Abstract
Purpose
To analyze the effects of various amount of peripheral defect of the acetabulum in terms of the stability of the acetabular cup with screw fixation.
Materials and Methods
The porous coated acetabular cup of Trilogy was press-fitted with 3 screw fixations after 2 mm undersized reaming into the distal part of 100 bovine femurs, but with insufficient peripheral coverage of 0%, 5%, 10%, 15% and 20% respectively. The micromotions of the acetabular cups were measured under the axial loads of 50, 100, 150, 200, 250, and 300 Kp, and torques were measured at 1° to 5° of cup rotation under applied rotational force in 10 specimens in each of five groups.