Kim and Yi: Comparison of Intraocular Pressures According to Position Using Icare Rebound Tonometer

Abstract

Purpose

To evaluate changes in intraocular pressure (IOP) according to position using a portable rebound tonometer.

Methods

We measured the IOP values of 20 healthy volunteers (40 eyes) in the sitting, supine, right lateral decubitus and left decubitus positions with a portable rebound tonometer, and then analyzed using the Wilcoxon signed rank test. IOP in sitting position was also measured with a non-contact tonometer and a Goldmann applanation tonometer, and analyzed with Kruskal-Wallis test and Spearman correlation analysis. Agreement among the 3 tonometers was calculated using the Bland-Altman method.

Results

The IOP measured with rebound tonometer in the supine position was significantly higher than in the sitting position (p = 0.002). However, there was no significant difference in IOP between the supine and decubitus positions. In the decubitus position, there was no significant difference in IOP between the dependent and non-dependent eyes. IOP measurement using the rebound tonometer showed positive correlation with that of the noncontact and Goldmann applanation tonometers.

Conclusions

In normal subjects, IOP measurement obtained with a rebound tonometer in the supine position was significantly higher than in the sitting position, but there was no significant difference in IOP between the supine and decubitus positions. A rebound tonometer may be useful for patients whose intraocular pressure measurement with Goldmann applanation tonometer or non-contact tonometer is impossible. When using a portable rebound tonometer in bed-ridden or pediatric patients, we should pay attention to the interpretation of IOP in the supine position.

