEOL Inc., Japan).

Methods

Urine albumin concentrations were measured by the Albumin in Urine/CSF FS kit using a BioMajesty JCA-BM6010/C analyzer. Imprecision, linearity, and carry-over were measured according to the Clinical Laboratory and Standards Institute documents EP10 and EP9. The assay was compared with the ALB-T TQ Gen.2 (Roche, Germany) assay on a Cobas8000 C702 (Roche, Germany), the Tina-Quant Albumin (Roche, Switzerland) assay on a Hitachi7600-210 (Hitachi, Japan), and an Abbott urine albumin assay (Abbott Laboratories, USA) on a TBA 200FR (Toshiba, Japan) using 50 random urine samples.

Results

Within-run and total imprecision were 0.551-1.023% and 0.551-1.214%, respectively. Linearity ranged from 6.31 to 30.60 mg/dL, and functional sensitivity was 0.5 mg/dL. Results from the Albumin in Urine/CSF FS kit showed good correlation with the ALB-T TQ Gen.2 (r=0.987) and the Tina-Quant Albumin assays (r=0.991). However, the four assays categorized 18 of 50 urine samples into different albuminuria groups.

Conclusions

Albumin in Urine/CSF FS testing on a BioMajesty JCA-BM6010/C analyzer showed good linearity, functional sensitivity, precision, and correlation with the ALB-T TQ Gen.2 and Tina-Quant Albumin assays. However, because some samples were categorized into different albuminuria groups by the different assays, further studies on the standardization of albuminuria assays are needed.

Figures and Tables

Fig. 1

Regression plots of results from Albumin in Urine/CSF FS testing on a BioMajesty JCA-BM6010/C analyzer. (A) Correlation of results from Albumin in Urine/CSF FS (DiaSys) testing on a JCA-BM6010/C analyzer and ALB-T TQ Gen.2 (Roche) testing on a Cobas8000 C702 showed a slope of 0.8628 and an intercept of 2.344 (N=44). (B) Correlation of results from Albumin in Urine/CSF FS (DiaSys) testing on a JCA-BM6010/C analyzer and Tina-Quant Albumin (Roche) testing on a Hitachi 7600-210 showed a slope of 1.056 and an intercept of 4.808 (N=44). (C) Correlation of results from Albumin in Urine/CSF FS (DiaSys) testing on a JCA-BM6010/C analyzer and Microalbumin (Abbott) testing on a TBA-200FR showed a slope of 7.134 and an intercept of -87.43.

lmo-6-64-g001
Fig. 2

Bland-Altman plots of albumin concentrations from Albumin in Urine/CSF FS (DiASys) testing on a BioMajesty JCA-BM6010/C analyzer. (A) Comparison of ALB-T TQ Gen.2 (Roche) testing on a Cobas8000 C702 and Albumin in Urine/CSF FS (DiaSys) testing on a JCA-BM6010/C. (B) Comparison of Tina-Quant Albumin (Roche) testing on a Hitachi 7600-210 and Albumin in Urine/CSF FS (DiaSys) testing on a JCA-BM6010/C. (C) Comparison of Microalbumin (Abbott) testing on a TBA-200FR and Albumin in Urine/CSF FS (DiaSys) testing on a JCA-BM6010/C. Thick solid lines show the means of the paired differences, and thin solid lines represent lines of identity. Dashed lines show the upper and lower 95% limits of agreement (LoA).

lmo-6-64-g002
Table 1

Within-run and total imprecisions of the Albumin in Urine/CSF FS (DiaSys) kit determined by a BioMajesty JCA-BM6010/C (JEOL) analyzer

lmo-6-64-i001
Within-run imprecision (CV%) Total imprecision (CV%) Manufacturer's claimed imprecision (CV%)* [7] Desirable specification for imprecision (CV%)** [9]
Low level (10.91 mg/dL) 1.023 1.214 0.86-2.21 18
Mid-level (49.50 mg/dL) 0.551 0.568
High level (101.1 mg/dL) 0.551 0.551

*Manufacturer's claimed imprecision is indicated in reagent information [7]; **Desirable specification for imprecision is indicated on the Westgard website [9].

Table 2

Linearity and recovery rate of the Albumin in Urine/CSF FS (DiaSys) kit determined by a BioMajesty JCA-BM6010/C (JEOL) analyzer

lmo-6-64-i002
Analyte Test range Observed linear range Linear range claimed by the manufacturer* Slope Intercept R2 Recovery (%)
Urinary albumin (mg/dL) 6.31-30.60 6.31-30.60 0.1-35.0 1.001 0.179 0.999 100.0-103.0

*Manufacturer's claimed linear range is indicated in reagent information [7].

