Park and Kim: Evaluation of iQ200 Automated Urine Microscopy Analyzer

Abstract

Background

Microscopic examination of urine sediment is one of the most commonly performed tests in the clinical laboratory. However, manual microscopic sediment examination is labor-intensive, time-consuming and imprecise. In this study, we evaluated the analytical performance and clinical usefulness of a recently introduced image-based automated urinalysis system, Iris iQ200 (Iris Diagnostics, USA).

Methods

We assessed the iQ200 for linearity, precision and carryover rate using patient's samples and quality control materials. On 337 urine samples, urine sediment analyses performed by the iQ200 were compared with manual microscopy results.

Results

The iQ200 showed a good linearity (r2>0.99) for all cellular components analyzed. Within-run and total CVs on urine specimens and quality control samples were less than 10% except for within-run CV for the samples with low concentration of the squamous epithelial cells. The carryover rates were 0.21% for RBCs and 1.92% for WBCs. The agreement rates within one grade between the iQ200 and manual microscopy for RBCs, WBCs, and squamous epithelial cells were 93.8%, 94.2% and 96.9%, respectively.

Conclusions

Since the iQ200 showed a reliable analytical performance and good concordance with manual microscopy, it could be useful in the clinical practice as a screening procedure.

REFERENCES

1.Fuller CE., Threatte GA., Henry JB. Basic examination of urine. Henry JB, editor. Clinical diagnosis and management by laboratory methods. 20th ed.Philadelphia: WB Saunders;;2001. p. 367–402.
2.Ferris JA. Comparison and standardization of the urine microscopic examination. Lab Med. 1983. 14:659–62.
crossref
3.Huussen J., Koene RA., Hilbrands LB. The (fixed) urinary sediment, a simple and useful diagnostic tool in patients with haematuria. Neth J Med. 2004. 62:4–9.
4.Elin RJ., Hosseini JM., Kestner J., Rawe M., Ruddel M., Nishi HH. Comparison of automated and manual methods for urinalysis. Am J Clin Pathol. 1986. 86:731–7.
crossref
5.Carlson DA., Statland BE. Automated urinalysis. Clin Lab Med. 1988. 8:449–61.
crossref
6.Wah DT., Wises PK., Butch AW. Analytic performance of the iQ200 automated urine microscopy analyzer and comparison with manual counts using Fuchs-Rosenthal cell chambers. Am J Clin Pathol. 2005. 123:290–6.
crossref
7.Lamchiagdhase P., Preechaborisutkul K., Lomsomboon P., Srisuchart P., Tantiniti P., Khan-u-Ra N, et al. Urine sediment examination: a comparison between the manual method and the iQ200 automated urine microscopy analyzer. Clin Chim Acta. 2005. 358:167–74.
crossref
8.Chien TI., Kao JT., Liu HL., Lin PC., Hong JS., Hsieh HP, et al. Urine sediment examination: a comparison of automated urinalysis systems and manual microscopy. Clin Chim Acta. 2007. 384:28–34.
crossref
9.Shayanfar N., Tobler U., von Eckardstein A., Bestmann L. Automated urinalysis: first experiences and a comparison between the Iris iQ200 urine microscopy system, the Sysmex UF-100 flow cytometer and manual microscopic particle counting. Clin Chem Lab Med. 2007. 45:1251–6.
crossref
10.Okada H., Sakai Y., Kawabata G., Fujisawa M., Arakawa S., Hamaguchi Y, et al. Automated urinalysis. Evaluation of the Sysmex UF-50. Am J Clin Pathol. 2001. 115:605–10.
11.National Committee for Clinical Laboratory Standards. Evaluation of the linearity of quantitative measurement procedures: a statistical approach; approved guideline. NCCLS document EP6-A. Wayne, PA: National Committee for Clinical Laboratory Standards;2003.
12.National Committee for Clinical Laboratory Standards. Evaluation of the precision performance of clinical chemistry devices; approved guideline. NCCLS document EP5-A. Wayne, PA: National Committee for Clinical Laboratory Standards;1999.
13.Pesce MA. Laboratory automation. Kaplan LA, Pesce AJ, editors. Clinical Chemistry. 4th ed.Baltimore: Mosby;2003. p. 294.
14.Ottiger C., Huber AR. Quantitative urine particle analysis: integrative approach for the optimal combination of automation with UF-100 and microscopic review with KOVA cell chamber. Clin Chem. 2003. 49:617–23.
crossref
15.Roe CE., Carlson DA., Daigneault RW., Statland BE. Evaluation of the Yellow IRIS. An automated method for urinalysis. Am J Clin Pathol. 1986. 86:661–5.
crossref
16.Deindoerfer FH., Gangwer JR., Laird CW., Ringold RR. “The Yellow IRIS” urinalysis workstation–the first commercial application of “automated intelligent microscopy”. Clin Chem. 1985. 31:1491–9.
crossref
17.Wargotz ES., Hyde JE., Karcher DS., Hitlan JP., Wilkinson DS. Urine sediment analysis by the Yellow IRIS automated urinalysis workstation. Am J Clin Pathol. 1987. 88:746–8.
18.Kim DC., Yoo MY., Jo SS., Park J., Yoon Y., Kim JQ. Laboratory evaluation of fully automated urine cell analyzer Sysmex UF-100. J Clin Pathol & Quality Control. 2001. 23:299–306. (김동찬, 유영미, 조성석, 박준완, 윤여민, 김진규. 자동요침사분석기 UF-100의평가. 임상병리와정도관리 2001;23: 299-306.).

