Kim, Lee, Yu, Lee, Moon, and Sohn: Comparative study of removal torque of 3 different hydroxyapatite coated implants in the femur of rabbits

Abstract

Introduction

This study compared the strength of osseointegration as determined by the resistance to reverse torque rotation of three different hydroxyapatite coated implants in the rabbit femur model.

Materials and Methods

Three hydroxyapatite coated implants (HAPTITE), Tapered Screw-Vent (TSV) and BioTite-H – were used. A total of 40 implants were placed in the femur of 20 adult male rabbits. The animals were divided into two groups. In group A (n=10); one HAPTITE was placed into each right femur and one TSV was placed into each left femur. In group B (n=10); one HAPTITE was placed into each right femur and one BioTite-H was placed into each left femur. Five rabbits of each group were sacrificed at 4 and 8 weeks. The implants were removed by reverse torque rotation using a digital torque-measuring device. A total of 40 implants in 20 rabbits were used for the removal torque measurements.

Results

In the Group A, 4 weeks after implant placement, the mean removal torque for the HAPTITE and TSV was 70.7±31.6 N cm and 28.9± 15.1 N cm, respectively. Eight weeks after implant placement, the mean removal torque for the HAPTITE and TSV was 87.9±26.2 N cm and 54.9± 22.4 N cm, respectively. In the Group B, 4 weeks after implant placement, the mean removal torque for the HAPTITE and BioTite-H was 58.0±29.6 N cm and 37.7±14.1 N cm, respectively. Eight weeks after implant placement, the mean removal torque for the HAPTITE and BioTite-H was 91.4± 47.1 N cm and 30.8±9.8 N cm. HAPTITE showed a higher removal torque than the other implants.

Conclusion

These results suggest that HAPTITE increases the strength of osseointegration significantly as determined by the resistance to reverse torque rotation.

