Lee and Song: High field strength magnetic resonance imaging of musculoskeletal diseases

Abstract

Musculskeletal magnetic resonance imaging (MRI) applications are making the transition rapidly from 1.5 tesla (T) to 3T. The higher signal-to-noise ratio (SNR) that is available with a 3T MRI system allows for greater spatial resolution and provides the potential to improve the diagnostic capability of musculoskeletal MRI. With the use of 3T systems, one can enhance the SNR, spatial resolution, and contrast-to-noise ratio of intrinsic joint structures such as osseous, tendinous, cartilaginous, and ligamentous structures, which makes them more discernable and amenable to proper radiologic assessment. The SNR gain and coil technology advances allow for a smaller voxel-size and parallel imaging, reducing the acquisition time without significant signal loss. Three-dimensional (3D) fast spin echo sequences with isotropic resolution reduce partial volume artifacts through the acquisition of thin continuous sections and enable free 3D-multiplanar-reformatting without loss of image quality. This technique may be a promising method to replace currently used 2D sequences in clinical practice. In addition to current clinical applications, 3T MRI will contribute to the development of new molecular and functional MRI techniques.

Figures and Tables

Figure 1
Coronal 3 tesla magnetic resonance images of knee in 69-year-old women show surgically confirmed avulsive tear of posterior root of medial meniscus. 2D coronal intermediateweighted image (A) shows faint intermediate signal intensity (arrow) in posterior medial meniscal root. However, definitely bandlike area of high signal intensity (arrow) is noted in posterior medial meniscal root on reformatted coronal image of 3D space isotropic intermediate-weighted density (B).
jkma-53-1074-g001
Figure 2
A 37-year-old man with focal cartilage defect and subchondral fracture in lateral tibial plateau. 2D coronal intermediate-weighted image (A) and reformatted coronal image of 3D isotropic intermediate-weighted sequence (B) of the knee demonstrate focal chondromalacia (arrow) clearly at 3 tesla magnetic resonance imaging.
jkma-53-1074-g002
Figure 3
A 22-year-old woman with osteonecrosis of right femoral head and secondary osteoarthritis. Fat-suppressed, coronal intermediate-weighted image at 3 tesla magnetic resonance imaging shows well diffuse irregularity of hyaline cartilage within the hip joint due to secondary osteoarthritis.
jkma-53-1074-g003
Figure 4
A 20-year-old man with healing process of subchondral insufficiency fracture at bilateral femoral heads. Coronal T1 VIBE images of both hip joints demonstrate normal acetabular labrum (arrow) within right hip joint (A) and deformed and detached labrum (arrow) of left hip joint (B).
jkma-53-1074-g004
Figure 5
A 53-year-old male patient with partial thickness tear of supraspinatus tendon. A focal defect at articular sur-face of supraspinatus tendon (arrow) was well delineated on post-arthrographic, fat-suppressed coronal T1-weighted image at 3 tesla system.
jkma-53-1074-g005
Figure 6
A 15-year-old girl with wrist pain. Focal discontinuity (arrow) of triangular fibrocartilage is well noted on fat-suppressed coronal intermediate-weighted image of 3 tesla magnetic resonance imaging.
jkma-53-1074-g006

References

1. Mosher TJ. Musculoskeletal imaging at 3T: current techniques and future applications. Magn Reson Imaging Clin N Am. 2006. 14:63–76.
crossref
2. Ramnath RR. 3T MR imaging of the musculoskeletal system (Part II): clinical applications. Magn Reson Imaging Clin N Am. 2006. 14:41–62.
crossref
3. Kladny B, Glückert K, Swoboda B, Beyer W, Weseloh G. Comparison of low-field (0.2 Tesla) and high-field (1.5 Tesla) magnetic resonance imaging of the knee joint. Arch Orthop Trauma Surg. 1995. 114:281–286.
crossref
4. Cotten A, Delfaut E, Demondion X, Lapégue F, Boukhelifa M, Boutry N, Chastanet P, Gougeon F. MR imaging of the knee at 0.2 and 1.5 T: correlation with surgery. AJR Am J Roentgenol. 2000. 174:1093–1097.
