Jung, Kim, Choi, Shin, Lee, and Cho: Performance of the Elecsys HIV combi PT Assay Compared to the ARCHITECT HIV Ag/Ab Combo Assay

Abstract

Background

The aim of this study was to compare the analytical sensitivity and specificity of the recently updated 4th generation Elecsys HIV combi PT assay (Roche Diagnostics GmbH, Germany) to those of the ARCHITECT HIV Ag/Ab Combo assay (Abbott Laboratories, Germany).

Methods

A total of 2,003 fresh random clinical samples, 4 HIV seroconversion panels, a WHO International Standard p24 antigen sensitivity panel, 5 HIV-1 subtype viral lysates, and 5 HIV-1 subtype antibodies were tested in comparative studies with the Elecsys HIV combi PT and ARCHITECT HIV Ag/Ab Combo assays. Samples were assayed with both tests on the same day. The MP Diagnostics HIV Western Blot 2.2 Assay, the Elecsys HIV p24 Ag Test and Confirmatory Test, and the COBAS AmpliPrep/COBAS TaqMan HIV-1 Test were performed as supplementary tests.

Results

Both the Elecsys and ARCHITECT assays detected viral antigens in all four seroconversion panels on the same bleed days, and had lower limits of detection of <1 IU/mL with the p24 antigen sensitivity panel. The ARCHITECT assay showed slightly higher sensitivity in detecting viral antigens with some HIV-1 subtype viral lysates, while the Elecsys assay was more sensitive in detecting each of the 5 HIV-1 subtype antibodies. Both assays detected 5/5 HIV+ clinical samples correctly. The analytical specificities of the Elecsys and ARCHITECT assays were 99.90% and 99.80%, respectively.

Conclusions

The Elecsys HIV combi PT assay performed comparably to the ARCHITECT HIV Ag/Ab Combo assay. Thus, the Elecsys HIV combi PT assay is suitable for diagnostic testing in university hospital settings.

Figures and Tables

Table 1
Characteristics of clinical samples and analytical sensitivity-evaluation samples
lmo-4-157-i001

Abbreviations: dil, dilutions; DM, diabetes mellitus.

Table 2
Method summary of the Elecsys HIV Combi PT and ARCHITECT HIV Ag/Ab Combo assays
lmo-4-157-i002

Abbreviations: Ag, antigen; Ab, antibody; Rx, reaction; MAb, monoclonal antibody.

Table 3
Sensitivities of the Elecsys HIV Combi PT and ARCHITECT HIV Ag/Ab Combo assays
lmo-4-157-i003

*The analytical sensitivity of either assay for p24 antigen detection (IU/mL) was determined using GraphPad Prism 5.0. Maximum dilution of HIV-1 antigen or antibody detected; WHO International Standard NIBSC Code: 90/636.

Abbreviation: d, days.

Table 4
Performance of the Elecsys HIV Combi PT and ARCHITECT HIV Ag/Ab Combo assays
lmo-4-157-i004

*Two samples (one false positive by Elecsys HIV Combi PT assay and one false positive by ARCHITECT HIV Ag/Ab Combo assay) without confirmatory test results were not included in the specificity analysis.

Table 5
Summary of HIV testing results and clinical findings for eight discrepant cases between the Elecsys and ARCHITECT assays
lmo-4-157-i005

Abbreviations: s/co, signal-to-cutoff ratio; WB, Western blot; N/A, not applicable; TND, target not detected; ND, not done; CKD, chronic kidney disease.

