Abstract
Purpose
Modular acetabular components have the potential of fretting and back side wear leading to osteolysis. The present study evaluated the outcomes of total hip arthroplasties (THAS) using a nonmodular cementless acetabular component with those of press-fit titanium mesh coatings fixed.
Materials and Methods
Eighty-two primary THAS in 75 patients were performed consecutively between April 1995 and August 1998. Sixty-two hips were eligible for follow up for an average 62 months. Mean patients age was 59 years. The clinical evaluation was performed using the modified Harris hip score (HHS), inguinal pain and limping. The radiologic evaluation included, component stability, polyethylene wear rate, pelvic osteolysis and the remodeling pattern around the acetabular component.
Results
Mean HHS improved from 50.2 points preoperatively to 90.5 points at the most recent follow up. All components were stable radiologically. The linear polyethylene wear rate was 0.07 mm/year. No pelvic osteolysis was found. Polar gaps found in 25 hips (40%) in postoperative radiographs and disappeared during follow up in 22 hips (88%). One hip was revised due to chronic unreduced dislocation.