Lee, Park, Yoon, Lee, and Cho: Evaluation of ExiPrep16 Automated System for the Extraction of Nucleic Acids from Nasopharyngeal Swabs for the Detection of Respiratory Viruses

Abstract

Background

Automated nucleic acid extraction offers a standardized sample treatment method, low error rate, and avoids sample nucleic acid contamination for use in molecular diagnostics. Here, we evaluated the performance of automated ExiPrep16 system (Bioneer Co.) in comparison with the manual Viral Gene-spin Viral DNA/RNA Extraction kit (VGspin; iNtRON Biotechnology Inc.) for the detection of respiratory viruses from nasopharyngeal flocked swabs.

Methods

To compare the agreement rate and analytical sensitivity between ExiPrep16 and VGspin, previously collected 78 patient samples and 11 pooled samples of each respiratory viruses and their serially diluted samples (until 1/108), were tested by multiplex reverse-transcriptase PCR (Seeplex RV 12 ACE Detection kit; SeeGene Inc.). In addition, we repeatedly analyzed the threshold cycle of the pooled and 1/103 dilution of adenovirus (ADV) and influenza virus A (Flu-A) by using real-time PCR to evaluate the precision and crossover of the ExiPrep16 system.

Results

The analytical sensitivity of the ExiPrep16 was comparable to that of VGspin, and the highest detectable dilution varied in the range of 1/10 to 1/106 depending on the viruses. The total, overall positive and negative percent agreements of ExiPrep16 in comparison with VGspin were 95.7%, 96.2%, and 95.2%, respectively. The mean (CV%) of pooled and 1/103 dilution of ADV were, respectively, 19.2 cycle (2.1%) and 31.6 cycle (4.3%) and those for Flu-A were 22.6 cycle (3.1%) and 35.5 cycle (2.6%). No carryover was detected.

Conclusions

Compared to the manual VGspin, ExiPrep16 ensured nucleic acid extraction for efficient detection of respiratory viruses.

Figures and Tables

Table 1
Comparison of the analytical sensitivity between manual Viral gene-spin Viral DNA/RNA Extraction kit (VGspin) and automatic ExiPrep16 system
lmo-3-227-i001

+: thick band, w+: thin band confirmed repeatedly, ww+: thin band, but cannot be confirmed repeatedly.

Abbreviations: ADV, adenovirus; Flu-A, influenza virus A; Flu-B, influenza virus B; PIV, parainfluenza virus; RSV, respiratory syncytial virus; hMPV, human metapneumovirus; hRV, human rhinovirus.

Table 2
Agreement rate between manual VGspin kitand automatic ExiPrep16 system for the detection of respiratory viruses in flocked nasopharyngeal swabs
lmo-3-227-i002

*Measurement of agreement (K) between ExiPrep16 and VGspin was 0.914; including 1 case of RSV-A and 1 case of hRV-A; including 1 case of hMPV and 1 case of ADV.

Table 3
Reproducibility of Ct values analysed by real-time PCR using nucleic acid extracted by automatic ExiPrep16 system
lmo-3-227-i003

Abbreviations: Ct, threshold cycle; CI, confidence interval.

Notes

This article is available from http://www.labmedonline.org

References

1. Mahony JB. Detection of respiratory viruses by molecular methods. Clin Microbiol Rev. 2008; 21:716–747.
crossref
2. Lee JH, Shin SR, Cho JH. Evaluation of direct immunofluorescence test with PCR for detection of novel influenza A (H1N1) virus during 2009 pandemic. Yonsei Med J. 2011; 52:680–682.
crossref
3. Chan KH, Yam WC, Pang CM, Chan KM, Lam SY, Lo KF, et al. Comparison of the NucliSens easyMAG and Qiagen BioRobot 9604 nucleic acid extraction systems for detection of RNA and DNA respiratory viruses in nasopharyngeal aspirate samples. J Clin Microbiol. 2008; 46:2195–2199.
crossref
4. Espy MJ, Uhl JR, Sloan LM, Buckwalter SP, Jones MF, Vetter EA, et al. Real-time PCR in clinical microbiology: applications for routine laboratory testing. Clin Microbiol Rev. 2006; 19:165–256.
crossref
5. Riemann K, Adamzik M, Frauenrath S, Egensperger R, Schmid KW, Brockmeyer NH, et al. Comparison of manual and automated nucleic acid extraction from whole-blood samples. J Clin Lab Anal. 2007; 21:244–248.
crossref
6. Beuselinck K, van Ranst M, van Eldere J. Automated extraction of viral-pathogen RNA and DNA for high-throughput quantitative real-time PCR. J Clin Microbiol. 2005; 43:5541–5546.
crossref
7. Aldous WK, Pounder JI, Cloud JL, Woods GL. Comparison of six methods of extracting Mycobacterium tuberculosis DNA from processed sputum for testing by quantitative real-time PCR. J Clin Microbiol. 2005; 43:2471–2473.
crossref
8. Thakur R, Sarma S, Goyal R. Comparison of DNA extraction protocols for Mycobacterium Tuberculosis in diagnosis of tuberculous meningitis by real-time polymerase chain reaction. J Glob Infect Dis. 2011; 3:353–356.
9. Kohdera U, Kino M, Ito M. Detection of adenovirus DNA in throat swabs and blood by SYBR green real-time PCR assay in patients with adenovirus-associated tonsillitis. Jpn J Infect Dis. 2006; 59:394–396.
10. Clinical and Laboratory Standards Institute. User protocol for evaluation of qualitative test performance; approved guideline. 2nd ed, EP12-A2. Wayne, PA: Clinical and Laboratory Standards Institute;2008.
11. Clinical and Laboratory Standards Institute. Evaluation of precision performance of quantitative measurement methods; approved guideline. 2nd ed, EP5-A2. Wayne, PA: Clinical and Laboratory Standards Institute;2004.
12. Yoon KH, Cho JH. Detection of respiratory viruses in children by multiplex reverse transcriptase PCR, direct immunofluorescence assay, and shell vial culture. Korean J Clin Microbiol. 2009; 12:110–115.
crossref
13. Yoo SJ, Kuak EY, Shin BM. Detection of 12 respiratory viruses with two-set multiplex reverse transcriptase-PCR assay using a dual priming oligonucleotide system. Korean J Lab Med. 2007; 27:420–427.
crossref
14. Cho JH, Lee HS, Lee KE, Park DS, Lee YJ, Moon HB, et al. Evaluation of a quantitative RealArt HBV LC PCR assay for hepatitis B virus by real-time PCR. Korean J Clin Micobiol. 2007; 10:25–31.
TOOLS
Similar articles