Journal List > Lab Med Online > v.5(2) > 1057297

Lee and Cha: Performance Evaluation of BAROZEN H, a Networking Blood Glucose Monitoring System for Medical Institutions

Abstract

Background

We evaluated the analytical performance of Barozen H (i-SENS Inc., Korea), a new glucometer equipped with networking function for medical institutions, according to the ISO 15197:2003 and ISO/DIS 15197:2011 guidelines.

Methods

We measured the precision of 10 Barozen H glucometers, in terms of repeatability and intermediate precision, and determined their accuracy relative to that of automatic chemistry analyzer AU5421 (Beckman Coulter, USA). Three other glucometers-Precision PCx (Abbott, USA), Glucocard Sigma (Arkray, Japan), and SureStep Flexx (Johnson & Johnson, USA) were also evaluated, and their accuracies and hematocrit interferences were compared.

Results

The standard deviation and coefficient of variation of Barozen H for repeatability and intermediate precision were 0.11-0.15 mmol/L and 2.3-3.6%, respectively. With respect to accuracy, in accordance with ISO 15197:2003 criteria, Barozen H yielded 98.0% of results within ±0.83 mmol/L or ±20%. Further, per the ISO/DIS 15197:2011 criteria, 95.2% of results were within ±0.83 mmol/L or ±15%; Barozen H was the only glucometer satisfying the more stringent ISO/DIS 15197:2011 criteria. Error grid analysis showed that all results from Barozen H were in zone A, indicating its excellent clinical accuracy. Hematocrit, ranging from 20% to 60% did not cause any significant interference.

Conclusions

Barozen H showed excellent analytical performance, and it was the most clinically accurate glucometer tested. It can be expected to provide reliable results satisfying ISO/DIS 15197:2011 as well as ISO 15197:2003 criteria.

Figures and Tables

Fig. 1

Difference plots of Barozen H and 3 other glucometers. (A) Barozen H, (B) Precision PCx, (C) Glucocard Sigma, (D) SureStep Flexx. The X-axis represents the mean value of the duplicates from the AU5421 chemistry analyzer. The Y-axis shows the difference between the result of each glucometer and that of AU5421. The dotted lines represent the acceptance criteria of ISO/DIS 15197:2011 (±0.83 mmol/L and ±15%).

lmo-5-69-g001
Fig. 2

Consensus error grids for Barozen H and 3 other glucometers. (A) Barozen H, (B) Precision PCx, (C) Glucocard Sigma, (D) SureStep Flexx. The X-axis represents the mean value of the duplicates from the AU5421 chemistry analyzer. The Y-axis shows the results of each glucometer. The percentage values in the box represent the proportion of results belonging to each zone.

lmo-5-69-g002
Fig. 3

Interference of the hematocrit in 4 glucometers. (A) Barozen H, (B) Precision PCx, (C) Glucocard Sigma, (D) SureStep Flexx. The X-axis shows the hematocrit (%) for each sample. The Y-axis shows the bias between each glucometer and the AU5421 chemistry analyzer as a reference measurement procedure. Each bias was presented in difference (mg/dL) at low glucose concentration (30-50 mg/dL, 1.7-2.8 mmol/L) and %difference (%) at mid (96-144 mg/dL, 5.3-8.0 mmol/L) and high (280-420 mg/dL, 15.5-23.3 mmol/L) glucose concentration.

lmo-5-69-g003
Table 1

Precision of Barozen H

lmo-5-69-i001
Repeatability Intermediate precision
Level 1 Level 2 Level 3 Level 4 Level 5 Level 1 Level 2 Level 3
Mean (mmol/L) 1.87 4.08 6.79 11.02 20.47 2.38 7.19 18.58
SD (mmol/L) 0.11 0.15 0.21 0.39 0.52 0.11 0.26 0.43
CV (%) 5.9 3.7 3.0 3.6 2.5 4.4 3.5 2.3
Table 2

System accuracy according to ISO 15197:2003

lmo-5-69-i002
Glucometers <4.2 mmol/L ≥4.2 mmol/L Total
Within ±0.28 mmol/L Within ±0.56 mmol/L Within ±0.83 mmol/L Within ±5% Within ±10% Within ±15% Within ±20%
Barozen H 25/30* (83.3%) 29/30 (96.7%) 30/30 (100%) 133/220 (60.5%) 189/220 (85.9%) 207/220 (94.1%) 215/220 (97.7%) 245/250 (98.0%)
Precision PCx 25/32 (78.1%) 32/32 (100%) 32/32 (100%) 95/218 (43.6%) 164/218 (75.2%) 203/218 (93.1%) 214/218 (98.2%) 246/250 (98.4%)
Glucocard Sigma 20/27 (74.1%) 27/27 (100%) 27/27 (100%) 103/223 (46.2%) 167/223 (74.9%) 201/223 (90.1%) 210/223 (94.2%) 237/250 (94.8%)
SureStep Flexx 16/33 (48.5%) 25/33 (75.8%) 32/33 (97.0%) 113/217 (52.1%) 171/217 (78.8%) 198/217 (91.2%) 214/217 (98.6%) 246/250 (98.4%)

*Number and percentage of acceptable results among all results in each interval.

