Abstract
Purpose
To evaluate the flow resistance though the tube in a glaucoma drainage implant using various tube ligation methods.
Methods
To measure the flow resistance, the following tube ligation methods were designed and tested: Type 1: no ligation of Ahmed valve tube. Type 2: ligation of Ahmed valve tube with 8-0 Vicryl. Type 3: ligation of Molteno tube and a 6-0 Vicryl as an intra-luminal stent with 8-0 Vicryl. Type 4: ligation of Ahmed valve tube and three strands of 8-0 nylon as extra-luminal stents with 8-0 Vicryl. Type 5: ligation of Ahmed valve tube and a 6-0 Prolene as an extra-luminal stent with 8-0 Vicryl followed by removal of the Prolene for partial ligation.
Results
The pressure was maintained under 0.143 mm Hg in type 1. In type 2, the pressure increased to 6.688 mm Hg and dropped to approximately 6.384 mm Hg. In type 3, the pressure was maintained at 5.396 mm Hg which decreased to 3.572 mm Hg after stent removal. In type 4, the pressure was maintained at 5.700 mm Hg which dropped to 5.472 mm Hg after the 1st stent removal, to 5.016 mm Hg after the 2nd stent removal and to 4.180 mm Hg after the 3rd stent removal. In type 5, the pressure increased to 6.384 mm Hg and decreased to 5.624 mm Hg.
References
1. el-Sayyad F, el-Maghraby A, Helal M, Amayem A. The use of re-leasable sutures in Molteno glaucoma implant procedures to reduce postoperative hypotony. Ophthalmic Surg. 1991; 22:82–4.
2. Lee HJ, Choi KR. Clinical experience with the ahmed glaucoma valve implant in refractory glaucoma. J Korean Ophthalmol Soc. 2006; 47:933–40.
3. Kim DH, Park CK, Ahn MD. Clinical results of ahmed valve implantation in the aspects of complications. J Korean Ophthalmol Soc. 2003; 44:888–95.
4. Lee SH, Ma KT, Hong YJ. Outcome of ahmed valve implantation in refractory glaucoma. J Korean Ophthalmol Soc. 2007; 48:83–90.
5. Zorab A. The reduction of tension in chronic glaucoma. Ophthalmoscope. 1912; 10:258–61.
6. Molteno AC. New implant for drainage in glaucoma. Clinical trial. Br J Ophthalmol. 1969; 53:606–15.
7. Lloyd MA, Baerveldt G, Heuer DK, et al. Initial clinical experience with the baerveldt implant in complicated glaucomas. Ophthalmology. 1994; 101:640–50.
8. Schocket SS, Lakhanpal V, Richards RD. Anterior chamber tube shunt to an encircling band in the treatment of neovascular glaucoma. Ophthalmology. 1982; 89:1188–94.
9. Krupin T, Podos SM, Becker B, Newkirk JB. Valve implants in filtering surgery. Am J Ophthalmol. 1976; 81:232–5.
10. Joseph NH, Sherwood MB, Trantas G, et al. A one-piece drainage system for glaucoma surgery. Trans Ophthalmol Soc U K. 1986; 105:657–64.
11. Coleman AL, Hill R, Wilson MR, et al. Initial clinical experience with the ahmed glaucoma valve implant. Am J Ophthalmol. 1995; 120:23–31.
13. Allen RC, Bellows AR, Hutchinson BT, Murphy SD. Filtration surgery in the treatment of neovascular glaucoma. Ophthalmology. 1982; 89:1181–7.
14. Englert JA, Freedman SF, Cox TA. The Ahmed valve in refractory pediatric glaucoma. Am J Ophthalmol. 1999; 127:34–42.
15. Huang MC, Netland PA, Coleman AL, et al. Intermediate-term clinical experience with the Ahmed Glaucoma Valve implant. Am J Ophthalmol. 1999; 127:27–33.
16. Ayyala RS, Zurakowski D, Smith JA, et al. A clinical study of the Ahmed glaucoma valve implant in advanced glaucoma. Ophthalmology. 1998; 105:1968–76.
17. Coleman AL, Smyth RJ, Wilson MR, Tam M. Initial clinical experience with the Ahmed Glaucoma Valve implant in pediatric patients. Arch Ophthalmol. 1997; 115:186–91.
18. Kee C. Prevention of early postoperative hypotony by partial ligation of silicone tube in Ahmed glaucoma valve implantation. J Glaucoma. 2001; 10:466–9.
19. Lee J, Kim Y, Kee C. The effect of partial occlusion of silicone tube on opening and closing pressure and flow rate. J Korean Ophthalmol Soc. 2000; 41:2709–14.
20. Lee JJ, Park KH, Kim DM, Kim TW. Clinical outcomes of ahmed glaucoma valve implantation using tube ligation and removable external stents. Korean J Ophthalmol. 2009; 23:86–92.
21. Prata JA Jr, Mérmoud A, LaBree L, Minckler DS. In vitro and in vivo flow characteristics of glaucoma drainage implants. Ophthalmology. 1995; 102:894–904.
22. Francis BA, Cortes A, Chen J, Alvarado JA. Characteristics of glaucoma drainage implants during dynamic and steady-state flow conditions. Ophthalmology. 1998; 105:1708–14.
23. Topouzis F, Coleman AL, Choplin N, et al. Follow-up of the origi-nal cohort with the Ahmed glaucoma valve implant. Am J Ophthalmol. 1999; 128:198–204.
24. Krupin T, Ritch R, Camras CB, et al. A long Krupin-Denver valve implant attached to a 180 degrees scleral explants for glaucoma surgery. Ophthalmology. 1988; 95:1174–80.
25. Huh KC, Kee CW. A clinical analysis of the ahmed glaucoma valve implant with or without partial ligation of silicone tube. J Korean Ophthalmol Soc. 2000; 41:2611–7.