Lee, Lee, Kim, Kim, and Tchah: Comparison of Flap Thickness Measured with Ultrasound Subtraction Method, Direct Method, and Optical Coherence Tomography

Abstract

Purpose

To compare the flap thicknesses measured with the ultrasound (US) subtraction method, the direct method, and optical coherence tomography (OCT) in femtosecond laser-assisted LASIK eyes.

Methods

The present study included 65 eyes of 37 patients who underwent femtosecond laser-assisted LASIK surgery. Flap thickness was measured intraoperatively using the direct and subtraction methods with US pachymetry and post-operatively using anterior segment OCT. Flap thickness measurements were compared among the three methods.

Results

There was no significant difference between the flap thickness measured using the US subtraction method and OCT (p = 0.511). The flap thickness values obtained by these two different methods also correlated well with each other (γ= 0.820, p < 0.001). However, flap thickness measured using the US direct method was significantly different (p < 0.001) with a low agreement value compared to the other two methods. Furthermore, the US direct method showed a lower measurement success rate compared to the US subtraction method (54% vs. 95%).

Conclusions

Flap thickness measured using the US subtraction method and OCT showed different values compared to that measured using the US direct method. Because of the lower measurement success rate and agreement value, the US direct method is not capable of being used in substitute for the US subtraction method.

References

1. Knorz M. Flap and interface complications in LASIK. Curr Opin Ophthalmol. 2002; 13:242–5.
crossref
2. Seiler T, Koufala K, Richter G. Iatrogenic keratectasia after laser in situ keratomileusis. J Refract Surg. 1998; 14:312–7.
crossref
3. Stulting R, Carr J, Thompson K, Waring G. Complications of laser in situ keratomileusis for the correction of myopia. Ophthalmology. 1999; 106:13–20.
crossref
4. Solomon K, Donnenfeld E, Sandoval H, et al. Flap thickness accuracy comparison of 6 microkeratome models. J Cataract Refract Surg. 2004; 30:964–77.
5. Talamo J, Meltzer J, Gardner J. Reproducibility of flap thickness with IntraLase FS and Moria LSK-1 and M2 microkeratomes. J Refract Surg. 2006; 22:556–61.
crossref
6. Schumer DJ, Bains HS. The Nidek MK-2000 microkeratome system. J Refract Surg. 2001; 17:S250–1.
crossref
7. Sutton G, Hodge C. Accuracy and precision of LASIK flap thickness using the IntraLase femtosecond laser in 1000 consecutive cases. J Refract Surg. 2008; 24:802–6.
crossref
8. Giledi O, Mulhern M, Espinosa M, et al. Reproducibility of LASIK flap thickness using the Hansatome microkeratome. J Cataract Refract Surg. 2004; 30:1031–7.
crossref
9. Flanagan GW, Binder PS. Precision of flap measurements for laser in situ keratomileusis in 4428 eyes. J Refract Surg. 2003; 19:113–23.
crossref
10. Flanagan G, Binder PS. Estimating residual stromal thickness before and after laser in situ keratomileusis. J Cataract Refract Surg. 2003; 29:1674–83.
crossref
11. Muallem MS, Yoo SH, Romano AC, et al. Flap and stromal bed thickness in laser in situ keratomileusis enhancement. J Cataract Refract Surg. 2004; 30:2295–302.
crossref
12. Eisner R, Binder P. Technique for measuring laser in situ keratomileusis flap thickness using the IntraLase laser. J Cataract Refract Surg. 2006; 32:556–8.
crossref
13. Stahl JE, Durrie DS, Schwendeman FJ, Boghossian AJ. Anterior segment OCT analysis of thin IntraLase femtosecond flaps. J Refract Surg. 2007; 23:555–8.
crossref
14. Chen HJ, Xia YJ, Zhong YY, et al. Anterior segment optical coherence tomography measurement of flap thickness after myopic LASIK using the Moria one use-plus microkeratome. J Refract Surg. 2010; 26:403–10.
crossref
15. Kim HJ, Oh SH, Lee DH, et al. Reproducibility of IntraLASIK flap thickness measured with optical coherence tomography. J Korean Ophthalmol Soc. 2007; 48:1630–5.
crossref
16. Friedlaender M. LASIK surgery using the IntraLase femtosecond laser. Int Ophthamol Clin. 2006; 46:145–53.
crossref
17. Cheng AC, Ho T, Lau S, et al. Measurement of LASIK flap thickness with anterior segment optical coherence tomography. J Refract Surg. 2008; 24:879–84.
crossref
18. Li EY, Mohamed S, Leung CK, et al. Agreement among 3 methods to measure corneal thickness: ultrasound pachymetry, Orbscan II, and Visante anterior segment optical coherence tomography. Ophthalmology. 2007; 114:1842–7.
crossref
19. Zhao PS, Wong TY, Wong WL, et al. Comparison of central corneal thickness measurements by visante anterior segment optical coherence tomography with ultrasound pachymetry. Am J Ophthalmol. 2007; 143:1047–9.
crossref
20. Wildner K, Muller M, Dawczynski J, Strobel J. Comparison of the corneal thickness as measured by Visante anterior segment OCT versus ultrasound technique. Klin Monbl Augenheilkd. 2007; 224:832–6.
21. Li Y, Netto MV, Shekhar R, et al. A longitudinal study of LASIK flap and stromal thickness with high-speed optical coherence tomography. Ophthalmology. 2007; 114:1124–32.
crossref
22. Nagy ZZ, Resch M, Suveges I. Ultrasound evaluation of flap thickness, ablation depth, and corneal edema after laser in situ keratomileusis. J Refract Surg. 2004; 20:279–81.
crossref
23. Thomas J, Wang J, Rollins AM, Sturm J. Comparison of corneal thickness measured with optical coherence tomography, ultrasonic pachymetry, and a scanning slit method. J Refract Surg. 2006; 22:671–8.
crossref
24. Dawson DG, Holley GP, Geroski DH, et al. Ex vivo confocal microscopy of human LASIK corneas with histologic and ultra-structural correlation. Ophthalmology. 2005; 112:634–44.
crossref
25. Wylegał a E, Teper S, Nowiή ska AK, et al. Anterior segment imaging: Fourier-domain optical coherence tomography versus time-domain optical coherence tomography. J Cataract Refract Surg. 2009; 35:1410–4.