Table 3

System accuracy according to ISO/DIS 15197:2011

lmo-5-69-i003
Glucometers <5.55 mmol/L ≥ 5.55 mmol/L Total
Within ±0.28 mmol/L Within ±0.56 mmol/L Within ±0.83 mmol/L Within ± 5% Within ± 10% Within ±15%
Barozen H 45/60* (75.0%) 54/60 (90.0%) 56/60 (93.3%) 117/190 (61.6%) 164/190 (86.3%) 182/190 (95.8%) 238/250 (95.2%)
Precision PCx 37/60 (61.7%) 53/60 (88.3%) 58/60 (96.7%) 87/190 (45.8%) 145/190 (76.3%) 178/190 (93.7%) 236/250 (94.4%)
Glucocard Sigma 37/60 (61.7%) 51/60 (85.0%) 56/60 (93.3%) 89/190 (46.8%) 145/190 (76.3%) 174/190 (91.6%) 230/250 (92.0%)
SureStep Flexx 31/60 (51.7%) 46/60 (76.7%) 54/60 (90.0%) 98/190 (51.6%) 152/190 (80.0%) 177/190 (93.2%) 231/250 (92.4%)

*Number and percentage of acceptable results among all results in each interval.

Notes

This article is available from http://www.labmedonline.org

References

1. The Diabetes Control and Complications Trial Research Group. The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus. N Engl J Med. 1993; 329:977–986.
2. UK Prospective Diabetes Study (UKPDS) Group. Intensive blood-glucose control with sulphonylureas or insulin compared with conventional treatment and risk of complications in patients with type 2 diabetes (UKPDS 33). Lancet. 1998; 352:837–853.
3. Sacks DB, Arnold M, Bakris GL, Bruns DE, Horvath AR, Kirkman MS, et al. Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus. Diabetes Care. 2011; 34:e61–e99.
crossref
4. Burtis CA, Ashwood ER, Tietz NW. Tietz textbook of clinical chemistry. 3rd ed. Philadelphia: W.B. Saunders;1999.
5. Park AJ, Kim HR, Lee MK. [Networking experience of point-of-care test glucometer]. Korean J Lab Med. 2006; 26:294–298.
crossref
6. The International Organization for Standardization. In vitro diagnostic test systems-requirements for blood glucose monitoring systems for self-testing in managing diabetes mellitus. ISO/TC 212/SC. International Standard ISO 15197. Geneva, Switzerland: ISO;2003.
7. The International Organization for Standardization. In vitro diagnostic test systems-requirements for blood glucose monitoring systems for self-testing in managing diabetes mellitus. ISO/TC 212/SC. Draft International Standard ISO/DIS 15197. Geneva, Switzerland: ISO;2011.
8. Parkes JL, Slatin SL, Pardo S, Ginsberg BH. A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose. Diabetes Care. 2000; 23:1143–1148.
crossref
9. Tonyushkina K. Glucose meters: a review of technical challenges to obtaining accurate results. J Diabetes Sci Technol. 2009; 3:971–980.
crossref
10. Sacks DB, Arnold M, Bakris GL, Bruns DE, Horvath AR, Kirkman MS, et al. Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus. Clin Chem. 2011; 57:e1–e47.
crossref
11. The International Organization for Standardization. In vitro diagnostic test systems -- Requirements for blood glucose monitoring systems for self-testing in managing diabetes mellitus. ISO/TC 212/SC. International Standard ISO 15197. Geneva, Switzerland: ISO;2013.
12. An D, Chung HJ, Lee HW, Lee W, Chun S, Min WK. [Analytical performance evaluation of glucose monitoring system following ISO15197]. Korean J Lab Med. 2009; 29:423–429.
crossref
13. Hussain K. The inaccuracy of venous and capillary blood glucose measurement using reagent strips in the newborn period and the effect of haematocrit. Early Hum Dev. 2000; 57:111–121.
crossref
14. Tang Z, Lee JH, Louie RF, Kost GJ. Effects of different hematocrit levels on glucose measurements with handheld meters for point-of-care testing. Arch Pathol Lab Med. 2000; 124:1135–1140.
crossref
TOOLS
Similar articles