Notes

This article is available from http://www.labmedonline.org

References

1. Weber B, Fall EH, Berger A, Doerr HW. Reduction of diagnostic window by new fourth-generation human immunodeficiency virus screening assays. J Clin Microbiol. 1998; 36:2235–2239.
crossref
2. Gürtler L, Muhlbacher A, Michl U, Hofmann H, Paggi GG, Bossi V, et al. Reduction of the diagnostic window with a new combined p24 antigen and human immunodeficiency virus antibody screening assay. J Virol Methods. 1998; 75:27–38.
crossref
3. Yun KA, Oh HB, Oh SY, Hwang BK. Evaluation of Vironostika HIV Uni-Form II Ag/Ab enzyme immunoassay kit for simultaneous detection of antigen and antibody of HIV. Korean J Blood Transfus. 2000; 11:1–11.
4. Kang HJ, Yoo KH, Kim HS, Cho HC. Evaluation of Abbott fourth generation HIV antigen and antibody assays. Korean J Lab Med. 2006; 26:39–44.
crossref
5. Park KI, Kim JM, Kim HS. Evaluation of a simultaneous HIV antigen and antibody detection test in Korean population. Yonsei Med J. 2001; 42:185–193.
crossref
6. Kwon JA, Yoon SY, Lee CK, Lim CS, Lee KN, Sung HJ, et al. Performance evaluation of three automated human immunodeficiency virus antigen-antibody combination immunoassays. J Virol Methods. 2006; 133:20–26.
crossref
7. Ly TD, Laperche S, Brennan C, Vallari A, Ebel A, Hunt J, et al. Evaluation of the sensitivity and specificity of six HIV combined p24 antigen and antibody assays. J Virol Methods. 2004; 122:185–194.
crossref
8. Malm K, von Sydow M, Andersson S. Performance of three automated fourth-generation combined HIV antigen/antibody assays in large-scale screening of blood donors and clinical samples. Transfus Med. 2009; 19:78–88.
crossref
9. Weber B, Gurtler L, Thorstensson R, Michl U, Muhlbacher A, Burgisser P, et al. Multicenter evaluation of a new automated fourth-generation human immunodeficiency virus screening assay with a sensitive antigen detection module and high specificity. J Clin Microbiol. 2002; 40:1938–1946.
crossref
10. Weber B, Orazi B, Raineri A, Thorstensson R, Burgisser P, Muhlbacher A, et al. Multicenter evaluation of a new 4th generation HIV screening assay Elecsys HIV combi. Clin Lab. 2006; 52:463–473.
11. Yeom JS, Jun G, Chang Y, Sohn MJ, Yoo S, Kim E, et al. Evaluation of a new fourth generation enzyme-linked immunosorbent assay, the LG HIV Ag-Ab Plus, with a combined HIV p24 antigen and anti-HIV-1/2/O screening test. J Virol Methods. 2006; 137:292–297.
crossref
12. Powers KA, Ghani AC, Miller WC, Hoffman IF, Pettifor AE, Kamanga G, et al. The role of acute and early HIV infection in the spread of HIV and implications for transmission prevention strategies in Lilongwe, Malawi: a modelling study. Lancet. 2011; 378:256–268.
crossref
13. WHO. Service delivery approaches to HIV testing and counselling (HTC): a strategic policy framework. 2012. Updated on July 2012. http://www.who.int/hiv/pub/vct/htc_framework/en/index.html.
14. EU. Commission Decision of 27 November 2009 amending Decision 2002/364/EC on common technical specifications for in vitro diagnostic medical devices. 2009. Updated on Dec 2009. http://eurlex.europa.eu/LexUriServ/LexUriServ.do?uri=OJ:L:2009:318:0025:0040:en:PDF.
15. Korea centers for disease control and prevention. Summary of HIV/AIDS notification report in Korea. 2012. Updated on Sep 2013. http://www.cdc.go.kr/CDC/info/CdcKrInfo0301.jsp?menuIds=HOME001-MNU1132-MNU1138-MNU0037-MNU1380&cid=21627.
16. UNAID. A global view of HIV infection. Report on the global AIDS epidemic. Updated on Nov 2009. http://www.unaids.org/documents/20101123_2010_HIV_Prevalence_Map_em.pdf.
17. Cha YJ. Quality Assessment of immune serology results report. 2012. Updated on Feb 2013. http://www.lab-qa.org/sub/catalog.php?CatNo=43&Mode=view&no=257.
18. Kim S, Lee JH, Choi JY, Kim JM, Kim HS. False-positive rate of a "fourth-generation" HIV antigen/antibody combination assay in an area of low HIV prevalence. Clin Vaccine Immunol. 2010; 17:1642–1644.
crossref
19. Shima-Sano T, Yamada R, Sekita K, Hankins RW, Hori H, Seto H, et al. A human immunodeficiency virus screening algorithm to address the high rate of false-positive results in pregnant women in Japan. PLoS One. 2010; 5:e9382.
crossref
20. Song EY, Hur M, Roh EY, Park MH, Moon HW, Yun YM. Performances of four fourth-generation human immunodeficiency virus-1 screening assays. J Med Virol. 2012; 84:1884–1888.
crossref
21. Mühlbacher A, Schennach H, van Helden J, Hebell T, Pantaleo G, Burgisser P, et al. Performance evaluation of a new fourth-generation HIV combination antigen-antibody assay. Med Microbiol Immunol. 2013; 202:77–86.
crossref
22. Tao CM, Cho Y, Ng KP, Han X, Oh EJ, Zainah S, et al. Validation of the Elecsys(R) HIV combi PT assay for screening and reliable early detection of HIV-1 infection in Asia. J Clin Virol. 2013; 58:221–226.
crossref
23. Roche Diagnostics GmbH. Elecsys HIV combi PT package insert. 2011.
24. Abbott Laboratories. ARCHITECT HIV Ag/Ab combo package insert. 2009.
25. National Institute of Health, Immune pathology center, AIDS tumor virus team. Molecular epidemiology of HIV in Korea. Updated on Aug 2012. http://www.cdc.go.kr/CDC/info/CdcKrInfo0301.jsp?menuIds=HOME001-MNU1132-MNU1138-MNU0037-MNU1380&cid=12203.
26. Ly TD, Plantier JC, Leballais L, Gonzalo S, Lemee V, Laperche S. The variable sensitivity of HIV Ag/Ab combination assays in the detection of p24 Ag according to genotype could compromise the diagnosis of early HIV infection. J Clin Virol. 2012; 55:121–127.
crossref
27. Jurriaans S, Back NK, Wolthers KC. Ten years of HIV testing with fourth generation assays: the Amsterdam experience. J Clin Virol. 2011; 52:Suppl 1. S67–S69.
crossref