Fig. 1.
Linearities of RBCs (A), WBCs (B), squamous epithelial cells (C), and preserved RBCs (D) analyzed by iQ200.
kjlm-28-267f1.tif
Table 1.
Grading system of DKUH for urine microscopy
  Cells/HPF
Grade 1 <1
Grade 2 1-4
Grade 3 5-9
Grade 4 10-29
Grade 5 30-1/2 of field of view
Grade 6 >1/2 of field of view

Abbreviations: DKUH, Dankook university hospital; HPF, high power field.

Table 2.
Precision of iQ200 on pooled urine specimens
Cell type N Mean (counts/μL) SD Within-run CV (%)
RBCs 30 34 2.9 8.3
  30 85 3.9 4.6
  30 240 5.3 2.3
WBCs 30 38 3.2 7.8
  30 90 4.3 4.7
  30 183 6.6 3.7
SECs 30 21 2.3 10.6
  30 72 5.4 7.7

Abbreviation: SECs, squamous epithelial cells.

Table 3.
Precision of iQ200 on quality control samples
Level N Mean (counts/μL) SD CV (%)
Within-run Total
Low 80 50.1 3.7 7.39 8.46
Middle 80 499.5 9.3 1.87 2.04
High 80 1002.1 8.0 0.80 0.97
Table 4.
Performance of iQ200 in comparison with manual microscopy
Cell type N. tested N (%)
HI LI Agreement
RBCs 325 13 (4.0) 7 (2.2) 305 (93.8)
WBCs 325 17 (5.2) 2 (0.6) 306 (94.2)
SECs 325 3 (0.9) 7 (2.2) 315 (96.9)

Data includes the number of cases within one grading difference.

Abbreviations: HI, Higher iQ200 cases with two or more grades than manual microscopy; LI, Lower iQ200 cases with two or more grades than manual microscopy; SECs, squamous epithelial cells.

Table 5.
Summary of RBC results between iQ200 and manual microscopy
iQ200 grading Manual microscopy grading (DKUH)
1+ 2+ 3+ 4+ 5+ 6+ Sum
1+ 88           88
2+ 31 21 19 2     73
3+ 4 12 20 17 1   54
4+ 1 3 3 20 4 4 35
5+   1 1 12 6 11 31
6+       3 4 37 44
Sum 124 37 43 54 15 52 325

Abbreviation: DKUH, Dankook university hospital.

Table 6.
Summary of WBC results between iQ200 and manual microscopy
iQ200 grading Manual microscopy grading (DKUH)
1+ 2+ 3+ 4+ 5+ 6+ Sum
1+ 61 1         62
2+ 65 29 2       96
3+ 5 60 28 15 1   109
4+ 1 4 4 17 3 1 30
5+   2 3 7 4 2 18
6+       2 3 5 10
Sum 132 96 37 41 11 8 325

Abbreviation: DKUH, Dankook university hospital.