Abstract
We evaluated one of the new non-contact tonometers, the Pulsair non-contact tonometer, to assess its accuracy and reliability. Measurements on 101 eyes were used to assess its accuracy against Goldmann tonometry measurements, and another 24 eyes were used to assess its reliability (reproducibility). The results showed the instrument to be highly accurate and reliable in measurement.