Abstract
Purpose
To investigate the correlations of keratometry and corneal high order aberrations (HOAs) between the Keratron Scout videokeratoscope and the Pentacam HR Scheimpflug camera.
Methods
From December 2012 to February 2013, keratometry and corneal HOAs were determined using the Keratron Scout videokeratoscope and Pentacam HR Scheimpflug camera in 23 healthy individuals (46 eyes).
Results
Average keratometry showed high correlation with 95% confidence interval −0.155 ± 0.37 between the Keratron Scout and Pentacam HR. When comparing HOAs of the Keratron Scout and total corneal HOAs of Pentacam HR, total root-meansquare (RMS), and spherical aberration were positively correlated between the 2 devices (r = 0.67, p < 0.001 and r = 0.74, p < 0.001, respectively). When comparing HOAs of Keratron Scout and anterior corneal HOAs of Pentacam HR, total RMS, spherical aberration and coma were positively correlated between the 2 devices (r = 0.62, p < 0.001, r = 0.81, p < 0.001, and r = 0.52, p = 0.047, respectively).
References
1. Bühren J, Kook D, Yoon G, Kohnen T. Detection of subclinical keratoconus by using corneal anterior and posterior surface aberrations and thickness spatial profiles. Invest Ophthalmol Vis Sci. 2010; 51:3424–32.
2. Jafri B, Li X, Yang H, Rabinowitz YS. Higher order wavefront aberrations and topography in early and suspected keratoconus. J Refract Surg. 2007; 23:774–81.
3. Kim HS, Kim SW, Ha BJ, et al. Ocular aberrations and contrast sensitivity in eyes implanted with aspheric and spherical intraocular lenses. J Korean Ophthalmol Soc. 2008; 49:1256–62.
4. Lee SM, Lee MJ, Kim MK, et al. Comparison of changes in higher-order aberrations between conventional and wavefront-guided LASEK. J Korean Ophthalmol Soc. 2007; 48:1028–35.
5. Kim SI, Oh JJ, Oh TH, et al. Higher-order aberrations and visual acuity with wavefront-guided and wavefront-optimized ablation in laser keratorefractive surgery. J Korean Ophthalmol Soc. 2014; 55:480–5.
6. Tripoli NK, Cohen KL, Holmgren DE, Coggins JM. Assessment of radial aspheres by the Arc-step algorithm as implemented by the Keratron keratoscope. Am J Ophthalmol. 1995; 120:658–64.
7. Buehl W, Stojanac D, Sacu S, et al. Comparison of three methods of measuring corneal thickness and anterior chamber depth. Am J Ophthalmol. 2006; 141:7–12.
8. Lackner B, Schmidinger G, Pieh S, et al. Repeatability and reproducibility of central corneal thickness measurement with Pentacam, Orbscan, and ultrasound. Optom Vis Sci. 2005; 82:892–9.
9. Lackner B, Schmidinger G, Skorpik C. Validity and repeatability of anterior chamber depth measurements with Pentacam and Orbscan. Optom Vis Sci. 2005; 82:858–61.
10. O'Donnell C, Maldonado-Codina C. Agreement and repeatability of central thickness measurement in normal corneas using ultrasound pachymetry and the OCULUS Pentacam. Cornea. 2005; 24:920–4.
11. Barkana Y, Gerber Y, Elbaz U, et al. Central corneal thickness measurement with the Pentacam Scheimpflug system, optical low-coherence reflectometry pachymeter, and ultrasound pachymetry. J Cataract Refract Surg. 2005; 31:1729–35.
12. Miranda MA, O'Donnell C, Radhakrishnan H. Repeatability of corneal and ocular aberration measurements and changes in aberrations over one week. Clin Exp Optom. 2009; 92:253–66.
13. Visser N, Berendschot TT, Verbakel F, et al. Evaluation of the comparability and repeatability of four wavefront aberrometers. Invest Ophthalmol Vis Sci. 2011; 52:1302–11.
14. Alpins NA. A new method of analyzing vectors for changes in astigmatism. J Cataract Refract Surg. 1993; 19:524–33.
15. MacRae S, Fujieda M. Slit skiascopic-guided ablation using the Nidek laser. J Refract Surg. 2000; 16:S576–80.
16. Savini G, Carbonelli M, Barboni P, Hoffer KJ. Repeatability of automatic measurements performed by a dual Scheimpflug analyzer in unoperated and post-refractive surgery eyes. J Cataract Refract Surg. 2011; 37:302–9.
17. Kawamorita T, Nakayama N, Uozato H. Repeatability and reproducibility of corneal curvature measurements using the Pentacam and Keratron topography systems. J Refract Surg. 2009; 25:539–44.
18. Tripoli NK, Cohen KL, Holmgren DE, Coggins JM. Assessment of radial aspheres by the Arc-step algorithm as implemented by the Keratron keratoscope. Am J Ophthalmol. 1995; 120:658–64.
Table 1.
Values | |
---|---|
Age (years) | 24.4 ± 4.24 |
UCVA (log MAR) | 0.16 ± 0.08 |
BCVA (log MAR) | 0.92 ± 0.09 |
Sphere (diopter) | −3.76 ± 1.54 |
Cylinder (diopter) | −1.48 ± 0.86 |