Journal List > Korean J Lab Med > v.27(3) > 1011392

Lee and Lee: Evaluation of SD BIOLINE Tetanus Kit

Abstract

Background

Point-of-care-testing (POCT) kits for tetanus toxoid antibody are used in emergency departments to evaluate the immunization status of patients with tetanus. The objective of this study was to evaluate the analytical performance and the utility of SD BIOLINE tetanus kit (Standard Diagnostic Inc., Yongin, Korea), as a POCT.

Methods

A total of 326 peripheral blood specimens (whole blood, 319; serum, 326) from healthy subjects and patients were used. SD BIOLINE tetanus kit was evaluated for precision, accuracy, effect of specimens, operator variance, and the total processing time. The results from SD BIOLINE tetanus kit were compared with those from 2 quantitative ELISA kits.

Results

Compared with ELISA kits, SD BIOLINE tetanus kit revealed a sensitivity of 88–97%, specificity of 87–92%, positive predictive value of 81–89%, negative predictive value of 90–98%, and kappa agreement of 0.78–0.82. SD BIOLINE tetanus kit also showed an excellent precision and a high accuracy. It showed a high concordance rate between whole blood and serum specimens. The total processing time of SD BIOLINE tetanus kit was 30–40 min.

Conclusions

SD BIOLINE tetanus kit showed an excellent analytical performance. With its rapid turnaround time and the ease of handling and interpretation, SD BIOLINE tetanus kit seems appropriate for the evaluation of tetanus immunization status as a POCT device. However, education for operators and standardized guidelines for result interpretation should be emphasized.

References

1. Humeau Y, Doussau F, Grant NJ, Poulain B. How botulinum and tetanus neurotoxins block neurotransmitter release. Biochimie. 2000; 82:427–46.
2. Kim HJ, Kang GH, Kim GW, Kim SC, Kim YS, Kim CH, et al. Korean multicenter study of tetanus antibody titer (KoMUST). J Korean Soc Emerg Med. 2005; 16:667–76.
3. Hsu SS, Groleau G. Tetanus in the emergency department: a current review. J Emerg Med. 2001; 20:357–65.
crossref
4. Coplu N, Esen B, Gozalan A, Miyamura K, Yoshida I, Kurtoglu D, et al. Tetanus antibody assay combining in-house ELISA and particle agglutination test and its serosurvey application in a province in Turkey. Jpn J Infect Dis. 2004; 57:97–102.
5. Caglar K, Karakus R, Aybay C. Determination of tetanus antibodies by a double-antigen enzyme-linked immunosorbent assay in individuals of various age groups. Eur J Clin Microbiol Infect Dis. 2005; 24:523–8.
crossref
6. Hong HA, Ke NT, Nhon TN, Thinh ND, van der Gun JW, Hendriks JT, et al. Validation of the combined toxin-binding inhibition test for determination of neutralizing antibodies against tetanus and diphtheria toxins in a vaccine field study in Viet Nam. Bull World Health Organ. 1996; 74:275–82.
7. Hendriksen CF, vd Gun JW, Nagel J, Kreeftenberg JG. The toxin binding inhibition test as a reliable in vitro alternative to the toxin neutralization test in mice for the estimation of tetanus antitoxin in human sera. J Biol Stand. 1988; 16:287–97.
crossref
8. Wahren E, Bengtsson C, Bjorksten B, Mark A, Granstrom M. Capture EIA for determination of IgE to diphtheria and tetanus toxins in comparison to RIA. Vaccine. 1998; 16:873–6.
crossref
9. Booth JR, Nuttall PA. A rapid automated latex screen for tetanus toxoid antibodies. Vox Sang. 1978; 34:239–40.
crossref
10. Sedgwick AK, Ballow M, Sparks K, Tilton RC. Rapid quantitative microenzyme-linked immunosorbent assay for tetanus antibodies. J Clin Microbiol. 1983; 18:104–9.
crossref
11. Nichols JH. Quality in point-of-care testing. Expert Rev Mol Diagn. 2003; 3:563–72.
crossref
12. Colombet I, Saguez C, Sanson-Le Pors MJ, Coudert B, Chatellier G, Espinoza P. Diagnosis of tetanus immunization status: multicenter assessment of a rapid biological test. Clin Diagn Lab Immunol. 2005; 12:1057–62.
crossref
13. Yoon YS, Kim EC, Lee WH, Chung SW, Yi JH, Park IC, et al. Utility of Tetanus Quick Stick test for selective tetanus prophylaxis. J Korean Soc Emerg Med. 2004; 15:95–101.

Table 1.
Performance of SD BIOLINE tetanus kit in comparison with two quantative ELISA kits
SD BIOLINE tetanus kit SD BIOLINE tetanus kit vs ELISA 1
SD BIOLINE tetanus kit vs ELISA2
Sensitivity % Specificity % PPV % NPV % Kappa agreement (95%CI) Sensitivity % Specificity % PPV % NPV % Kappa agreement (95%CI)
in whole blood 86.7 91.8 88.6 90.4 0.79 94.8 89.2 83.3 96.7 0.82
(n=319)         (0.72–0.86)         (0.75–0.88)
in serum 88.3 89.9 86.4 91.4 0.78 96.5 87 80.7 97.5 0.8
(n=326)         (0.71–0.85)         (0.74–0.87)

ELISA 1, ELISA anti-tetanus IgG assay (SCIMEDX Co); ELISA2, IBL tetanus Ab. ELISA (IBL IMMUNO BIOLOGICAL LABORATORIES).

Abbreviations: PPV, positive predictive value; NPV, negative predictive value; CI, confidence interval.

Table 2.
Testing results of 3 analytical methods around cut-off (0.1 IU/mL)
Specimen No. ELISA 1 (IU/mL) ELISA 2 (IU/mL) SD BIOLINE tetanus kit
by operator 1 by operator 2
1 0.101 0.113 Negative Negative
2 <0.100 0.124 Negative Negative
3 0.144 <0.100 Negative Negative
4 0.118 <0.100 Negative Negative
5 0.104 <0.100 Positive Negative
6 0.135 0.107 Positive Negative
7 <0.100 0.123 Positive Positive
8 0.100 0.143 Positive Positive
9 0.100 <0.100 Positive Positive
10 0.105 0.127 Positive Positive
11 0.115 0.106 Positive Positive
12 0.128 0.102 Positive Positive
13 0.136 0.120 Positive Positive
14 0.138 0.147 Positive Positive
15 0.145 0.150 Positive Positive

Abbreviations: See Table 1.