5. Loew R, Kreitner KF, Runkel M, Zoellner J, Thelen M. MR arthrography of the shoulder: comparison of low-field (0.2 T) vs high-field (1.5 T) imaging. Eur Radiol. 2000. 10:989–996.
crossref
6. Shellock FG, Bert JM, Fritts HM, Gundry CR, Easton R, Crues JV 3rd. Evaluation of the rotator cuff and glenoid labrum using a 0.2-Tesla extremity magnetic resonance (MR) system: MR results compared to surgical findings. J Magn Reson Imaging. 2001. 14:763–770.
crossref
7. Kreitner KF, Romaneehsen B, Krummenauer F, Oberholzer K, Müller LP, Düber C. Fast magnetic resonance imaging of the knee using a parallel acquisition technique (mSENSE): a prospective performance evaluation. Eur Radiol. 2006. 16:1659–1666.
crossref
8. Zuo J, Li X, Banerjee S, Han E, Majumdar S. Parallel imaging of knee cartilage at 3 Tesla. J Magn Reson Imaging. 2007. 26:1001–1009.
crossref
9. Lichy MP, Wietek BM, Mugler JP 3rd, Horger W, Menzel MI, Anastasiadis A, Siegmann K, Niemeyer T, Königsrainer A, Kiefer B, Schick F, Claussen CD, Schlemmer HP. Magnetic resonance imaging of the body trunk using a single-slab, 3-dimensional, T2-weighted turbo-spin-echo sequence with high sampling efficiency (SPACE) for high spatial resolution ima-ging: initial clinical experiences. Invest Radiol. 2005. 40:754–760.
crossref
10. Norris DG. High field human imaging. J Magn Reson Imaging. 2003. 18:519–529.
crossref
11. Fullerton GD, Cameron IL, Ord VA. Orientation of tendons in the magnetic field and its effect on T2 relaxation times. Radiology. 1985. 155:433–435.
crossref
12. Kijowski R, Davis KW, Woods MA, Lindstrom MJ, De Smet AA, Gold GE, Busse RF. Knee joint: comprehensive assessment with 3D isotropic resolution fast spin-echo MR imaging: diagnostic performance compared with that of conventional MR imaging at 3.0 T. Radiology. 2009. 252:486–495.
crossref
13. Magee T, Williams D. 3.0-T MRI of meniscal tears. AJR Am J Roentgenol. 2006. 187:371–375.
crossref
14. Ramnath RR, Magee T, Wasudev N, Murrah R. Accuracy of 3-T MRI using fast spin-echo technique to detect meniscal tears of the knee. AJR Am J Roentgenol. 2006. 187:221–225.
crossref
15. Brody JM, Hulstyn MJ, Fleming BC, Tung GA. The meniscal roots: gross anatomic correlation with 3-T MRI findings. AJR Am J Roentgenol. 2007. 188:W446–W450.
crossref
16. Ha TP, Li KC, Beaulieu CF, Bergman G, Ch'en IY, Eller DJ, Cheung LP, Herfkens RJ. Anterior cruciate ligament injury: fast spin-echo MR imaging with arthroscopic correlation in 217 examinations. AJR Am J Roentgenol. 1998. 170:1215–1219.
crossref
17. Vellet AD, Lee DH, Munk PL, Hewett L, Eliasziw M, Dunlavy S, Vidito L, Fowler PJ, Miniaci A, Amendola A. Anterior cruciate ligament tear: prospective evaluation of diagnostic accuracy of middle- and high-field-strength MR imaging at 1.5 and 0.5 T. Radiology. 1995. 197:826–830.
crossref
18. Barry KP, Mesgarzadeh M, Triolo J, Moyer R, Tehranzadeh J, Bonakdarpour A. Accuracy of MRI patterns in evaluating anterior cruciate ligament tears. Skeletal Radiol. 1996. 25:365–370.
crossref
19. Umans H, Wimpfheimer O, Haramati N, Applbaum YH, Adler M, Bosco J. Diagnosis of partial tears of the anterior cruciate ligament of the knee: value of MR imaging. AJR Am J Roentgenol. 1995. 165:893–897.
crossref
20. Servant CT, Ramos JP, Thomas NP. The accuracy of magnetic resonance imaging in diagnosing chronic posterior cruciate ligament injury. Knee. 2004. 11:265–270.
crossref
21. Vahey TN, Broome DR, Kayes KJ, Shelbourne KD. Acute and chronic tears of the anterior cruciate ligament: differential features at MR imaging. Radiology. 1991. 181:251–253.
crossref
22. Gross ML, Grover JS, Bassett LW, Seeger LL, Finerman GA. Magnetic resonance imaging of the posterior cruciate ligament: clinical use to improve diagnostic accuracy. Am J Sports Med. 1992. 20:732–737.
crossref
23. Suh JS, Cho JH, Shin KH, Kim SJ. Chondromalacia of the knee: evaluation with a fat-suppression three-dimensional SPGR imaging after intravenous contrast injection. J Magn Reson Imaging. 1996. 6:884–888.
crossref
24. Gold GE, Han E, Stainsby J, Wright G, Brittain J, Beaulieu C. Musculoskeletal MRI at 3.0 T: relaxation times and image contrast. AJR Am J Roentgenol. 2004. 183:343–351.
crossref
25. Schröder RJ, Fischbach F, Unterhauser FN, Weiler A, Felix R, Bruhn H. Value of various MR sequences using 1.5 and 3.0 Tesla in analyzing cartilaginous defects of the patella in an animal model. Rofo. 2004. 176:1667–1675.