Figure 1.
Postoperative flap thickness measurement with anterior segment optical coherence tomography.
jkos-52-524f1.tif
Figure 2.
Scattergram showing the correlation of flap thickness measured by ultrasound (US) subtraction method, US direct method, and optical coherence tomography (OCT). (A) Correlation between US subtraction method and US direct method. (B) Correlation between US subtraction method and OCT. (C) Correlation between US direct method and OCT.
jkos-52-524f2.tif
Figure 3.
Bland-Altman plots between each two measurement methods. The middle line is the mean and the lines on the side represent the upper and lower 95% limits of agreement (LoA). (A) Ultrasound (US) subtraction method and US direct method. (B) US subtraction method and optical coherence tomography (OCT). (C) US direct method and OCT.
jkos-52-524f3.tif
Table 1.
Comparison of the measured flap thickness
Method n Flap thickness (mean ± SD, μm) Range (μm) p-value
US* subtraction pachymeter 32 118.3 ± 14.4 86.3-154.8 0.005
US direct pachymeter 127.4 ± 14.3 103.0-158.5
US subtraction pachymeter 28 125.9 ± 16.8 96.7-167.3 0.511
OCT 124.7 ± 10.6 111.0-153.0
US direct pachymeter 19 131.4 ± 14.5 103.0-158.5 0.007
OCT 120.0 ± 10.3 101.0-141.0

* US = ultrasound;

OCT = optical coherence tomography.