References

1. Goldmann H, Schmidt T. Applanation tonometry. Ophthalmologica. 1957; 134:221–42.
2. Whitacre MM, Stein R. Sources of error with use of Goldmanntype tonometers. Surv Ophthalmol. 1993; 38:1–30.
crossref
3. Rosentreter A, Neuburger M, Jordan JF, et al. Factors influencing applanation tonometry - a practical approach. Klin Monbl Augenheilkd. 2011; 228:109–13.
4. Lee K, Lee JY, Moon JL, Park MH. Comparison of Icare rebound tonometer with Goldmann applanation tonometry. J Korean Ophthalmol Soc. 2013; 54:296–302.
crossref
5. Martinez-de-la-Casa JM, Garcia-Feijoo J, Castillo A, Garcia-Sanchez J. Reproducibility and clinical evaluation of rebound tonometry. Invest Ophthalmol Vis Sci. 2005; 46:4578–80.
crossref
6. Fernandes P, Diaz-Rey JA, Queiros A, et al. Comparison of the ICare rebound tonometer with the Goldmann tonometer in a normal population. Ophthal Physiol Opt. 2005; 25:436–40.
7. Garcia-Resua C, Gonzalez-Meijome JM, Gilino J, Yebra-Pimentel E. Accuracy of the new ICare rebound tonometer vs. other portable tonometers in healthy eyes. Optom Vis Sci. 2006; 83:102–7.
8. Vincent SJ, Vincent RA, Shields D, Lee GA. Comparison of intraocular pressure measurement between rebound, noncontact and Goldmann applanation tonometry in treated glaucoma patients. Clin Experiment Ophthalmol. 2012; 40:e163–70.
crossref
9. Takenaka J, Mochizuki H, Kunihara E, et al. Intraocular pressure measurement using rebound tonometer for deviated angels and positions in human eyes. Curr Eye Res. 2012; 37:109–14.
10. Jablonski KS, Rosentreter A, Gaki S, et al. Clinical use of a new position-independent rebound tonometer. J Glaucoma. 2013; 22:763–7.
crossref
11. Anderson DR, Grant WM. The influence of position on intraocular pressure. Invest Ophthalmol. 1973; 12:204–12.
12. Kontiola AI. A new induction-based impact method for measuring intraocular pressure. Acta Ophthalmol Scand. 2000; 78:142–5.
crossref
13. Kontiola AI, Goldblum D, Mittag T, Danias J. The induction/impact tonometer: a new instrument to measure intraocular pressure in the rat. Exp Eye Res. 2001; 73:781–5.
crossref
14. Krieglstein GK, Waller WK. Goldmann applanation versus hand-applanation and schiötz indentation tonometry. Albrecht Von Graefes Arch Clin Exp Ophthalmol. 1975; 194:11–6.
crossref
15. Baskett JS, Goen TM, Terry JE. A comparison of Perkins and Goldmann applanation tonometry. J Am Optom Assoc. 1986; 57:832–4.
16. Lee JY, Yoo C, Jung JH, et al. The effect of lateral decubitus position on intraocular pressure in healthy young subjects. Acta Ophthalmologica. 2012; 90:e68–72.
crossref
17. Kim HS, Park KH, Jeoung JW. Can we measure the intraocular pressure when the eyeball is against the pillow in the lateral decubitus position. Acta Ophthalmol. 2013; 91:e502–5.
crossref
18. Lee TE, Yoo C, Kim YY. Effects of different sleeping postures on intraocular pressure and ocular perfusion pressure in healthy young subjects. Ophthalmology. 2013; 120:1565–70.
crossref
19. Almubrad TM, Ogbuehi KC. On repeated corneal applanation with the Goldmann and two non-contact tonometers. Clin Exp Optom. 2010; 93:77–82.
crossref
20. Gaton DD, Ehrenberg M, Lusky M, et al. Effect of repeated applanation tonometry on the accuracy of intraocular pressure measurements. Curr Eye Res. 2010; 35:475–9.
crossref
21. Smith TJ, Lewis J. Effect of inverted body position intraocular pressure. Am J Ophthalmol. 1985; 99:617–8.
22. Malihi M, Sit AJ. Effect of head and body position on intraocular pressure. Ophthalmology. 2012; 119:987–91.
crossref
23. Blondeau P, Tétrault JP, Papamarkakis C. Diurnal variation of episcleral venous pressure in healthy patients: a pilot study. J Glaucoma. 2001; 10:18–24.
crossref
24. Miyamoto S, Tambara K, Tamaki S, et al. Effects of right lateral decubitus position on plasma norepinephrine and plasma atrial natriuretic peptide levels in patients with chronic congestive heart failure. Am J Cardiol. 2002; 89:240–2.
crossref
25. Miyamoto S, Fujita M, Sekiguchi H, et al. Effects of posture on cardiac autonomic nervous activity in patients with congestive heart failure. J Am Coll Cardiol. 2001; 37:1788–93.
crossref
26. Parsley J, Powell RG, Keightley SJ, Elkington AR. Postural response of intraocular pressure in chronic open-angle glaucoma following trabeculectomy. Br J Ophthalmol. 1987; 71:494–6.
crossref
27. Williams BI, Peart WS, Letley E. Abnormal intraocular pressure control in systemic hypertension and diabetic mellitus. Br J Ophthalmol. 1980; 64:845–51.
crossref
28. Kim KN, Jeoung JW, Park KH, et al. Effect of lateral decubitus position on intraocular pressure in glaucoma patients with asymmetric visual field loss. Ophthalmology. 2013; 120:731–5.
crossref
29. Lee CM, Yu YC. Intraocular pressure measurement with the noncontact tonometer and rebound tonometer through plano soft contact lenses. J Korean Ophthalmol Soc. 2012; 53:662–7.
crossref
30. Rosentreter A, Athanasopoulos A, Schild AM, et al. Rebound, applanation, and dynamic contour tonometry in pathologic corneas. Cornea. 2013; 32:313–8.
crossref
31. Suman S, Agrawal A, Pal VK, Pratap VB. Rebound tonomter: Ideal tonometer for measurement of accurate intraocular pressure. J Glaucoma. 2013; [E-pub ahead of print].
32. Lee JH, Seong MC, Kang MH, et al. Comparison of rebound tonometer, non-contact tonometer, goldmann applanation tonometer and the relationship to central corneal thickness. J Korean Ophthalmol Soc. 2012; 53:988–95.
crossref
33. Lee K, Lee JY, Moon JI, Park MH. Comparison of Icare rebound tonometer with Goldmann applanation tonometry. J Korean Ophthalmol Soc. 2013; 54:296–302.
crossref
34. Cook JA, Botello AP, Elders A, et al. Systemic review ofthe agreement of tonometers with Goldmann applanation tonometry. Ophthalmology. 2012; 119:1552–7.