26. Kijowski R, Blankenbaker DG, Davis KW, Shinki K, Kaplan LD, De Smet AA. Comparison of 1.5-and 3.0-T MR imaging for evaluating the articular cartilage of the knee joint. Radiology. 2009. 250:839–848.
crossref
27. Tardieu M, Lazennec JY, Christel P, Brasseur JL, Roger B, Grenier P. Normal and pathological MRI aspects of the posterolateral corner of the knee. J Radiol. 1995. 76:605–609.
28. Sims WF, Jacobson KE. The posteromedial corner of the knee: medial-sided injury patterns revisited. Am J Sports Med. 2004. 32:337–345.
29. Newberg AH, Newman JS. Imaging the painful hip. Clin Orthop Relat Res. 2003. (406):19–28.
crossref
30. Mintz DN, Hooper T, Connell D, Buly R, Padgett DE, Potter HG. Magnetic resonance imaging of the hip: detection of labral and chondral abnormalities using noncontrast imaging. Arthroscopy. 2005. 21:385–393.
crossref
31. Hodler J, Trudell D, Pathria MN, Resnick D. Width of the articular cartilage of the hip: quantification by using fatsuppression spin-echo MR imaging in cadavers. AJR Am J Roentgenol. 1992. 159:351–355.
crossref
32. Rubin SJ, Totterman SM, Meyers SP, Hartley DF. Magnetic resonance imaging of the hip with a pelvic phased-array surface coil: a technical note. Skeletal Radiol. 1998. 27:77–82.
crossref
33. Keeney JA, Peelle MW, Jackson J, Rubin D, Maloney WJ, Clohisy JC. Magnetic resonance arthrography versus arthroscopy in the evaluation of articular hip pathology. Clin Orthop Relat Res. 2004. (429):163–169.
crossref
34. Leunig M, Podeszwa D, Beck M, Werlen S, Ganz R. Magnetic resonance arthrography of labral disorders in hips with dysplasia and impingement. Clin Orthop Relat Res. 2004. (418):74–80.
crossref
35. Magee T, Williams D. 3.0-T MRI of the supraspinatus tendon. AJR Am J Roentgenol. 2006. 187:881–886.
crossref
36. Magee T. 3-T MRI of the shoulder: is MR arthrography necessary? AJR Am J Roentgenol. 2009. 192:86–92.
crossref
37. Palmer WE, Brown JH, Rosenthal DI. Rotator cuff: evaluation with fat-suppressed MR arthrography. Radiology. 1993. 188:683–687.
crossref
38. Bencardino JT, Beltran J, Rosenberg ZS, Rokito A, Schmahmann S, Mota J, Mellado JM, Zuckerman J, Cuomo F, Rose D. Superior labrum anterior-posterior lesions: diagnosis with MR arthrography of the shoulder. Radiology. 2000. 214:267–271.
crossref
39. Jee WH, McCauley TR, Katz LD, Matheny JM, Ruwe PA, Daigneault JP. Superior labral anterior posterior (SLAP) lesions of the glenoid labrum: reliability and accuracy of MR arthrography for diagnosis. Radiology. 2001. 218:127–132.
crossref
40. Magee T, Shapiro M, Williams D. Comparison of high-field-strength versus low-field-strength MRI of the shoulder. AJR Am J Roentgenol. 2003. 181:1211–1215.
crossref
41. Timmerman LA, Schwartz ML, Andrews JR. Preoperative evaluation of the ulnar collateral ligament by magnetic resonance imaging and computed tomography arthrography: evaluation in 25 baseball players with surgical confirmation. Am J Sports Med. 1994. 22:26–31.
crossref
42. Carrino JA, Morrison WB, Zou KH, Steffen RT, Snearly WN, Murray PM. Noncontrast MR imaging and MR arthrography of the ulnar collateral ligament of the elbow: prospective evaluation of two-dimensional pulse sequences for detection of complete tears. Skeletal Radiol. 2001. 30:625–632.
