Namkoong, Hong, Kim, and Park: Reliability on Intra-Laboratory and Inter-Laboratory Data of Hair Mineral Analysis Comparing with Blood Analysis

Abstract

Background

Nowadays, although its clinical value remains controversial institutions utilize hair mineral analysis. Arguments about the reliability of hair mineral analysis persist, and there have been evaluations of commercial laboratories performing hair mineral analysis.

Objective

The objective of this study was to assess the reliability of intra-laboratory and inter-laboratory data at three commercial laboratories conducting hair mineral analysis, compared to serum mineral analysis.

Methods

Two divided hair samples taken from near the scalp were submitted for analysis at the same time, to all laboratories, from one healthy volunteer. Each laboratory sent a report consisting of quantitative results and their interpretation of health implications. Differences among intra-laboratory and interlaboratory data were analyzed using SPSS version 12.0 (SPSS Inc., USA).

Results

All the laboratories used identical methods for quantitative analysis, and they generated consistent numerical results according to Friedman analysis of variance. However, the normal reference ranges of each laboratory varied. As such, each laboratory interpreted the patient's health differently. On intra-laboratory data, Wilcoxon analysis suggested they generated relatively coherent data, but laboratory B could not in one element, so its reliability was doubtful. In comparison with the blood test, laboratory C generated identical results, but not laboratory A and B.

Conclusion

Hair mineral analysis has its limitations, considering the reliability of inter and intra laboratory analysis comparing with blood analysis. As such, clinicians should be cautious when applying hair mineral analysis as an ancillary tool. Each laboratory included in this study requires continuous refinement from now on for inducing standardized normal reference levels.

INTRODUCTION

Scalp hair has been widely accepted for the evaluation of toxic element exposure and has been performed by clinical laboratories since it was first used to assess systemic levels of the elements in 19291. Researchers have found correlations between essential elements and diseases, metabolic disorders, environmental exposures, and nutritional status2-7.
However, according to several studies that assessed the reliability of commercial laboratories performing hair mineral analysis, it has been suggested that using hair to measure essential element is meaningless8-13.
The usefulness of hair mineral analysis remains controversial. Nevertheless, much technical progress has been made in the laboratory field in the last 10 years. The aim of this study was to evaluate the performance of current commercial laboratories performing hair mineral analysis in terms of intra-laboratory and inter-laboratory reliability, compared to the analysis of serum minerals.

MATERIALS AND METHODS

Enrolled laboratories

At the time of this study, there were three commercial institutions conducting hair mineral analysis in Korea. Two of them are based on Korea, and one is a branch of a large laboratory headquartered in the USA, which receives samples from Korea for analysis. All three laboratories were included in this study.

Donor selection

Hair was donated by a 23 year-old healthy, black-haired, Asian man. He had no medical or drug history. Routine blood tests and serum levels of copper, zinc, iron and magnesium were within the normal limits. Electrocardiogram (EKG) and chest radiograph were normal. This study was approved by the Institutional Review Board of the Dankook University Hospital, and the hair donor provided informed consent prior to participation.

Methods of hair analysis

Hair was cut adjacent to the scalp from several areas of the parieto-occipital regions, using clean stainless steel scissors. Hairs measuring 3 cm from the scalp were included. Collected hairs were intermingled, weighed, and divided into 6 groups according to the amounts required by each laboratory, ranging from 60~80 mg. For each laboratory, two divided samples were submitted at the same time with different IDs.

Statistical analysis

Differences among laboratories were analyzed by Friedman analysis of variance, using the mean value of two results at each laboratory. Differences between intra-laboratory data were analyzed by the Wilcoxon signed rank sum test. In all analyses, SPSS for windows (version 12.0; SPSS Inc., Chicago, IL, USA) was used. Data were valued within a confidence interval of 95%. A p-value of less than 0.05 was considered statistically significantly different (p<0.05).

