Journal List > J Korean Ophthalmol Soc > v.55(12) > 1009855

Shin, Lee, and Chung: Comparison of Keratometry and Corneal Higher Order Aberrations between Scout Videokeratoscope and Pentacam Scheimpflug Camera

Abstract

Purpose

To investigate the correlations of keratometry and corneal high order aberrations (HOAs) between the Keratron Scout videokeratoscope and the Pentacam HR Scheimpflug camera.

Methods

From December 2012 to February 2013, keratometry and corneal HOAs were determined using the Keratron Scout videokeratoscope and Pentacam HR Scheimpflug camera in 23 healthy individuals (46 eyes).

Results

Average keratometry showed high correlation with 95% confidence interval −0.155 ± 0.37 between the Keratron Scout and Pentacam HR. When comparing HOAs of the Keratron Scout and total corneal HOAs of Pentacam HR, total root-meansquare (RMS), and spherical aberration were positively correlated between the 2 devices (r = 0.67, p < 0.001 and r = 0.74, p < 0.001, respectively). When comparing HOAs of Keratron Scout and anterior corneal HOAs of Pentacam HR, total RMS, spherical aberration and coma were positively correlated between the 2 devices (r = 0.62, p < 0.001, r = 0.81, p < 0.001, and r = 0.52, p = 0.047, respectively).

Conclusions

Although the 2 devices are based upon different principles, the Keratron Scout videokeratoscope and Pentacam HR Scheimpflug camera showed positive correlations in keratometry, total RMS, spherical aberration, and coma. Both devices may be useful for clinical applications.

References

1. Bühren J, Kook D, Yoon G, Kohnen T. Detection of subclinical keratoconus by using corneal anterior and posterior surface aberrations and thickness spatial profiles. Invest Ophthalmol Vis Sci. 2010; 51:3424–32.
crossref
2. Jafri B, Li X, Yang H, Rabinowitz YS. Higher order wavefront aberrations and topography in early and suspected keratoconus. J Refract Surg. 2007; 23:774–81.
crossref
3. Kim HS, Kim SW, Ha BJ, et al. Ocular aberrations and contrast sensitivity in eyes implanted with aspheric and spherical intraocular lenses. J Korean Ophthalmol Soc. 2008; 49:1256–62.
crossref
4. Lee SM, Lee MJ, Kim MK, et al. Comparison of changes in higher-order aberrations between conventional and wavefront-guided LASEK. J Korean Ophthalmol Soc. 2007; 48:1028–35.
crossref
5. Kim SI, Oh JJ, Oh TH, et al. Higher-order aberrations and visual acuity with wavefront-guided and wavefront-optimized ablation in laser keratorefractive surgery. J Korean Ophthalmol Soc. 2014; 55:480–5.
crossref
6. Tripoli NK, Cohen KL, Holmgren DE, Coggins JM. Assessment of radial aspheres by the Arc-step algorithm as implemented by the Keratron keratoscope. Am J Ophthalmol. 1995; 120:658–64.
crossref
7. Buehl W, Stojanac D, Sacu S, et al. Comparison of three methods of measuring corneal thickness and anterior chamber depth. Am J Ophthalmol. 2006; 141:7–12.
crossref
8. Lackner B, Schmidinger G, Pieh S, et al. Repeatability and reproducibility of central corneal thickness measurement with Pentacam, Orbscan, and ultrasound. Optom Vis Sci. 2005; 82:892–9.
crossref
9. Lackner B, Schmidinger G, Skorpik C. Validity and repeatability of anterior chamber depth measurements with Pentacam and Orbscan. Optom Vis Sci. 2005; 82:858–61.
crossref
10. O'Donnell C, Maldonado-Codina C. Agreement and repeatability of central thickness measurement in normal corneas using ultrasound pachymetry and the OCULUS Pentacam. Cornea. 2005; 24:920–4.
11. Barkana Y, Gerber Y, Elbaz U, et al. Central corneal thickness measurement with the Pentacam Scheimpflug system, optical low-coherence reflectometry pachymeter, and ultrasound pachymetry. J Cataract Refract Surg. 2005; 31:1729–35.
crossref
12. Miranda MA, O'Donnell C, Radhakrishnan H. Repeatability of corneal and ocular aberration measurements and changes in aberrations over one week. Clin Exp Optom. 2009; 92:253–66.
crossref
13. Visser N, Berendschot TT, Verbakel F, et al. Evaluation of the comparability and repeatability of four wavefront aberrometers. Invest Ophthalmol Vis Sci. 2011; 52:1302–11.
crossref
14. Alpins NA. A new method of analyzing vectors for changes in astigmatism. J Cataract Refract Surg. 1993; 19:524–33.
crossref
15. MacRae S, Fujieda M. Slit skiascopic-guided ablation using the Nidek laser. J Refract Surg. 2000; 16:S576–80.
crossref
16. Savini G, Carbonelli M, Barboni P, Hoffer KJ. Repeatability of automatic measurements performed by a dual Scheimpflug analyzer in unoperated and post-refractive surgery eyes. J Cataract Refract Surg. 2011; 37:302–9.
crossref
17. Kawamorita T, Nakayama N, Uozato H. Repeatability and reproducibility of corneal curvature measurements using the Pentacam and Keratron topography systems. J Refract Surg. 2009; 25:539–44.
crossref
18. Tripoli NK, Cohen KL, Holmgren DE, Coggins JM. Assessment of radial aspheres by the Arc-step algorithm as implemented by the Keratron keratoscope. Am J Ophthalmol. 1995; 120:658–64.
crossref
19. Piñero DP, Nieto JC, Lopez-Miguel A. Characterization of corneal structure in keratoconus. J Cataract Refract Surg. 2012; 38:2167–83.
crossref
20. Bühren J, Kühne C, Kohnen T. Defining subclinical keratoconus using corneal first-surface higher-order aberrations. Am J Ophthalmol. 2007; 143:381–9.
crossref

