Kim, Ryu, Kim, and Yang: Effects of Transient Prone Position on Vault and Anterior Chamber Angle in ICL Implanted Patients

Abstract

Purpose

To evaluate the effects of transient prone position on vault and anterior chamber angle parameters in ICL implanted patients.

Methods

40 eyes of 20 ICL implanted patients with at least 1 month of follow-up were included in the present study. The central ICL vault and anterior chamber parameters including angle opening distance at 500 (AOD500) were measured with the Visante anterior segment optical coherence tomography (OCT) (Carl Zeiss, Dublin, CA) in both the sitting and prone positions by tilting the OCT 90 degrees in the vertical axis and having the patient fixate downwards towards the floor.

Results

The mean central vault was 0.55 ± 0.21 mm (SD) and 0.59 ± 0.21 mm (SD) in the sitting and prone positions, respectively (p < 0.0001). The nasal and temporal AOD500 were 0.26 ± 0.11 mm and 0.28 ± 0.08 mm, respectively in the sitting position, which decreased to 0.24 ± 0.10 mm and 0.26 ± 0.08 mm in the prone position, however, both were not statistically significant (p = 0.08, p = 0.09). AOD500 was inversely correlated with vault (r = -0.47; p = 0.0024). There were no significant correlations between increase of vault and anterior chamber depth or white to white nor ICL vault.

Conclusions

Transient prone positioning of ICL implanted patients can induce a significant increase in ICL vault.

Figures and Tables

Figure 1
Box plot showing the change in vault (A), nasal AOD500 (B) and temporal AOD500 (C) between prone position and sitting position.
jkos-53-761-g001
Figure 2
Scatter diagrams with line of equality showing the changes in nasal AOD500 (A) and temporal AOD500 (B) induced by transient prone position.
jkos-53-761-g002
Table 1
General characteristics of 20 patients
jkos-53-761-i001

Values are presented as number or mean ± SD.

Table 2
General characteristics of 40 eyes of 20 patients
jkos-53-761-i002

Values are presented as mean ± SD.

ACD = anterior chamber depth; CCT = central corneal thickness.

Table 3
Comparison of parameters between sitting and prone position
jkos-53-761-i003

Paired t-test was done.

Values are presented as mean ± SD.

AOD500 = angle opening distance at 500 µm.

Table 4
Pearson correlation matrix of parameters
jkos-53-761-i004

ACD = anterior chamber depth; WTW = white to white by topography; UBM = sulcus to sulcus by UBM; AOD500 = angle opening distance at 500 µm.

*p<0.05; p<0.01; p<0.005.

References

1. Ieong A, Hau SC, Rubin GS, Allan BD. Quality of life in high myopia before and after implantable Collamer lens implantation. Ophthalmology. 2010. 117:2295–2300.
2. Rayner SA, Bhikoo R, Gray T. Spherical implantable collamer lenses for myopia and hyperopia: 126 eyes with 1-year follow up. Clin Experiment Ophthalmol. 2010. 38:21–26.
3. Kamiya K, Shimizu K, Igarashi A, et al. Four-year follow-up of posterior chamber phakic intraocular lens implantation for moderate to high myopia. Arch Ophthalmol. 2009. 127:845–850.
4. Khalifa YM, Moshirfar M, Mifflin MD, et al. Cataract development associated with collagen copolymer posterior chamber phakic intraocular lenses: clinicopathological correlation. J Cataract Refract Surg. 2010. 36:1768–1774.
5. Khalifa YM, Goldsmith J, Moshirfar M. Bilateral explantation of Visian Implantable Collamer Lenses secondary to bilateral acute angle closure resulting from a non-pupillary block mechanism. J Refract Surg. 2010. 26:991–994.
6. Chung TY, Park SC, Lee MO, et al. Changes in iridocorneal angle structure and trabecular pigmentation with STAAR implantable collamer lens during 2 years. J Refract Surg. 2009. 25:251–258.
7. Kojima T, Maeda M, Yoshida Y, et al. Posterior chamber phakic implantable collamer lens: changes in vault during 1 year. J Refract Surg. 2010. 26:327–332.
8. Alfonso JF, Lisa C, Abdelhamid A, et al. Three-year follow-up of subjective vault following myopic implantable collamer lens implantation. Graefes Arch Clin Exp Ophthalmol. 2010. 248:1827–1835.
9. Kamiya K, Shimizu K, Kawamorita T. Changes in vaulting and the effect on refraction after phakic posterior chamber intraocular lens implantation. J Cataract Refract Surg. 2009. 35:1582–1586.
10. Petternel V, Köppl CM, Dejaco-Ruhswurm I, et al. Effect of accommodation and pupil size on the movement of a posterior chamber lens in the phakic eye. Ophthalmology. 2004. 111:325–331.
11. Lindland A, Heger H, Kugelberg M, Zetterström C. Vaulting of myopic and toric Implantable Collamer Lenses during accommodation measured with Visante optical coherence tomography. Ophthalmology. 2010. 117:1245–1250.
12. Bechmann M, Ullrich S, Thiel MJ, et al. Imaging of posterior chamber phakic intraocular lens by optical coherence tomography. J Cataract Refract Surg. 2002. 28:360–363.
13. Mérula RV, Cronemberger S, Diniz Filho A, Calixto N. New comparative ultrasound biomicroscopic findings between fellow eyes of acute angle closure and glaucomatous eyes with narrow angle. Arq Bras Oftalmol. 2008. 71:793–798.
14. Marchini G, Pagliarusco A, Toscano A, et al. Ultrasound biomicroscopic and conventional ultrasonographic study of ocular dimensions in primary angle-closure glaucoma. Ophthalmology. 1998. 105:2091–2098.
15. Ramani KK, Mani B, Ronnie G, et al. Gender variation in ocular biometry and ultrasound biomicroscopy of primary angle closure suspects and normal eyes. J Glaucoma. 2007. 16:122–128.
16. Friedman DS, Gazzard G, Foster P, et al. Ultrasonographic biomicroscopy, Scheimpflug photography, and novel provocative tests in contralateral eyes of Chinese patients initially seen with acute angle closure. Arch Ophthalmol. 2003. 121:633–642.
TOOLS
Similar articles