Abstract

There are few reports of the failure mode of a dynamic hip screw with a trochanteric stabilizing plate (TSP) in the treatment of unstable proximal femur fractures. We present 3 cases of fixation failure of a TSP used to treat unstable intertrochanteric fractures. All cases had the same type of failure mode, which represented a fatigue fracture of the distal screw at the junction of a plate and screw. Non-union occurred in 2 cases, and one case was an incidental finding 4 years after union. Biomechanical analysis and SEM (scanning electron micrograph) examination was also performed. Adding TSP to the plate might increase the level of shear stress and slip motion in the distal screw followed by a fatigue fracture. SEM of the surface of the broken screw revealed a multiple cycle and low stress fatigue fracture. Recommendations are made regarding the implant design and surgical technique.

Keywords: Unstable intertrochanteric fracture; Trochanteric stabilizing plate; Metal failure

TOOLS
Similar articles

Effect of a Trochanter-stabilizing Plate in Unstable IntertrochantericFractures : A Clinical and Biomechanical Study

The Treatment of Unstable Intertrochanteric Fractures with Dynamic Hip Screw and Trochanter Stabilizing Plate

Treatment of Comminuted Trochanteric Fractures with Dynamic Hip Screw and DHS Trochanter Stabilizing Plate

Factors affecting Fixation Failure in Treatment of Intertrochanteric Fracture of the Femur by Dynamic Compression Hip Screw

Treatment of Unstable Intertrochanteric Fracture Using Short Barrel Compression Hip Screws and a Trochanteric Stabilizing Plate