RESULTS

Features of the laboratories

The features of the laboratories are summarized in Table 1. All 3 laboratories analyzed hair samples with inductively coupled plasma mass spectrometry (ICP-MS). The advantage of ICP-MS is its lower detection threshold compared to inductively coupled plasma atomic emission spectrometry (ICP-AES). The required quantity for analysis at each laboratory was 60, 60 and 80 mg, respectively. Laboratory A analyzed 30 elements, and B and C analyzed 26 and 38 elements, respectively. In total, 24 elements including aluminum, arsenic, barium, bismuth, cadmium, calcium, chromium, cobalt, copper, iron, lead, lithium, magnesium, manganese, molybdenum, mercury, phosphorus, potassium, selenium, sodium, sulfur, uranium, vanadium, and zinc were commonly analyzed by all 3 laboratories. Laboratories provided numerical results and interpretation, which included an evaluation of health status, prediction of possible illnessess in the future, and recommendation on dietary habit. Two laboratories (B, C) added promotions of branded supplements according to the results.
Each laboratory presented a normal reference range for each analyzed element. They were summarized in Table 2. In many elements, all 3 laboratories used different reference ranges, based on studies. As seen in Table 2, the inter-laboratory differences of the reference range for several elements were sufficiently wide in that they did not overlap.

Reliability of inter-laboratory data

The laboratories presented their results on split hair samples as illustrated in Table 3. Statistically, the Friedman test of variance indicated no significant difference between different numerical results from 6 samples (p=0.513).
In the interpretation of their own results, the laboratories did not reach a general consensus although their numerical results were coherent, due to different reference ranges. As summarized in Table 4, laboratory A commented that the donor lacked iron and selenium. Laboratory B identified a cobalt, copper, selenium, zinc and phosphorus deficiency. According to laboratory C, the donor was only deficient in molybdenum. As a result, these differences resulted in the laboratories recommending different nutritional supplements and treatments to the hair donor.

Reliability of intra-laboratory data

The laboratories did not reproduce identical results, but statistically they were coherent numerical results according to the Wilcoxon signed rank sum test (p=0.513).
On interpretation, laboratory B failed to produce consistent results. B detected a cobalt, copper, selenium, and zinc deficiency in one hair sample, and a deficiency in cobalt, copper, selenium, and phosphorus in another one.

Comparison of blood test and hair mineral analysis

The laboratories indicated that the donor might experience symptoms such as chronic fatigue, dizziness, insomnia, depression, and anxiety. They also claimed he was at increased risk of developing osteomalasia, skin disease, arthralgia, anemia, hypercholesterolemia, alopecia and myalagia. In actuality, the donor did not experience health problems, and his blood test, EKG, and chest radiograph were unremarkable. Specifically, his serum levels of copper, zinc, iron, and magnesium was normal, but his iron level was below the normal limit according to laboratory A. Laboratory B concluded that he was copper and zinc deficient (Table 5).

DISCUSSION

Hair analysis provides good evidence in the fields of forensic and clinical toxicology, doping control, and occupational medicine, but do not demonstrate clear evidence in the fields of general health screening. While urine and blood tests show the recent and current body status, hair represents a longer time frame, which can incorporate many years1. Since the elements are present in the hair at higher levels, more sensitive and accurate analysis results can be expected9. As such, it has been suggested that hair mineral analysis can be a cost-effective means of screening an individual, or monitoring a patient during putative toxic exposure or its therapy10.
However, the analysis of elements in hair remains controversial. There exists little or no evidence to support hair as a true biological marker. Analytical accuracy and lack of standardization have been cited as the main challenge. In the field of dermatology, trichothiodystrophy is a well known candidate disease for hair mineral analysis, but other skin diseases including hair diseases are not. In 1985, Barrett8 asserted that commercial use of hair analysis is unscientific and economically wasteful after assessing the reliability of commercial hair mineral test results. Seidel et al.11 concluded that commercial laboratories failed to generate consistent results in 2001. Seidel et al.11 also called the varied normal reference ranges into question. In a similar study conducted in Germany in 2002, the accuracy of hair mineral analysis was refuted12.
In a matter of accuracy, it displayed different results in our study compared to those previously. All 3 commercial laboratories generated coherent results on identical hair samples in repeated analyses. There were imperceptible numerical gaps, but were deemed insignificant by statistical analysis. This may partly be due to identical analysis methods, unlike those studied previously. The non-standardized reference ranges remained a major challenge. In inter-laboratory comparison, each laboratory gave different interpretations on the same patient, because they applied different normal ranges. Insufficient elements were detected and possible predicted future health problems varied, and each laboratory recommended different nutritional supplements.
In intra-laboratory comparison, laboratory A, C made the same interpretations but laboratory B did not. Laboratory B needs to improve its detection consistency.
Compared with serum copper, zinc, iron and magnesium, laboratory A gave a different interpretation in iron and B gave one in copper, and zinc. Laboratory C made an identical interpretation in all three elements. These differences represent a challenge in the evaluation of a patient's health. More research is required to evaluate for the correlation between clinical state, hair mineral analysis and blood test.
For now, it appears that we can apply hair analysis carefully as an ancillary tool to evaluate patients, in addition to history taking, physical examination, diet, lifestyle, the environment, and other laboratory data because each laboratory can generate precise numerical results1,14. However, hair mineral analysis cannot be a primary tool for the assessment of individual health without a well-established normal reference range, because the interpretation of results varies greatly according to the wide diverse reference range of each laboratory. It remains a challenge to establish 'normal' reference ranges for minor and trace elements in human hair due to the natural variance of hair compositions9. Thus, well-designed large-scale studies are needed to standardize international reference ranges for hair mineral elements.