References

1. Bra�nemark PI, Adell R, Breine U, Hansson BO, Lindstro¨m J, Ohlsson A. Intraosseous anchorage of dental prostheses. I. Experimental studies. Scand J Plast Reconstr Surg. 1969; 3:81–100.
2. Hutton JE, Heath MR, Chai JY, Harnett J, Jemt T, Johns RB, et al. Factors related to success and failure rates at 3-year follow-up in a multicenter study of overdentures supported by Branemark implants. Int J Oral Maxillofac Implants. 1995; 10:33–42.
3. Albrektsson T, Zarb G, Worthington P, Eriksson AR. The longterm efficacy of currently used dental implants: a review and proposed criteria of success. Int J Oral Maxillofac Implants. 1986; 1:11–25.
4. Shirakura M, Fujii N, Ohnishi H, Taguchi Y, Ohshima H, Nomura S, et al. Tissue response to titanium implantation in the rat maxilla, with special reference to the effects of surface conditions on bone formation. Clin Oral Implants Res. 2003; 14:687–96.
crossref
5. Franchi M, Fini M, Martini D, Orsini E, Leonardi L, Ruggeri A, et al. Biological fixation of endosseous implants. Micron. 2005; 36:665–71.
crossref
6. Predecki P, Auslaender BA, Stephan JE, Mooney VL, Stanitski C. Attachment of bone to threaded implants by ingrowth and mechanical interlocking. J Biomed Mater Res. 1972; 6:401–12.
7. Cook SD, Kay JF, Thomas KA, Jarcho M. Interface mechanics and histology of titanium and hydroxylapatite-coated titanium for dental implant applications. Int J Oral Maxillofac Implants. 1987; 2:15–22.
8. Vercaigne S, Wolke JG, Naert I, Jansen JA. Histomorphometrical and mechanical evaluation of titanium plasma-spray-coated implants placed in the cortical bone of goats. J Biomed Mater Res. 1998; 41:41–8.
crossref
9. Buser D, Schenk RK, Steinemann S, Fiorellini JP, Fox CH, Stich H. Influence of surface characteristics on bone integration of titanium implants. A histomorphometric study in miniature pigs. J Biomed Mater Res. 1991; 25:889–902.
crossref
10. Cochran DL, Schenk RK, Lussi A, Higginbottom FL, Buser D. Bone response to unloaded and loaded titanium implants with a sandblasted and acid-etched surface: a histometric study in the canine mandible. J Biomed Mater Res. 1998; 40:1–11.
crossref
11. S � balle K, Hansen ES, Brockstedt-Rasmussen H, Pedersen CM, Bu¨nger C. Hydroxyapatite coating enhances fixation of porous coated implants. A comparison in dogs between press fit and noninterference fit. Acta Orthop Scand. 1990; 61:299–306.
12. Block MS, Kent JN, Kay JF. Evaluation of hydroxylapatite-coated titanium dental implants in dogs. J Oral Maxillofac Surg. 1987; 45:601–7.
crossref
13. Wennerberg A, Albrektsson T, Andersson B, Krol JJ. A histomorphometric and removal torque study of screw-shaped titanium implants with three different surface topographies. Clin Oral Implants Res. 1995; 6:24–30.
14. Kumar M, Dasarathy H, Riley C. Electrodeposition of brushite coatings and their transformation to hydroxyapatite in aqueous solutions. J Biomed Mater Res. 1999; 45:302–10.
crossref
15. Baker D, London RM, O'Neal R. Rate of pull-out strength gain of dual-etched titanium implants: a comparative study in rabbits. Int J Oral Maxillofac Implants. 1999; 14:722–8.
16. Masuda T, Yliheikkila¨ PK, Felton DA, Cooper LF. Generalizations regarding the process and phenomenon of osseointegration. Part I. In vivo studies. Int J Oral Maxillofac Implants. 1998; 13:17–29.
17. Cooper LF, Masuda T, Yliheikkila¨ PK, Felton DA. Generalizations regarding the process and phenomenon of osseointegration. Part II. In vitro studies. Int J Oral Maxillofac Implants. 1998; 13:163–74.
18. Rich A, Harris AK. Anomalous preferences of cultured macrophages for hydrophobic and roughened substrata. J Cell Sci. 1981; 50:1–7.
crossref
19. Johansson C, Albrektsson T. Integration of screw implants in the rabbit: a 1-year follow-up of removal torque of titanium implants. Int J Oral Maxillofac Implants. 1987; 2:69–75.
20. Sakakura CE, Margonar R, Holzhausen M, Nociti FH Jr, Alba RC Jr, Marcantonio E Jr. Influence of cyclosporin A therapy on bone healing around titanium implants: a histometric and biome-chanic study in rabbits. J Periodontol. 2003; 74:976–81.
crossref
21. Sennerby L, Thomsen P, Ericson LE. A morphometric and bio-mechanic comparison of titanium implants inserted in rabbit cortical and cancellous bone. Int J Oral Maxillofac Implants. 1992; 7:62–71.
22. Ivanoff CJ, Widmark G, Johansson C, Wennerberg A. Histologic evaluation of bone response to oxidized and turned titanium micro-implants in human jawbone. Int J Oral Maxillofac Implants. 2003; 18:341–8.
23. Roberts WE, Garreto LP. Bone physiology and metabolism. Misch CE, editor. Contemporary implant dentistry. 2nd ed.St. Louis: Mosby Co.;1999. p. 234–5.
24. Misch CM. Hydroxylapatite-coated implants. Design considerations and clinical parameters. N Y State Dent J. 1993; 59:36–41. Erratum in: N Y State Dent J 1993;59: 12.
25. Davies JE, Lowenberg B, Shiga A. The bone-titanium interface in vitro. J Biomed Mater Res. 1990; 24:1289–306.
26. Yang Y, Bumgardner JD, Cavin R, Carnes DL, Ong JL. Osteoblast precursor cell attachment on heat-treated calcium phosphate coatings. J Dent Res. 2003; 82:449–53.
crossref
27. Cle`ries L, Ferna′ndez-Pradas JM, Sardin G, Morenza JL. Dissolution behaviour of calcium phosphate coatings obtained by laser ablation. Biomaterials. 1998; 19:1483–7.
crossref

Fig. 1.
Digital torque meter used to measure the peak value of resistance to reverse torque rotation in N cm.
jkaoms-37-49f1.tif
Fig. 2.
Mean removal torque values (N cm) of implants 4 weeks after implantation.(TSV: Tapered Screw-Vent)
jkaoms-37-49f2.tif
Fig. 3.
Mean removal torque values (N cm) of implants 8 weeks after implantation.(TSV: Tapered Screw-Vent)
jkaoms-37-49f3.tif
Table 1.
Experimental design
  Right femur (n=20) Left femur (n=20) N Number of Rabbits
Group A HAPTITE (n=10) TSV (n=10) 10
Group B HAPTITE (n=10) BioTite-H (n=10) 10

(TSV: Tapered Screw-Vent)

Table 2.
Removal torque values in Newton's
Group Weeks Implant Removal torque (N cm) Mean±SD
A 4 HAPTITE 121.0 63.5 40.5 50.0 78.5 70.7±31.6
    TSV 50.0 8.5 34.0 25.0 27.0 28.9±15.1
  8 HAPTITE 55.0 114.5 66.5 109.0 94.5 87.9±26.2
    TSV 36.5 72.5 46.5 84.5 34.5 54.9±22.4
B 4 HAPTITE 47.5 108.5 32.0 56.5 45.5 58.0±29.6
    BioTite-H 39.5 53.0 14.5 42.0 39.5 37.7±14.1
  8 HAPTITE 36.5 86.5 153.0 122.5 58.5 91.4±47.1
    BioTite-H 29.5 44.5 33.0 17.0 30.0 30.8±9.8

(SD: standard deviation, TSV: Tapered Screw-Vent)

TOOLS
Similar articles