Abstract
Background
To achieve consistency in poikilocytes grading in peripheral blood cell examinations, we made an image-based differential count (IDC) software to measure the degree of abnormalities in individual red blood cells (RBCs) and relative fractions of poikilocytes.
Methods
Thirty peripheral blood samples were analyzed. Smear slides were examined on a microscope with charge-coupled device (CCD) camera. To verify this program, we compared the IDC results with the results of manual differential counting (MDC). Relative fractions of schistocytes, echinocytes, and elliptocytes were measured by IDC and MDC. The error rate of IDC was measured by confirming the final processed images of IDC. Correlations of IDC and MDC results were compared using linear regression analysis and the time required for each test was measured. For presentation of the mathematical decision criteria of poikilocytes, IDC algorithms for recognizing schistocytes, echinocytes, and elliptocytes were made using simple geometrical or mathematical formulas.
Figures and Tables
Table 2
Schistocyte grading system: +/- ≤1%, 1% <1+ ≤3%, 3% <2+ ≤6%, 6% <3+ ≤12%, 4+ >12%. Echinocyte grading system: +/- ≤10%, 10% <1+ ≤25%, 25% <2+ ≤50%, 50% <3+ ≤75%, 4+ >75%. Elliptocyte grading system: +/- ≤6%, 6% <1+ ≤20%, 20% <2+ ≤50%, 50% <3+ ≤75%, 4+ >75%.
Abbreviations: MDC, manual differential counting; IDC, image-based differential counting.
References
1. Bull BS. Lichtman MA, editor. Morphology of the erythron. Williams hematology. 2006. 7th ed. New York: McGraw-Hill;377–382.
2. Hyun BH, Gulati GL, Ashton JK. Differential leukocyte count: manual or automated, what should it be? Yonsei Med J. 1991. 32:283–291.
3. Gulati GL, editor. Blood cell morphology grading guide. 2009. 1st ed. Chicago: American Society of Clinical Pathology Press;1–33.
4. Briggs C, Longair I, Slavik M, Thwaite K, Mills R, Thavaraja V, et al. Can automated blood film analysis replace the manual differential? An evaluation of the CellaVision DM96 automated image analysis system. Int J Lab Hematol. 2009. 31:48–60.
5. Rümke CL. The imprecision of the ratio of two percentages observed in differential white blood cell counts: a warning. Blood Cells. 1985. 11:137–140.
6. Beucher S. Watersheds of functions and picture segmentation. Proc IEEE Int Conf Acoust Speech Signal Process. 1982. 7:1928–1931.
7. Ceelie H, Dinkelaar RB, Van Gelder W. Examination of peripheral blood films using automated microscopy; evaluation of Diffmaster Octavia and Cellavision DM96. J Clin Pathol. 2007. 60:72–79.
8. Lesesve JF, Salignac S, Alla F, Defente M, Benbih M, Bordigoni P, et al. Comparative evaluation of schistocyte counting by an automated method and by microscopic determination. Am J Clin Pathol. 2004. 121:739–745.
9. Sir JU, Kim TY. A new digital image processing technique for mixed pattern analysis of anti-neutrophil cytoplasmic antibody test. J Korean Rheum Assoc. 2004. 11:333–341.