Figures and Tables

Table 1
Characteristics of the selected laboratories
ad-25-67-i001

MS: mass spectrometry.

Table 2
Reference value of each laboratory
ad-25-67-i002

NA: not analyzed.

Table 3
Results of generated at each laboratory
ad-25-67-i003

NA: not analyzed. *Above or below normal reference range.

Table 4
Elements below the normal reference range at each laboratory
ad-25-67-i004
Table 5
Comparison of the result between blood test and hair mineral analysis
ad-25-67-i005

*Below normal reference range.

ACKNOWLEDGMENT

The present research was conducted by the research fund of Dankook University in 2010.

References

1. Bass DA, Hickock D, Quig D, Urek K. Trace element analysis in hair: factors determining accuracy, precision, and reliability. Altern Med Rev. 2001. 6:472–481.
2. Suliburska J, Bogdański P, Pupek-Musialik D, Krejpcio Z. Dietary intake and serum and hair concentrations of minerals and their relationship with serum lipids and glucose levels in hypertensive and obese patients with insulin resistance. Biol Trace Elem Res. 2011. 139:137–150.
crossref
3. Kim OY, Baek SH, Lee YJ, Lee KH. Association of increased hair calcium levels and enhanced augmentation index (AIx): a marker of arterial stiffness. Biol Trace Elem Res. 2010. 138:90–98.
crossref
4. Park SB, Choi SW, Nam AY. Hair tissue mineral analysis and metabolic syndrome. Biol Trace Elem Res. 2009. 130:218–228.
crossref
5. Yasuda H, Yoshida K, Segawa M, Tokuda R, Tsutsui T, Yasuda Y, et al. Metallomics study using hair mineral analysis and multiple logistic regression analysis: relationship between cancer and minerals. Environ Health Prev Med. 2009. 14:261–266.
crossref
6. Rautio A, Kunnasranta M, Valtonen A, Ikonen M, Hyvärinen H, Holopainen IJ, et al. Sex, age, and tissue specific accumulation of eight metals, arsenic, and selenium in the European hedgehog (Erinaceus europaeus). Arch Environ Contam Toxicol. 2010. 59:642–651.
crossref
7. Grabeklis AR, Skalny AV, Nechiporenko SP, Lakarova EV. Indicator ability of biosubstances in monitoring the moderate occupational exposure to toxic metals. J Trace Elem Med Biol. 2011. 25:Suppl 1. S41–S44.
crossref
8. Barrett S. Commercial hair analysis. Science or scam? JAMA. 1985. 254:1041–1045.
crossref
9. Miekeley N, Dias Carneiro MT, da Silveira CL. How reliable are human hair reference intervals for trace elements? Sci Total Environ. 1998. 218:9–17.
crossref
10. Druyan ME, Bass D, Puchyr R, Urek K, Quig D, Harmon E, et al. Determination of reference ranges for elements in human scalp hair. Biol Trace Elem Res. 1998. 62:183–197.
crossref
11. Seidel S, Kreutzer R, Smith D, McNeel S, Gilliss D. Assessment of commercial laboratories performing hair mineral analysis. JAMA. 2001. 285:67–72.
crossref
12. Drasch G, Roider G. Assessment of hair mineral analysis commercially offered in Germany. J Trace Elem Med Biol. 2002. 16:27–31.
crossref
13. Steindel SJ, Howanitz PJ. The uncertainty of hair analysis for trace metals. JAMA. 2001. 285:83–85.
crossref
14. Hamilton T, Schweinsberg F. The reliability of conclusions based on hair mineral analysis in individual diagnostic. Versicherungsmedizin. 2004. 56:136–140.
TOOLS
ORCID iDs

Byung Cheol Park
https://orcid.org/http://orcid.org/0000-0002-5449-8313

Similar articles