Abstract
Purpose
To assess the mode of a polyethylene liner failure in total hip replacement using a Harris-Galante II acetabular component.
Materials and Methods
A retrospective review was performed and 128 consecutive primary total hip arthroplasties were analyzed in 111 patients who had a second-generation Harris-Galante cup. All patients were followed up for a minimum of five years, with a mean follow up of 7.4 years (range, 5 to 12.9 years). A total of 26 cups (20.2%) were revised at the time of the latest follow up. The medical records, radiographs, surgical notes, and retrieved components were analyzed in order to determine the precise causes of the revision.
Results
Twenty-two cases (17.3%) were found to have experienced problems with the polyethylene liner. Three types of liner failure were identified; metallosis associated with polyethylene liner wear (14 cases), excessive wear without metallosis necessitating a revision (6 cases), and polyethylene liner dissociation without gross evidence of wear (2 cases).