Park, Cho, and Lee: The Clinical Efficacy of the Haigis Formula Using A-Scan Contact Ultrasound Biometry

Abstract

Purpose

To investigate the accuracy of the Haigis formula compared to other formulas using contact ultrasound biometry.

Methods

This study was performed on 94 patients (114 eyes) who underwent cataract surgery in our hospital. Axial length (AXL) and anterior chamber depth (ACD) were measured using both A-scan and intraocular lens (IOL) Master® Patients were divided into three groups based on AXL; Group I (AXL < 22.5 mm), Group II (22.5 mm ≤ AXL < 25.5 mm), and Group III (AXL≥ 25.5 mm). Before cataract surgery, predicted refraction was calculated using the Haigis, SRK/T, Hoffer Q, and Holladay 1 formulas using both A-scan and IOL Master® measurements. Mean absolute error (MAE) were analyzed at one month after surgery using the various IOL formulas.

Results

Using contact ultrasound biometry, in Group I, MAE of Haigis was 0.80 ± 0.67 D and was significantly lower than that using SRK/T. In Group II, the Haigis MAE was 0.72 ± 0.55 D and was significantly lower than the results of all other formulas. In Group III, the Haigis MAE was 0.76 ± 1.13 D and not significantly different from the results of other formulas. Comparing MAE of A-scan to IOL Master® the Haigis formula showed 0.16 D higher error that decreased when the AXL was close to the normal range.

Conclusions

Using contact ultrasound biometry, the Haigis formula provided the best predictability of postoperative refractive outcome compared to other formulas in eyes with normal axial length.

References

1. Olsen T. Prediction of intraocular lens position after cataract extraction. J Cataract Refract Surg. 1986; 12:376–9.
crossref
2. Olsen T. Sources of error in intraocular lens power calculation. J Cataract Refract Surg. 1992; 18:125–9.
crossref
3. Bhatt AB, Schefler AC, Feuer WJ, et al. Comparison of predictions made by the intraocular lens master and ultrasound biometry. Arch Ophthalmol. 2008; 126:929–33.
crossref
4. Choi J, Choi SK. Accuracy of intraocular lens power calculation in diabetic patients. J Korean Ophthalmol Soc. 2010; 51:188–94.
crossref
5. Tehrani M, Krummenauer F, Blom E, Dick HB. Evaluation of the practicality of optical biometry and applanation ultrasound in 253 eyes. J Cataract Refract Surg. 2003; 29:741–6.
crossref
6. Giers U, Epple C. Comparison of A-scan device accuracy. J Cataract Refract Surg. 1990; 16:235–42.
crossref
7. Yi CH, Choi SH, Chung ES, Chung TY. Accuracy of the haigis formula based on axial length and anterior chamber depth. J Korean Ophthalmol Soc. 2011; 52:175–81.
crossref
8. Haigis W, Lege B, Miller N, Schneider B. Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis. Graefes Arch Clin Exp Ophthalmol. 2000; 238:765–73.
crossref
9. Eleftheriadis H. IOLMaster biometry: refractive results of 100 consecutive cases. Br J Ophthalmol. 2003; 87:960–3.
crossref
10. Rose LT, Moshegov CN. Comparison of the Zeiss IOLMaster and applanation A-scan ultrasound: biometry for intraocular lens calculation. Clin Experiment Ophthalmol. 2003; 31:121–4.
crossref
11. Shin JA, Chung SK. Comparison of the refractive results measured by ultrasound and partial coherence interferometers. J Korean Ophthalmol Soc. 2013; 54:723–7.
crossref
12. Hwang JS, Lee JH. Comparison of the IOL Master® and A-scan ultrasound: refractive results of 96 consecutive cases. J Korean Ophthalmol Soc. 2007; 48:27–32.
13. Hoffer KJ. The Hoffer Q formula: a comparison of theoretic and regression formulas. J Cataract Refract Surg. 1993; 19:700–12.
crossref
14. Haigis W. The Haigis formula. Shammas HJ, editor. Intraocular lens power calculations. Thorofare, NJ: Slack Inc.;2004. p. 5–57.
15. Kim BH, Wee WR, Kim MK. Analysis of factors that influence on accuracy of intraocular lens power calculation. J Korean Ophthalmol Soc. 2014; 55:173–81.
crossref
16. Dinc UA, Gorgun E, Oncel B, et al. Assessment of anterior chamber depth using Visante optical coherence tomography, slitlamp optical coherence tomography, IOL Master, Pentacam and Orbscan IIz. Ophthalmologica. 2010; 224:341–6.
crossref
17. Holladay JT, Prager TC, Chandler TY, et al. A three-part system for refining intraocular lens power calculations. J Cataract Refract Surg. 1988; 14:17–24.
crossref

