Abstract
Background and Objectives
Patients on oral anticoagulation therapy (OAT) require regular monitoring of the prothrombin time (PT) and dosage adjustment to prevent thromboembolic diseases without the risk of hemorrhage. Portable self PT monitors have been recently developed because the standard PT measurements are complicated and take considerable time. This study compared the International normalized ratio (INR) results that were obtained with using the CoaguChek XS device (Roche Diagnostic, Mannheim, Germany) with those obtained using a standard laboratory method Sysmex CA-1500 (Sysmex Corporation, Kobe, Japan) in the patients on OAT and also a healthy control group.
Subjects and Methods
100 outpatients on OAT and 20 healthy controls were enrolled on a volunteer basis after providing informed consent at the Dong-A University Hospital. The outpatients and the healthy control group provided us the INR measurements with using both the CoaguChek XS and the Sysmex CA-1500.
Results
The coefficients of variation for CoaguChek XS and Sysmex CA-1500 were less than 10%. The PT (INR) results of CoaguChek XS and Sysmex CA-1500 were 2.0±0.7 and 2.2±0.7, respectively (p<0.001). There was a good correlation between CoaguChek XS and Sysmex CA-1500 (r=0.974, p<0.001). On the regression analysis, the slope of the regression line was 0.9197 and the y-intercept was 0.0058. On the Bland-Altman analysis, the INR mean difference (bias) between the two methods (CoaguChek XS INR-Sysmex CA-1500 INR) was -0.2 and the limit of agreement was +0.168~-0.568.
Figures and Tables
References
1. Hylek EM, Regan S, Go AS, Hughes RA, Singer DE, Skates SJ. Clinical predictors of prolonged delay in return of the international normalized ratio to within the therapeutic range after excessive anticoagulation with warfarin therapy. Ann Intern Med. 2001. 135:393–400.
2. Jeong MH. The anticoagulant and antiplatelet therapy in cardiovascular Patients. Korean Circ J. 1999. 29:96–103.
3. Hart RG, Halperin JL. Atrial fibrillation and thromboembolism: a decade of progress in stroke prevention. Ann Intern Med. 1999. 131:688–695.
4. Njaastad AM, Abildgaard U, Lassen JF. Gains and losses of warfarin therapy as performed in an anticoagulation clinic. J Intern Med. 2006. 259:296–304.
5. Lee SM, Choue CW, Hong SJ, et al. The joint multicenter study on the atrial fibrillation in Korea. Korean Circ J. 2000. 30:646–652.
6. Choue JH. More aggressive drug therapy for the management of atrial fibrillation. Korean Circ J. 2002. 32:199–204.
7. Bland JM, Altman DG. Stratistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986. 1:307–310.
8. Khoschnewis S, Hannes FM, Tschopp M, Wuillemin WA. INR comparison between the CoaguChek Pro PTN and a standard laboratory method. Thromb Res. 2004. 113:327–332.
9. Atterman J, Andersen NT, Korsgaard H, Maegaard M, Hasenkam JM. Precision of INR measured with a Patient operated whole blood coagutometer. Thromb Res. 2003. 110:65–68.
10. Havrda DE, Hawk TL, Marvin CM. Accuracy and precison of the CoaguChek S versus laboratory INRs in a clinic. Ann Pharmacother. 2002. 36:769–775.
11. Cosmi B, Palareti G, Moia M, et al. Accuracy of a portable prothrombin time monitor(Coagucheck) in patients on chronic oral anticoagulant therapy: a prospective multicenter study. Thromb Res. 2000. 100:279–286.
12. Sirithunyanont C, Bhuripanyo K, Kangkagate C, Winyart W, Sirchaya P, Wangtip K. Accurarcy of international normalized ratio determined by protable venous-blood coagulation monitor versus a central laboratory. J Med Assoc Thai. 2003. 86:S67–S75.
13. Bereznicki LR, Jackson SL, Peterson GM, Jeffrey EC, Marsden KA, Jupe DM. Accuracy and clinical utility of the CoaguChek XS portable international normalized ratio monitor in a pilot study of warfarin home-monitoring. J Clin Pathol. 2007. 60:311–314.
14. Vacas M, Fernandez MA, Martinez-Brotons F, et al. Comparative study of a portable prothrombin time monitor employing three different systems in oral anticoagulant units. Haemostasis. 2001. 31:18–25.
15. Douketis JD, Lane A, Milne J, Ginsberg JS. Accuracy of a portable International Normalization Ratio monitor in outpatients receiving long-term oral anticoagulant therapy: comparison with a laboratory reference standard using clinically relevant criteria for agreement. Thromb Res. 1998. 92:11–17.
16. Hentrich DP, Fritschi J, Muller PR, Wuillemin WA. INR comparison between the CoaguChekR S and a standard laboratory method among patients with self-management of oral anticoagulation. Thromb Res. 2007. 119:489–495.
17. Cromheecke ME, Levi M, Colly LP, et al. Oral anticoagulation self-management and management by a specialst anticoagulation clinic: a randomised cross-over comparison. Lancet. 2000. 356:97–102.
18. Sawicki PT. A structured and self-management program for patients receiving oral anticoagulation: a randomized controlled trial. JAMA. 1999. 281:145–150.
19. Watzke HH, Forberg E, Svolba G, Jimenez-Boj E, Krinninger B. A prospective controlled trial comparing weekly self-testing and self-dosing with the standard management of patients on stable oral anticoagulation. Thromb Haemost. 2000. 83:661–665.
20. Heneghan C, Alonso-Coello P, Garcia-Alamino JM, Perera R, Meats E, Glasziou P. Self-monitoring of oral anticoagulation: a systematic review and meta-analysis. Lancet. 2006. 367:404–411.