I. INTRODUCTION
Accurate determination of the length of the root canal is one of the most important preconditions for successful root canal treatment
1). Establishing the length of the root canal system at the apical constriction is considered an ideal working length for endodontic treatment.
The traditional radiographic method has shortcomings
2), including the use of ionizing radiation, and the potential inaccuracy in root canal due to difference between radiographic apex and actual apex. In addition, technical problems such as film positioning, beam angulation, processing problems
3) have been the catalyst for the develop ment of alternative methods for the determination of root canal length. Numerous devices have appeared on the market since Sunada
4) first introduced electronic measurement of root canal length in 1962.
Unfortunately, many apex locators are inaccurate on root canals that contain moisture, vital pulp tissue, blood and other exudates or remnants of intracanal irrigants
5-
7). Recently, the 'ratio method' for measuring root canal length was introduced
8). In this method, the quotient of two simultaneously measured impedances of two different frequencies is calculated to reveal the position of an electrode (endodontic file) inside the root canal. The latest generations of apex locators are based on this principle and have many advantages when compared to earlier devices, especially because these instruments are reliable in both dry and wet canals.
Since the development of the most recent generation, electronic apex locators (EALs) have gained in popularity. This generation uses two frequencies and enables tooth length measurements in the presence of electrical conductive media in the root canals
9). Accuracy of the recent generation of EAL averages around from 83% to 93%
3,
5,
7,
10-
12). An EAL that further improves the accuracy rate is desirable and, if proven to be a reliable tool, could potentially replace, in many instances, the classic radiographic method for tooth length determination.
EALs are frequently used attached to a small size endodontic file, however, the effect of the relative diameters of the file on the measurement of the root canal length has not been clarified. Experimentally, researchers
3,
10,
13) often used files that appeared to match the canal diameter. But the manufacturer claims that the accuracy of measurement is allegedly not affected by the size of the canal or that of the measuring instrument
14). It was necessary to clarify whether the accuracy of measurement would be affected by the use of a small size instrument in large canal. Because working length can be changed through-out canal shaping, it is necessary to measure electronic length again before canal filling. In this aspect, it is worth whether the size of measuring file affects accuracy of the EAL.
The purpose of this study was to compare the accuracy of measurements in pre-enlarged canals with small size instruments using four apex locators and to compare the differences between the measurements, in enlarged canal, with small size instruments and instruments that match the actual canal diameter with each apex locators.
IV. DISCUSSION
In 1962, Sunada
4) developed the first apex locator based on the assumption that there is a constant resistance of 6.5 kΩ at a defined current between the periodontal membrane and oral mucosa. Since this introduction, EALs have gained in popularity, especially since the development of the most recent generation. This generation uses two frequencies and is based on the change in impedance of the probing electrode to tissue fluids. When a file tip is located away from the minor foramen, the impedance in the canal is negligible, but when the file reaches the immediate vicinity of the minor foramen, the magnitude of the impedance of the canal suddenly increases. As the file tip contacts the periapical tissue, such as the periodontal ligament, the impedance value rapidly decreases, indicating that the file is beyond the minor foramen.
The main shortcoming of early apex locators such as erroneous readings with electrolytes was overcome with the introduction of the ratio method and the subsequent development of the self-calibrating Root ZX
8). The ratio method works on the principle that two electric currents with different sine wave frequencies will have measurable impedances that can be measured and compared as a ratio regardless of the type of electrolyte in the canal. The capacitance of a root canal increases significantly at the apical constriction, and the quotient of the impedances reduces rapidly as the apical constriction is reached. Kobayashi and Suda
8), in 1994, showed that the ratio of different frequencies have definitive values, and that the ratio rate of change did not change with different electrolytes in the canal.
The Root ZX uses two different frequencies (8 kHz and 400 Hz) to simultaneously measure the impedances in the canal. The device then determines a quotient value by dividing the 8 kHz impedance value by the 400 Hz impedance value. The minor foramen is located when the quotient equals 0.67. Different types of fluids in the canal will give different impedance values. By using two frequencies, almost devices currently can be used in all types of fluids because the quotient (0.67) is always the same.
The Root ZX and Bingo 1020 use two separate frequencies 400 Hz and 8 kHz, but e-Magic Finder uses 500 Hz and 5 kHz. The manufacturers of Bingo 1020 claim that the combination of using only one frequency at a time and basing measurements on the root mean square values of the signals increases the measurement accuracy and the reliability of the device.
Several studies have assessed the accuracy of the Root ZX. An
in vivo study carried out by Shabahang et al.
20) produced values to a precision of 96.2%. Also, Nguyen et al.
19) showed that the accuracy of Root ZX was not affected by a wide apical foramen. An
in vitro study of the Bingo 1020 found it to be as reliable as the Root ZX
17). Tinaz et al.
16) found the Bingo 1020 to be as accurate as the Root ZX in an
in vitro study and easier for a beginner to use in preflared canals.
There were many studies about Root ZX and Bingo 1020, but were rarely about SmarPex and e-Magic Finder. So, this study was designed and performed.
The instruction manual for the Root ZX states: "the bar indicating the apical constriction of the root canal flashes on and off… which indicates that the tip of the file is in the vicinity of the apical foramen (an average of 0.2 - 0.3 mm past the apical constriction towards the apex)" A study by Weik et al.
21) demonstrated that the mean distance beyond the minor diameter was 0.19 mm, with - 0.5 mm to + 1.73 mm. Thus, we could obtain more definite length by reading electronic length in index indicating apex of each device.
The
in vitro model
2) for assessing the accuracy of EAL measurement utilized herein proved to be easy to fabricate and remained intact for the duration of the study. The rigid alginate firmly supported the teeth and permitted canal enlargement to be performed with the roots embedded. Thus, the operator remained 'blinded' as to the location of the file tip in the canal and had to rely on the EAL for controlling the apical extent of canal enlargement. In this respect, the alginate model appeared to simulate the clinical situation better than the previously used gelatin models
13,
22,
23).
In S10, the accuracy of EALs was revealed in the order of e-Magic Finder, Root ZX, Bingo 1020 and SmarPex. There was significant difference between e-Magic Finder and SmarPex. In L10 and L40, the accuracy of EALs was revealed in the order of e-Magic Finder, Bingo 1020, Root ZX and SmarPex. The e-Magic Finder and Bingo 1020 were more accurate than SmarPex. These results might come from the mechanical and electrical characteristics, e.g. microprocessor and internal calibration, of each device.
Nguyen et al.
19) reported the length measurement obtained with small and large size files were comparable in the enlarged canals. Similarly, all devices had not statistically significant differences between L10 and L40.
In addition, the ± 0.5 mm to the foramen range has been considered as the strictest acceptable clinical range
24,
25). Thus, measurements attained within this tolerance are considered accurate. By the way, in this study, SmarPex showed fairly lower rates of measurement within ± 0.5 mm of actual length than others.
The results of this study illustrate some variables that must be considered when using root canal length measuring devices. It seems users of apex locator in clinic need to consider the measurement accuracy of each device could be affected by various using condition such as file size, intracanal conductive media. Sometimes, by the measuring conditions, the device's measurements may be exaggerated to exceed the clinically tolerable limit.
Further studies about the accuracy of various apex locators in various using conditions are needed.