crossref
43. Magee T. Comparison of 3-T MRI and arthroscopy of intrinsic wrist ligament and TFCC tears. AJR Am J Roentgenol. 2009. 192:80–85.
crossref
44. Saupe N, Prüssmann KP, Luechinger R, Bösiger P, Marincek B, Weishaupt D. MR imaging of the wrist: comparison between 1.5-and 3-T MR imaging-preliminary experience. Radiology. 2005. 234:256–264.
crossref
45. Potter HG, Asnis-Ernberg L, Weiland AJ, Hotchkiss RN, Peterson MG, McCormack RR Jr. The utility of high-resolution magnetic resonance imaging in the evaluation of the triangular fibrocartilage complex of the wrist. J Bone Joint Surg Am. 1997. 79:1675–1684.
crossref
46. Yoshioka H, Ueno T, Tanaka T, Shindo M, Itai Y. High-resolution MR imaging of triangular fibrocartilage complex (TFCC): comparison of microscopy coils and a conventional small surface coil. Skeletal Radiol. 2003. 32:575–581.
crossref
47. Lenk S, Ludescher B, Martirosan P, Schick F, Claussen CD, Schlemmer HP. 3.0 T high-resolution MR imaging of carpal ligaments and TFCC. Rofo. 2004. 176:664–667.
crossref
48. Schmitt R, Christopoulos G, Meier R, Coblenz G, Fröhner S, Lanz U, Krimmer H. Direct MR arthrography of the wrist in comparison with arthroscopy: a prospective study on 125 patients. Rofo. 2003. 175:911–919.
49. Schweitzer ME, Brahme SK, Hodler J, Hanker GJ, Lynch TP, Flannigan BD, Godzik CA, Resnick D. Chronic wrist pain: spin-echo and short tau inversion recovery MR imaging and conventional and MR arthrography. Radiology. 1992. 182:205–211.
crossref
50. Haims AH, Schweitzer ME, Morrison WB, Deely D, Lange RC, Osterman AL, Bednar JM, Taras JS, Culp RW. Internal derangement of the wrist: indirect MR arthrography versus unenhanced MR imaging. Radiology. 2003. 227:701–707.
crossref
51. Zanetti M, Bräm J, Hodler J. Triangular fibrocartilage and intercarpal ligaments of the wrist: does MR arthrography improve standard MRI? J Magn Reson Imaging. 1997. 7:590–594.
crossref
52. Muhle C, Frank LR, Rand T, Yeh L, Wong EC, Skaf A, Dantas RW, Haghighi P, Trudell D, Resnick D. Collateral ligaments of the ankle: high-resolution MR imaging with a local gradient coil and anatomic correlation in cadavers. Radiographics. 1999. 19:673–683.
crossref
53. Kornaat PR, Reeder SB, Koo S, Brittain JH, Yu H, Andriacchi TP, Gold GE. MR imaging of articular cartilage at 1.5T and 3.0T: comparison of SPGR and SSFP sequences. Osteoarthritis Cartilage. 2005. 13:338–344.
crossref
54. Gold GE, Fuller SE, Hargreaves BA, Stevens KJ, Beaulieu CF. Driven equilibrium magnetic resonance imaging of articular cartilage: initial clinical experience. J Magn Reson Imaging. 2005. 21:476–481.
crossref
55. Schmid MR, Pfirrmann CW, Koch P, Zanetti M, Kuehn B, Hodler J. Imaging of patellar cartilage with a 2D multiple-echo data image combination sequence. AJR Am J Roentgenol. 2005. 184:1744–1748.
crossref
56. Dardzinski BJ, Schmithorst VJ, Holland SK, Boivin GP, Imagawa T, Watanabe S, Lewis JM, Hirsch R. MR imaging of murine arthritis using ultrasmall superparamagnetic iron oxide particles. Magn Reson Imaging. 2001. 19:1209–1216.
crossref
57. Lutz AM, Seemayer C, Corot C, Gay RE, Goepfert K, Michel BA, Marincek B, Gay S, Weishaupt D. Detection of synovial macrophages in an experimental rabbit model of antigen-induced arthritis: ultrasmall superparamagnetic iron oxide-enhanced MR imaging. Radiology. 2004. 233:149–157.
crossref
58. Mosher TJ, Dardzinski BJ. Cartilage MRI T2 relaxation time mapping: overview and applications. Semin Musculoskelet Radiol. 2004. 8:355–368.
crossref
TOOLS
Similar articles