Figure 1.
The comparison of average keratometry (K) between the Keratron Scout videokeratoscopy and Pentacam HR Scheimpflug Camera. The solid line represents ±0.37 standard deviation confidence interval (CI) (i.e. 95% of the points should fall within these lines).
jkos-55-1758f1.tif
Figure 2.
The correlation between corneal higher order aberrations (HOAs) of the Keratron Scout videokeratoscopy and total corneal HOAs of Pentacam HR Scheimpflug Camera. RMS = root-mean-square; SA = spherical aberration.
jkos-55-1758f2.tif
Figure 3.
The correlation between corneal higher order aberrations (HOAs) of Keratron Scout videokeratoscopy and anterior corneal HOAs of Pentacam HR Scheimpflug Camera. RMS = root-mean-square; SA = spherical aberration.
jkos-55-1758f3.tif
Table 1.
Demographics of 23 patients (M:F = 8:15)
Values
Age (years) 24.4 ± 4.24
UCVA (log MAR) 0.16 ± 0.08
BCVA (log MAR) 0.92 ± 0.09
Sphere (diopter) −3.76 ± 1.54
Cylinder (diopter) −1.48 ± 0.86

Values are presented as mean ± SD.

UCVA = uncorrected visual acuity; BCVA = best corrected visual acuity.

Table 2.
The comparison of cornea higher order aberrations (HOAs) between the Keratron Scout videokeratoscopy and Pentacam HR Scheimpflug Camera
Scout (corneal HOAs) Pentacam (corneal HOAs) p-value
Total RMS 0.41 ± 0.06 0.37 ± 0.05 0.054
Spherical aberration 0.24 ± 0.07 0.22 ± 0.08 0.061
Coma 0.20 ± 0.07 0.18 ± 0.05 0.005
Trefoil 0.16 ± 0.01 0.10 ± 0.01 <0.001
Scout (corneal HOAs) Pentacam (anterior corneal HOAs) p-value
Total RMS 0.41 ± 0.06 0.40 ± 0.07 0.31
Spherical aberration 0.26 ± 0.06 0.27 ± 0.08 0.153
Coma 0.20 ± 0.07 0.20 ± 0.07 0.173
Trefoil 0.18 ± 0.06 0.10 ± 0.01 <0.001

Values are presented as mean ± SD.

RMS = root-mean-square.

Table 3.
The correlation of cornea higher order aberrations (HOAs) between the Keratron Scout videokeratoscopy and Pentacam HR Scheimpflug Camera
Scout (corneal HOAs) Pentacam (corneal HOAs) Differences Correlation coefficient p-value
Total RMS 0.41 ± 0.06 0.37 ± 0.05 0.04 ± 0.05 0.67 <0.001
Spherical aberration 0.24 ± 0.07 0.22 ± 0.08 0.02 ± 0.05 0.74 <0.001
Coma 0.20 ± 0.07 0.18 ± 0.05 0.03 ± 0.09 0.32 0.189
Trefoil 0.16 ± 0.01 0.10 ± 0.01 0.07 ± 0.06 0.45 0.105
Scout (corneal HOAs) Pentacam (anterior corneal HOAs) Differences Correlation coefficient p-value
Total RMS 0.41 ± 0.06 0.40 ± 0.07 0.01 ± 0.06 0.62 <0.001
Spherical aberration 0.26 ± 0.06 0.27 ± 0.08 0.03 ± 0.05 0.81 <0.001
Coma 0.20 ± 0.07 0.20 ± 0.07 0.02 ± 0.09 0.52 0.047
Trefoil 0.18 ± 0.06 0.10 ± 0.01 0.07 ± 0.07 0.43 0.094

Values are presented as mean ± SD.

RMS = root-mean-square.

TOOLS