Table 3

Discrepant albuminuria classification according to four assays

lmo-6-64-i003
Sample No. Albumin in Urine/CSF FS (DiaSys) on JCA-BM6010/C ALB-T TQ Gen.2 (Roche) on Cobas8000 C702 Tina-Quant Albumin (Roche) on Hitachi 7600-210 Microalbumin (Abbott) on TBA-200FR
1 Very High Very High Very High High
2 High Low Normal High
4 Very High Very High High High
6 -* Very High Very High High
8 Very High Very High High High
10 Very High - - High
11 High High Normal -
14 Very High Very High Very High High
16 High - Normal High
28 High Low Low High
29 Very High Very High Very High High
33 High Low Low High
36 High Normal Normal High
42 Very High - Very High High
43 High High High Very High
47 - Very High Very High High
48 Very High High High High
49 Very High Very High Very High High

HighNormal, ACR <10 mg/g; Low albuminuria, 10 mg/g≤ ACR<30 mg/g; High albuminuria, 30 mg/g≤ ACR<300 mg/g; Very high albuminuria, ACR ≥300 mg/g. Among the 50 urine samples listed above, 18 urine samples were classified into different albuminuria groups by the four assays. Urine creatinine levels were measured by Roche Cre (Roche Diagnostics, Basel, Switzerland) assay on a TBA-200FR platform.

*Dash means an outlier.

Notes

This article is available from http://www.labmedonline.org

References

1. Levey AS, Cattran D, Friedman A, Miller WG, Sedor J, Tuttle K, et al. Proteinuria as a surrogate outcome in CKD: report of a scientific workshop sponsored by the National Kidney Foundation and the US Food and Drug Administration. Am J Kidney Dis. 2009; 54:205–226.
crossref
2. The Korean Society for Laboratory Medicine. Laboratory Medicine. 5th ed. Seoul: Beommun Education;2014. p. 15p. 359p. 500–501.
3. Hillege HL, Fidler V, Diercks GF, van Gilst WH, de Zeeuw D, van Veldhuisen DJ, et al. Urinary albumin excretion predicts cardiovascular and noncardiovascular mortality in general population. Circulation. 2002; 106:1777–1782.
crossref
4. Kidney Disease Outcomes Quality Initiative. National kidney practice guidelines for chronic kidney disease: evaluation, classification, and stratification. Ann Intern Med. 2003; 139:137–147.
5. Clinical and Laboratory Standards Institute. Preliminary evaluation of quantitative clinical measurement procedures; approved guideline. CLSI document EP10-A3. 3rd ed. Wayne, PA: Clinical and Laboratory Standards Institute;2014.
6. Clinical and Laboratory Standards Institute. Method comparison and bias estimation using patient samples; approved guideline. CLSI document EP9-A2. 2nd ed. Wayne, PA: Clinical and Laboratory Standards Institute;2002.
7. DiaSys Diagnostics Systems GmbH. Reagent information of Albumin in Urine/CSF FS. 2011. Feb.
8. Cho JH, Lee CM, Park CM, Moon HW, Hur M, Yun YM, et al. Evaluation of the performance of Lumipulse G1200 for tumor marker assays. Lab Med Online. 2012; 2:131–138.
crossref
9. Ricos C, Alvarez V, Cava F, Garcia-Lario JV, Hernandez A, Jimenez CV, et al. Desirable specifications for total error, imprecision, and bias, derived from intra- and inter-individual biologic variation. updated in 2014. https://www.westgard.com/biodatabase1.htm#im.
10. Fraser CG, Harris EK. Generation and application of data on biological variation in clinical chemistry. Crit Rev Clin Lab Sci. 1989; 27:409–437.
crossref
11. Ricós C, Alvarez V, Cava F, García-Lario JV, Hernández A, Jiménez CV, et al. Current databases on biological variation: pros, cons and progress. Scand J Clin Lab Invest. 1999; 59:491–500.
crossref
12. National Kidney Foundation. K/DOQI Clinical practice guidelines for chronic kidney disease: evaluation, classification, and stratification. Am J Kidney Dis. 2002; 39:S1–S266.
13. Warram JH, Gearin G, Laffel L, Krolewski AS. Effect of duration of type I diabetes on the prevalence of stages of diabetic nephropathy defined by urinary albumin/creatinine ratio. J Am Soc Nephrol. 1996; 7:930–937.
crossref
14. Jensen JS, Feldt-Rasmussen B, Strandgaard S, Schroll M, Borch-Johnsen K. Arterial hypertension, microalbuminuria, and risk of ischemic heart disease. Hypertension. 2000; 35:898–903.
crossref
15. Architect. Reagent information of Microalbumin. 2004. Dec.