Um, Lee, Kim, Kim, Tchach, and Kim: Comparison of Six-Month Outcomes in Eyes with Two Hydrophilic Aspheric Intraocular Lenses after Cataract Surgery

Abstract

Purpose

To compare the six-month clinical outcomes of Tek-Lens II model 872 (Tekia, Inc.) and Akreos MI-60 (Bausch & Lomb, Inc.) hydrophilic aspheric intraocular lenses (IOLs) implanted in-the-bag.

Methods

After phacoemulsification was performed by a single surgeon (JY Kim), two different hydrophilic aspheric IOLs were implanted: the Tek-Lens II IOL in 57 eyes and the MI60 IOL in 49 eyes. The best corrected visual acuity (BCVA), refractive error (RE), total high-order aberration (HOA), anterior chamber depth (ACD), and contrast sensitivity (CS) were measured preoperatively and one, three, and six months postoperatively. All parameters were analyzed using the independent t-test to compare the two IOL groups.

Results

There was no significant difference in the BCVA, RE, or HOA between the two groups at any of the follow-up visits. Three months postoperatively, the ACD of the TEK-Lens II group as measured by corneal topography was significantly shallower than that of the MI60 group (3.67 ± 0.52 mm vs. 4.10 ± 0.40 mm; p = 0.008), and most of the photopic CS in the Tek-Lens II group was significantly higher than that in the MI60 group (all p < 0.05).

Conclusions

The Tek-Lens II IOL showed comparable clinical outcomes with a proven hydrophilic aspheric IOL for the postoperative six months.

Figures and Tables

Figure 1
Comparison of postoperative contrast sensitivity (CS) in photophic and mesopic conditions. (A, C, E) CS in photopic condition at postoperative 1, 3, 6 months respectively. (B, D, F) CS in mesopic condition at postoperative 1, 3, 6 months respectively (*p < 0.05).
jkos-53-767-g001
Figure 2
Comparison of surgery-induced astigmatism (SIA) between two groups.
jkos-53-767-g002
Figure 3
Photographs of two IOLs. (A) Tek-Lens II model 872™, (B) Akreos MI60™.
jkos-53-767-g003
Table 1
Comparison of preoperative characteristics between two groups
jkos-53-767-i001

Values are presented as mean ± SD or number.

BCVA = best corrected visual acuity; log MAR = logarithm of the minimum angle of resolution; AL = axial length; IOL = intraocular lens.

Table 2
Comparison of postoperative outcomes by 6 months after cataract surgeries in which the Tek-Lens II model 872™ or Akreos MI60™ IOL was implanted
jkos-53-767-i002

Values are presented as mean ± SD.

BCVA = best corrected visual acuity; log MAR = logarithm of the minimum angle of resolution; SE = spherical equivalent; RE = refractive error; HOR = high order aberration; SA = spherical aberration; VC = vertical coma; HC = horizontal coma; ACD = anterior chamber depth; SIA = surgery-induced astigmatism.

Table 3
Characteristics of the two IOLs in this study
jkos-53-767-i003

IOL = intraocular lens.

References

1. Ginsburg AP, Evans DW, Sekule R, Harp SA. Contrast sensitivity predicts pilots' performance in aircraft simulators. Am J Optom Physiol Opt. 1982. 59:105–109.
2. Pesudovs K, Hazel CA, Doran RM, Elliott DB. The usefulness of Vistech and FACT contrast sensitivity charts for cataract and refractive surgery outcomes research. Br J Ophthalmol. 2004. 88:11–16.
3. Applegate RA, Hilmantel G, Howland HC, et al. Corneal first surface optical aberrations and visual performance. J Refract Surg. 2000. 16:507–514.
4. Bhattacharjee H, Bhattacharjee K, Medhi J. Visual performance: Comparison of foldable intraocular lenses. J Cataract Refract Surg. 2006. 32:451–455.
5. Mierdel P, Kaemmerer M, Mrochen M, et al. Ocular optical aberrometer for clinical use. J Biomed Opt. 2001. 6:200–204.
6. Holladay JT, Piers PA, Koranyi G, et al. A new intraocular lens design to reduce spherical aberration of pseudophakic eyes. J Refract Surg. 2002. 18:683–691.
7. Son SW, Seo JW, Shin SJ, Chung SK. Comparison of the stability between three-piece and single-piece aspheric intraocular lenses. J Korean Ophthalmol Soc. 2010. 51:1584–1589.
8. Jeong JH, Kim MK, Wee WR, Lee JH. Comparison of optical performances in eyes implanted with aspheric and spherical intraocular lenses after cataract surgery. J Korean Ophthalmol Soc. 2010. 51:1445–1452.
9. Mun GH, Im SK, Park HY, Yoon KC. Comparison of clinical results between two spherical aberraion-free intraocular lenses. J Korean Ophthalmol Soc. 2010. 51:670–676.
10. Bae HW, Kim EK, Kim TI. Spherical aberration, contrast sensitivity and depth of focus with three aspherical intraocular lenses. J Korean Ophthalmol Soc. 2009. 50:1639–1644.
11. Botelho PJ, Johnson LN, Arnold AC. The effect of aspirin on the visual outcome of nonarteritic anterior ischemic optic neuropathy. Am J Ophthalmol. 1996. 121:450–451.
12. Artal P, Berrio E, Guirao A, Piers P. Contribution of the cornea and internal surfaces to the change of ocular aberrations with age. J Opt Soc Am A Opt Image Sci Vis. 2002. 19:137–143.
13. Chalita MR, Chavala S, Xu M, Krueger RR. Wavefront analysis in post-LASIK eyes and its correlation with visual symptoms, refraction, and topography. Ophthalmology. 2004. 111:447–453.
14. Rawer R, Stork W, Spraul CW, Lingenfelder C. Imaging quality of intraocular lenses. J Cataract Refract Surg. 2005. 31:1618–1631.
15. Bach M, Wesemann W, Kolling G, et al. [Photopic contrast sensitivity. Local contrast perception]. Ophthalmologe. 2008. 105:46–48.
16. Kaufman PL, Alm A, Adler FH. Adler's Physiology of the Eye : Clinical Application. 2003. 10th ed. St. Louis: Mosby.
17. Packer M, Fine IH, Hoffman RS. Contrast sensitivity and measuring cataract outcomes. Ophthalmol Clin North Am. 2006. 19:521–533.
18. Shandiz JH, Derakhshan A, Daneshyar A, et al. Effect of cataract type and severity on visual acuity and contrast sensitivity. J Ophthalmic Vis Res. 2011. 6:26–31.
19. Hayashi K, Hayashi H. Comparison of the stability of 1-piece and 3-piece acrylic intraocular lenses in the lens capsule. J Cataract Refract Surg. 2005. 31:337–342.
20. Wirtitsch MG, Findl O, Menapace R, et al. Effect of haptic design on change in axial lens position after cataract surgery. J Cataract Refract Surg. 2004. 30:45–51.
21. Shin CJ, Lee JE, Lee JH, et al. Clinical outcomes after microincision cataract surgery and in-the-bag implantation of a new intraocular lens. J Korean Ophthalmol Soc. 2010. 51:677–683.
22. Alió J, Rodríguez-Prats JL, Galal A, Ramzy M. Outcomes of microincision cataract surgery versus coaxial phacoemulsification. Ophthalmology. 2005. 112:1997–2003.
TOOLS
Similar articles