Figure 1.
Comparison of intraocular pressures according to body position with rebound tonometer (*p < 0.05, Wilcoxon signed rank test).
jkos-55-1049f1.tif
Figure 2.
Comparison of intraocular pressure (IOP) measured with non-contact tonometer and Icare rebound tonometer (r = 0.636, p = 0.000). IOP = intraocular pressure; RB = Icare rebound tonometer; NCT = non-contact tonometer.
jkos-55-1049f2.tif
Figure 3.
Comparison of intraocular pressure measured with Goldmann applanation tonometer and Icare rebound tonometer (r = 0.662, p = 0.000). IOP = intraocular pressure; RB = Icare rebound tonometer; GAT = Goldmann applanation tonomter.
jkos-55-1049f3.tif
Figure 4.
Bland-Altman plot. The differences between IOP measured by RB and NCT are plotted against the mean of the two measurements. IOP = intraocular pressure; RB = rebound tonometer; NCT = non-contact tonometer; SD = standard deviation.
jkos-55-1049f4.tif
Figure 5.
Bland-Altman plot. The differences between IOP measured by RB and GAT are plotted against the mean of the two measurements. IOP = intraocular pressure; RB = rebound tonometer; GAT = Goldmann applanation tonomter; SD = standard deviation.
jkos-55-1049f5.tif
Table 1.
Demographic data, IOP measurements (mm Hg) with 3 tonometers
Age (years) 25.95 ± 2.28
Sex (M:F) 14:6
CCT (μm) 537.8 ± 29.68
IOP NCT (mm Hg) 16.2 ± 3.13
IOP GAT (mm Hg) 15.72 ± 2.49
Sitting IOP RB (mm Hg) 15.82 ± 2.87
Supine IOP RB (mm Hg) 16.85 ± 3.08
Right lateral decubitus IOP RB (mm Hg) 16.73 ± 2.84
Left lateral decubitus IOP RB (mm Hg) 16.80 ± 2.80

Values are presented as mean ± SD.

CCT = central corneal thickness; IOP = intraocular pressure; NCT = non-contact tonometer; GAT = Goldmann applnation tonomter; RB = rebound tonomter.

Table 2.
Intraocular pressure measurements with rebound tonometer during postural change
Right eye (mm Hg) Left eye (mm Hg) p-value*
Sitting 15.85 ± 2.93 15.79 ± 2.89 0.718
Supine 16.91 ± 3.09 16.80 ± 3.14 0.862
Right lateral decubitus 16.75 ± 3.12 16.71 ± 2.61 0.968
Left lateral decubitus 16.62 ± 2.87 16.98 ± 2.79 0.698

Values are presented as mean ± SD.

* Wilcoxon signed rank test.

Table 3.
Statistical comparison of intraocular pressure between different body position
Position p-value*
Right lateral decubitus
Left lateral decubitus
Right eye (dependent eye) Left eye Right eye Left eye (dependent eye)
Supine
  Right eye 0.616 0.398
  Left eye 0.121 0.331

* Wilcoxon signed rank test.

Table 4.
Spearman's correlation coefficient among 3 tonometers
RB NCT
NCT 0.636 (p = 0.000)
GAT 0.662 (p = 0.000) 0.474 (p = 0.002)

NCT = non-contact tonometer; GAT = Goldmann applnation tonomter; RB = rebound tonomter.