Journal List > J Korean Acad Prosthodont > v.55(2) > 1034916

Lee: Removal torque of a two-piece zirconia abutment with a novel titanium component in an internal connection implant system after dynamic cyclic loading

Abstract

Purpose

The aim of this study was to evaluate screw removal torque of the two-piece zirconia abutment with the novel titanium component compared to the conventional one-piece titanium abutment in the internal connection implant before and after dynamic cyclic loading.

Materials and methods

Two types of the abutment assemblies with internal connection were prepared and divided into the groups; titanium abutment-titanium abutment screw assemblies as control, and zirconia abutment-titanium socket-titanium abutment screw assemblies as experimental group. A total of 12 abutments and implants were used of six assemblies each group. Each assembly was tightened to 30 Ncm. A cyclic load of 300 N at an angle of 30 degrees in reference to the loading axis was applied until one million cycles or failure. The removal torque values (RTVs) of the abutment screws were measured with a digital torque gauge before and after cyclic loading. The RTVs of the pre load and post load were analyzed with t-test, and P-values < .05 were considered statistically significant.

Results

The assemblies of both groups survived all after the dynamic cyclic loading test without screw loosening. The statistically significant differences were found between the mean RTVs before and after the cyclic loading in both groups (P < .05). The RTV differences for the control and the experimental group were -7.25 ± 1.50 Ncm and -7.33 ± 0.93 Ncm, respectively. Statistical analysis revealed that the RTV differences in both groups were not significantly different from each other (P > .05).

Conclusion

Within the limitation of this study, the two-piece zirconia abutment with the titanium component did not show a significant RTV difference of the abutment screw compared to the titanium abutment after dynamic cyclic loading. (J Korean Acad Prosthodont 2017;55:151-5)

REFERENCES

1.Prestipino V., Ingber A. Esthetic high-strength implant abutments. Part I. J Esthet Dent. 1993. 5:29–36.
2.Manicone PF., Rossi Iommetti P., Raffaelli L. An overview of zirconia ceramics: basic properties and clinical applications. J Dent. 2007. 35:819–26.
crossref
3.Yildirim M., Fischer H., Marx R., Edelhoff D. In vivo fracture resistance of implant-supported all-ceramic restorations. J Prosthet Dent. 2003. 90:325–31.
crossref
4.Apholt W., Bindl A., Lüthy H., Mörmann WH. Flexural strength of Cerec 2 machined and jointed InCeram-Alumina and InCeram-Zirconia bars. Dent Mater. 2001. 17:260–7.
crossref
5.Adatia ND., Bayne SC., Cooper LF., Thompson JY. Fracture resistance of yttria-stabilized zirconia dental implant abutments. J Prosthodont. 2009. 18:17–22.
crossref
6.Sailer I., Sailer T., Stawarczyk B., Jung RE., Hämmerle CH. In vitro study of the influence of the type of connection on the fracture load of zirconia abutments with internal and external implant-abutment connections. Int J Oral Maxillofac Implants. 2009. 24:850–8.
7.Gehrke P., Alius J., Fischer C., Erdelt KJ., Beuer F. Retentive strength of two-piece CAD/CAM zirconia implant abutments. Clin Implant Dent Relat Res. 2014. 16:920–5.
crossref
8.Chun HJ., Yeo IS., Lee JH., Kim SK., Heo SJ., Koak JY., Han JS., Lee SJ. Fracture strength study of internally connected zirconia abutments reinforced with titanium inserts. Int J Oral Maxillofac Implants. 2015. 30:346–50.
crossref
9.Theoharidou A., Petridis HP., Tzannas K., Garefis P. Abutment screw loosening in single-implant restorations: a systematic review. Int J Oral Maxillofac Implants. 2008. 23:681–90.
10.Taylor TD. Prosthodontic problems and limitations associated with osseointegration. J Prosthet Dent. 1998. 79:74–8.
crossref
11.Stüker RA., Teixeira ER., Beck JC., da Costa NP. Preload and torque removal evaluation of three different abutment screws for single standing implant restorations. J Appl Oral Sci. 2008. 16:55–8.
crossref
12.Dhingra A., Weiner S., Luke AC., Ricci JL. Analysis of dimensional changes in the screw and the surface topography at the interface of a titanium screw and a zirconia abutment under cyclic loading: an in vitro study. Int J Oral Maxillofac Implants. 2013. 28:661–9.
crossref
13.Siamos G., Winkler S., Boberick KG. Relationship between implant preload and screw loosening on implant-supported prostheses. J Oral Implantol. 2002. 28:67–73.
14.Tsuge T., Hagiwara Y. Influence of lateral-oblique cyclic loading on abutment screw loosening of internal and external hexagon implants. Dent Mater J. 2009. 28:373–81.
crossref
15.Craig RG. Restorative dental materials,. 9th ed.Mosby, St. Louis;1993. p. 75–7.
16.Att W., Kurun S., Gerds T., Strub JR. Fracture resistance of single-tooth implant-supported all-ceramic restorations: an in vitro study. J Prosthet Dent. 2006. 95:111–6.
crossref
17.Dittmer MP., Dittmer S., Borchers L., Kohorst P., Stiesch M. Influence of the interface design on the yield force of the implant-abutment complex before and after cyclic mechanical loading. J Prosthodont Res. 2012. 56:19–24.
crossref
18.Khraisat A., Abu-Hammad O., Dar-Odeh N., Al-Kayed AM. Abutment screw loosening and bending resistance of external hexagon implant system after lateral cyclic loading. Clin Implant Dent Relat Res. 2004. 6:157–64.
crossref
19.Khraisat A., Hashimoto A., Nomura S., Miyakawa O. Effect of lateral cyclic loading on abutment screw loosening of an external hexagon implant system. J Prosthet Dent. 2004. 91:326–34.
crossref
20.Gracis S., Michalakis K., Vigolo P., Vult von Steyern P., Zwahlen M., Sailer I. Internal vs. external connections for abutments/re-constructions: a systematic review. Clin Oral Implants Res. 2012. 23:202–16.
crossref
21.Delben JA., Barão VA., Ferreira MB., da Silva NR., Thompson VP., Assunção WG. Influence of abutment-to-fixture design on reliability and failure mode of all-ceramic crown systems. Dent Mater. 2014. 30:408–16.
crossref
22.Kim JS., Raigrodski AJ., Flinn BD., Rubenstein JE., Chung KH., Mancl LA. In vitro assessment of three types of zirconia implant abutments under static load. J Prosthet Dent. 2013. 109:255–63.
crossref
23.Dixon DL., Breeding LC., Sadler JP., McKay ML. Comparison of screw loosening, rotation, and deflection among three implant designs. J Prosthet Dent. 1995. 74:270–8.
crossref

Fig. 1.
The implant/abutment assemblies used in the study, (A) internal connection type titanium implant, (B) control group: internally connected titanium abutment with a titanium screw, (C) experimental group: externally connected zirconia abutment with a titanium socket and a titanium screw.
jkap-55-151f1.tif
Table 1.
Pre- and post-cyclic load reverse torque values (RTVs), and RTV differences
Group Pre-cyclic load RTV (Ncm) Post-cyclic load RTV (Ncm) RTV difference (Ncm)
Control 25.92 ± 0.58 18.67 ± 1.13 -7.25 ± 1.50
Experimental 25.08 ± 0.74 17.75 ± 0.88 -7.33 ± 0.93