Figure 1.
The perform ance of various intraocular lens (IOL) calculation formulas with ultrasound biometry. MAE = mean absolute error.
jkos-55-1793f1.tif
Figure 2.
Intraocular lens (IOL) power calculation error using A-scan in the various IOL calculation formulas. MAE = mean absolute error using A-scan (diopter).
jkos-55-1793f2.tif
Figure 3.
The linear regression analysis of absolute mean absolute error difference between ultrasound and intraocular lens (IOL) Master® using Haigis formula (Linear regression model: R2 = 0.170, p < 0.001). MAEdiff = | MAEA-scan – MAEIOLM |; MAEA-scan = mean absolute error of Haigis formula using A-scan; MAEIOLM = mean absolute error of Haigis formula using IOL Master®; AXL = axial length.
jkos-55-1793f3.tif
Table 1.
Patient preoperative demography
Group I (AXL < 22.5) Group II (22.5 ≤ AXL < 25.5) Group III (AXL ≥ 25.5) Total
No. of eyes 27 (23.7%) 55 (48.2%) 32 (28.1%) 114
Age (years) 71.6 ± 9.0 69.3 ± 7.8 56.2 ± 16.0 66.2 ± 12.5
Implanted IOL power (D) 22.57 ± 1.83 20.26 ± 2.14 10.33 ± 5.08 18.02 ± 5.85
A-scan
 AXL (mm) 22.06 ± 0.41 23.47 ± 0.57 27.95 ± 2.39 24.40 ± 2.66
 ACD (mm) 2.63 ± 0.47 2.76 ± 0.31 3.14 ± 0.51 2.83 ± 0.45
IOL Master®
 AXL (mm) 22.16 ± 0.39 23.61 ± 0.58 28.26 ± 2.48 24.57 ± 2.75
 ACD (mm) 2.88 ± 0.41 3.15 ± 0.31 3.69 ± 0.30 3.24 ± 0.45
 Keratometry (D) 45.83 ± 1.29 44.08 ± 1.38 43.75 ± 1.59 44.40 ± 1.63
p-value
 AXL <0.001* <0.001* <0.001* <0.001*
 ACD <0.001* <0.001* <0.001* <0.001*

Values are presented as mean ± SD unless otherwise indicated; Group: Classification of patients was based on the IOL Master® measurements.

AXL = axial length; IOL = intraocular lens; D = diopter; ACD = anterior chamber depth.

* Statistically significant;

Wilcoxon's signed rank test between A-scan & IOL Master® measurement.

Table 2.
Comparison of IOL power calculation error between A-scan and IOL Master®
A-scan IOL Master® p-value
Group I-MAE (D)
 Haigis 0.80 ± 0.67 0.80 ± 0.88 0.572
 SRK/T 1.09 ± 0.77 0.94 ± 0.86 0.007*
 Holladay 1 0.94 ± 0.75 0.85 ± 0.80 0.064
 Hoffer Q 0.91 ± 0.76 0.88 ± 0.83 0.373
Group II-MAE (D)
 Haigis 0.72 ± 0.55 0.56 ± 0.51 0.001*
 SRK/T 1.04 ± 0.58 0.77 ± 0.55 <0.001*
 Holladay 1 0.89 ± 0.57 0.65 ± 0.55 <0.001*
 Hoffer Q 0.80 ± 0.59 0.57 ± 0.54 <0.001*
Group III-MAE (D)
 Haigis 0.76 ± 1.13 0.85 ± 1.37 0.806
 SRK/T 0.86 ± 1.07 0.85 ± 1.28 0.232
 Holladay 1 0.74 ± 1.17 0.95 ± 1.41 0.199
 Hoffer Q 0.72 ± 1.20 0.95 ± 1.43 0.063

Values are presented as mean ± SD.

IOL = intraocular lens; MAE = mean absolute error; D = diopter.

* Statistically significant;

Wilcoxon's signed rank test between A-scan & IOL Master® measurement.

Table 3.
IOL power calculation error with various formulas by ultrasound biometry
MAEA-scan p-value
Group I
 MAE A-scan (D)
  Haigis 0.80 ± 0.67
  SRK/T 1.09 ± 0.77 0.004*
  Holladay 1 0.94 ± 0.75 0.044
  Hoffer Q 0.91 ± 0.76 0.034
 Overall p-value <0.001*
Group II
 MAE A-scan (D)
  Haigis 0.72 ± 0.55
  SRK/T 1.04 ± 0.58 <0.001*
  Holladay 1 0.89 ± 0.57 <0.001*
  Hoffer Q 0.80 ± 0.59 0.006*
 Overall p-value <0.001*
Group III
 MAE A-scan (D)
  Haigis 0.76 ± 1.13
  SRK/T 0.86 ± 1.07 0.181
  Holladay 1 0.74 ± 1.17 0.570
  Hoffer Q 0.72 ± 1.20 0.125
 Overall p-value 0.140

Values are presented as mean ± SD.

IOL = intraocular lens; MAE A-scan = mean absolute error using A-scan; D = diopter.

* Statistically significant;

Post-hoc analysis with Wilcoxon's signed rank test between Haigis and other formulas (significant p-value is p < 0.013);

Friedman's test among all formulas.